K120680 · Spine View, Inc. · HRX · Jun 26, 2012 · Orthopedic
Device Facts
Record ID
K120680
Device Name
ENSPIRE DISCECTOMY SYSTEM
Applicant
Spine View, Inc.
Product Code
HRX · Orthopedic
Decision Date
Jun 26, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.1100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The enSpire™ Discectomy System is intended for use in cutting, grinding and aspirating intervertebral disc material during discectorny procedures in the cervical, thoracic and lumbar spine.
Device Story
The enSpire™ Discectomy System is a single-use, battery-powered, handheld surgical device used by physicians in operating rooms or medical suites. It features a working shaft with a distal serrated cutting window and an internal auger mechanism. During operation, the motor-driven auger rotates against entrapped tissue and the fixed cutting window to cut and grind intervertebral disc material. The excised debris is then aspirated through the working shaft into an integrated collection chamber. The device is used percutaneously, endoscopically, or in open surgery, with visualization provided by direct view, fluoroscopy, or other imaging modalities. By enabling controlled tissue removal, the device assists in discectomy procedures, potentially benefiting patients by facilitating spinal decompression.
Clinical Evidence
Bench testing only. No clinical data provided. In-vitro studies evaluated cannula compatibility, switch durability, deployment/retraction, shaft length, device durability, travel limiter function, tensile strength, articulation, visualization, peak operating temperature, electromagnetic compatibility, electrical safety, packaging, shipping, sterility, shelf life, and biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity). Testing confirmed tissue volume/material removal without breach of annulus or endplates.
Technological Characteristics
Single-use, battery-powered (9V or 18V) mechanical discectomy device. Features a straight or articulating working shaft with a non-expandable serrated cutting window. Materials: Stainless steel. Sterilization: Gamma. Connectivity: None. Operates via mechanical auger rotation for tissue excision and aspiration. Complies with IEC 60601-1 for electrical safety.
Indications for Use
Indicated for patients undergoing discectomy procedures in the cervical, thoracic, and lumbar spine requiring cutting, grinding, and aspiration of intervertebral disc material.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
{0}------------------------------------------------
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### A. Name, Address, Phone and Fax Number of Applicant
K120680
Spine View, Inc. 48810 Kato Road, Suite 100E Fremont, CA 94538 Phone: (510) 623-1931 (510) 490-1753 Fax:
#### B. Contact Person
Diana DeGregorio Lincé Consulting Regulatory Affairs Consultant (925) 980-8047 dianadegregorio@comcast.net
Alternate Contact: Mbithi Muthini Director Quality and Regulatory (510) 743-5090 mmuthini@spineview.com
#### C. Date Prepared
March 5, 2012
#### D. Device Name
Trade Name: enSpire™ Discectomy System Common Name: Arthroscope & Accessory Classification Name: Arthroscope & Accessories (21 CFR §888.1100, Product Code HRX)
#### E. Predicate Devices
The modified enSpire™ Discectomy System is substantially equivalent to the Spine View enSpire™ Discectomy System cleared under K110992 on October 21, 2011.
{1}------------------------------------------------
Spine View, Inc.
#### F. Device Description
The modified enSpire™ Discectomy System is a single-use discectomy device that is designed to cut and grind intervertebral disc material. An auger mechanism retrieves the excised debris and ejects it into a collection chamber.
The modified enSpire™ Discectorny System is supplied as a sterile, single patient use, disposable device.
#### G. Intended Use
The modified enSpire™ Discectomy System is intended for use in cutting, grinding and aspirating intervertebral disc material during discectorny procedures in the cervical, thoracic and fumbar spine.
#### H. Technological Comparison
The modified enSpire™ Discectorny System has similar features compared to the predicate devices in the table below.
