CELLO BALLOON GUIDE CATHETER (6F), 950 MM AND 1020 MM LENGTH
K120781 · Fuji Systems Corp. · DQY · Aug 9, 2012 · Cardiovascular
Device Facts
Record ID
K120781
Device Name
CELLO BALLOON GUIDE CATHETER (6F), 950 MM AND 1020 MM LENGTH
Applicant
Fuji Systems Corp.
Product Code
DQY · Cardiovascular
Decision Date
Aug 9, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The CELLO Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of intravascular catheters into a selected blood vessel in the peripheral and neuro vasculature systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures.
Device Story
CELLO Balloon Guide Catheter is a coaxial-lumen, braid-reinforced, variable stiffness catheter; features distal silicone rubber balloon and radiopaque markers for visualization. Used in hospitals by trained personnel for peripheral and neurovascular procedures. Device facilitates insertion/guidance of intravascular catheters; balloon provides temporary vascular occlusion during angiography. Supplied with dilator. Operates via manual manipulation by physician; provides mechanical support and occlusion. Benefits include improved access to tortuous vasculature and controlled vascular occlusion during diagnostic or interventional procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by nonclinical bench testing including surface inspection, force at break, leakage, tip configuration, hub compliance (ISO 594-1/2), balloon burst pressure, fatigue, and biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemolysis, complement activation, and in-vivo thrombogenicity).
Technological Characteristics
Coaxial-lumen, braid-reinforced catheter; materials: polyurethane, polyamide, stainless steel, PFA, silicone rubber balloon. Dimensions: 6F-9F. Radiopaque markers. Sterilization: Ethylene oxide. Hubs comply with ISO 594-1/2. No software or electronic components.
Indications for Use
Indicated for patients requiring intravascular catheter insertion and guidance in peripheral and neurovascular systems, and temporary vascular occlusion during angiographic procedures.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
# 510(k) Summary, K120781
## Manufacturer
Fuji Systems Corporation 23-14, Hongo 3-Chome, Bunkyo, Tokyo 113-0033 Japan Phone: +81-5489-1915
## Contact
Yoshihiko Semba General Manager, Director International Division
## Date of Preparation
July 25, 2012
## Product Names
AUG - 9 2012
| Generic Name | Classification | Product<br>Code | Trade Name | Model No. |
|--------------------------|--------------------------------|-----------------|------------------------------------|---------------|
| | 21 CFR<br>870.1250<br>Class II | DQY | CELLO<br>Balloon Guide<br>Catheter | 1610560 (6 F) |
| | | | | 1610561 (6 F) |
| Percutaneous<br>Catheter | | | | 1610570 (7 F) |
| | | | | 1610571 (7 F) |
| | | | | 1610580 (8 F) |
| | | | | 1610590 (9 F) |
# Predicate Device
Concentric Balloon Guide Catheter, K021899
{1}------------------------------------------------
#### Device Description
The CELLO Balloon Guide Catheter is a coaxial-lumen, braid-reinforced, variable stiffness catheter with two radiopaque markers on both the distal and proximal ends of the balloon and a bifurcated luer hub on the proximal end. A compliant silicone balloon is mounted on the distal end. Balloon Guide Catheter dimensions and recommended inflation volumes are indicated on the product label. Each catheter is supplied with an appropriately-size Dilator.
The CELLO Balloon Guide Catheters are used in hospitals or other health care facilities which are equipped with trained personnel and specialized equipment to perform peripheral and/or neurovascular procedures.
The CELLO Balloon Guide Catheter is used for facilitating the insertion and guidance of intravascular catheters into selected blood vessels in the peripheral and neurovasculature. The tip of the catheter features a balloon of silicone rubber. Radiopaque markers identify the proximal and distal ends of the balloon. The shaft is a dual lumen type with coaxial structure featuring a built-in braided stainless steel coil.
The materials of construction resemble those used in many other similar catheters. The shaft is made of polyurethane, polyamide, stainless steel and PFA; the balloon is silicone. Patient contact is of limited duration, less than 24 hrs.
#### Intended Use
The CELLO Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of intravascular catheters into a selected blood vessel in the peripheral and neurovasculature systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures.
