Peer-reviewed scientific dental journals; Prospective multicenter clinical study data; Histological and histomorphometric human clinical reports
The sponsor provided a summary of published clinical literature and a prospective clinical report to demonstrate the safety and effectiveness of the TiXos dental implants by evaluating human bone response and clinical performance in routine clinical settings.
Clinical literature; Prospective multicenter study; Histomorphometric study; Human bone response
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Prospective Clinical Evaluation of 201 Direct Laser Metal Forming Implants: Results From a 1 Year Multicenter Study; Prospective Multicenter Study; Follow-up/Duration: 1 year; Study Period: Published 2011
Patients receiving Direct Laser Metal Forming (DLMF) implants; Sample Size: 201 implants; Number of Sites: Multicenter
Not applicable for this study
Clinical performance/safety
Single-Tooth Direct Laser Metal Forming (Dlmf) Titanium Implants: Results From A 1- Year Prospective Multicenter Study; Prospective Multicenter Study; Follow-up/Duration: 1 year
Patients receiving single-tooth DLMF implants; Number of Sites: Multicenter
Not applicable for this study
Clinical performance
Indications for Use
TiXos dental implants are permanent devices for single use only that are intended to be surgically placed in the bone of the mandibular and/or maxillary dental arches of the patient in order to restore their original dental function and aesthetic features by providing support for fixed and/or removal prosthesis. TiXos dental implants can support single tooth, multiple tooth, or total prosthesis restorations and are compatible with the following dental abutments manufactured by Leader Italia that were previously cleared under K062931.
Device Story
TiXos dental implants are permanent, single-use titanium endosseous implants; fabricated via Direct Laser Metal Forming (DLMF) using titanium powder; resulting in porous surface with open intercommunicating cavities. Implants are surgically placed in mandibular or maxillary bone by dental professionals to support fixed or removable prostheses. Device integrates with Leader Italia prosthetic abutments (cleared under K062931). Gamma sterilized. Clinical benefit includes restoration of dental/masticatory function and aesthetics.
Clinical Evidence
Evidence includes a prospective summary report and radiographic images from a European clinical study, plus multiple peer-reviewed publications. Studies evaluated bone response, histomorphometry, and 1-year outcomes for implants loaded with prosthetic abutments. Data supports safety and effectiveness for osseointegration and functional restoration.
Technological Characteristics
Material: Ti-6Al-V4. Fabrication: Direct Laser Metal Forming (DLMF). Surface: Porous with intercommunicating cavities. Sterilization: Gamma. Configurations: Cylindrical and S-Type. Connection: Internal and External hex. Standards: ISO 14801 (fatigue), ISO 10993-5 (cytotoxicity).
Indications for Use
Indicated for patients requiring dental restoration via single, multiple, or total tooth prosthesis supported by endosseous implants placed in mandibular or maxillary arches.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "LEADER" spelled out in individual boxes. Each letter is in a separate box, and the boxes are arranged in a row. The word is written in a simple, sans-serif font. The background of the boxes is white, and the letters are black.
# 5. 510(K) Summary
#### Submission Correspondent
Company Name:
Company Address:
Company Contact:
Country:
Phone:
Fax:
Submission Date:
Website:
Submission Sponsor
Company Name:
Company Address:
Country: Phone:
Fax:
:
Website:
Image /page/0/Picture/17 description: The image shows the logo for Emergo Group, which is displayed above a world map. The text "EMERGO GROUP" is in all caps. Below the logo and map is the text "K120797".
5 2012 SEP
Emergo Group, Inc.
611 West 5th Street Third Floor Austin, TX 78701
Stuart R. Goldman Senior Consultant project.management@emergogroup.com
USA
512.327.9997
512.327.9998
March 8, 2012
www.emergogroup.com
Leader Italia, Srl
Via Aquileja 43/B Ciniseilo Balsamo (MI) 20092
Italy
39 02 618651
39 02 61290676
www.leaderitalia.it
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "LEADER" spelled out in individual boxes. Each letter is separated by a vertical line. The letters are in a sans-serif font and are all capitalized. The background of the boxes is white, and the letters are black.
Image /page/1/Picture/1 description: The image shows the logo for Emergo Group. The logo consists of the words "EMERGO GROUP" in a sans-serif font, with a stylized globe symbol in between the two words. A faded world map is in the background.
