ABUTMENT FOR COMPETITIVE IMPLANT NOBEL ACTIVE CONNECTION
K120776 · Biomet 3i · NHA · Sep 11, 2012 · Dental
Device Facts
Record ID
K120776
Device Name
ABUTMENT FOR COMPETITIVE IMPLANT NOBEL ACTIVE CONNECTION
Applicant
Biomet 3i
Product Code
NHA · Dental
Decision Date
Sep 11, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
BIOMET 3i Dental Abutments are intended for use as accessories to endosseous dental implants to support a prosthetic device in a partially or completely edentulous patient. A dental abutment is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be screw retained or cement retained to the abutment.
Device Story
The Encode Patient Specific Abutment is a dental abutment accessory for endosseous implants. It features an anti-rotational, internal engagement interface designed to connect specifically with Nobel Active Internal Connection Implants. The device is manufactured for use by dental laboratories, which provide limited design input to match patient-specific anatomical requirements. The abutment supports either screw-retained or cement-retained prosthetic devices (artificial teeth). By providing a customized interface between the implant and the prosthesis, the device restores patient chewing function and esthetics. It is intended for use in clinical dental settings.
Clinical Evidence
Bench testing only. Dynamic fatigue testing was performed according to ISO 14801:2007 to evaluate mechanical properties, including worst-case scenarios such as 30-degree pre-angled abutments. All test articles met predetermined acceptance criteria.
Technological Characteristics
Material: Ti-6Al-4V ELI (abutment and screw). Design: Anti-rotational, internal engagement feature. Platforms: 3.5mm, 4.3mm, 5.0mm. Biocompatible. Sterilization: Not specified. Connectivity: None (mechanical component).
Indications for Use
Indicated for partially or completely edentulous patients requiring single or multiple tooth prosthetic support in the mandible or maxilla, using endosseous dental implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
Encode Patient Specific Abutment for Straumann Bone Level Connection (K112730)
Nobel Active Internal Connection Implant (K071370)
Submission Summary (Full Text)
{0}------------------------------------------------
K120776
SEP 11 2012
### 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR & 807.93
Submitter: BIOMET 3i 4555 Riverside Drive Palm Beach Gardens, FL 33410
Establishment Reg. 1038806 Number:
Prepared By: Mayank Choudhary Regulatory Affairs Specialist BIOMET 3i 4555 Riverside Drive Palm Beach Gardens, FL 33410 Tel. 561-776-6831 Fax. 561-514 6316 Email mayank.choudhary@BIOMET.com
Jacquelyn Hughes Contact for Questions/Review: Director Clinical and Regulatory Affairs BIOMET 3! 4555 Riverside Drive Palm Beach Gardens, FL 33410 Tel. 561-776-6700 Fax. 561-514 6316 Email jacquelyn.hughes@BIOMET.com
Encode Patient Specific Abutment for Nobel Active Connection Trade/Proprietary Name:
Common/Usual Dental Implant Abutment Name:
Classification Abutment, implant, dental, endosseous/ Dental Panel Name/ FDA Reviewing Branch:
Device Class II - 21 CFR \$872.3630 / NHA Classification/Code:
K112730 - Encode Patient Specific Abutment for Straumann Bone Level Connection Predicate Device K071370 - Nobel Active Internal Connection Implant Manufacturer:
{1}------------------------------------------------
## Purpose of the SPECIAL 510(k) notice:
The reason for this Special 510k submission is to request clearance for a modification to a device that has been cleared under the 510(k) process referred to herein as Encode Patient Specific Abutment for Nobel Active Connection. Dental Implant Abutment are referenced under 21 CFR §872.3630 and are considered Class II devices.
Device Description: BIOMET 3i Encode Patient Specific Abutment for Nobel Active Connection can be compared to current Encode Patient Specific Abutment for Straumann Bone level Connection. The change to this device is only to the connection that is intended to interface which is identical to the interface found on the Nobel Active Internal Connection Implant. This feature does not modify the intended functionality of the implant/abutment system. This device is also intended for dental laboratories to provide limited design input to match the anatomical requirements of the patient.
