K120909 · Medtronic Advanced Energy, LLC · GEI · Nov 2, 2012 · General, Plastic Surgery
Device Facts
Record ID
K120909
Device Name
EC 2.7 ENDOSCOPIC CUTTER
Applicant
Medtronic Advanced Energy, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
Nov 2, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EC 2.7 Endoscopic Cutter is intended to be inserted through the working channel of a flexible bronchoscope having an instrument channel diameter of 2.8 mm minimum, a working length of 600 mm, activated by a foot-switch connected to a qualified electrosurgical generator, and utilized in electrosurgical procedures involving removal/cutting of soft tissues (excision, incision, vaporization, ablation) while also providing electrosurgical coagulation and hemostasis.
Device Story
Disposable, single-patient-use monopolar RF surgical endotherapy device; inserted through working channel of flexible bronchoscope. Comprised of cutting/coagulation electrode on flexible multi-lumen shaft; includes electrical adaptor for connection to qualified electrosurgical generator (e.g., ValleyLab Force FX). Operated by physician via generator foot-switch. Device delivers RF energy to soft tissue for excision, incision, vaporization, ablation, and hemostasis. Used in clinical settings for upper airway and tracheobronchial tree procedures. Output allows physician to remove obstructions and control bleeding during bronchoscopy, potentially improving patient airway patency.
Clinical Evidence
No clinical data was necessary. Substantial equivalence supported by bench testing, human cadaveric studies, and in vivo animal data collection.
Technological Characteristics
Monopolar RF electrosurgical instrument; polymeric flexible multi-lumen shaft; 750 mm active length; 2.68 mm max diameter. Sterilized via ETO. Compliant with ISO 11135-1, ISO 11607, ISO 10993 (biocompatibility), ANSI/AAMI ES 60601-1, IEC 60601-2-2, and IEC 60601-2-18.
Indications for Use
Indicated for cutting soft tissue obstructions in upper airways and tracheobronchial tree and providing electrosurgical hemostasis in patients undergoing bronchoscopic procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Medtronic Advanced Energy LLC Traditional 510(k)
EC 2.7 Endoscopic Cutter
*NOV* 2 *2012*
#### K120909 510(K) SUMMARY
EC 2.7 Endoscopic Cutter - 510(k) Summary
Submitter:
Medtronic Advanced Energy, LLC
Contact:
Martin J. Leighton Tel: 603-742-1515 Fax: 603-742-1488
November 2, 2012 Date Summary Prepared:
Device Trade Name: EC 2.7 Endoscopic Cutter
Common Name: Electrosurgical cutting and coagulation accessories and Endoscopic electrosurgical unit and accessories
Classification Name: Electrosurgical cutting and coagulation device and accessories
21 CFR §878.4400 Classification Code:
Product Code GEI
Equivalent Device(s): Olympus CD-6C-1 Coagulation Electrode (K971321)
The EC 2.7 Endoscopic Cutter is a disposable, single-Device Description: patient-use monopolar RF (radio-frequency) surgical endotherapy device intended to be inserted through the working channel of a flexible bronchoscope. The EC 2.7 Endoscopic Cutter is terminally sterilized via ETO. It is comprised of a separately packaged, cylindrically shaped, disposable monopolar cutting and coagulation electrode affixed to the distal end of a polymeric, flexible multi lumen shaft , and a separately packaged, single-patient-use, electrical adaptor with cabling, intended for connection to the monopolar output port of a Covidien® Corporation, ValleyLab™ Force FX™ or FX-C Electrosurgical Generator. The ValleyLab™ Force FX™ was cleared to market via 510(k) K944602.
{1}------------------------------------------------
EC 2.7 Endoscopic Cutter
Medtronic Advanced Energy LLC Traditional 510(k)
Intended use:
Indications for Use:
Non-clinical Performance Data: The Endoscopic Cutter System is intended to be energized for use by the foot-switch accessory supplied with the electrosurgical generator. The EC 2.7 has an active length of approximately 750 mm and a maximum diameter crossing profile of 2.68 mm. It is sized such that may be delivered through the working channel of a flexible bronchoscope 2.8 mm diameter working channel and working length of 600 mm.
