K120604 · Eltech S.R.L · ILY · May 25, 2012 · Physical Medicine
Device Facts
Record ID
K120604
Device Name
K-LASER
Applicant
Eltech S.R.L
Product Code
ILY · Physical Medicine
Decision Date
May 25, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Each device is indicated for emitting energy in the Infrared Spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with arthritis and promoting relaxation of the muscle tissue and to temporarily increase local blood circulation.
Device Story
K-Laser Cube family (1, 2, 3, 4) are table-top infrared laser devices; emit coherent light in continuous or pulse modes at wavelengths 800nm, 905nm, and 970nm. Operated by clinicians; features touch screen interface for function management, handpiece for light delivery, and battery pack for portable use. Devices elevate tissue temperature via infrared energy; intended for temporary relief of minor muscle/joint pain, spasms, arthritis-related stiffness, and increased local blood circulation. Benefits include non-invasive pain management and muscle relaxation.
Clinical Evidence
Bench testing only. No clinical data provided. Device performance verified against applicable FDA performance standards for electronic and light-emitting products.
Technological Characteristics
Table-top infrared laser; wavelengths 800nm, 905nm, 970nm; peak power 10W-20W. Continuous or pulse mode operation. Battery-powered or AC-powered. Includes touch screen interface, handpiece, and software-controlled emission. Conforms to 21 CFR 1010, 1040.10, and 1040.11.
Indications for Use
Indicated for patients requiring topical heating to relieve minor muscle/joint pain, muscle spasms, pain/stiffness from arthritis, promote muscle relaxation, and temporarily increase local blood circulation.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows handwritten text. The text at the top reads "K120604". Below that, the text reads "Page 1 of 3". The handwriting is somewhat stylized, but the text is still legible.
MAY 2 5 2012
#### SUMMARY OF CTIVENESS
### SUBMITTED BY:
ELTECH s.r.l.
Via Castagnole, 20/H
31100 Treviso, Italy
Phone: 0039 422 210 430
Fax: 0039 422 297 137
### CONTACT:
## BE MANN UE CAEFIY A
#### Official Correspondent:
dr. Richard Albright
K-Laser USA
1185 W. Main Street, Franklin, TN 37064
Phone: 866-595-7749
#### US Agent:
dr. Richard Albright ,PRES K-Laser USA 1185 W. Main Street, Franklin, TN 37064 Phone: 866-595-7749 Fax: 615-261-3535 Email: ralbright@k-laserusa.com
#### DEVICE NAME (Trade/common, and classification): 1.
Proprietary name: K-LASER Common/usual name: K-Laser Cube 1, K-Laser Cube 2, K-Laser Cube 3, K-Laser Cube 4 Classification name: Infrared Lamp Classification: Class II Regulation Nos.: 21 CFR 890.5500 Product Codes: ILY
.
": .
:
1-
{1}------------------------------------------------
K120604
Page 2 of 3
#### PREDICATE DEVICES: 2.
SUBMISSION
The device under submission is substantially equivalent to the predicate devices:
K091497 (K-1200);
K061656 (Laser-D68);
The device under submission is a family of laser emit a beam of coherent light in either continuous wave or pulse mode at the following wavelengths:
K-Laser Cube 1: 905nm; peak power: 10W; K-Laser Cube 2: 800nm, 970nm; peak power: 15W; K-Laser Cube 3: 800nm, 905nm, 970; peak power: 20W; K-Laser Cube 4: 905nm; peak power: 10W.
# 3. PERFORMANCE STANDARDS: substantially equiv ser
The device conforms to the applicable requirements of 21 CFR section 1010 (Performance Standards for Electronic Products: General) and 21 CFR sections 1040.10 and 1040.11 (Performance Standards for Light-Emitting Products).
## 4. DESCRIPTION:
The devices under submission are a family of laser that emits a beam of coherent light in either continuous wave or pulse mode at the wavelengths previously described.
Each device is a table device, easy to transport, usable also without electrical net, thanks to a battery pack. It is composed of a touch screen for managing all the device functions, an emitter, an handpiece for the delivery of light, software and an on/off button to activate and deactivate the infrared emission.
ﺍﻟﺘﻲ ﺗ · '1'1' 1' 1' 11. 11 . . . . . re or pulse made at the wavelengths pre
{2}------------------------------------------------
120604
Page 3 of
#### INTENDED USE/ INDICATIONS FOR USE: 5.
Each device is indicated for emitting energy in the Infrared Spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with arthritis and promoting relaxation of the muscle tissue and to temporarily increase local blood circulation.
#### SUBSTANTIAL EQUIVALENCE (SE) RATIONALE: 6.
The devices under submission share the same intended use, similar design and functional features as the predicate devices without raising any issues of safety or effectiveness. Therefore, the devices under submission are substantially equivalent to the predicate devices K091497 (K-1200), K061656 (Laser-D68).
#### accard to contling chergy in the are 7. SAFETY AND EFFECTIVENESS: 1 , . .
There are no substantive differences between the product defined in this 510(k) submission and the predicate devices. They are similar to the technologies that are currently used in other similar medical devices. They were developed and documented under Eltech's mature Quality Managerment System, under The Quality System Regulation, 21 CFR Part 820, under design/change control, and verified/validated to applicable standards/guidance documents.
RA ( C ) RA ( ( / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / Besides, Eltech's Quality Assurance System is certified by CERMET, notified body n. CE 0476, according to Annex II of 93/42 EEC Directive, transposed in Italy by Dlgs. n. 46 of 24 February 1997.
The devices are safe and effective when used as indicated in specific applications under a clinician's supervision/therapy program.
.. , EFECTIVENESS: Date: 25th April 2012
Signature:
Francesco Zanata Eltech s.r.l. President
schily P.5Stardinge 57510111 15 Cell.
. : .
,
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized image of an eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The eagle is facing to the right, and its wings are curved upwards. The text is in all capital letters and is evenly spaced around the circle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Eltech S.R.L. % K-Laser USA Mr. Richard Albright President 1185 West Main Street Franklin, Tennessee 37064
2012 2 గ్రా
Re: K120604
Trade/Device Name: K-Laser Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: IL Y Dated: February 28, 2012 Received: February 28, 2012
Dear Mr. Albright:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 - Mr. Albright
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Eunel Keith
r Mark N. Melkerson Director Division of Surgical. Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE
## Indications for Use:
SECTION 4: STAT 1. KFT K-laser Cube 1,2,3, and 4 device is indicated for emitting energy in the Infrared Spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary . relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with arthritis and promoting relaxationy?the\musple.tissue and to temporarily increase local blood circulation.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
........ relax.com of the muses
Neil R. Dale for mxm
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Device
510(k) Number K120604
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.