← Product Code [LLZ](/productcode/LLZ) · K120637

# SYNAPSE 3D BRAIN PERFUSION (K120637)

_Fujifilm Medical System U.S.A., Inc. · LLZ · May 8, 2012 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K120637

## Device Facts

- **Applicant:** Fujifilm Medical System U.S.A., Inc.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** May 8, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

Synapse 3D Brain Perfusion is medical imaging software used with Synapse 3D Base Tools that is intended to provide trained medical professionals with tools to aid them in reading, interpreting, reporting, and treatment planning. Synapse 3D Brain Perfusion accepts DICOM compliant medical images acquired from CT and MR. This product is not intended for use with or for the primary diagnostic interpretation of Mammography images. Addition to Synapse 3D Base Tools, Synapse 3D Brain Perfusion provides the parameter images by post-processing with dynamic scanned CT cerebral arteriography images and magnetic resonance images (MR) with contrast agent to aid the assessment of cerebral blood flow. The parameter images are Cerebral Blood Volume (CBV), Cerebral Blood Flow (CBF), Mean Transit Time (MTT), and Time To Peak (TTP).

## Device Story

Synapse 3D Brain Perfusion is medical imaging software for post-processing dynamic CT cerebral arteriography and contrast-enhanced MR images. It calculates and maps cerebral hemodynamic parameters: Cerebral Blood Volume (CBV), Cerebral Blood Flow (CBF), Mean Transit Time (MTT), and Time To Peak (TTP). Used in clinical settings by trained medical professionals, the software integrates with the Synapse Workstation or Synapse Cardiovascular system. It functions as an add-on to Synapse 3D Base Tools, which provides standard image manipulation (windowing, zoom, measurement). Clinicians use the generated parameter maps to assess cerebral blood flow, aiding in interpretation and treatment planning. The device does not perform primary diagnosis but supports clinical decision-making by visualizing perfusion data.

## Clinical Evidence

Bench testing only. Verification and validation activities, including testing against Software Requirements Specifications and a Traceability Matrix, were performed to establish performance, functionality, and reliability.

## Technological Characteristics

Software-based medical image processing application. Operates on DICOM-compliant CT and MR images. Provides post-processing for hemodynamic parameter mapping (CBV, CBF, MTT, TTP). Integrates with Synapse Workstation (v3.2.1+) and Synapse Cardiovascular. Classified as moderate concern per FDA software guidance.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Synapse 3D Cerebral Analysis ([K103687](/device/K103687.md))
- IB Neuro v1.0 ([K080762](/device/K080762.md))

## Reference Devices

- Synapse 3D Base Tools ([K120361](/device/K120361.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

FUJIFILM Medical Systems U.S.A., Inc. Synapse 3D Brain Perfusion (V3.0) 510(k) 12120637

MAY - 8 2012

# 5. 510(k) Summary

### Date Prepared:

February 28, 2012

### Submitter's Information:

FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, Connecticut 06902

|          | Telephone: (301) 251-1092 |
|----------|---------------------------|
| Fax:     | (203) 602-3785            |
| Contact: | Jyh-Shyan Lin             |

### Device Trade Name:

Synapse 3D Brain Perfusion

#### Device Common Name:

Medical Image Processing and Analysis Software

#### Regulation Number:

21 CFR 892 2050

### Device Classification:

Class II

### Device Classification Name:

Picture Archiving and Communications System (PACS)

### Panel:

Radiology

### Product Code:

LLZ

#### Date Received:

TBD

### Decision Date:

TBD

#### Decision:

TBD

{1}------------------------------------------------

### FUJIFILM Medical Systems U.S.A., Inc. Synapse 3D Brain Perfusion (V3.0) 510(k)

### Predicate Device:

- . Synapse 3D Cerebral Analysis. FUJIFILM Medical Systems U.S.A., Inc. (K103687)
- IB Neuro v1.0. Imaging Biometrics, LLC (K080762) �

### Description of the Device.

Synapse 3D Brain Perfusion (V3.0) is the updated version of previously-cleared Synapse 3D Cerebral Analysis software (cleared by CDRH via K103687 on 03/04/2011).

Synapse 3D Brain Perfusion is used in addition to Synapse 3D Base Tools (K120361, cleared on April 6, 2012) to analyze the images acquired from CT and MR. Synapse 3D Brain Perfusion is intended to provide trained medical professionals with tools to aid them in reading, interpreting, reporting, and treatment planning of DICOM compliant medical images. This product is not intended for use with or for the primary diagnostic interpretation of mammography images.

Synapse 3D Cerebral Analysis (V3.0) is an application that analyzes the changes of the cerebral blood flow on the dynamic scanned CT cerebral arteriography and the MR images.