# Device Classification
| Device Sponsor: | Leader Italia, Srl |
|------------------------------|----------------------------------------|
| Device Trade Name: | TiXos Implant System (dental implants) |
| Product Classification Name: | Implant, Endosseous, Root-Form |
| Product Code: | DZE |
| Regulation Number: | 872.3640 |
| Classification Panel: | Dental Devices |
| Regulatory Class: | Class 2 |
#### Predicate Device
IMPLUS Implant System (K062931) manufactured by Leader Italia, Srl
#### Indications for Use
TiXos dental implants are permanent devices for single use only that are intended to be surgically placed in the bone of the mandibular and/or maxillary dental arches of the patient in order to restore their original dental function and aesthetic features by providing support for fixed and/or removal prosthesis. TiXos dental implants can support single tooth, multiple tooth, or total prosthesis restorations and are compatible with the following dental abutments manufactured by Leader Italia that were previously cleared under K062931.
| ABUTMENT CODE | ABUTMENT NAME |
|------------------------------------------|------------------------------------------------------|
| Tixos Internal Hex Ø 3.3 | Platform Size (mm): 3.5 |
| 01ATI33 | Straight Abutment Ø 4.5 |
| 01ATIF33 | Frictioning Abutment Ø 4.5 |
| 01AS33 | Shoulder Abutment Ø 4.5 |
| 01ASF33 | Shoulder Frictioning Abutment Ø 4.5 |
| 01AP1533 | 15° Angled Abutment Ø 4.5 |
| 01AP601533 | 15° Angled Abutment Ø 6 |
| 01AP15F33 | 15° Angled Frictioning Abutment Ø 4.5 |
| 01AP2533 | 25° Angled Abutment Ø 4.5 |
| 01AP602533 | 25° Angled Abutment Ø 6 |
| 01AP25F33 | 25° Angled Frictioning Abutment Ø 4.5 |
| 01AB332, 01AB333<br>01AB334, 01AB335 | Ball Abutment Ø 4.5 |
| 01ABA332, 01ABA333<br>01ABA334, 01ABA335 | Ball Abutment with Antirotation Base Ø 4.5 |
| 01ATS331, 01ATS332 | Tissue Abutment Ø 4.5 for Screw-<br>Retained Denture |
| 01AATI331A,<br>01AATI332A | Antirotation Abutment |
| Tixos Internal Hex Ø 3.75 | Platform Size (mm): 4 |
| 01ATIC4 | Straight Abutment Ø 4.5 |
| 01ATIC6 | Straight Abutment Ø 6 |
| ABUTMENT CODE | ABUTMENT NAME |
| Tixos Internal Hex Ø 4.5 | Platform Size (mm): 5 |
| 01ATI5 | Straight Abutment Ø 5 |
| 01ATIF5 | Frictioning Abutment Ø 5 |
| 01AP1560 | 15° Angled Abutment Ø 6 |
| 01AB502, 01AB503<br>01AB504 | Ball Abutment Ø 5 |
| 01AATI451A | Antirotation Abutment |
| 01ATI5 | Straight Abutment Ø 5 |
| Tixos Internal Hex Ø 5.5 | Platform Size (mm): 6 |
| 01AB602, 01AB603<br>01AB604 | Ball Abutment Ø 6 |
| 01ATI6055 | Straight Abutment Ø 6 |
| 01ATI60F55 | Frictioning Abutment Ø 6 |
| 01AATI551A | Antirotation Abutment |
| Tixos External Hex Ø 3.3/3.75 | Platform Size (mm): 4.1 |
| 01ATIX33 | Straight Abutment Ø 4.5 |
| 01ATI60X33 | Straight Abutment Ø 6 |
| 01ATIXF33 | Straight Frictioning Abutment Ø 4.5 |
| 01AXS33 | Shoulder Abutment Ø 4.5 |
| 01AXS6033 | Shoulder Abutment Ø 6 |
| 01AXSF33 | Shoulder Frictioning Abutment Ø 4.5 |
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "LEADER" spelled out using individual letter tiles. Each letter is printed in black ink on a white tile. Vertical black lines separate each letter, creating a distinct visual break between them.
