K120895 · Intouch Health, Inc. · DRG · May 24, 2012 · Cardiovascular
Device Facts
Record ID
K120895
Device Name
REMOTE PRESENCE SYSTEM
Applicant
Intouch Health, Inc.
Product Code
DRG · Cardiovascular
Decision Date
May 24, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2910
Device Class
Class 2
Indications for Use
The Remote Presence System is a clinical communications tool that provides a means of transmitting, receiving, and storing real-time audio and video, and patient data. The Remote Presence System may also be used in conjunction with 510(k)-cleared devices that transmit patient biometric data, including vital signs information. The Remote Presence System transmits and receives information over a highspeed connection between patients, health professionals, and critical transport teams. The Remote Presence System can be used in communications for active patient monitoring in high acuity clinical environments where immediate clinical action may be required, e.g., pre-, peri-operative and postsurgical, cardiovascular, neurological, pre-natal, psychological, and critical care assessments and examinations. Clinical judgment and experience are required to review and interpret the information transmitted.
Device Story
Telecommunications platform enabling real-time audio/video/data transmission between patients and remote clinicians. System comprises Control Station (PC) and endpoint (mobile cart, handheld, or boom-mounted). Inputs include live video/audio feeds and data from 510(k)-cleared medical devices (e.g., electronic stethoscope). Operator uses mouse/joystick to control endpoint movement or manually positions device. Connectivity via broadband internet, 802.11 wireless, or cellular. Output displayed on monitors for clinician review. Facilitates remote clinical assessment, monitoring, and decision-making in hospitals, clinics, and transport settings. Benefits include expanded access to specialist care in high-acuity environments.
Clinical Evidence
Bench testing only. Validation testing performed on system hardware and software improvements. Independent testing confirmed effectiveness of electronic stethoscope communication channel.
Technological Characteristics
Telecommunications platform; wired/wireless (802.11, cellular) connectivity; RS-232, Serial, USB, Bluetooth interfaces for external devices. Powered by AC or internal batteries with integrated chargers. VGA display monitors. Proprietary software for data collection/transmission. Operates via Session Initiation Protocol or proprietary protocols.
Indications for Use
Indicated for patients in high acuity clinical environments (pre/peri/post-surgical, cardiovascular, neurological, pre-natal, psychological, critical care) requiring active monitoring and assessment by healthcare professionals via remote communication.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
Predicate Devices
InTouch Remote Presence Robotic System, Model RP-7 (K073710)
{0}------------------------------------------------
K120895
#### 5. 510(k) Summary
MAY 24 2012
| Name of 510(k) sponsor: | InTouch Health, Inc. |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 6330 Hollister Ave.<br>Goleta, CA 93117 |
| Contact information: | Steve Sidwell<br>Director of Regulatory Affairs & Quality Assurance<br>InTouch Health<br>6330 Hollister Ave.<br>Goleta, CA 93117<br>Phone: 805 562 8686 (ext. 254)<br>Fax: 805 562 8663 |
| Date summary prepared: | March 22, 2012 |
| Proprietary name of device: | Remote Presence System- RP-7i®, RP-B (a/k/a BoomBot and<br>KSEA VisitOR1); RP-Lite®; KSEA VisitOR1 Cart; RP-<br>Vantage®; and RP-Xpress™ |
| Generic/classification name: | Transmitters and Receivers, Physiological Signal,<br>Radiofrequency |
| Product code (classification): | 21 C.F.R. § 870.2910, Product Code DRG; Class II |
| Legally Marketed Predicate Device: | InTouch Remote Presence Robotic System, Model RP-7;<br>K073710; April 11, 2008 |
# Device Description and Technological Characteristics:
The Remote Presence System is a telecommunications platform that enables real-time videoconferencing and clinical communications, and provides a means for transmitting, receiving, and storing real-time audio and video, and patient data. The Remote Presence System consists of a Control Station ("CS"), (i.e., desktop or laptop computer) and an end point, which may be controlled by an input device (i.e., mouse or joystick) that the operator uses to control the movement of the end point in the remote location (e.g., RP-70), or manually located by the user (e.g., RP-Lite®, KSEA VisitOR1 Cart, RP-Vantage®, or RP-Xpress™), or used in a restricted clinical environment, such as an operating room, where it is boom mounted (e.g., BoomBot (aka RP-B and KSEA VisitOR1)). The end point and CS are each equipped with various combinations of cameras, displays, microphones and speakers, depending upon the specific device, which facilitate two-wav audio-video communication. Optional accessories include Class II devices, including an integrated electronic stethoscope, which are used for the same purpose for which they received 510(k) clearance. Communication between the CS and the end-point is established wia broadband Internet and an 802.11 wireless network or a broadband cellular connection.