### Indications for Use:
BIOMET 3i Dental Abutments are intended for use as accessories to endosseous dental implants to support a prosthetic device in a partially or completely edentulous patient. A dental abutment is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be screw retained or cement retained to the abutment.
| | Bellatek™ Patient Specific Abutment (EDA and EDAX) | |
|---------------|----------------------------------------------------|---------------|
| Manufacturer | Implant | Platform (mm) |
| Biomet 3i | Biomet3i Certain Osseotite, Nanotite | 3.4 |
| | | 4.1 |
| | | 5 |
| | | 6 |
| Biomet 3i | Biomet3i Ex-Hex Osseotite, Nanotite | 3.4 |
| | | 4.1 |
| | | 5 |
| | | 6 |
| Nobel Biocare | Internal Connection Nobel Replace Implant | 3.5 |
| | | 4.3 |
| | | 5 |
| | | 6 |
| | Internal Connection Nobel Active Implant | 3.5 |
{2}------------------------------------------------
| | | 4.3 |
|-----------|----------------------|-----|
| | | 5.0 |
| | | 3.3 |
| Straumann | Nobel Active Implant | 4.1 |
| | | 4.8 |
Technological Characteristics: The predicates and Encode Patient Specific Abutment for Nobel Active Connection have a number of very similar and equivalent design/ technological characteristics, as follows:
| Element of<br>Comparison | Proposed BIOMET 3i Patient<br>Specific Abutment for Nobel<br>Active Connection | Predicate OEM Nobel Active Internal<br>Connection Implants | Predicate Encode Patient<br>Specific Abutments for<br>Straumann Bone Level |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | TBD | K071370 | K112730 |
| Indications | The 3i Patient-Specific Dental<br>Abutment is intended for use as<br>an accessory to an endosseous<br>dental implant to support a<br>prosthetic device in a partially<br>or edentulous patient. It is<br>intended for use to support<br>single and multiple tooth<br>prostheses, in the mandible or<br>maxilla. The prostheses can be<br>screw or cement retained to<br>the abutment. | Nobel Biocare's NobelActive implants<br>are endosseous implant intended to<br>be surgically placed in the bone of the<br>upper or lower jaw arches to provide<br>support for prosthetic devices, such<br>as an artificial tooth, in order to<br>restore patient esthetics and chewing<br>function. Nobel Biocare's NobelActive<br>implants are indicated for single or<br>multiple unit restorations in splinted<br>or non-splinted applications. Nobel<br>Biocare's NobelActive implants may<br>be placed immediately and put into<br>immediate function provided that<br>initial stability requirements detailed<br>in the manual are satisfied. | The 3i Patient-Specific<br>Dental Abutment is<br>intended for use as an<br>accessory to an<br>endosseous dental<br>implant to support a<br>prosthetic device in a<br>partially or edentulous<br>patient. It is intended for<br>use to support single and<br>multiple tooth<br>prostheses, in the<br>mandible or maxilla. The<br>prostheses can be screw<br>or cement retained to the<br>abutment. |
| Mating<br>Connection | Anti-rotational, internal<br>engagement feature<br>compatible with Nobel Active<br>Implants | Anti-rotational, internal engagement<br>feature for Nobel Active Internal<br>Connection Implants | Anti-rotational, internal<br>engagement feature<br>compatible with<br>Straumann Bone Level |
| Statistical<br>tolerance<br>analysis | Yes | OEM confidential data | Yes |
{3}------------------------------------------------
| Compatible<br>Platforms | 3.5mm, 4.3mm, 5.0mm | 3.5mm, 4.3mm, 5.0mm | 3.3m, 4.1mm, 4.8mm |
|-------------------------|---------------------|---------------------|--------------------|
| Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Biocompatible | Yes | Yes | Yes |
| Screw Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
Performance Data: BIOMET 31 has conducted Design Verification Testing according to ISO 14801:2007 "Dentistry -- Dynamic Fatigue Test for Endosseous Dental Implants" on the Encode Patient Specific Abutment for Nobel Active Connection under this submission. All testing conducted met the acceptance criteria and evaluated the worst case scenario including 30° pre-angled abutments as compared to predicate BIOMET 31 designs commercially in the marketplace. Performance testing data indicates that changes to the predicate device are safe and effective for its intended use and it demonstrate to be substantially equivalent. Bench Testing conducted outlined in the FDA Guidance referenced above demonstrates that the proposed device meets the mechanical properties recommendations by FDA.
### N/A Clinical Data:
The following FDA Guidance Document for this type of product was utilized in this Performance submission: "Guidance for Industry and FDA Staff: - Special Controls Class II -Standards: Root Form Endosseous Dental Abutment/ Implant". Also, Testing was conducted following ISO standard 14801:2007 Dentistry -- Implants - "Dynamic fatigue test for endosseous dental implants". The test articles met all predetermined acceptance criteria.
Substantial The Encode Patient Specific Abutment for Nobel Replace Connection included in this submission have the same intended use, indications for use, technological Equivalence: characteristics, and principles of operation as previously cleared Encode Patient Specific abutment for Straumann Bone Level connection per the 510(k) number referenced in the Predicate Devices section above.