| 01ATICF4 | Frictioning Abutment Ø 4.5 |
|-----------------------------|---------------------------------------|
| 01ASC4 | Shoulder Abutment Ø 4.5 |
| 01ASC6 | Shoulder Abutment Ø 6 |
| 01ASCF4 | Shoulder Frictioning Abutment Ø 4.5 |
| 01APC1540 | 15° Angled Abutment Ø 4.5 |
| 01APC601540 | 15° Angled Abutment Ø 6 |
| 01APC2540 | 25° Angled Abutment Ø 4.5 |
| 01APC602540 | 25° Angled Abutment Ø 6 |
| 01APC15F40 | 15° Angled Frictioning Abutment Ø 4.5 |
| 01APC25F40 | 25° Angled Frictioning Abutment Ø 4.5 |
| 01ATIM4 | Multi-function Abutment |
| 01ABC452, 01ABC453 | Ball Abutment Ø 4.5 |
| 01ABC454, 01ABC455 | Ball Abutment with Antirotation Base |
| 01ABA402, 01ABA403 | |
| 01ABA404, 01ABA405 | Ø 4.5 |
| 01AC1A, 01AC2A | Antirotation Abutment |
| | |
| 01APX1533 | 15° Angled Abutment Ø 4.5 |
| 01APX2533 | 25° Angled Abutment Ø 4.5 |
| 01ABX332, 01ABX333 | Ball Abutment Ø 4.5 |
| 01ABX334, 01ABX335 | |
| 01ABXA2, 01ABXA3 | Ball Abutment with Antirotation Base |
| 01ABXA4, 01ABXA5 | |
| 01AATIX331A,<br>01AATIX332A | Antirotation Abutment |
| Tixos External Hex Ø 5 | Platform Size (mm): 4.1 |
| 01AATIX501A | Antirotation Abutment |
| 01ATIX5 | Straight Abutment Ø 5 |
| 01ATIXF5 | Straight Frictioning Abutment Ø 5 |
# Device Description
Leader Italia's line of TiXos dental implants are titanium endosseous implants that are fabricated by a Eeader frails 3 line of 17xos detal Forming ("DLMF"). This fabrication process results in the production of process cance Direct asser wisee characterized by open intercommunicating cavities and networks. The implant with a porous surface only placed in the bone of the mandibular and/or maxillary dental implants are intended to be interest to be their fixed and/or removal prosthesis, and to restore their arches of the putient to provental and/or masticatory function. TiXos dental implants are permanent original devices for single use and can support single tooth, multiple tooth, or total prosthesis dental devices for single ass and our on on on on endexclusively with Leader Italia's line of restorations: These implants and surgical instruments. TiXos dental implants are gamma sterilized, have a five prostheticulabatinents and submit wo basic model configurations called Cylindrical and S-Type, and a number of different sizes and diameters as shown below.
| TiXos Implants | |
|----------------------------|--------------------------------------|
| | Models & Sizes |
| TiXos Cylindrical Line | |
| Internal Connection | Ø 3.3 mm / L 8, 10, 11.5, 13, 16 mm |
| | Ø 3.75 mm / L 8, 10, 11.5, 13, 16 mm |
| | Ø 4.5 mm / L 8, 10, 11.5, 13, 16 mm |
| | Ø 5.5 mm / L 8, 10, 11.5, 13 mm |
| External Connection - Long | Ø 3.3 mm / L 10, 11.5, 13, 16 mm |
| | Ø 3.75 mm / L 8, 10, 11.5, 13, 16 mm |
| | Ø 5.0 mm / L 8, 10, 11.5, 13 mm |
| TiXos S-Type Line | |
| S-Type | Ø 3.75 mm / L 10, 11.5, 13 mm |
| | Ø 4.5 mm / L 10, 11.5, 13 mm |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "LEADER" spelled out using individual letters. Each letter is printed in black ink on a white rectangular card. Vertical black lines separate each letter, creating distinct boxes for each character.
Image /page/3/Picture/1 description: The image shows the logo for Emergo Group. The logo consists of the words "EMERGO" and "GROUP" in a sans-serif font, with a stylized globe symbol in between. A faded world map is in the background.
### Predicate Device Comparison
Aside from the slight differences in sizes (i.e., diameter and length) between TiXos and the predicate implants, TiXos implants are manufactured using Leader Italia's Direct Laser Metal Forming process as opposed to the more conventional machining and acid etching techniques the company uses to produce its IMPLUS implants.
