# OEC 9900 ELITE (K120613)

_Ge Healthcare Surgery GE Oec Medical Systems, Inc. · OXO · Apr 6, 2012 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K120613

## Device Facts

- **Applicant:** Ge Healthcare Surgery GE Oec Medical Systems, Inc.
- **Product Code:** [OXO](/productcode/OXO.md)
- **Decision Date:** Apr 6, 2012
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1650
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The OEC® 9900 Elite Mobile Fluoroscopy System is designed to provide fluoroscopic and spot-film images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, vascular, cardiac, critical care and emergency room procedures.

## Device Story

Mobile fluoroscopic imaging system; assists physicians during surgical/interventional procedures. System consists of C-arm gantry and workstation. X-ray generator in C-arm base produces high voltage; X-ray tube emits radiation directed at patient; image intensifier captures transmitted X-rays. Camera converts intensifier output to digital images displayed on workstation monitor. Operator adjusts C-arm rotation/translation, X-ray technique, image size, and orientation to account for anatomical variation. Used in clinical environments (OR, ER, etc.) by physicians and operators. Provides real-time visualization and records of anatomy/pathology to localize surgical regions, guide procedures, and document outcomes. Benefits include improved visualization for orthopedic, vascular, and cardiac interventions.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

Mobile fluoroscopic X-ray system. Components: C-arm gantry, workstation, X-ray generator, X-ray tube, image intensifier, camera. Energy source: High voltage X-ray. Connectivity: Workstation-based display/storage. Sterilization: Not applicable (device is capital equipment).

## Regulatory Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- OEC® 9900 Elite ([K082781](/device/K082781.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## APR - 6 2012

Image /page/0/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circle. The letters and the circle are black, creating a strong contrast. The logo is a well-known symbol associated with the multinational conglomerate corporation.

# GE Healthcare 510(k) Premarket Notification Submission

K120613

#### 510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

February 24, 2012 Date:

Submitter:

GE Healthcare Surgery 384 Wright Brothers Drive Salt Lake City, UT 84116

Primary Contact Person:

Gerald Buss Director Regulatory Affairs GE Healthcare Surgery Phone: (801) 517-6591 Fax: (801) 517-6566

Secondary Contact Person:

Karen Russell Regulatory Affairs Leader GE Healthcare Surgery Phone: (801) 536-4930 Fax: (801) 517-6566

Device: (Trade Name): Common/Usual Name: Classification Names: Product Code:

OEC® 9900 Elite

Mobile Fluoroscopic Imaging System

21 CFR 892.1650 Image-intensified fluoroscopic x-ray system

90OXO and 90JAA

K082781 OEC® 9900 Elite Predicate_Device(s):

Device Description:

The OEC® 9900 Elite is a system used to assist trained physicians. The system is used to provide X-Ray images while the physician performs a medical procedure. Images from the system help the physician to visualize the patients' anatomy. This visualization helps to localize surgical regions of interest and The images provide real-time visualization and pathology. records of pre-surgical anatomy, in vivo-surgical activity and post surgical outcomes.

The system is composed of two primary physical elements. The first is referred to as the "C-Arm" because of its "C" shaped image gantry; the second referred to as the "Workstation" because this is the primary user interface to the system.

{1}------------------------------------------------

# GE Healthcare 510(k) Premarket Notification Submission

The system is used in different surgical procedures. Orthopedic physicians may use the system to help perform hip replacements and reconstructions of badly fractured bones. Orthopedic physicians may use the system to help perform hip replacements and reconstructions of badly fractured bones. Vascular physicians may use the system to perform blood flow studies.

Cardiologists may use the system to help see if there are blockages in some of the key arteries supplying blood to the heart. The procedures that these physicians perform are broadly referred to as "Clinical Applications". The system is controlled and run in a clinical environment.

The system employs X-Rays as its imaging technology. An X-Ray Generator located in the base of the C-Arm creates high voltage. High voltage is carried to the X-Ray tube across a set of cables. The X-Ray tube emits X-Rays that are directed toward the patient under the control of the operator. The Physician defines the desired view for the specific clinical procedure and directs the operator. The X-Rays pass through the patient and are captured by the image intensifier (II). Image intensifier images are captured by a camera and displayed on the image monitor located on the Workstation. The Physician or system operator view the images as they are displayed and thev may choose to store the images for later review.

In order to perform these procedures different views of the human anatomy are required, so the system is designed with the ability to rotate and translate the C-Arm's image gantry to obtain different viewing angles. In addition since there is variation in thickness and density of the anatomy the system operator has the ability to adjust the X-Ray Generator technique, image size and orientation

urologic, orthopedic, neurologic, vascular, cardiac, critical care

to account for the anatomical differences. Intended Use: The OEC® 9900 Elite Mobile Fluoroscopy System is designed to provide fluoroscopic and spot-film images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy,

Technology:

The modified OEC® 9900 Elite device employs the same fundamental scientific technology as the predicate device.

and emergency room procedures.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circle. There are decorative swirls around the outside of the circle. The logo is black and white.

00004

# GE Healthcare 510(k) Premarket Notification Submission

#### Determination of Substantial Equivalence:

The OEC® 9900 Elite and its application comply with voluntary standards as detailed in Section 9, of this premarket submission. The modifications from the predicate device OEC® 9900 Elite were completed in accordance with GE Healthcare Surgery's quality management system design controls. Engineering testing and standards compliance testing were successfully conducted and did not raise any new safety questions or concerns or identify new risks. Information is included with this 510(k) submission that supports this determination.

Conclusion:

GE Healthcare considers the modified GE OEC® 9900 Elite to be safe, as effective, and performance is substantially equivalent to the predicate device OEC® 9900 Elite (K082781).

{3}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Mr. Gerald Buss Director Regulatory Affairs GE Healthcare Surgery 384 Wright Brothers Drive SALT LAKE CITY UT 84116

Re: K120613

Trade/Device Name: GE OEC 9900 Elite Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, OXO and JAA Dated: March 15. 2012 Received: March 16, 2012

Dear Mr. Buss:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

APR - 6 2012

{4}------------------------------------------------

#### Page 2

Enclosure

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed needicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

{5}------------------------------------------------

GE Healthcare 510(k) Premarket Notification Submission

### INDICATIONS FOR USE

510(k) Number (if known):

Device Name: GE OEC 9900 Elite

Indications for Use:

The OEC® 9900 Elite is designed to provide fluoroscopic and spot-film images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, vascular, cardiac, critical care and emergency room procedures.

Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Radiological Devices
Office ofi In Vitro Diagnostic Device Evaluation and Safety
K120613

BIOK

---

**Source:** [https://fda-staging.innolitics.com/device/K120613](https://fda-staging.innolitics.com/device/K120613)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).