← Product Code [FLL](/productcode/FLL) · K120711

# NON-CONTACT THERMOMETER (K120711)

_K-Jump Health Co., Ltd. · FLL · Nov 19, 2012 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K120711

## Device Facts

- **Applicant:** K-Jump Health Co., Ltd.
- **Product Code:** [FLL](/productcode/FLL.md)
- **Decision Date:** Nov 19, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2910
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Pediatric

## Indications for Use

The Non-contact Thermometer is intended to measure the human body temperature for children and adults without contact to human body. It can be used by consumers in household environment and doctor in clinics as reference.

## Device Story

Non-contact thermometer detects infrared energy radiated from human body; infrared sensor converts energy into temperature reading. Device is compact, battery-powered, and connects to smartphone via headphone jack; temperature displayed on smartphone screen. Used in household or clinical settings by consumers or physicians. Provides non-invasive temperature measurement for clinical reference.

## Clinical Evidence

Bench testing only. Device validated for compliance with ANSI/AAMI ES60601-1 (safety), ANSI/AAMI/IEC 60601-1-2 (EMC), and ASTM E1965-98 (performance/accuracy). No clinical trials performed.

## Technological Characteristics

Infrared sensor for radiation detection; battery-powered (CR2032); dimensions 59 x 22 x 61 mm; weight 25g. Connectivity via smartphone headphone jack. Complies with ASTM E1965-98, ANSI/AAMI ES60601-1, and ANSI/AAMI/IEC 60601-1-2. Measurement range 32.2°C–43.3°C (90.0°F–109.9°F); accuracy ±0.2°C (0.4°F).

## Regulatory Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

## Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; 
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and 
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

## Predicate Devices

- Non-Contact Infrared Thermometer, Model KI-8280 ([K102947](/device/K102947.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### 510(k) Summary of Safety and Effectiveness

1. Submitter

K-jump Health Co., Ltd.

No. 56. Wu Kung 5th Rd. New Taipei Industrial Park New Taipei City 24890, Taiwan

Contact: JM Lin, Regulatory Affairs Representative +886-2-2299-1378 ext. 237 Phone: Fax: +886-2-2299-1385 jm(a)kjump.com.tw (preferred) Email:

### 2. Name of Device

Common/Usual Name: Non-contact Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Thermometer, Electronic, Clinical Regulatory Class: II Product Code: FLL

#### 3. Predicate Device

| Device Name                                                                   | 510(k) Number | Decision Date |
|-------------------------------------------------------------------------------|---------------|---------------|
| K-jump Health Co., Ltd.<br>Non-Contact Infrared Thermometer,<br>Model KI-8280 | K102947       | 2/25/2011     |

### 4. Device Description

The Non-contact Thermometer is a compact, small and light-weight device, which is powered by a battery and can be connected to a smartphone through headphone jack, for detecting body temperature without contact to human body. The measured temperature will be displayed on the screen of a smartphone.

#### 5. Indications for Use

The Non-contact Thermometer is intended to measure the human body temperature for children and adults without contact to human body. It can be used by consumers in household environment and doctor in clinics as reference.

તું

May 1, 2012

NOV 1 9 2012

{1}------------------------------------------------

#### 6. Technological Characteristics

The Non-contact Thermometer uses an infrared sensor to detect the infrared energy radiated from objects. As the intensity of the radiated energy will depend on the temperature of the air, solid, liquid or human body, the detected infrared energy can be acquired and calculated to temperature reading. The technological characteristics of Noncontact Thermometer have no difference with predicate device.

#### 7. Comparison to Predicate Device

The Non-contact Thermometer is substantially equivalent to the predicate devices, K102947, the Non-contact infrared Thermometer. The comparison of their technological characteristics is summarized in the table below.

