InstaPACS™InstaRISPACS ™ is a software device (DICOM Gateway Application, Enterprise Server, and Workstation) used for viewing and manipulating digital medical images. Digital images (including mammography and data from various sources (including CT, MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or imaging sources) can be displayed, processed, stored and communicated across computer networks using this software. InstaPACS™/instaRISPACS™ can be integrated with an institution's existing Hospital Information System (HIS) or Radiology Information System (RIS) based on the study of the System, providing seamless access to reports for fullyintegrated electronic patient records. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 MPixel resolution and meets other technical specifications reviewed and accepted by FDA.
Device Story
InstaPACS™/InstaRISPACS™ is a software-based Picture Archiving and Communication System (PACS) for medical imaging. It ingests digital images from various modalities (CT, MR, US, RF, CR, DR) via a DICOM Gateway Application. The system consists of a central server for storage and web-based distribution, and a client-side workstation for image access. Radiologists and clinicians use the workstation to view, manipulate (window/level, annotate), and store images. The system integrates with existing Hospital Information Systems (HIS) or Radiology Information Systems (RIS) to provide seamless access to patient records. It facilitates remote viewing by radiologists over the web. The device does not perform automated diagnosis; it serves as a tool for human interpretation by qualified healthcare professionals. It benefits patients by enabling efficient image management, storage, and timely access to diagnostic information across clinical networks.
Clinical Evidence
Bench testing only. Validation testing was performed to evaluate all input/output functions and software operations in accordance with a System Validation Test Plan. Results demonstrated that all predetermined acceptance criteria were met.
Technological Characteristics
Software-based PACS; DICOM-compliant; supports web-based image viewing and manipulation; integrates with HIS/RIS; no patient-contacting components; does not control life-sustaining equipment; requires 5 MPixel resolution monitors for mammography interpretation.
Indications for Use
Indicated for viewing and manipulating digital medical images, including mammography, CT, MR, US, RF, CR, and DR, for clinical interpretation. Contraindicated for primary interpretation of lossy compressed mammographic images and digitized film screen images. Mammographic interpretation requires an FDA-approved monitor with at least 5 MPixel resolution.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
# 510(k) Summary of Safety
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
Date Prepared:
February 19, 2012
Submitter's Information: 21 CFR 807.92(a)(1)
Mr. Sanieev. MD and CEO MEDDIFF Technologies Pvt. Ltd. Salarpuria Palladium, 3rd floor, #2021. 100 ft Road. HAL 2ndStage Bangalore - 560 008, Karnataka - India Tel: +91-80-65350192 Fax: +91-80-66885510
Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2)
- Trade Name: Common Name: Device Classification: Product Code:
InstaPACS™/InstaRISPACS™ Picture Archiving Communications System 892.2050 System, Image Processing LLZ
#### Predicate Device: 21 CFR 807. 92(a)(3)
InstaPACS™/InstaRISPACS™ is substantially equivalent to:
| Device Classification Name | system, image processing, radiological |
|-----------------------------------|----------------------------------------|
| 510(k) Number | K080334 |
| Device Name | InstaRAD |
| Applicant | MEDSPHERE TECHNOLOGIES PVT LTD. |
| Regulation Number | 892.2050 |
| Classification Product Code | LLZ |
| Decision Date | 02/21/2008 |
| Decision | substantially equivalent (SE) |
| Classification Advisory Committee | Radiology |
#### Device Description: 21 CFR 807 92(a)(4)
InstaPACS™/InstaRISPACS™ is a software application used for viewing and manipulating medical images. Digital images and data from various sources (including CT, MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or imaging sources) can be displayed, processed, stored and communicated across computer networks using this software. When viewing images, users can perform adjustments of window width and level, annotation, and various image
APR 1 9 2012
k120718
1
{1}------------------------------------------------
manipulations. In addition, the device can be integrated with an institution's existing Hospital Information System or Radiology Information System (based on the study), providing seamless access to reports for fully-integrated electronic patient records.
InstaPACS™/InstaRISPACS™ allows multiple centers or hospitals to send their images to a central server where Radiologists can view images over Web. It consists of following components:
- . Central Server - runs the Web Server and Image Server and provide study and image data to doctors. The server includes a DICOM Gateway Application which is deployed at remote centers or hospitals. It receives images from modalities over LAN and uploads them to central server
- WorkStation User at the client side can access the study list in the . browser. They can select the patient and download the images.
#### Indications for Use: 21 CFR 807 92(a){5)
InstaPACS™InstaRISPACS ™ is a software device (DICOM Gateway Application, Enterprise Server, and Workstation) used for viewing and manipulating digital medical images. Digital images (including mammography and data from various sources (including CT, MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or imaging sources) can be displayed, processed, stored and communicated across computer networks using this software. InstaPACS™/instaRISPACS™ can be integrated with an institution's existing Hospital Information System (HIS) or Radiology Information System (RIS) based on the study of the System, providing seamless access to reports for fullyintegrated electronic patient records. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 MPixel resolution and meets other technical specifications reviewed and accepted by FDA.
#### Technological Characteristics: 21 CFR 807 92(a)(6)
InstaPACS™/InstaRISPACS ™ device is a software product that handles digital medical images. The device does not contact the patient, nor does it control any life sustaining devices. Diagnosis is not performed by the software but by Radiologists, Clinicians and referring Physicians. A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed.
The new device and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application, and intended use. The new device does not raise any new potential safety risks and is equivalent in performance to the existing legally marketed devices.
#### Nonclinical Testing:
The complete system configuration has been assessed and tested at the factory and has passed all predetermined in-house testing criteria. The Validation Test Plan was designed to evaluate all input functions, output functions, and actions performed by the InstaPACS™InstaRISPACS ™ software in each operational mode and followed the process documented in the System Validation Test Plan.
{2}------------------------------------------------
Nonclinical testing results are provided in the 510(k). Validation testing indicated that as required by the risk analysis, designated individuals performed all verification and validation activities and that the results demonstrated that the predetermined acceptance criteria were met.
### Conclusion: 21 CFR 807 92(b)(1)
The 510(k) Pre-Market Notification for InstaPACS™/InstaRISPACS ™ contains adequate information, data, and nonclinical test results to enable FDA - CDRH to determine substantial equivalence to the predicate device.
The subject device will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey. The new device and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application, and intended use does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices.
Nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the predicate device.
Therefore, InstaPACS™/InstaRISPACS™ is substantially equivalent to the predicate device.
3
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird design, with three distinct wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. The text is in all caps and appears to be in a sans-serif font.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
MEDDIFF Technologies Pvt. Ltd. % Mr. Carl Alletto Official Correspondent 1600 Manchester Way CORINTH TX 76210
Re: K120718
Trade/Device Name: InstaPACSTM/InstaRISPACSTM Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: April 7, 2012 Received: April 12, 2012
## Dear Mr. Alletto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
APR 1 9 2012
{4}------------------------------------------------
### Page 2
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K120718
# Indications for Use
510(k) Number (if known):
Device Name:
InstaPACS™/InstaRISPACS™
Indications for Use:
InstaPACS™InstaRISPACS ™ is a software device (DICOM Gateway Application, Enterprise Server, and Workstation) used for viewing and manipulating digital medical images. Digital images (including mammography and data from various sources (including CT, MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or imaging sources) can be displayed, processed, stored and communicated across computer networks using this software. InstaPACS™InstaRISPACS™ can be integrated with an institution's existing Hospital Information System (HIS) or Radiology Information System (RIS) based on the study of the System, providing seamless access to reports for fully-integrated electronic patient records.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) (Division Sign-Off) Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
**510(k) Number** K120718
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.