Both TiXos and IMPLUS implants are single use, permanent dental devices that are surgically placed in either the mandibular and/or maxillary dental arches of the patient; both implants are cleaned, enckaged and gamma sterilized in identical manners; use the same abutments that have been designed and manufactured by Leader Italia, and have identical implant-to-abutment interface connections; and can be implanted using the same surgical instruments.
Some select properties and characteristics of the subject and predicate devices are compared side-byside as shown below.
| Substantial Equivalence Comparison<br>Selected Properties Of IMPLUS vs. TiXos Dental Implants | | |
|-----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | Leader Italia<br>IMPLUS Implants | Leader Italia<br>TiXos Implants |
| 510(k): | K062931 | Pending |
| Product Code: | DZE | DZE |
| Regulation Number: | 872.3640 | 872.3640 |
| Regulation Name: | Implant, Endosseous, Root-Form | Implant, Endosseous, Root-Form |
| Implant Material: | Ti grade 4 | Ti-6Al-V4 |
| Raw Material Form: | Rod Stock | Powder |
| Implant Fabrication Method: | Traditional CNC Machining | Direct Laser Metal Forming |
| Surface Formed By: | Acid Etching | Direct Laser Metal Forming |
| Implant Geometries: | Various | Various |
| Designed For Prosthetic<br>Abutments: | Yes | Yes |
| Implant/abutment Interface: | Internal and External | Internal, External, and<br>Integrated Abutment |
| Implant Diameters (mm): | 3.3, 3.75, 4.5, 4.75, 5.0<br>5.5, 5.75, 6 | 3.3, 3.75,<br>4.5, 5.0, 5.5 |
| Implant Lengths (mm): | 8, 10, 11.5, 13, 16, 20 | 8, 10, 11.5, 13, 16 |
| Single Use: | Yes | Yes |
| Sterile: | Yes | Yes |
| Intended Use: | Titanium endosseous implants to<br>be surgically placed in the bone<br>of the mandibular and/or<br>maxillary dental arches in order<br>to provide support for fixed | Titanium endosseous implants to<br>be surgically placed in the bone<br>of the mandibular and/or<br>maxillary dental arches in order<br>to provide support for fixed |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "LEADER" spelled out in individual boxes. Each letter is separated by a vertical line. The letters are all capitalized and in a sans-serif font. The background of the boxes is white.
Image /page/4/Picture/1 description: The image shows the logo for Emergo Group. The logo consists of the words "EMERGO" and "GROUP" in a sans-serif font, with a stylized symbol in between. The symbol appears to be a stylized "E" within a circle. A faded world map is in the background.
| | and/or removal prosthesis to<br>restore the original aesthetic<br>features, dental and/or<br>masticatory function to the<br>patient after osseointegration<br>has taken place. | and/or removal prosthesis to<br>restore the original aesthetic<br>features, dental and/or<br>masticatory function to the<br>patient after osseointegration<br>has taken place. |
|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
## Non-Clinical Data - Bench Testing & Device Validations
As part of demonstrating the safety and effectiveness of TiXos dental implants and in showing substantial equivalence to its IMPLUS implants, Leader Italia submitted a number of its dental implants for fatigue testing in accordance with ISO 14801, Dentistry - Implants - Dynamic Fatigue Test for Endosseous Dental Implants, where their implants were tested in dry (i.e., air) environment. Testing was performed on the TiXos dental implants to simulate worst-case loading conditions and was compared to the predicate device and found it be similar.
Further, TiXos dental implants also underwent extensive validation activities for biocompatibility, cleaning, packaging, shelf-life and sterilization in accordance with Guidance for Industry and FDA Staff -Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments, and all applicable recognized FDA consensus standards for dental implants, including but not limited to, ISO 10993-5, Biological Evaluation of Medical Devices - Part 5: Test for in vitro Cytotoxicity.