Like the predicate device, the Remote Presence System provides a real-time link between the patient and the healthcare professional. This link occurs over a wired or wireless broadband connection and includes real-time audio and video to facilitate communication between the patient, side healthcare professionals, and remote healthcare professionals. Also like the predicate device, the Remote Presence System provides connections for the transfer of data from 510(k)-cleared devices
{1}------------------------------------------------
K120895 P2/4
# 510(k) Submission Remote Presence System
between the patient and the healthcare professional. Like the predicate device, these 510(k) cleared devices are not controlled or manipulated through the Remote Presence System, and consequently, no additional risk is presented.
Expanding on the predicate device, the Remote Presence System introduces new cart-based and hand-held models, providing improved system flexibility by adapting to additional clinical environments (e.g., operating rooms, emergency departments, intensive care units, patient transport, etc.).
Redundant safeguards are designed into the Remote Presence System to address risks associated with additional models, and hardware and software improvements. The effectiveness of these improvements was demonstrated by the validation testing performed on the system. The communication channel used by the electronic stethoscope was also proven effective by independent tests.
The performance data discussed in this 510(k) application demonstrate that the Remote Presence System is as safe and effective, and performs as well as or better than the predicate device.
#### Intended Use:
The Remote Presence System is a clinical communications tool that provides a means of transmitting, receiving, and storing real-time audio and video, and patient data. The Remote Presence System may also be used in conjunction with 510(k)-cleared devices that transmit patient biometric data, including vital signs information. The Remote Presence System transmits and receives information over a highspeed connection between patients, health professionals, and critical transport teams. The Remote Presence System can be used in communications for active patient monitoring in high acuity clinical environments where immediate clinical action may be required, e.g., pre-, peri-operative and postsurgical, cardiovascular, neurological, pre-natal, psychological, and critical care assessments and examinations. Clinical judgment and experience are required to review and interpret the information transmitted.
### Comparison with Predicate Device
A substantial equivalence table comparing the InTouch Remote Presence System to the predicate device is provided below.
{2}------------------------------------------------
510(k) Submission
Remote Presence System
| Table<br>5.1:<br>Substance | Equivalent<br>Quantity | Comparison |
|----------------------------|------------------------|------------|
| d-Tartaric | 1 | 1 |
| dl-Tartaric | 2 | 2 |
| L-Tartaric | 1 | 1 |
| Meso-Tartaric | 1 | 3 |
| | New Device | Predicate Device |
|-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) # | To be assigned | K073710 |
| Company | InTouch Health | InTouch Health |
| Name/Model # | Remote Presence System | Remote Presence Robotic System, Model RP-7 |
| Indications for use | The Remote Presence System is a clinical communications<br>tool that provides a means of transmitting, receiving, and<br>storing real-time audio and video, and patient data. The<br>Remote Presence System may also be used in conjunction<br>with 510(k)-cleared devices that transmit patient biometric<br>data, including vital signs information. The Remote Presence<br>System transmits and receives information over a high-speed<br>connection between patients, health professionals, and<br>critical transport teams. The Remote Presence System can<br>be used in communications for active patient monitoring in<br>high acuity clinical environments where immediate clinical<br>action may be required, e.g., pre-, peri-operative and post-<br>surgical, cardiovascular, neurological, pre-natal,<br>psychological, and critical care assessments and<br>examinations. Clinical judgment and experience are required<br>to review and interpret the information transmitted. | The RP-7 is a communications tool that provides a means of<br>transmitting, receiving, and storing real-time audio and video,<br>and patient data. The RP-7 also can be used in conjunction<br>with 510(k)-cleared devices that transmit patient biometric data,<br>including vital signs information. The RP-7 transmits and<br>receives information over a high speed connection between<br>patients and health professional, and is intended to be used in a<br>hospital or clinic environment. Clinical judgment and experience<br>are required to review and interpret the information transmitted. |