The EC 2.7 Endoscopic Cutter is intended to be inserted through the working channel of a flexible bronchoscope having an instrument channel diameter of 2.8 mm minimum, a working length of 600 mm, activated by a foot-switch connected to a qualified electrosurgical generator, and utilized in electrosurgical procedures involving removal/cutting of soft tissues (excision, incision, vaporization, ablation) while also providing electrosurgical coagulation and hemostasis.
The EC 2.7 Endoscopic Cutter is a single use electrosurgical instrument designed to be used with flexible bronchoscopes and qualified electrosurgical units. It is indicated for cutting of soft tissue obstructions in upper airways and tracheobronchial tree, and provision of electrosurgical hemostasis during such procedures.
Performance testing per standardized methods and internal test protocols, including bench testing, human cadaveric studies and in vivo animal data collection, was conducted and provides support that the EC 2.7 Endoscopic Cutter is substantially equivalent to currently marketed predicate devices. Additional comparative performance testing was conducted inconsideration of other devices with similar intended indications and intended use.
Non-clinical performance test protocols and reports of results are identified in Sections 14 through 19 of this 510(k) submission.
Test protocols and reports of results demonstrate that, in consideration of its intended and indications for use, the performance, design, labeling, packaging and sterilization of the EC 2.7 Endoscopic Cutter is compliant with the following standards:
- ANSI/AAMI ISO 11135-1:2007: Sterilization of health care products - Ethylene Oxide - Part 1: Requirements for the development, validation and
{2}------------------------------------------------
- ISO 11607-1:2006, Packaging for terminally sterilized medical devices -Part 1: Requirements for materials, sterile barrier
- ISO 11607-2:2006: Packaging for terminally sterilized . medical devices -Part 2: Validation requirements for forming, sealing and assembly processes, including Annex B (informative) listing of standardized test methods and procedures
- ASTM F1980-07; Shelf-life and accelerated aging . techniques for standard evaluation of packaging performance
- ISO 10993-4:2002, Biological evaluation of medical . devices - Part 4: Selection of tests for interactions with blood
- ISO 10993-5:2009, Biological evaluation of medical ● devices - Part 5: Tests for In Vitro Cytotoxicity.
- ISO 10993-7:2008, Biological evaluation of medical . devices - Part 7: Ethylene oxide sterilization residuals
- ISO 10993-10:2002 (A1:2006), Biological evaluation . of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity.
- ISO 10993-11:2006, Biological evaluation of medical . devices - Part 11: Tests for systemic toxicity.
- ANSI AAMI ES 60601-1: 2005 3rd Edition: Medical . electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2007, Medical Electrical Equipment -Part 1-2: General Requirements for Safety Collateral Standard Electromagnetic Compatibility Requirements and Tests.
- IEC 60601-2-2:2009, 5th Edition, Medical electrical . equipment - Part 2-2: Particular requirements for the safety of high frequency surgical equipment
- IEC 60601-2-18:2009, 3th Edition, Particular . requirements for the basic safety and essential performance of endoscopic equipment
Clinical Performance Data:
Clinical data was not necessary to support that the EC 2.7 is substantially equivalent to currently marketed predicate devices.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized depiction of an eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Letter Dated: November 2, 2012
Medtronic Advanced Energy, LLC % Mr. Martin J. Leighton Senior Regulatory Affairs Specialist 180 International Drive Portsmouth, New Hampshire 03801
Re: K120909
Trade/Device Name: EC 2.7 Endoscopic Cutter Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: October 25, 2012 Received: October 26, 2012
Dear Mr. Leighton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
### Page 2 - Mr. Martin J. Leighton
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
### 510(k) Number: K120909
Device Name: EC 2.7 Endoscopic Cutter
Indications For Use: The EC 2.7 Endoscopic Cutter is a single use electrosurgical instrument designed to be used with flexible bronchoscopes and qualified electrosurgical units. It is indicated for cutting of soft tissue obstructions in upper airways and tracheobronchial tree, and provision of electrosurgical hemostasis during such procedures.
Prescription Use X (21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
K120909
Mark M. Miller
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
Page 1 of 1
510(k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.