# Clinical Data
As part of demonstrating the safety and effectiveness of its TiXos dental implants, Leader Italia submitted a prospective summary report, and radiographic images, to the FDA that were part of its European clinical study and represent follow-up data for their implants at different time intervals after the implants were loaded with their prosthetic abutments. Results of this clinical information have been published in a number of peer reviewed scientific dental journals as shown below.
| Tixos Implants | |
|-------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Summary of In-Vivo Clinical Articles | |
| Title | Publication |
| Early Human Bone Response to Laser Metal Sintering Surface Topography:<br>A Histological Report | Journal of Oral Implantology<br>Vol. XXXVI/No. Two, 2010 |
| Influence of Direct Laser Fabrication Implant Topography on Type IV Bone:<br>A Histomorphometric Study In Humans | Journal of Biomedical Materials<br>Research, Part A, 9 July, 2009 |
| SEM and X-Ray Dispersive Spectrometry Evaluation of Direct Laser Metal<br>Sintering Surface and Human Bone Interface: A Case Series | Lasers In Medical Science, 2011 |
| Single-Tooth Direct Laser Metal Forming (Dlmf) Titanium Implants: Results<br>From A 1- Year Prospective Multicenter Study | Lasers in Medical Science |
| Report on a Screw on Sintered Titanium, Removed After Two Months of<br>Osseointegration | The University of Insubria, Varese,<br>Italy - Department of Human<br>Morphology, 2008 |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "LEADER" spelled out using individual letter tiles. Each letter is printed in black ink on a white tile. Vertical black lines separate each letter, creating a distinct visual break between them. The tiles are arranged horizontally, forming the complete word.
Image /page/5/Picture/1 description: The image shows the logo for Emergo Group. The logo consists of the company name in capital letters, with a stylized globe symbol in the center. The logo is set against a faded map of the world.
| Prospective Clinical Evaluation of 201 Direct Laser Metal Forming Implants:<br>Results From a 1 Year Multicenter Study | Lasers in Medical Science, accepted<br>on February 2011 |
|---------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| Dental implants from laser fusion of titanium microparticles: from research<br>to clinical applications | Journal of Osseointegration |
| Immediate Loading of Mandibular Overdentures Supported by Unsplinted<br>Direct Laser Metal-Forming Implants: Results From a 1-Year Prospective<br>Study | Journal of Periodontol |
#### Substantially Equivalence
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device, or has the same intended use and different technological characteristics, but it can be demonstrated that the new device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness compared to the predicate device.
It has been shown in this 510(k) submission that the differences between the TiXos and IMPLUS dental implants do not raise any questions regarding its safety and effectiveness. Leader Italia's TiXos Implant System, as designed and manufactured, is therefore determined to be substantially equivalent to their IMPLUS Implant System previously cleared under K062931.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle, emphasizing the department's name and national affiliation.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Leader Italia S.R.L. C/O Mr. Stuart R. Goldman Senior Consultant Emergo Group, Incorporated 611 West 5TH Street, Third Floor Austin, Texas 78701
Re: K120792
Trade/Device Name: TiXos Implant System (Dental Implants) Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: August 1, 2012 Received: August 3, 2012
Dear Mr. Goldman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
SEP 5 2012
{7}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
For
Anthony D. Watson, B.S., M.S., M.B.A. · Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications For Use
510(k) Number (if known): 120792
TiXos dental implants are permanent devices for single use only that are intended to be surgically placed in the bone of the mandibular and/or maxillary dental arches of the patient in order to restore their m inc bental function and aesthetic features by providing support for fixed and/or removal prosthesis. onginal dental implants can support single tooth, multiple tooth, or total prosthesis restorations and are r noo deliter with the following dental abutments manufactured by Leader Italia that were previously cleared under K062931.
