Flexitime Fast & Scan is an addition-cross-linking polyvinyl siloxane impression material for all inlay, crown & bridges, endentulous and partial impressions. The Flexitime Fast & Scan range of products are prepared without requiring additional surface treatment for optical scanning in dental scanners designed for scanning impression materials, such as the 3shape D700 laser scanners.
Device Story
Addition-cross-linking polyvinyl siloxane impression material; used for dental impressions (inlay, crown, bridge, edentulous, partial). Contains titanium dioxide to enable direct optical scanning by red laser dental scanners (e.g., 3shape D700) without surface powdering. Delivered in cartridges or containers; applied by dental professionals in clinical settings. Provides hydrophilic properties and mechanical stability for accurate dental models. Benefits include streamlined digital workflow by eliminating surface treatment steps.
Clinical Evidence
No clinical data. Safety and effectiveness supported by bench testing, biocompatibility evaluation per ISO 10993, and risk management analysis. Product performance verified against ISO 4823 standards.
Technological Characteristics
Addition-cross-linking polyvinyl siloxane impression material; contains titanium dioxide for optical scannability. Compliant with ISO 4823. Delivered in 50ml cartridges, 380ml cartridges, or 300ml containers. Hydrophilic formulation; extra-oral working time up to 1.5 minutes; intra-oral setting time of 2.0 minutes.
Indications for Use
Indicated for dental impressions including inlay, crown, bridge, edentulous, and partial impressions. For use by dental professionals.
Regulatory Classification
Identification
Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.
Flexitime Fast & Scan Impression Material (K102770)
Submission Summary (Full Text)
{0}------------------------------------------------
Heraeus
# 510(k) Summary
JUL 25 2012
510(k) summary of safety and effectiveness information for Flexitime Fast & Scan in accordance with 21 CFR 807.92.
# 1. Submitter Information
| Company Name: | Heraeus Kulzer, LLC |
|------------------|-----------------------------------------|
| Company Address: | 300 Heraeus Way<br>South Bend, IN 46614 |
| Company Phone: | (574) 299-5421 |
Contact Person: Jamie Mearna
Date Summary prepared: December 28, 2011(Revised June 29, 2012)
# 2. Device Identification
Trade /Proprietary Name:
- Flextime Fast & Scan Impression Materials to include
- Flexitime Fast & Scan Light Flow �
- ◆ Flexitime Fast & Scan Medium Flow,
- Flexitime Fast & Scan Dynamix Putty �
- . Flexitime Fast & Scan Easy Putty
- Flexitime Fast & Scan Dynamix Heavy Tray .
| Common Name: | Material, Impression (21 CFR 872.3660) |
|-----------------|----------------------------------------|
| Classification: | Class II |
| FDA Code: | Dental ELW. |
# 3. Legally Market Devices to which Substantial Equivalence is claimed:
Heraeus Kulzer, GmBH - Flexitime Impression Material (K000629) Heraeus Kulzer, GmBH - Flexitime Fast & Scan Impression Material (K102770)
# 4. Device Description
Flexitime Fast & Scan Impression Materials are addition-cross-linking polyvinyl siloxane materials. The Flexitime Fast & Scan Impression Materials family consists of Light Flow, Medium Flow, Dynamix Putty, Easy Putty and Dynamix Heavy Tray. Flexitime Light Flow, Medium Flow, Heavy Tray are delivered in 50 ml cartridges while Flexitime Fast & Scan Dynamix Putty and Dynamix Heavy Tray are delivered in 380 ml Cartridges and Flexitime Fast & Scan Easy Putty is delivered in a 300 ml container.
The Flexitime Fast & Scan assortment is characterized by the addition of Titanium Dioxide (in order to ensure scannability) and is technically characterized by an extra-oral working time of up to 1.5 minutes and a short time in mouth of 2.0 minutes. The materials were developed to ensure hydrophilic characteristics for optimal impression
{1}------------------------------------------------
taking in the wet surroundings of the mouth combined with good mechanical properties. Also, they ensure scannability with state of the art red laser light impression scanners.
Flexitime Fast & Scan products are part of the Flexitime System.
# 5. Intended Use
Flexitime Fast & Scan is an addition-cross-linking polyvinyl siloxane impression material for all inlay, crown & bridges, endentulous and partial impressions.
