← Product Code [LLZ](/productcode/LLZ) · K120679

# SYNAPSE 3D NODULE ANALYSIS (K120679)

_Fujifilm Medical Systems U.S.A, Inc. · LLZ · Jun 1, 2012 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K120679

## Device Facts

- **Applicant:** Fujifilm Medical Systems U.S.A, Inc.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Jun 1, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

Synapse 3D Nodule Analysis is medical imaging software used with Synapse 3D Base Tools that is intended to provide trained medical professionals with tools to aid them in reading, interpreting, reporting, and treatment planning. Synapse 3D Nodule Analysis accepts DICOM compliant medical images acquired from CT and PT. This product is not intended for use with or for the primary diagnostic interpretation of Mammography images. Addition to Synapse 3D Base Tools, Synapse 3D Nodule Analysis is intended to: use non-contrast and contrast enhanced computed tomography (CT) images, provide custom workflows and UI, and reporting functions for nodule analysis including boundary detection of nodules based on the location specified by the user, measurement of nodules, temporal comparison of nodule images, and fusion of a CT nodule image on a PT nodule image. use positron emission tomography (PT) images, provide custom workflows and UI, and reporting functions for nodule analysis including SUV measurement of nodules with sphere VOI based on the location specified by the user, temporal comparison of nodule images, and fusion of a PT nodule image on a CT nodule image.

## Device Story

Synapse 3D Nodule Analysis is an optional software module for the Synapse 3D platform; operates on Windows server/client configurations. Inputs: DICOM-compliant CT and PT images. Functionality: provides custom workflows, UI, and reporting for nodule analysis; performs boundary detection (user-specified location), nodule measurement, temporal image comparison, and image fusion (CT/PT). For PET images, calculates standardized uptake value (SUV) using sphere volume of interest (VOI). Used by trained medical professionals in clinical settings to assist in treatment planning and response evaluation (RECIST/PERCIST). Output: visualized analysis, measurements, and reports; supports clinical decision-making by providing quantitative data and comparative views of nodule progression.

## Clinical Evidence

Bench testing only. Verification and validation activities, including testing against Software Requirements Specifications and Traceability Matrix, were performed to establish performance, functionality, and reliability.

## Technological Characteristics

Software-based medical image processing module; runs on commercial general-purpose Windows-compatible hardware. Connectivity: DICOM-compliant. Software class: Moderate concern per FDA guidance. Features: boundary detection, nodule measurement, temporal comparison, image fusion, and SUV calculation.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Advantage Windows Lung Analysis: GE Healthcare (GE Medical Systems) ([K042694](/device/K042694.md))

## Reference Devices

- Synapse 3D Base Tools (V3.0) ([K120361](/device/K120361.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K120679

FUJIFILM Medical Systems U.S.A., Inc. Synapse 3D Nodule Analysis 510(k)

# 5. 510(k) Summary

# Date Prepared:

February 28, 2012

### Submitter's Information:

FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, Connecticut 06902

Telephone: (301) 251-1092 Fax: (203) 602-3785 Contact: Jyh-Shyan Lin

### Device Trade Name:

Synapse 3D Nodule Analysis

### Device Common Name:

Medical Image Processing and Analysis Software

### Requlation Number:

21 CFR 892.2050

#### Device Classification:

Class II

### Device Classification Name :

Picture Archiving and Communications System (PACS)

### Panel:

Radiology

### Product Code:

LLZ

### Date Received:

TBD

### Decision Date :

TBD

### Decision :

TBD

# JUN - 1 2012

{1}------------------------------------------------

## FUJIFILM Medical Systems U.S.A. Synapse 3D Nodule Analysis 510(k)

### ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

· Advantage Windows Lung Analysis: GE Healthcare (GE Medical Systems) (K042694).

### Description of the Device

Synapse 3D Nodule Analysis is an optional software module that works with Synapse 3D Base Tools (V3.0) (K120361, cleared on April 6, 2012). Synapse 3D Nodule Analysis, Synapse 3D Base Tools (V3.0), as well as other optional software modules, all belong to the Synapse 3D product family.

Synapse 3D is the medical application software running on Windows server/client configuration installed on a commercial general-purpose Windows-compatible computer. It offers software tools which can be used by trained medical professionals to interpret medical images obtained from various medical devices, to create reports, or to develop treatment plans.

The key features in Synapse 3D Nodule Analysis include boundary detection of nodules, measurement of nodules, temporal comparison of nodule images, and fusion of CT and PT nodule images on the non-contrast and contrast enhanced CT or PT images.

