The GM11006200 Titanium Fletcher-Style Applicator Sets are designed for intracavitary radiotherapy treatments in the uterus, cervix, endometrium and vagina. They are compatible with Varian afterloaders and can be used in combination with the appropriate accessories. The devices are intended to be used by trained and qualified personnel such as Radiation Oncologists, Physicians, Radiologists, Dosimetrists, Medical Physicists, and Nurses/MTRAs/Radiology Technicians/Radiographers in a hospital environment. The Fletcher Style Applicator Sets (defined geometry) are intended to treat cancer of the uterus, cervix, endometrium and vagina.
Device Story
Titanium Fletcher-style Applicator Set; used for intracavitary brachytherapy; treats cancer of uterus, cervix, endometrium, vagina. Device consists of intrauterine probe with stopper and colpostats with adjustable spread. Operates as passive mechanical accessory for remote-controlled radionuclide afterloaders (GammaMed plus, GammaMed 12i(t), VariSource). Used in hospital environments by radiation oncologists, medical physicists, and trained clinical staff. Provides defined geometry for radiation source delivery; enables precise dose distribution. Features CT compatibility and MR conditional status (3 Tesla). Benefits patients via improved imaging compatibility and material durability compared to legacy stainless steel/tungsten designs.
Indicated for brachytherapy treatment of cancer of the uterus, cervix, endometrium, and vagina in patients requiring intracavitary radiotherapy.
Regulatory Classification
Identification
A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.
{0}------------------------------------------------
### PREMARKET NOTIFICATION
### 510(k) Summary
### Titanium Fletcher-Style Applicator Set
As required by 21 CFR 807.92
Submitter's Name:
Varian Medical Systems 3100 Hansen Way, m/s E110 Palo Alto CA94304
Contact Name: Vy Tran Phone: 650/424.5731 Fax:650/842.5040 Date: 07 October 2011
Proprietary Name:
Classification Name:
Common/Usual Name:
Predicate Devices:
Device Description:
Indications for Use:
GM11006200 Titanium Fletcher-Style Applicator Set-Defined Geometry
Remote controlled radionuclide applicator system 21CFR892.5700 Class H
GM11006200 Titanium Fletcher-Style Applicator Set
K980590Henschke Type GYN Applicator Set
The GM11006200 Titanium Fletcher-Style Applicator Sets are designed for intracavitary radiotherapy treatments in the uterus, cervix, endometrium and vagina. They are compatible with Varian afterloaders and can be used in combination with the appropriate accessories.
The devices are intended to be used by trained and qualified personnel such as Radiation Oncologists, Physicians, Radiologists, Dosimetrists, Medical Physicists, and Nurses/MTRAs/Radiology Technicians/Radiographers in a hospital environment.
The Fletcher Style Applicator Sets (defined geometry) are intended to treat cancer of the uterus, cervix, endometrium and vagina.
Technological Characteristics:
| FEATURE AND/OR | CLEARED DEVICE | NEW DEVICE |
|------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SPECIFICATION OF | FEATURE/SPECIFICATION | |
| NEW/MODIFIED DEVICE | | |
| Predicate Device<br>Clearance Number: | Henschke Type GYN Applicator<br>Set K980590 | N/A |
| Compatible Afterloader | VariSource | GammaMed plus<br>GammaMed 12i(t)<br>VariSource |
| Intended use | The Varian VariSource Remote<br>High Dose Rate Afterloader<br>(system including applicators<br>and accessories) is a device<br>intended to be used by trained<br>and licensed medical personnel<br>to provide radiation<br>brachytherapy. The VariSource<br>Henschke Type GYN applicator<br>which is the subject of this<br>510(k) is a component of the<br>VariSource system. | The GammaMed Titanium<br>Fletcher-style Applicator Set<br>with defined geometry was<br>developed to treat cancer of the<br>uterus, cervix, endometrium<br>and vagina. |
| Indications for Use | The Henschke Type Applicator<br>is indicated for use in any case<br>where High Dose Rate (HDR)<br>radiation treatment of the<br>cervix and uterus is accepted<br>clinical practice. | The GammaMed Titanium<br>Fletcher-style Applicator Set<br>with defined geometry was<br>developed to treat cancer of the<br>uterus, cervix, endometrium<br>and vagina. |
| Design | Rigid intra-uterine tandem and<br>a pair of interlocking colpostats<br>with Henschke type ovoids. The<br>hemispsherical ovoids are<br>segemented and will accept<br>either tungsten or plastic<br>inserts so that shielding is<br>optional. Distance between<br>ovoids is adjustable. Three<br>tandems are provided: one<br>straight and two with a<br>minimum and maximum<br>curvature of the distal end to<br>accommodate intrauterine tilt. | Intrauterine Probe with stopper<br>30 °, length: 40mm, 60mm,<br>80mm (optional 30mm, 50mm,<br>70mm<br>Colpostats: 20mm, 25mm,<br>30mm spread |
| Materials | Stainless Steel, Polysulphone,<br>Tungsten | PEEK Titanium, stainless steel,<br>PPSU. |
| Packing | individual | individual |
| Sterility | Provided Non sterile | Provided Non sterile |
| Sterilization method | EtO Gas sterilization or Steam<br>sterilization | Steam sterilization |
| Sterilization conditions | 30minutes ±2 minutes at<br>250°F±5°F (121°C±1°C). | 15minutes @121°C<br>5 minutes @134°C<br>18 minutes @134°C |
| Biocompatibility | Biocompatibility not<br>mentioned. | Full biocompatibility |
| Anatomical sites | uterus, cervix, | uterus, cervix, endometrium,<br>vagina |
| Compatibility with the<br>environment and other<br>devices | Not stated to be CT compatible<br>or MR Conditional for 3 Tesla | CT compatible<br>MR Conditional for 3 Tesla |
| Rectal Retractor | Rectal Retractor not included as<br>optional accessory | Rectal Retractor included as<br>optional accessory |
K120731
JUN - 6 2012
{1}------------------------------------------------
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
{2}------------------------------------------------
and the control control control control controlled in the consideration of the consistential
1. Sales of the same of the same
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which consists of a stylized human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the emblem. The emblem is black and white.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
- 6 2012
Ms. Vy Tran Vice President, Regulatory Affairs and Quality Systems Varian Medical Systems, Inc. 3100 Hansen Way PALO ALTO CA 94304
Re: K120731
Trade/Device Name: Titanium Fletcher-style Applicator Set-Defined Geometry Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: II Product Code: JAQ Dated: March 7, 2012 Received: March 9, 2012
### Dear Ms. Tran:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
Indications for Use Form
510(k) Number (if known):
## Device Name: Titanium Fletcher-style Applicator Set-Defined Geometry
Indications for Use:
The Titanium Fletcher-style Applicator Set with defined geometry was developed for brachytherapy treatment of cancer of the uterus, cervix, endometrium and vagina.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safe 510(k)
Page 1 of _ 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.