The Hybrid ArcticGel™ Pad is indicated as a standalone cooling device for the reduction of core temperature of adult patients where clinically indicated (e.g. heat stroke or hyperthermia). Additionally, the Hybrid ActicGel™ Pad is indicated as an adjunct to patient temperature control with the Arctic Sun® Temperature Management System. The Hybrid ArcticGel™ Pad is intended for use by health care professionals in all clinical environments, including, but not limited to in- hospital as well as EMS/transport settings.
Device Story
Hybrid ArcticGel Pad functions as a dual-mode thermal regulation device. In Ice Gel Mode, pre-frozen pads (stored at -3°C to -20°C) are applied directly to patient skin; heat transfer occurs via conduction through hydrogel adhesive into the ice gel compartment, providing cooling for 60-90 minutes. In Water Flow Mode, pads connect to the Arctic Sun Temperature Management System; temperature-controlled water circulates through the pad to maintain target patient temperature. Used by healthcare professionals in hospitals or EMS/transport settings. Output is patient core temperature modulation. Benefits include early initiation of cooling in pre-hospital settings and continuous temperature management in clinical environments.
Clinical Evidence
Bench testing and porcine skin model testing performed. Biocompatibility testing (ISO 10993-05/10) confirmed hydrogel adhesive is non-cytotoxic, non-irritating, and non-sensitizing. Porcine model demonstrated that minimum mean skin temperature remains above freezing, reaching minimum values within 15 minutes and rising over 90 minutes; placement was well-tolerated. Comparative performance testing confirmed cooling and temperature control performance equivalent to predicates.
Technological Characteristics
Noninvasive surface thermal regulation pad. Materials: polymer construction with biocompatible medical adhesive. Sensing/Actuation: conductive heat transfer via hydrogel layer. Energy Source: passive (pre-frozen ice) or active (circulating water via Arctic Sun system). Connectivity: fluid line connection to Arctic Sun control module. Sterilization: not specified. Software: none.
Indications for Use
Indicated for adult patients requiring core temperature reduction (e.g., heat stroke, hyperthermia) or temperature control via the Arctic Sun Temperature Management System. Intended for use by healthcare professionals in hospital and EMS/transport settings.
Regulatory Classification
Identification
A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.
{0}------------------------------------------------
JUL 20 2012
Image /page/0/Picture/1 description: The image shows the title "510(k) SUMMARY" followed by the date the document was prepared. The date is listed as May 15, 2012. The text is in black font and is centered on the page. The text is easy to read.
### Device Name
Arctic Sun Hybrid ArcticGel Pad
# Manufacturer Name , Address and Contact Information
Medivance. Inc. 321 South Taylor Avenue, Suite 200 Louisville, CO 80027 Lynne Aronson, Director Regulatory Affaris Contact: 303-926-1917 Phone: 720-880-5400 Fax:
# Common, Classification & Proprietary Names
Common Name: Classification Name: Proprietary Name: Classification: Classification Panel: Classification Regulation: Product Code:
# patient temperature management system system, thermal regulating Arctic Sun Hybrid ArcticGel Pad Class II Cardiovascular Devices 870.5900 DWJ
K010338 K100071
### Predicate Devices
- Medivance Arctic Sun ArcticGel Pad
- 페 EMCOOLS Pad
# Device Description
The Arctic Sun Hybrid ArcticGel Pads are a modified version of the Arctic Sun ArcticGel Pads.
The Hybrid ArcticGel Pad is designed to operate in two different modes, Ice Gel Mode in which is the pre-frozen pads are used as standalone devices to initiate patient cooling, and Water Flow Mode in which the pads are connected to the Arctic Sun Temperature Management System which is used to control the patient at a target temperature.
These dual modes provide the opportunity for early initiation of cooling in the pre-hospital setting or the hospital emergency department followed by continued temperature management with the Arctic Sun once the ice gel is thawed.
The Hybrid ArcticGel Pads are pre-frozen in a freezer operating between -3°C and -20°C (27°F and -4°F). The release liners are removed from the hydrogel adhesive, and the frozen Hybrid ArcticGel Pads are applied to the patient's skin. Between 4 and 8 pads are applied to the patient, depending on the patient's size and patient temperature reduction required. Heat from the patient is conducted through the hydrogel and pad layers into the ice gel compartment. The heat from the skin melts the ice gel. The duration of Ice Gel Mode is approximately 60 to 90 minutes from the time of application to complete melting of the Ice Gel.
Medivance, Inc.
510(k) K120849 Hybid ArcticGel Pad
ﻬﻢ ﺍﻟﻤﺴﺘﻘﺒﺔ ﻭﺍﻟﻤﻘﺎﺑﻠﺔ
Image /page/0/Picture/20 description: The image contains text that appears to be the beginning of a word or phrase. The text says "app" and "Ge". Without more context, it is difficult to determine the meaning of the text. The text is in a simple, sans-serif font.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image contains the logo for Medivance. The logo features a stylized graphic element to the left of the word "Medivance" in a bold, sans-serif font. Below the company name is the tagline "Committed to Restoring Life" in a smaller, italicized font. The overall design is clean and professional.