{2}------------------------------------------------
K120781
P. 3 of 6
. ..
| Device | Proposed | Predicate | Consideration |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Device Name | CELLO Balloon Guide<br>Catheter | Concentric Balloon Guide<br>Catheter | (N/A) |
| Device<br>Description * | The CELLO Balloon Guide<br>Catheter is a coaxial-lumen,<br>braid-reinforced, variable<br>stiffness catheter with two<br>radiopaque markers on<br>both the distal and proximal<br>ends of the balloon and a<br>bifurcated luer hub on the<br>proximal end. A compliant<br>silicone balloon is mounted<br>on the distal end. Balloon<br>Guide Catheter dimensions<br>and the recommended<br>balloon inflation volumes<br>are indicated on product<br>label.<br><br>A dilator is provided with<br>each catheter. | Concentric and Merci<br>Balloon Guide catheters are<br>coaxial lumen, braid-<br>reinforced, variable stiffness<br>catheters with a radiopaque<br>marker on the distal end<br>and a bifurcated luer hub<br>on the proximal end. A<br>compliant balloon is flush<br>mounted on the distal end.<br>Balloon Guide catheter<br>dimensions and maximum<br>recommended balloon<br>inflation volume are<br>indicated on product label.<br><br>If indicated on product<br>label, a dilator is provided. | Similar<br>technological<br>characteristics |
| Size | 6F, 7F, 8F, 9F | 8F, 9F | 8F and 9F are<br>the same |
| Effective length | 920 mm to 1020 mm | 800 to 950 | Overlap from<br>920 to 950 mm |
| Material/Shaft | Polyurethane, Polyamide,<br>Stainless steel, PFA | Polyurethane, Stainless<br>steel | Similar<br>materials |
| Material/Balloon | Silicone rubber | Silicone rubber | Same |
| Sterilization | Ethylene oxide | Ethylene oxide | Same |
Comparison of Technological Characteristics, compared to the Predicate
*Verbatim from each device's IFU.
-
'
{3}------------------------------------------------
# Nonclinical Tests, CELLO Balloon Guide Catheters
| Test | Results | Conclusion |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Surface | Free from extraneous matter, process and surface defects. | Pass |
| Force at Break | All joints meet reliability/confidence requirements in statistical confidence limits test | Pass |
| Freedom from Leakage | Shall not leak liquid when inflated.<br>Air shall not leak into the hub during aspiration. | Pass |
| Tip Configuration | Distal tip smooth, rounded, tapered or similarly finished. | Pass |
| Hubs | Comply with ISO 594-1 and ISO 594-2 | Pass |
| Freedom from Leakage and Damage on Inflation | No leakage or evidence of damage, such as herniation or bursting of the shaft or balloon. | Pass |
| Dimensional Verification | All dimensions meet reliability/confidence requirements in statistical confidence limits test | Pass |
| Balloon Preparation, Deployment and Retraction | Catheters can be advanced to intended sites within a realistic tortuous path model where other devices can be deployed distally and retracted; all devices could be retracted without damage. | Pass |
| Balloon Rated Burst Pressure (RBP) | All balloons meet reliability/confidence requirements in statistical confidence limits test | Pass |
| Balloon Fatigue | All balloons withstand 20 cycles of inflation | Pass |
| Balloon Compliance | All balloons have predictable change in size with pressure | FIO* |
| Balloon Inflation-Deflation Time | All balloons inflate and deflate predictably | FIO* |
| Flexibility and Kink | All catheters meet reliability/confidence requirements in statistical confidence limits test | Pass |
| Torque Strength | All models can be torqued at least 180° when the distal tip is fixed in a tortuous path model | FIO* |
| Radiopacity | Radiopaque markers are visible during angiography under a variety of conditions | FIO* |
·
* For Information Only
{4}------------------------------------------------
# Nonclinical Tests, Dilator
ﺮ
| Test | Results | Conclusion |
|-------------------------------------------------|----------------------------------------------------------------------------------------------|------------|
| Surface | Free from extraneous matter, process and surface<br>defects. | Pass |
| Force at Break | All joints meet reliability/confidence requirements<br>in statistical confidence limits test | Pass |
| Freedom from Leakage | Shall not leak liquid when inflated.<br>Air shall not leak into the hub during aspiration. | Pass |
| Tip Configuration | Distal tip smooth, rounded, tapered or similarly<br>finished. | Pass |
| Hubs | Comply with ISO 594-1 and ISO 594-2 | Pass |
| Freedom from Leakage and<br>Damage on Inflation | No leakage or evidence of damage, such as<br>herniation or bursting of the shaft or balloon. | Pass |
# Biocompatibility, CELLO Balloon Guide Catheters
.