| Intended use | Telemedicine system | Telemedicine system |
| Intended users | Healthcare professional, inpatient, outpatient | Healthcare professional, inpatient |
| Site of use | Hospital, clinic, patient transport | Hospital, clinic |
| Data collection<br>software | Proprietary software | Proprietary software |
| Communication<br>method with<br>remote care<br>management<br>system | Wired or wireless broadband connection | Wired broadband connection |
| | New Device | Predicate Device |
| 510(k) # | To be assigned | K073710 |
| Company | InTouch Health | InTouch Health |
| Name/Model # | Remote Presence System | Remote Presence Robotic System, Model RP-7 |
| Types of devices that can be interfaced (wired or wirelessly) to receiver hub | Electronic Stethoscope (K034046) and other cleared medical devices that transmit patient data. | Electronic Stethoscope (K034046) |
| Implementation method of collecting data from device | External communication device | External communication device |
| Sensor software | Unchanged | Unchanged |
| Connectivity | Wired, wireless to hub | Wireless to hub |
| Communication method of hub with devices | RS-232, Serial communication, USB, Bluetooth® | RS-232, Serial communication |
| Communications protocol | Proprietary or Session Initiation Protocol | Proprietary |
| Wireless frequency | 802.11 A, B, or G (varies based on the customer) | 802.11 A, B, or G (varies based on the customer) |
| Power source | AC or batteries with AC-DC battery chargers built in | Batteries with AC-DC battery chargers built in |
| Display | VGA Monitors on computers and end points | VGA Monitor on computers |
| Video conferencing | 2-way video conferencing via a broadband internet or cellular connection | 2-way video conferencing via a broadband internet connection |
Page 8
DB1/ 69029658.3
{3}------------------------------------------------
:
Page 9
DB1/ 69029658.3
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's three main goals: health, human services, and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES" are arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
# MAY 2 4 2012
InTouch Health, Inc. c/o Mr. Steve Sidwell Director of Regulatory Affairs & Quality Assurance 6330 Hollister Ave. Goleta, CA 93117
K120895 Re:
> Trade/Device Name: Remote Presence System RP-7i®, RP-B (a/k/a BoomBot and KSEAVisitOR1 ); RP-Lite®; KSEA VisitOR1 Cart; RPVantage®; and RP-Xpress™
Regulation Number: 21 CFR 870.2910
Regulation Name: Radiofrequency Physiological Signal Transmitters and Receivers, Regulatory Class: Class II (two) Product Codes: DRG Dated: March 22, 2012 Received: March 26, 2012
Dear Mr. Sidwell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 – Mr. Steve Sidwell
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours
Bram D. Zuckerman, M.D.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K120895
# 510(k) Submission Remote Presence System
#### 4. Indications for Use Statement
· Applicant: InTouch Health, Inc. 510(k) Number. Not assigned. Device Name: Remote Presence System
Indications for Use: The Remote Presence System is a clinical communications tool that provides a means of transmitting, receiving, and storing real-time audio and video, and patient data. The Remote Presence System may also be used in conjunction with 510(k)-cleared devices that transmit patient biometric data including vital signs information. The Remote Presence System transmits and receives information over a high speed connection between patients, health professionals and critical transport teams. The Remote Presence System can be used in communications for active patient monitoring in . high acuity clinical environments where immediate clinical action may be required, e.g., pre-, perioperative and post-surgical, cardiovascular, neurological, pre-natal, psychological and critical care assessments and examinations. Clinical judgment and experience are required to review and interpret the information transmitted.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number *K120895*
CONFIDENTIAL
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.