To convert to Water Flow Mode, pad lines are irreversibly snapped onto the inlet and outlet water ports of the Hybrid ArcticGel Pads. The pad line connectors are connected to the Arctic Sun fluid delivery line. Up to 6 Hybrid ArcticGel Pads may be connected to the Arctic Sun. Temperature controlled water from the Arctic Sun control module can then be circulated through the pad. The water passes through the water flow compartment, allowing heat to be transferred from the patient's skin across the hydrogel adhesive and into the circulating water.
#### Indications for Use
The Hybrid ArcticGe!™ Pad is indicated as a standalone cooling device for the reduction of core temperature of adult patients where clinically indicated (e.g. heat stroke or hyperthermia).
Additionally, the Hybrid ActicGel™ Pad is indicated as an adjunct to patient temperature control with the Arctic Sun® Temperature Management System.
The Hybrid ArcticGel™ Pad is intended for use by health care professionals in all clinical environments, including, but not limited to in- hospital as well as EMS/transport settings.
### Substantial Equivalence
The Arctic Sun Hybrid ArcticGel Pad was shown to be substantially equivalent in intended use, design, technological characteristics, materials and system features and functions to the predicate devices.
The Hybrid ArcticGel Pad and the predicated ArcticGel Pads and EMCOOLS Pads are noninvasive surface temperature management devices. The Hybrid ArcticGel Pads and the EMCOOLS Pads may be pre-frozen and used as standalone devices for use in patient temperature reduction. Additionally, the Hybrid ArcticGel Pads and the predicate ArcticGel Pads may be used in with the Artic Sun Temperature Management System for use in continuous patient temperature monitoring and control.
The Hybrid ArcticGel Pads and the predicate devices are manufactured from polymer materials that are common to many disposable medical devices, and utilize a biocompatible medical adhesive material to adhere the pads to the patient's skin. The Hybrid ArcticGel Pads and the predicate devices are of similar sizes and weights, and provide similar cooling capacities and durations.
#### Testing
Biocompatibility testing performed in accordance with ISO 10993-05 and 10993-10 demonstrated the hydrogel patient adhesive to be non-cytotoxic, non-irritating and nonsensitizing.
Comparative performance testing demonstrated that the cooling and temperature control performance of the Hybrid ArcticGel Pads is substantially equivalent to that of the predicate devices.
Medivance, Inc.
Page 17
{2}------------------------------------------------
ം ഇക്ക
Image /page/2/Picture/1 description: The image shows the logo for Medivance. The logo consists of a stylized swoosh shape followed by the word "Medivance" in a bold, sans-serif font. Below the company name is the tagline "Committed to Restoring Life" in a smaller font size.
Safety tests performed in the porcine skin model demonstrated that the minimum mean skin temperature remains above freezing, reaches its minimum value within the first 15 minutes after pad application, and gradually rises over the course of the 90 minute cooling treatment. The placement of the frozen Hybrid ArcticGel Pads on the pigs' skin of pigs was welltolerated.
#### Conclusions
Based upon the testing and comparison to the predicate devices, the Arctic Sun Hybrid Arctic Gel Pad performs as intended and raises no new safety or effectiveness issues.
Image /page/2/Picture/6 description: The image shows a grainy, black and white picture. The image is mostly white, with black speckles and shapes scattered throughout. The black shapes are concentrated on the left side of the image, forming a vertical column.
Medivance, Inc.
ದ ಸಾವ್ಯ
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -- WO66-G609 Silver Spring, MD 20993-0002
JUL 2002012
Medivance, Inc. c/o Ms. Lynne Aronson Director, Regulatory Affairs 321 South Taylor Avenue, Suite 200 Louisville, CO 80027
Re: K120849
Hybrid ArcticGel Pad Regulation Number: 21 CFR 870.5900 Regulation Name: System, Thermal Regulating Regulatory Class: Class II Product Code: DWJ Dated: May 21, 2012 Received: May 22, 2012
Dear Ms. Aronson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{4}------------------------------------------------
Page 2 - Ms. Lynne Aronson
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
M.S. Hilliam
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# STATEMENT OF INDICATIONS FOR USE (FDA FORM)
Statement of Indications for Use (FDA Form)
510(k):
K120849
Device:
Hybrid ArcticGel Pad
Indications for Use:
The Hybrid ArcticGel™ Pad is indicated as a standalone cooling device for the reduction of core temperature of adult patients where clinically indicated (e.g. heat stroke or hyperthermia).
Additionally, the Hybrid ActicGel™ Pad is indicated as an adjunct to patient temperature control with the Arctic Sun® Temperature Management System.
The Hybrid ArcticGel™ Pad is intended for use by health care professionals in all clinical environments, including, but not limited to in- hospital as well as EMS/transport settings.
Prescription Use X Use (Per 21 CFR 801.109)
> (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ym. Yillebe
Division of Cardiovascular Devices
510(k) Number k120849
(Optional Format 1-2-96)
Over-the-Counter
Medivance, Inc.
510(k) K120849 Hybid ArcticGel Pad
Page 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.