| Test | Results | Conclusion |
|-----------------------------------------------------|-------------------------------|-------------------------------------------|
| L29 MEM Elution Test-ISO | Non-Cytotoxic | Pass |
| Kligman Maximization Test | No Sensitization | Pass |
| Intracutaneous Injection Test | Non-Irritant | Pass |
| Systemic Injection Test-ISO | No acute toxicity | Pass |
| Rabbit Pyrogen Test (Material<br>Mediated) - ISO | Non-pyrogenic | Pass |
| Hemolysis- Direct Contact | Non-hemolytic | Pass |
| Hemolysis-Rabbit Blood-ASTM<br>Indirect Contact | Non-hemolytic | Pass |
| Complement Activation-Direct<br>Contact | Non-Complement<br>Activation | Pass |
| Complement Activation Assay-ISO<br>Indirect Contact | Non-Complement<br>Activation | Pass |
| In-vivo Thrombogenicity | Patent vessels<br>No thrombus | Better thrombo-resistance<br>than control |
{5}------------------------------------------------
### Biocompatibility, Dilator
| Test | Results | Conclusion |
|-----------------------------------------------------|-------------------------------|-------------------------------------------|
| L29 MEM Elution Test-ISO | Non-Cytotoxic | Pass |
| Kligman Maximization Test | No Sensitization | Pass |
| Intracutaneous Injection Test | Non-Irritant | Pass |
| Systemic Injection Test-ISO | No acute toxicity | Pass |
| Rabbit Pyrogen Test (Material<br>Mediated)-ISO | Non-pyrogenic | Pass |
| Hemolysis- Direct Contact | Non-hemolytic | Pass |
| Hemolysis-Rabbit Blood--ASTM<br>Indirect Contact | Non-hemolytic | Pass |
| Complement Activation-Direct<br>Contact | Non-Complement<br>Activation | Pass |
| Complement Activation Assay-ISO<br>Indirect Contact | Non-Complement<br>Activation | Pass |
| In-vivo Thrombogenicity | Less thrombus than<br>control | Better thrombo-resistance<br>than control |
#### Clinical Tests
No clinical testing was submitted to support this premarket notification.
#### Conclusions
The conclusions drawn from the nonclinical tests demonstrate that the CELLO Balloon Guide Catheter is substantially equivalent to the legally marketed predicate device.
The design of the CELLO Balloon Guide catheter resembles that of the predicate with regard to its principle of operation: each catheter shaft includes an internal stainless steel braid surrounded by polymer. The balloon is made from silicone rubber and its position can be identified during angiography because of radiopaque markers.
The intended use of the CELLO Balloon Guide Catheter is substantially equivalent to that of the predicate.
The patient population of the CELLO Balloon Guide Catheter is substantially equivalent to that of the predicate.
The CELLO Balloon Guide Catheter is substantially equivalent to the predicate device with respect to the anatomical sites where the device is used.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its wings or body.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
> AUG 9 2012
Fuji Systems Corporation c/o Paul Mason, PhD Consultant 3250 Second Avenue San Diego, CA 92103
Re: K120781
CELLO Balloon Guide Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: July 26, 2012 Received: July, 27, 2012
#### Dear Dr. Mason:
We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section 210(x) premance is substantially equivalent (for the indications referenced above and have determined the devices marketed in interstate
for use stated in the enclosure) to legally marketed press Davice Amendments for use stated in the encrosule) to regary mansted producal Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Federal Food. Drug commerce prior to May 28, 1970, the chaculture with the provisions of the Federal Food. Drug, devices that have been reciassified in accordance while approval application (PMA).
and Cosmetic Act (Act) that do not require approval approval of the Act. The and Cosmetic Act (Act) that do not require apperal controls provisions of the Act. The Act. The Act. The You may, therefore, market the device, subject to the geniret for amual registration, listing of
general controls provisions of the Act include requirements for annual regist general controls provisions of the Act merade requirement.
devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it
and be If your device is classified (see adver) mio cimer sits it (epions affecting your device can be
may be subject to additional controls. Exist. Rich 24 - Rollion FDA may may be subject to additional controls: "Laisting Title 21, Parts 800 to 898. In addition, FDA may
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In a found in the Code of Federal Regulations, This Levice in the Federal Register.
publish further announcements concerning your and any also any and publish further announcements concerning your ac tree determination does not mean
Please be advised that FDA's issuance of a substantial equivalies with other requirements of Please be advised that FDA's issuance of a substaintial equivalents and the requirements of the Act
that FDA has made a determination that your device complies with other req
{7}------------------------------------------------
## Page 2 - Paul Mason, PhD
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicc-rclated adverse events) (21 CFR 803); good manufacturing
practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
CMA Wilhelm
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use Statement
# Indications for Use
510(k) Number (if known): K120781
Device Name: CELLO Balloon Guide Catheter
Indications for Use:
The CELLO Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of intravascular catheters into a selected blood vessel in the peripheral and neuro vasculature systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures.
Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
4
CmS. Willeke
(Division Sign-Off) Division of Cardiovascular Devices
K120781 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.