- Brain Perfusion (CT): Unchanged from the cleared version K103687 . Brain Perfusion is an application analyzing the changes of the cerebral blood flow on the dynamic scanned CT cerebral arteriography images. CBV (Cerebral Blood Volume), CBF (Cerebral Blood Flow), MTT (Mean Transit Time), and TTP (Time To Peak) are calculated and mapped on images.
· Brain Perfusion (MR):

Brain Perfusion is an application analyzing the changes of the cerebral blood flow on the magnetic resonance images (MR) with contrast agent. CBV (Cerebral Blood Volume), CBF (Cerebral Blood Flow), MTT (Mean Transit Time), and TTP (Time To Peak) are calculated and mapped on images.

Common image processing functions (such as window width and window level, zooming, panning, flip, rotate, adding annotations on an image, measurement of lengths, areas, etc.),are available to support the cerebral analysis of the CT and MR images. These functions belong to and are provided by Synapse 3D Base Tools (K120361) that is used with Synapse 3D Brain Perfusion (V3.0) (this submission).

The Brain Perfusion (CT) application is unchanged from the cleared version (K103465) and the Brain Perfusion (MR) is the added application.

Synapse 3D Brain Perfusion with Synapse 3D Basic/Base Tools can be integrated with our cleared Fujifilm's Synapse Workstation, version 3.2.1 and above, and can be used as a part of a Synapse system. Synapse 3D Brain Perfusion also can be integrated with Fujifilm's Synapse Cardiovascular for cardiology purposes.

{2}------------------------------------------------

### FUJIFILM Medical Systems U.S.A., Inc. Synapse 3D Brain Perfusion (V3.0) 510(k)

### Indication for Use

Synapse 3D Brain Perfusion is medical imaging software used with Synapse 3D Base Tools that is intended to provide trained medical professionals with tools to aid them in reading, interpreting, reporting, and treatment planning. Synapse 3D Brain Perfusion accepts DICOM compliant medical images acquired from CT and MR.

This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.

Addition to Synapse 3D Base Tools, Synapse 3D Brain Perfusion provides the parameter images by post-processing with dynamic scanned CT cerebral arteriography images and magnetic resonance images (MR) with contrast agent to aid the assessment of cerebral blood flow. The parameter images are Cerebral Blood Volume (CBV), Cerebral Blood Flow (CBF), Mean Transit Time (MTT), and Time To Peak (TTP).

#### Technological Characteristics

Synapse 3D Brain Perfusion introduces no new safety or efficacy issues other than those already indentified with the predicate devices. The results of the Hazard Analysis combined with the appropriate preventive measures taken indicate that the device is of moderate concern as per the May 11, 2005 issue of the "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices."

#### Testing

Synapse 3D Brain Perfusion is tested successfully with reference to its Software Requirements Specification, as well as design verification and validation documents and Traceability Matrix document. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the Synapse 3D Brain Perfusion software, which is found to be safe and effective and substantially equivalent to the currently-cleared predicate device.

#### Conclusion

This 510(k) premarket notification submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. We conclude the subject device to be as safe and effective as the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the emblem.

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Mr. Jyh-Shyan Lin Senior Manager, Regulatory, Ouality and Clinical Affairs FUJIFILM Medical Systems, Inc. 419 West Avenue STAMFORD CT 06902

MAY - 8 2012

Re: K120637

Trade/Device Name: Synapse 3D Brain Perfusion Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 28, 2012 Received: March 1, 2012

#### Dear Mr. Lin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{4}------------------------------------------------

#### Page 2

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please,go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Indications for Use

510(k) Number (if known):

Device Name: Synapse 3D Brain Perfusion

Indications for Use:

Synapse 3D Brain Perfusion is medical imaging software used with Synapse 3D Base Tools that is intended to provide trained medical professionals with tools to aid them in reading, interpreting, reporting, and treatment planning. Synapse 3D Brain Perfusion accepts DICOM compliant medical images acquired from CT and MR.

This product is not intended for use with or for the primary diagn ostic interpretation of Mammography images.

Addition to Synapse 3D Base Tools, Synapse 3D Brain Perfusion provides the parameter images by post-processing with dynamic scanned CT cerebral arteriography images and magnetic resonance images (MR) with contrast agent to aid the assessment of cerebral blood flow. The parameter images are Cerebral Blood Volume (CBV), Cerebral Blood Flow (CBF), Mean Transit Time (MTT), and Time To Peak (TTP).

Prescription Use × AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division of Radiological Device Office of in V

510K

Page 1 of

---

**Source:** [https://fda-staging.innolitics.com/device/K120637](https://fda-staging.innolitics.com/device/K120637)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