The Flexitime Fast & Scan range of products are prepared without requiring additional surface treatment for optical scanning in dental scanners designed for scanning impression materials, such as the 3shape D700 laser scanners.
# 6. Technological Characteristics
The physical properties of the additional Flexitime Fast& Scan assortments: Dynamix Heavy Tray and Easy Putty are like Flexitime Fast and Scan (K102770) and Flexitime (K000629) products in the fact that they are also in compliance with ISO 4823.
The new assortment of Flextime Fast and Scan materials (like the original Flexitime Fast & Scan K102770) can be prepared without requiring additional surface treatment for optical scanning in dental scanners designed for scanning impression materials.
# 7, Nonclinical Testing
In accordance with US FDA Medical Device Regulations, any medical device is required to be evaluated by the legal medical device manufacturer regarding its clinical performance and safety. This includes an evaluation of biocompatibility in accordance with FDA recognized standard, EN ISO 10993-1.
The biological compatibility of Flexitime Fast & Scan was verified in accordance with the international standards.
The biocompatibility of Flexitime Fast & Scan in the aforementioned indication was documented in a biocompatibility evaluation report and the benefit/risk relation has been judged as positive.
#### 8. Clinical Testing
In accordance with US FDA Medical Device Regulations, any medical device is required to be evaluated by the legal medical device manufacturer regarding its clinical performance and c variation by the regal nical evaluation, which is intended to critically evaluate the clinical benefits of the medical device in comparison to tis potential risks. Therefore, any clinical benefits of the medical compulsory risk management process according to FDA recognized evaluation 15 part of the son critical findings must further be considered in the current risk Standard, int process of the medical device manufacturer responsible for the evaluated device.
{2}------------------------------------------------
On this background, the clinical evaluation was performed. This critical evaluation followed the procedures outlined in the corresponding clinical evaluation plan.
Considering the evaluated scientific data and technical results for Flextime Fast & Scan it is concluded that the products can be expected to exhibit the claimed technical performance and that potential undesirable clinical effects and risks seem well controlled and accepted, when weighed against their benefit to dentistry. Therefore, a positive benefit versus risk ratio can be stated by the experts for Flexitime Fast & Scan assortment, provided that the products are applied in accordance with the intended use as outlined in the manufacturer's instructions for use.
# 9. Summary of Conclusion
The new Flexitime Fast & Scan assortment is substantially equivalent to the original Flexitime Fast & Scan and Flexitime. All of the products are indicated for taking impression materials of suited techniques. In addition, the new assortment of Flexitime Fast & Scan is scannable without prior powdering as is required with the predicate device, Flexitime Fast and Scan.
A biocompatibility evaluation has been performed by a toxicologist for Flextime Fast & Scan, and it was confirmed that the product meets the requirements of FDA recognized standard, ISO 10993 Standard and it was concluded that the safety of Flexitime Fast & Scan is equivalent to that of the predicate device.
The risk analysis was carried out for Flexitime Fast & Scan and it was concluded that the safety of the Flexitime Fast & Scan device, for the intended use, is substantially equivalent to the predicate device. Flexitime Fast & Scan and the predicate device both have the same indications for use, warnings and contraindications.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUL 25 2012
Mr. Jamie Mearna Quality Assurance & Regulatory Associate Heraeus Kulzer, LLC 300 Heraeus Way South Bend, Indiana 46614
Re: K120914
Trade/Device Name: Flexitime Fast & Scan Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: II Product Code: ELW Dated: July 09, 2012 Received: July 11, 2012
Dear Mr. Mearna:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
# Page 2 - Mr. Mearna:
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR. 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
For
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K/26 914
Device Name: Flexitime Fast & Scan
Indications for Use:
Flexitime Fast & Scan is an addition-cross-linking polyvinyl siloxane impression material for all inlay, crown and bridges, endentulous and partial impression.
The Flexitime Fast & Scan range of products are prepared without requiring additional surface treatment for optical scanning in dental scanners designed for scanning impression materials, such as the 3 shape D700 laser scanners.
Prescription Use V . Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <img alt="Signature" src="signature.png"/> |
|---------------|----------------------------------------------|
| | (Division Sign-Off) |
| | Division of Anesthesiology, General Hospital |
| | Infection Control, Dental Devices |
| 10(k) Number: | k120914 |
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.