| Characteristics Art   | ( Non-contact Thermometer (   | Non-contact Infrare<br>nermometer |
|-----------------------|-------------------------------|-----------------------------------|
| 510(k) Number         | K120711                       | K102947                           |
| Intended Use          | Similar                       | Similar                           |
| Technology            |                               |                                   |
| Measurement method    | Infrared radiation detection  | Infrared radiation detection      |
| Function              |                               |                                   |
| Measuring Range       | Forehead Mode                 | Forehead Mode                     |
|                       | 32.2℃~43.3℃,(90.0°F~109.9°F)  | 32.2℃~43.3℃, (90.0°F~109.9°F)     |
| C/F Switchable        | Yes                           | Yes                               |
| Performance           |                               |                                   |
| Display Resolution    | 0.1℃ (0.1 °F)                 | 0.1℃ (0.1 °F)                     |
| Measuring accuracy    | Forehead Mode, ±0.2°C (0.4°F) | Forehead Mode, ±0.2℃(0.4°F).      |
| Low battery detection | Yes                           | Yes                               |
| Environmental         |                               |                                   |
| Operating condition   | 10℃~40℃ (50.0~104°F)          | 10℃~40℃ (50.0~104°F)              |
|                       | 15%-95% RH                    | 15%-95% RH                        |
| Storage condition     | -20℃~55℃ (-13.0~131°F)        | -20℃~55℃ (-13.0~131°F)            |
|                       | 15%-95% RH                    | 15%-95% RH                        |
| Power                 | One CR2032                    | Two AAA batteries                 |
| Physical              |                               |                                   |
| Dimensions            | 59 x 22 x 61 mm               | 138 x 90 x 45 mm                  |
| Weight                | 25g (with battery)            | 125g (with batteries)             |

f

{2}------------------------------------------------

#### 8. Performance Summary

The performance of the Non-contact Thermometer is verified and validated to comply with following recognized standards.

- 1. ANSI/AAMI ES60601-1:2005 Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Essential Performance
The Non-contact Thermometer complies with to applicable ANSI/AAMI ES60601-1 requirements including general requirements, protection against electrical hazards, protection against mechanical hazards, protection against excessive temperatures, hazardous situations and fault conditions, and constructions.

- 2. ANSI/AAMI/IEC 60601-1-2:2007 Medical Electrical Equipment Part 1-2: General Requirements for Safety; Electromagnetic Compatibility
The Non-contact Thermometer complies with applicable ANSVAAMI/IEC 60601-1-2 requirements including radiated emission test, electrostatic discharge immunity test, radiated RF electromagnetic field immunity test, and power frequency magnetic field immunity test.

- 3: ASTM E1965-98 Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
The Non-contact Thermometer complies applicable ASTM E1965-98 requirements including displayed temperature range, maximum permissible laboratory error, ambient conditions, low power supply operation, display and human interface, constructions.

#### 9. Conclusions

The Non-contact Thermometer has the similar intended use, same fundamental scientific technology, similar technological characteristics with the predicate device. Moreover, both devices comply with similar safety and performance standards. All information described above can demonstrate the Non-contact Thermometer is substantial equivalent to the predicate device.

{3}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

#### November 19, 2012

Mr. JM Lin

Regulatory Affairs Representative K-Jump Health Company, Limited Number 56, Wu Kung 5th Road New Taipei Industrial Park New Taipei City, Taiwan 24890

Re: K120711

Trade/Device Name: Non-Contact Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: October 30, 2012 Received: October 31, 2012

Dear Mr. Lin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

#### Page 2 - Mr. Lin

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Evilla " " " " " = " =

DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, cn=Anthony D. Watson. 0.9.2342.19200300.100.1.1=1300 092402

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices

Office of Device Evaluation

Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K-jump®

# Indications for Use

510(k) Number (if known): 112-07 U

Device Name: Non-contact Thermometer

Indications For Use:

The Non-contact Thermometer is intended to measure the human body temperature for children and adults without contact to human body. It can be used by consumers in household environment and doctor in clinics as reference.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Kwame O. Ulmer ---------------------------------------------------------------------------------------------------------------------------------------------------------------

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page 1 of 1

510(k) Number: K120711

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for K-Jump. The logo is in black and features a stylized lowercase "k" connected to the word "jump". A registered trademark symbol is present to the right of the word "jump".

# Indications for Use

K120711 510(k) Number (if known):_

Device Name: Non-contact Thermometer

Indications For Use:

The Non-contact Thermometer is intended to measure the human body temperature for children and adults without contact to human body. It can be used by consumers in household environment and doctor in clinics as reference.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use × (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Rhd C. Chay 0 11/9/12

(Division Sign-Off) (Division Sign-Only Division Control, Dental Devices

Page 1 of __ 1

x 120711 510(k) Number:_

---

**Source:** [https://fda-staging.innolitics.com/device/K120711](https://fda-staging.innolitics.com/device/K120711)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