· Refer to the following substantial equivalence data table:
| Element of<br>Comparison | Proposed BIOMET 3i Patient<br>Specific Abutment for Nobel<br>Active Connection | Predicate OEM Nobel Active<br>Implants | Predicate Encode Patient<br>Specific Abutments for<br>Straumann Bone Level |
|--------------------------|--------------------------------------------------------------------------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------|
| 510(k) Number | TBD | K071370 | K112730 |
| Indications | The 3i Patient-Specific Dental<br>Abutment is intended for use as | Nobel Biocare's NobelActive implants<br>are endosseous implant intended to | The 3i Patient-Specific<br>Dental Abutment is |
3/13/2012
{4}------------------------------------------------
| | an accessory to an endosseous dental implant to support a prosthetic device in a partially or edentulous patient. It is intended for use to support single and multiple tooth prostheses, in the mandible or maxilla. The prostheses can be screw or cement retained to the abutment. | be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. Nobel Biocare's NobelActive implants are indicated for single or multiple unit restorations in splinted or non-splinted applications. Nobel Biocare's NobelActive implants may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied. | intended for use as an accessory to an endosseous dental implant to support a prosthetic device in a partially or edentulous patient. It is intended for use to support single and multiple tooth prostheses, in the mandible or maxilla. The prostheses can be screw or cement retained to the abutment. |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Mating<br>Connection | Anti-rotational, internal engagement feature compatible with Nobel Active Implants | Anti-rotational, internal engagement feature for Nobel Active Internal Connection Implants | Anti-rotational, internal engagement feature compatible with Straumann Bone Level |
| Statistical<br>tolerance<br>analysis | Yes | OEM confidential data | Yes |
| Compatible<br>Platforms | 3.5mm, 4.3mm, 5.0mm | 3.5mm, 4.3mm, 5.0mm | 3.3mm, 4.1mm, 4.8mm |
| Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Biocompatible | Yes | Yes | Yes |
| Screw Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
Conclusion: Encode Patient Specific Abutment for Nobel Active Connection and predicate designs have the same intended use, indications for use, similar technological characteristics, and principles of operation. The major technological difference between the Encode Patient Specific Abutment for Nobel Active Connection and its predicates is:
> The change is only to the connection that is intended to interface which is identical to the interface found on the Nobel Active Internal Connection Implant.
The information demonstrates that the proposed device is substantially equivalent to the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains an image of an eagle in the center. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are written around the edge of the circle. The eagle is facing to the right and has its wings spread.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SEP 11 2012
Biomet 3i Ms. Tamara West Senior Regulatory Affairs Specialist 4555 Riverside Drive Palm Beach Gardens, Florida 33410
Re: K120776
Trade/Device Name: Encode Patient Specific Abutment for Nobel Active Connection Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: August 13, 2012 Received: August 15, 2012
Dear Ms. West:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
### Page 2- Ms. West
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use .
.
510(k) Number (if known): k120776
# Device Name: Encode Patient Specific Abutment for Nobel Active Connection
Indications for Use: BIOMET 3i Dental Abutments are intended for use as accessories to endosseous dental implants to support a prosthetic device in a partially or completely edentylous patient. A dental abutment is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be screw retained or cement retained to the abutment.
| Bellatek™ Patient Specific Abutment (EDA and EDAX) | | |
|----------------------------------------------------|-------------------------------------------|---------------|
| Manufacturer | Implant | Platform (mm) |
| Biomet 3i | Biomet3i Certain Osseotite, Nanotite | 3.4 |
| | | 4.1 |
| | | 5 |
| | | 6 |
| Biomet 3i | Biomet3i Ex-Hex Osseotite, Nanotite | 3.4 |
| | | 4.1 |
| | | 5 |
| | | 6 |
| Nobel Biocare | Internal Connection Nobel Replace Implant | 3.5 |
| | | 4.3 |
| | | 5 |
| | | 6 |
| Nobel Biocare | Internal Connection Nobel Active Implant | 3.5 |
| | | 4.3 |
| | | 5.0 |
{8}------------------------------------------------
| Straumann | Nobel Active Implant | 3.3 |
|-----------|----------------------|-----|
| | | 4.1 |
| | | 4.8 |
Prescription Use _____________________________________________________________________________________________________________________________________________________________ Ov er-The-Counter Use _
(Part 21 CFR 801 Subpart D)
(Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Runne
(Division Sign-Off) (Division Sign-Off)
Division of Anestheslology, General Hospital
Sheether Control Dental Devices Division of Anesthesions .
Infection Control, Dental Devices
**510(k) Number:**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.