K120711 · K-Jump Health Co., Ltd. · FLL · Nov 19, 2012 · General Hospital
Device Facts
Record ID
K120711
Device Name
NON-CONTACT THERMOMETER
Applicant
K-Jump Health Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Nov 19, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Non-contact Thermometer is intended to measure the human body temperature for children and adults without contact to human body. It can be used by consumers in household environment and doctor in clinics as reference.
Device Story
Non-contact thermometer detects infrared energy radiated from human body; infrared sensor converts energy into temperature reading. Device is compact, battery-powered, and connects to smartphone via headphone jack; temperature displayed on smartphone screen. Used in household or clinical settings by consumers or physicians. Provides non-invasive temperature measurement for clinical reference.
Clinical Evidence
Bench testing only. Device validated for compliance with ANSI/AAMI ES60601-1 (safety), ANSI/AAMI/IEC 60601-1-2 (EMC), and ASTM E1965-98 (performance/accuracy). No clinical trials performed.
Technological Characteristics
Infrared sensor for radiation detection; battery-powered (CR2032); dimensions 59 x 22 x 61 mm; weight 25g. Connectivity via smartphone headphone jack. Complies with ASTM E1965-98, ANSI/AAMI ES60601-1, and ANSI/AAMI/IEC 60601-1-2. Measurement range 32.2°C–43.3°C (90.0°F–109.9°F); accuracy ±0.2°C (0.4°F).
Indications for Use
Indicated for measuring human body temperature in children and adults without physical contact. Suitable for use by consumers in household settings and by clinicians in medical clinics.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Non-Contact Infrared Thermometer, Model KI-8280 (K102947)
Submission Summary (Full Text)
{0}------------------------------------------------
### 510(k) Summary of Safety and Effectiveness
1. Submitter
K-jump Health Co., Ltd.
No. 56. Wu Kung 5th Rd. New Taipei Industrial Park New Taipei City 24890, Taiwan
Contact: JM Lin, Regulatory Affairs Representative +886-2-2299-1378 ext. 237 Phone: Fax: +886-2-2299-1385 jm(a)kjump.com.tw (preferred) Email:
### 2. Name of Device
Common/Usual Name: Non-contact Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Thermometer, Electronic, Clinical Regulatory Class: II Product Code: FLL
#### 3. Predicate Device
| Device Name | 510(k) Number | Decision Date |
|-------------------------------------------------------------------------------|---------------|---------------|
| K-jump Health Co., Ltd.<br>Non-Contact Infrared Thermometer,<br>Model KI-8280 | K102947 | 2/25/2011 |
### 4. Device Description
The Non-contact Thermometer is a compact, small and light-weight device, which is powered by a battery and can be connected to a smartphone through headphone jack, for detecting body temperature without contact to human body. The measured temperature will be displayed on the screen of a smartphone.
#### 5. Indications for Use
The Non-contact Thermometer is intended to measure the human body temperature for children and adults without contact to human body. It can be used by consumers in household environment and doctor in clinics as reference.
તું
May 1, 2012
NOV 1 9 2012
{1}------------------------------------------------
#### 6. Technological Characteristics
The Non-contact Thermometer uses an infrared sensor to detect the infrared energy radiated from objects. As the intensity of the radiated energy will depend on the temperature of the air, solid, liquid or human body, the detected infrared energy can be acquired and calculated to temperature reading. The technological characteristics of Noncontact Thermometer have no difference with predicate device.
#### 7. Comparison to Predicate Device
The Non-contact Thermometer is substantially equivalent to the predicate devices, K102947, the Non-contact infrared Thermometer. The comparison of their technological characteristics is summarized in the table below.