| Manufacturer | Spine View, Inc. | Spine View, Inc. |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | enSpire™ Discectomy System | Modified enSpire™ Discectomy System |
| 510(k) Number | K110992 | TBD |
| Indications for Use | The enSpire™ Discectomy System is intended for use in cutting, grinding and aspirating intervertebral disc material during discectomy procedures in the cervical, thoracic and lumbar spine. | Same |
| Product Code | HRX<br>Arthroscope, 21CFR888.1100, Class II, (Debrider) | Same |
| Principal Operator | Physician | Same |
| Use Location | Operating Room or Medical Suite | Same |
| Operating Principal | Percutaneous, endoscopic or open surgical Discectomy system with standard surgical accessories | Same |
| Functions of Included Devices | Dilatation<br>Access<br>Excision<br>Aspiration | Same |
| Mechanics of Action | Activating the device causes the cutting mechanism to expand and rotate and auger to rotate to cut and grind the disc material, which is then aspirated down the working shaft of the device and collected in the collection chamber | Activating the device causes the auger to rotate against the entrapped tissue and fixed cutting mechanism to cut and grind the disc material, which is then aspirated down the working shaft of the device and collected in the collection chamber |
| Target Anatomy | Cervical, Thoracic and Lumbar spinal segments | Same |
| Design Features | Auger housed in a tube with expandable rotating cutting assembly (spiral wire or curette style cutter) on the distal end. Driven by a battery-powered motor that is housed in a plastic handle at the proximal end of the device. | Auger housed in a tube with a non-expandable Serrated Cutting Window at the distal end. Driven by a battery-powered motor that is housed in a plastic handle at the proximal end of the device. |
| Manufacturer | Spine View, Inc. | Spine View, Inc. |
| Device Name | enSpire™ Discectomy System | Modified enSpire™ Discectomy System, |
| 510(k) Number | K110992 device. | TBD |
| | Cut debris can pass up the tube and into the<br>collection chamber located at the end of the<br>auger. It is inserted into the surgical site<br>either directly, via an introducer cannula or<br>Arthroscope. | Cut debris can pass up the tube and into the<br>collection chamber located at the end of the<br>auger. It is inserted into the surgical site<br>either directly, via an introducer cannula or<br>Arthroscope. |
| | The device contains a straight, curved or<br>articulating working shaft. | The device contains a straight or articulating<br>working shaft. |
| | Expandable spiral wire and a curette style<br>cutter<br>Sweep Diameter (Expanded): 0.280 - 0.390" | Non-Expandable Serrated Cutting Windows<br>SV3309: 0.075" x 0.240" x 2 windows<br>SV1107: 0.055" x 0.170" x 2 windows |
| Tip Materials | PEEK, Aramid Fiber, Polyimide, Stainless<br>Steel, and Tungsten | Stainless Steel |
| Sterile Packaging | The enSpire™ Discectomy System is placed<br>into a thermo formed tray with a<br>thermoformed insert lid, and sealed with a<br>Tyvek tray lid. The sealed tray is then placed<br>in a labeled chip board shelf carton. | Same |
| Sterilization Method | Gamma | Same |
| Biocompatible for<br>Intended Use | Yes | Yes |
| Single use | Yes | Yes |
| Configuration | Straight, Curved, Articulating | Straight or Articulating |
| Handle Design | Hand-held rotary device, in-line or pistol-grip<br>handle | Same |
| Profile | 0.046-0.165" OD working shaft with tip<br>0.280-0.400" OD deployed, 0.058-0.220" OD<br>unexpanded | 0.058-0.110" OD shaft diameter at tip |
| Working Length | 2 -22" working length | 2 -9" working length |
| Energy Type | Mechanical | Same |
| Power | Battery, 9V or 18V | Same |
| Meets Applicable<br>IEC60601-1 testing | Yes | Yes |
| User visualization/guidance | Direct visualization, fluoroscopic imaging or<br>other imaging modalities | Same |
{2}------------------------------------------------
Spine View, Inc.
# K120680
The technological characteristics and principals of operation of the modified enSpire™ Discectomy System are substantially equivalent to the named predicate device.
#### i. Summary of Non-Clinical Data
The modified enSpire™ Discectomy System performance characteristics were evaluated in the following in-vitro bench studies:
{3}------------------------------------------------
K120680
- Cannula Compatibility .
- . Enable Switch Durability
- . Deployment & Retraction
- Working Shaft Length .
- Device Durability .
- Travel Limiter Attachment .
- Travel Limiter .
- Tensile Strength .
- Articulation Function .
- . Articulation Angle
- . Visualization
- Peak Temperature during . Operation
- · Electromagnetic Compatibility and Electrical Safety
- · Packaging Testing
- · Shipping Testing
- · Sterility Testing
- · Shelf Life Testing
- · Biocompatibility:
- Cytotoxicity ం
- Sensitization o
- Irritation O
- Systemic Toxicity o
- Tissue Volume/Material Removal . No Breach of Annulus or Endplates
Results of the pre-clinical testing demonstrate that the materials chosen, the manufacturing process, and design of the modified enSpire™ Discectomy System meet the established specifications necessary for consistent performance during its intended use. In addition, the testing demonstrates the modified enSpire™ Discectomy System is substantially equivalent to the named predicate.
#### J. Summary of Data
.
The modified enSpire™ Discectomy System has been carefully compared to a legally marketed device with respect to intended use and technological characteristics. In addition, non-clinical testing was conducted to validate the performance of the device and ensure the modified enSpire™ Discectomy System performs as intended and meets the design specifications. The non-clinical performance testing and comparison to the predicate device demonstrate that the modified enSpire™ Discectomy System is substantially equivalent to the predicate device and does not raise new issues of safety or effectiveness.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Spine View, Incorporated % Ms. Diana DeGregorio Regulatory Affairs Consultant 48810 Kato Road Suite 100E Fremont, California 94538
Re: K120680
Trade/Device Name: enSpire™ Discectomy System Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope & Accessory Regulatory Class: Class II Product Code: HRX Dated: June 15, 2012 Received: June 18, 2012
Dear Ms. DeGregorio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
JUN 2 6 2012
{5}------------------------------------------------
Page 2 - Ms. Diana DeGregorio
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
for Math Dar
or
Mark N. Melke Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known): K_K12 06 80
Device Name:
enSpire™ Discectomy System
### Indications for Use:
The enSpire™ Discectomy System is intended for use in cutting, grinding and aspirating intervertebral disc material during discectorny procedures in the cervical, thoracic and lumbar spine.
Prescription Use × OR
Over-The-Counter Use
(per 21 CFR 801.109)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Niel R.P. Ogden for mkn
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
. 510(k) Number K120680
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.