K120799 · Yukon Medical, LLC · FPA · Mar 30, 2012 · General Hospital
Device Facts
Record ID
K120799
Device Name
ARISURE NEUTRAL VALVE
Applicant
Yukon Medical, LLC
Product Code
FPA · General Hospital
Decision Date
Mar 30, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Arisure™ Neutral Valve is a single use, sterile, non-pyrogenic device intended for use as an accessory to intravascular administration sets for the administration or withdrawal of fluids from a patient through a cannula placed in the vein or artery. The Arisure™ Neutral Valve may be used with low-pressure power injectors up to 300 psi.
Device Story
Arisure™ Neutral Valve is a sterile, single-use, luer-activated needle-free connector; used as an accessory to intravascular administration sets. Device consists of polycarbonate housings, retention ring, and pre-slit silicone piston lubricated with silicone oil. Principle of operation: male luer activation deforms the internal pre-slit silicone piston, opening a fluid pathway for administration or withdrawal of fluids; piston returns to original shape upon disconnection. Used in clinical settings by healthcare providers. Device facilitates fluid access to venous or arterial cannulas; supports low-pressure power injection up to 300 psi. Hemocompatibility testing confirms low hemolysis rates during blood withdrawal, comparable to 21GA needles. Benefits include needle-free access and secure fluid management.
Clinical Evidence
Bench testing only. Evaluated microbial ingress, liquid/vacuum leakage, flow rate, tensile strength, and disconnection bolus. Hemocompatibility testing compared blood withdrawal via Arisure™ Neutral Valve versus 21GA needle; both demonstrated hemolytic percentages below 2% (ASTM F 756-00), confirming non-hemolytic properties.
Technological Characteristics
Sterile, single-use, luer-activated needle-free connector. Materials: Polycarbonate (housings, retention ring), silicone rubber (piston), fluorosilicone (lubricant). Principle: Mechanical deformation of pre-slit elastomeric sleeve over internal post. Dimensions/form factor: Standard luer-compatible. Connectivity: Standalone mechanical interface. Sterilization: Not specified. No software/algorithm.
Indications for Use
Indicated for use as an accessory to intravascular administration sets for the administration or withdrawal of fluids from a patient through a cannula placed in the vein or artery. Compatible with low-pressure power injectors up to 300 psi.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
510(K) Notification
Yukon Medical, LLC
Arisure™ Nestral Valve
# 5.0 510(k) Summary
MAR 3 0 2012
ﺮ ":"
-
#### 5.1. Submitter
| Submitter's Name<br>and Address | Todd Korogi, President | |
|---------------------------------|--------------------------------------------------|--|
| | Yukon Medical, LLC | |
| | 4819-400 Emperor Blvd. | |
| | Durham, NC 27703 | |
| | Phone: 919-595-8250 | |
| | Email: tkorogi@yukonmedical.com | |
| Contact Person | Carl Dupper, Director of Engineering and Quality | |
| | Yukon Medical | |
| | 4819-400 Emperor Blvd. | |
| | Durham, NC 27703 | |
| | Phone: 919-595-8257 | |
| | Email: cdupper@yukonmedical.com | |
| Prepared By | James Fentress, Product Development Engineer | |
| | Gilero Biomedical | |
| | 4022-300 Stirrup Creek Dr. | |
| | Durham, NC 27703 | |
| | Phone: 919-595-8236 | |
| | Email: jfentress@gilero.com | |
| Date Prepared | 01 November 2011 | |
#### 5.2. Device Identification
| Trade Name | Arisure™ Neutral Valve |
|----------------|--------------------------------------------------------------------------------------|
| Common Name | Needle-free access device, needle-free connector, luer<br>access device |
| Classification | Intravascular Administration Set<br>Product Code - FPA<br>Regulation 21 CFR 880.5440 |
#### 5.3. Predicate Devices
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
| Device | Manufacturer | 510(k) |
|--------------------------------------------------------------|-------------------|---------|
| B3300<br>MicroCLAVE®<br>Connector; C1000<br>CLAVE® Connector | ICU Medical, Inc. | K970855 |
{1}------------------------------------------------
#### 5.4. Device Description
The Arisure™ Neutral Valve is a sterile, single-use, swab-able, normally closed, luer-activated, valved connector. The Arisure™ Neutral Valve consists of five primary components: lower housing, upper housing, retention ring, pre-slit silicone piston, and lubricating silicone oil. The lower housing, made of polycarbonate, consists of a male luer and a long feature that provides support for the silicone piston to move upon. The pre-slit silicone piston attaches to the lower housing. A series of round protrusions on the inside of the silicone piston provides sliding seals over the long feature on the lower housing. These sliding seals allow fluid to travel through the silicone piston when the piston is accessed with a male luer. The retention ring, made of polycarbonate, fits over the silicone piston and provides extra rigidity. The silicone piston and top housing together comprise the female end of the needle-free valve. The top housing, made of polycarbonate, fits over the top of the piston and is permanently affixed to the bottom housing. When pressed downward by a male luer, the slit in the piston opens, and the sliding seal on the inside of the piston opens a fluid pathway through the device. Silicone oil is used to lubricate the silicone piston. The Arisure™ Neutral Valve does not contain any natural rubber latex.
#### 5.5. Intended Use
The Arisure™ Neutral Valve is a single use, sterile, non-pyrogenic device intended for use as an accessory to intravascular administration sets for the administration or withdrawal of fluids from a patient through a cannula placed in the vein or artery. The Arisure™ Neutral Valve may be used with low-pressure power injectors up to 300 psi.