| ABUTMENT CODE | ABUTMENT NAME | ABUTMENT CODE | ABUTMENT NAME |
|-----------------------------------------------------|------------------------------------------------------|----------------------------------------------------------|--------------------------------------|
| Tixos Internal Hex Ø 3.3<br>Platform Size (mm): 3.5 | | Tixos Internal Hex Ø 4.5<br>Platform Size (mm): 5 | |
| 01ATI33 | Straight Abutment Ø 4.5 | 01ATIS | Straight Abutment Ø 5 |
| 01ATIF33 | Frictioning Abutment Ø 4.5 | 01ATIFS | Frictioning Abutment Ø 5 |
| 01AS33 | Shoulder Abutment Ø 4.5 | 01AP1560 | 15° Angled Abutment Ø 6 |
| 01ASF33 | Shoulder Frictioning Abutment Ø 4.5 | 01AB502, 01AB503<br>01AB504 | Ball Abutment Ø 5 |
| 01AP1533 | 15° Angled Abutment Ø 4.5 | 01AATI451A | Antirotation Abutment |
| 01AP601533 | 15° Angled Abutment Ø 6 | 01ATIS | Straight Abutment Ø 5 |
| 01AP15F33 | 15° Angled Frictioning Abutment Ø 4.5 | Tixos Internal Hex Ø 5.5<br>Platform Size (mm): 6 | |
| 01AP2533 | 25° Angled Abutment Ø 4.5 | 01AB602, 01AB603<br>01AB604 | Ball Abutment Ø 6 |
| 01AP602533 | 25° Angled Abutment Ø 6 | | |
| 01AP25F33 | 25° Angled Frictioning Abutment Ø 4.5 | 01ATI6055 | Straight Abutment Ø 6 |
| 01AB332, 01AB333<br>01AB334, 01AB335 | Ball Abutment Ø 4.5 | 01ATI60F55 | Frictioning Abutment Ø 6 |
| 01ABA332, 01ABA333<br>01ABA334, 01ABA335 | Ball Abutment with Antirotation Base<br>Ø 4.5 | 01AATI551A | Antirotation Abutment |
| 01ATS331, 01ATS332 | Tissue Abutment Ø 4.5 for Screw-<br>Retained Denture | Tixos External Hex Ø 3.3/3.75<br>Platform Size (mm): 4.1 | |
| 01AATI331A,<br>01AATI332A | Antirotation Abutment | 01ATIX33 | Straight Abutment Ø 4.5 |
| Tixos Internal Hex Ø 3.75<br>Platform Size (mm): 4 | | 01ATI60X33 | Straight Abutment Ø 6 |
| 01ATIC4 | Straight Abutment Ø 4.5 | 01ATIXF33 | Straight Frictioning Abutment Ø 4.5 |
| 01ATIC6 | Straight Abutment Ø 6 | 01AXS33 | Shoulder Abutment Ø 4.5 |
| 01ATICF4 | Frictioning Abutment Ø 4.5 | 01AXS6033 | Shoulder Abutment Ø 6 |
| 01ASC4 | Shoulder Abutment Ø 4.5 | 01AXSF33 | Shoulder Frictioning Abutment Ø 4.5 |
| 01ASC6 | Shoulder Abutment Ø 6 | 01APX1533 | 15° Angled Abutment Ø 4.5 |
| 01ASCF4 | Shoulder Frictioning Abutment Ø 4.5 | 01APX2533 | 25° Angled Abutment Ø 4.5 |
| 01APC1540 | 15° Angled Abutment Ø 4.5 | 01ABX332, 01ABX333<br>01ABX334, 01ABX335 | Ball Abutment Ø 4.5 |
| 01APC601540 | 15° Angled Abutment Ø 6 | 01ABXA2, 01ABXA3<br>01ABXA4, 01ABXA5 | Ball Abutment with Antirotation Base |
| 01APC2540 | 25° Angled Abutment Ø 4.5 | 01AATIX331A,<br>01AATIX332A | Antirotation Abutment |
| 01APC602540 | 25° Angled Abutment Ø 6 | Tixos External Hex Ø 5 | Platform Size (mm): 4.1 |
| 01APC15F40 | 15° Angled Frictioning Abutment Ø 4.5 | 01AATIX501A | Antirotation Abutment |
| 01APC25F40 | 25° Angled Frictioning Abutment Ø 4.5 | 01ATIX5 | Straight Abutment Ø 5 |
| 01ATIM4 | Multi-function Abutment | 01ATIXF5 | Straight Frictioning Abutment Ø 5 |
| 01ABC452, 01ABC453<br>01ABC454, 01ABC455 | Ball Abutment Ø 4.5 | | |
| 01ABA402, 01ABA403<br>01ABA404, 01ABA405 | Ball Abutment with Antirotation Base<br>Ø 4.5 | | |
Suer Reiser
gn-Om)
Anesthesiology, General Hospital
510(k) Number:
{9}------------------------------------------------
Over-The-Counter Use _
·
(21 CFR 801 Subpart C)
・
#### Prescription Use 区
AND/OR
(Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
.
.
.
.
.
. .
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.