| Characteristics Art | ( Non-contact Thermometer ( | Non-contact Infrare<br>nermometer |
|-----------------------|-------------------------------|-----------------------------------|
| 510(k) Number | K120711 | K102947 |
| Intended Use | Similar | Similar |
| Technology | | |
| Measurement method | Infrared radiation detection | Infrared radiation detection |
| Function | | |
| Measuring Range | Forehead Mode | Forehead Mode |
| | 32.2℃~43.3℃,(90.0°F~109.9°F) | 32.2℃~43.3℃, (90.0°F~109.9°F) |
| C/F Switchable | Yes | Yes |
| Performance | | |
| Display Resolution | 0.1℃ (0.1 °F) | 0.1℃ (0.1 °F) |
| Measuring accuracy | Forehead Mode, ±0.2°C (0.4°F) | Forehead Mode, ±0.2℃(0.4°F). |
| Low battery detection | Yes | Yes |
| Environmental | | |
| Operating condition | 10℃~40℃ (50.0~104°F) | 10℃~40℃ (50.0~104°F) |
| | 15%-95% RH | 15%-95% RH |
| Storage condition | -20℃~55℃ (-13.0~131°F) | -20℃~55℃ (-13.0~131°F) |
| | 15%-95% RH | 15%-95% RH |
| Power | One CR2032 | Two AAA batteries |
| Physical | | |
| Dimensions | 59 x 22 x 61 mm | 138 x 90 x 45 mm |
| Weight | 25g (with battery) | 125g (with batteries) |
f
{2}------------------------------------------------
#### 8. Performance Summary
The performance of the Non-contact Thermometer is verified and validated to comply with following recognized standards.
- 1. ANSI/AAMI ES60601-1:2005 Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Essential Performance
The Non-contact Thermometer complies with to applicable ANSI/AAMI ES60601-1 requirements including general requirements, protection against electrical hazards, protection against mechanical hazards, protection against excessive temperatures, hazardous situations and fault conditions, and constructions.
- 2. ANSI/AAMI/IEC 60601-1-2:2007 Medical Electrical Equipment Part 1-2: General Requirements for Safety; Electromagnetic Compatibility
The Non-contact Thermometer complies with applicable ANSVAAMI/IEC 60601-1-2 requirements including radiated emission test, electrostatic discharge immunity test, radiated RF electromagnetic field immunity test, and power frequency magnetic field immunity test.
- 3: ASTM E1965-98 Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
The Non-contact Thermometer complies applicable ASTM E1965-98 requirements including displayed temperature range, maximum permissible laboratory error, ambient conditions, low power supply operation, display and human interface, constructions.
#### 9. Conclusions
The Non-contact Thermometer has the similar intended use, same fundamental scientific technology, similar technological characteristics with the predicate device. Moreover, both devices comply with similar safety and performance standards. All information described above can demonstrate the Non-contact Thermometer is substantial equivalent to the predicate device.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
#### November 19, 2012
Mr. JM Lin
Regulatory Affairs Representative K-Jump Health Company, Limited Number 56, Wu Kung 5th Road New Taipei Industrial Park New Taipei City, Taiwan 24890
Re: K120711
Trade/Device Name: Non-Contact Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: October 30, 2012 Received: October 31, 2012
Dear Mr. Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Lin
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Evilla " " " " " = " =
DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, cn=Anthony D. Watson. 0.9.2342.19200300.100.1.1=1300 092402
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K-jump®
# Indications for Use
510(k) Number (if known): 112-07 U
Device Name: Non-contact Thermometer
Indications For Use:
The Non-contact Thermometer is intended to measure the human body temperature for children and adults without contact to human body. It can be used by consumers in household environment and doctor in clinics as reference.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kwame O. Ulmer ---------------------------------------------------------------------------------------------------------------------------------------------------------------
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
510(k) Number: K120711
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for K-Jump. The logo is in black and features a stylized lowercase "k" connected to the word "jump". A registered trademark symbol is present to the right of the word "jump".
# Indications for Use
K120711 510(k) Number (if known):_
Device Name: Non-contact Thermometer
Indications For Use:
The Non-contact Thermometer is intended to measure the human body temperature for children and adults without contact to human body. It can be used by consumers in household environment and doctor in clinics as reference.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use × (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Rhd C. Chay 0 11/9/12
(Division Sign-Off) (Division Sign-Only Division Control, Dental Devices
Page 1 of __ 1
x 120711 510(k) Number:_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.