The indications statement for the Arisure™ Neutral Valve differs from the indications statement of the predicate device (presented in 510(k) K970855) in that it additionally states that the Arisure™ Neutral Valve can be used to withdraw fluid from a patient, rather than just administer fluid. This difference is not critical to the intended therapeutic use of the device, and the difference does not affect the safety and effectiveness of the device when used as indicated.
The primary concern with the withdrawal of fluids from a patient through an cannula placed in the vein or artery of a patient is damage that can occur to the withdrawn fluid sample; damage to blood. In addition to the traditional tests for material hemocompatibility presented in the biocompatibility section (section 15), additional testing was also conducted on the Arisure™ Neutral Valve to determine whether the withdrawal of blood using the Arisure™ Neutral Valve caused damage to the blood. Section 15 details this test, which shows that when compared to blood withdrawn using a 21GA needle, the Arisure™ Neutral Valve has less hemolysis. A general threshold for assessing the hemolytic property of a given material is that hemolytic percentages below 2% are considered nonhemolytic (ASTM F 756-00). Both the 21GA needle and the Arisure™ Neutral Valve had hemolytic percentages below 2%.
{2}------------------------------------------------
In addition to the hemolytic testing conducted on the Arisure™ Neutral Valve, the predicate CLAVE® device is also used for the withdrawal of fluid. ICU Medical Instructions M1-1053 Rev01 illustrates the use of the CLAVE® device for withdrawing blood from a patient.
#### 5.6. Predicate Device Comparison (Technological Characteristics)
A direct comparison of the technical and performance characteristics of the Arisure™ Neutral Valve and the technical and performance characteristics of the predicate device demonstrates equivalency. Minor differences in subject and predicate device characteristics do not introduce any new safety or efficacy concerns.
| Table 5.6: Device Technological Comparison | | |
|---------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device Arisure™ Neutral<br>Valve | Predicate Device MicroCLAVE®<br>B3300 Connector and CLAVE®<br>C1000 Connector<br>(510(k):K970855) | |
| Materials | | |
| Neutral Valve Materials:<br>Lower Housing (Internal Conduit) -<br>Polycarbonate | Materials per 510(k): K970855<br>Internal Conduit- Polycarbonate | |
| Pre-slit Silicone Piston: silicone rubber | Silicone Seal[Silicone Piston] - Silicone<br>Rubber | |
| Lubricant - Fluorosilicone | Lubricant - Fluorosilicone | |
| Upper Housing - Polycarbonate | [upper]Housing - Polyester | |
| Retention Ring- Polycarbonate | | |
| Principles of<br>Operation | The female Luer valve opens to permit the<br>introduction or withdrawal of fluids when<br>accessed by a male Luer tip. This access<br>deforms a compressible element that returns<br>to its original shape through the mechanical<br>properties of the deformable element. | The female Luer valve opens to permit the<br>introduction or withdrawal of fluids when<br>accessed by a male Luer tip. This access<br>deforms a compressible element that returns<br>to its original shape through the mechanical<br>properties of the deformable element. |
| Technology<br>and Design | When activated by a male Luer, a pre-slit<br>elastomeric sleeve advances over an internal<br>post, opening a fluid pathway that connects<br>the female and male ends of the device. | When activated by a male Luer, a pre-slit<br>elastomeric sleeve advances over an internal<br>post, opening a fluid pathway that connects<br>the female and male ends of the device. |
{3}------------------------------------------------
#### 5.7. Discussion of Non-clinical tests
The following non-clinical tests have been done in support of the substantial equivalence of the Arisure™ Neutral Valve to the Predicate CLAVE® devices:
- Microbial Ingress Testing .
- Biocompatibility Testing .
- Liquid Leak (open position) .
- Liquid Leak (closed position) .
- Vacuum Leak (open position) .
- Vacuum Leak (closed position) .
- Flow Rate .
- Tensile Strength .
- Disconnection Bolus .
Results from non-clinical testing indicate that the Arisure™ Neutral valve is as safe and effective as the predicate device. Testing conducted that reference the design verification and validation testing of the Arisure™ Neutral valve meet pre-determined acceptance criteria for the device. Testing conducted as a direct comparison of the Arisure™ Neutral valve and predicate device demonstrate that the equivalence of the devices.
#### 5.8. Conclusion
The Arisure™ Neutral Valve is equivalent to the predicate device and is safe and effective for its intended use.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002
Yukon Medical, LLC C/O Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25TH Street N.W. Buffalo, Minnesota 55313
MAR 3 0 2012
Re: K120799
Trade/Device Name: Arisure™ Neutral Valve Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set. Regulatory Class: II Product Code: FPA Dated: March 15, 2012 Received: March 16, 2012
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2 - Mr. Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony Dimon
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# 4.0 Indications for Use Statement
510(k) Number (if known): K120799
Device Name: Arisure™ Neutral Valve
Indications for Use:
The Arisure™ Neutral Valve is a single use, sterile, non-pyrogenic device intended for use as an accessory to intravascular administration sets for the administration or withdrawal of fluids from a patient through a cannula placed in the vein or artery. The Arisure™ Neutral Valve may be used with low-pressure power injectors up to 300 psi.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Ruld C. Chign 3/29/2012
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
120799 510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.