Synapse 3D Nodule Analysis offers physicians the following clinical applications in addition to the features available from Synapse 3D Base Tools (V3.0) to analyze the image data retrieved from CT and PT devices.

- . Response Evaluation Criteria in Solid Tumors (RECIST) Tracker Use non-contrast and contrast enhanced computed tomography (CT) images, provide custom workflows, user interface (UI), and reporting functions for nodule analysis including boundary detection of nodules (based on the location specified by the user), measurement of nodules, temporal comparison of nodule images, and fusion of a CT nodule image on a PT nodule image.
- Positron Emission Response Criteria in Solid Tumors (PERCIST) Tracker . Use positron emission tomography (PT) images, provide custom workflows user interface (UI), and reporting functions for nodule analysis including standardized uptake value (SUV) measurement of nodules with sphere VOI (based on the location specified by the user), temporal comparison of nodule images, and fusion of a PT nodule image on a CT nodule image.

## Indication for Use

Synapse 3D Nodule Analysis is medical imaging software used with Synapse 3D Base Tools that is intended to provide trained medical professionals with tools to aid them in reading, interpreting, reporting, and treatment planning. Synapse 3D Nodule Analysis accepts DICOM compliant medical images acquired from CT and PT.

{2}------------------------------------------------

FUJIFILM Medical Systems U.S.A., Inc Synapse 3D Nodule Analysis 510(k)

This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.

Addition to Synapse 3D Base Tools, Synapse 3D Nodule Analysis is intended to:

- use non-contrast and contrast enhanced computed tomography (CT) images, provide A custom workflows and UI. and reporting functions for nodule analysis including boundary detection of nodules based on the location specified by the user, measurement of nodules, temporal comparison of nodule images, and fusion of a CT nodule image on a PT nodule image.
- 公 use positron emission tomography (PT) images, provide custom workflows and UI, and reporting functions for nodule analysis including SUV measurement of nodules with sphere VOI based on the location specified by the user, temporal comparison of nodule images, and fusion of a PT nodule image on a CT nodule image.

### Technological Characteristics

Synapse 3D Nodule Analysis introduces no new safety or efficacy issues other than those already indentified with the predicate devices. The results of the Hazard Analysis combined with the appropriate preventive measures taken indicate that the device is of moderate concern as per the May 11, 2005 issue of the "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices."

### Testing

Synapse 3D Nodule Analysis is tested successfully with reference to its Software Requirements Specification, as well as design verification and validation documents and Traceability Matrix document. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the Synapse 3D Nodule Analysis software, which is found to be safe and effective and substantially equivalent to the currently-cleared predicate device.

### Conclusion

This 510(k) premarket notification submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. We conclude the subject device to be as safe and effective as the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized design featuring three abstract shapes that resemble human figures or birds in flight.

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Mr. Jyh-Shyan Lin Senior Manager, Regulatory, Quality and Clinical Affairs FUJIFILM Medical Systems, U.S.A., Inc. 419 West Avenue STAMFORD CT 06902

JUN - 1 2012

Re: K120679

Trade/Device Name: Synapse 3D Nodule Analysis Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 28, 2012 Received: March 6, 2012

### Dear Mr. Lin:

We have reviewed your Section 510(k) premarket notification of intent to market the device . referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{4}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice mourements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket with anony four to FDA finding of substantial equivalence of your device to a legally marketed nonication. The I Drivinant ssification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 11 Jul deen before the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known):

Device Name: Synapse 3D Nodule Analysis

Indications for Use:

Synapse 3D Nodule Analysis is medical imaging software used with Synapse 3D Base Tools that is intended to provide trained medical professionals with tools to aid them in reading. interpreting, reporting, and treatment planning. Synapse 3D Nodule Analysis accepts DICOM compliant medical images acquired from CT and PT.

This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.

Addition to Synapse 3D Base Tools, Synapse 3D Nodule Analysis is intended to;

- > use non-contrast and contrast enhanced computed tomography (CT) images, provide custom workflows and Ul, and reporting functions for nodule analysis including boundary detection of nodules based on the location specified by the user, measurement of nodules, temporal comparison of nodule images, and fusion of a CT nodule image on a PT nodule image.
- use positron emission tomography (PT) images, provide custom workflows and UI, and A reporting functions for nodule analysis including SUV measurement of nodules with sphere VOI based on the location specified by the user, temporal comparison of nodule images, and fusion of a PT nodule image on a CT nodule image.

Prescription Use x AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

---

(Division Sign Off)

Division of Radiological Devices Office of In 510K

Page 1 of

---

**Source:** [https://fda-staging.innolitics.com/device/K120679](https://fda-staging.innolitics.com/device/K120679)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
