K120595 · Consensus Orthopedics, Inc. · LWJ · Oct 12, 2012 · Orthopedic
Device Facts
Record ID
K120595
Device Name
UNISYN PLUS HIP STEM
Applicant
Consensus Orthopedics, Inc.
Product Code
LWJ · Orthopedic
Decision Date
Oct 12, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3360
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis. Revision of failed femoral head replacement, hip arthroplasty or other hip procedures. A. Proximal femoral fractures. B. Avascular necrosis of the femoral head. C. Non-union of proximal femoral neck fractures. D. Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, pseudarthrosis conversion, and structural abnormalities. UNISYN assemblies with roughened and plasma coated bodies are intended for cemented or uncemented use. UNISYN assemblies with plasma/HA or HA coated bodies are intended for uncemented use only.
Device Story
Modular hip prosthesis system consisting of neck, body, and stem components. Subject device introduces modified 'Plus' stems with adjusted body thickness to improve contact with medullary canal during revision surgery. Stems available in neutral or +1 mm styles, 10-20 mm diameters, and 110-210 mm lengths. Used by orthopedic surgeons in clinical/surgical settings for hip arthroplasty. Implanted to restore joint function and stability in patients with degenerative or traumatic hip conditions. Design modifications maintain compatibility with existing UniSyn system components and surgical techniques.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, specifically distal fatigue performance evaluation per ISO 7206-4 on a worst-case stem configuration.
Technological Characteristics
Modular hip prosthesis system. Materials and design construct are identical to predicate components. Features include grit-blasted, plasma-coated, or HA-coated bodies. Stems available in neutral or +1 mm styles, 10-20 mm diameters, and 110-210 mm lengths. Distal fatigue performance tested per ISO 7206-4. No changes to sterilization or manufacturing processes.
Indications for Use
Indicated for patients with significantly impaired joints due to rheumatoid, osteoarthritis, or post-traumatic arthritis; revision of failed hip procedures; proximal femoral fractures; avascular necrosis; non-union of femoral neck fractures; or structural/developmental hip conditions (e.g., dysplasia, coxa vara/valga).
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum. This generic type of device includes designs which are intended to be fixed to the bone with bone cement (§ 888.3027) as well as designs which have large window-like holes in the stem of the device and which are intended for use without bone cement. However, in these latter designs, fixation of the device is not achieved by means of bone ingrowth.
{0}------------------------------------------------
## k 120 595
#### 2. 510(k) SUMMARY
OCT 1 2 2012
| Sponsor Name: | Consensus Orthopedics, Inc.<br>1115 Windfield Way, Suite 100<br>El Dorado Hills, CA 95762 | | |
|----------------------|---------------------------------------------------------------------------------------------------------------------------|--|--|
| 510(k) Contact: | Matthew M. Hull, RAC<br>Phone: 916 355-7156<br>e-mail: mhull@consensusortho.com | | |
| Date Prepared: | 11 October 2012 | | |
| Trade Name: | UNISYN Plus stem | | |
| Common Name: | Stem, modular hip prosthesis. | | |
| Device Class: | 2 | | |
| Classification Name: | Prosthesis, Hip, Semi-Constrained, Metal/Polymer,<br>Uncemented (21 CFR 888.3360, LWJ) | | |
| | Prosthesis, Hip, Semi-Constrained, Metal/Polymer,<br>Cemented (21 CFR 888.3350, JDI) | | |
| | Prosthesis, Hip, Semi-Constrained,<br>Metal/Ceramic/Polymer, Cemented or Non-Porous,<br>Uncemented (21 CFR 888.3353, LZO) | | |
#### Device Description:
The UniSyn Hip System was originally cleared as a modular hip system for Hayes Medical under the name "Triton", modifications to the system were cleared in a subsequent submission by Hayes under the UniSyn name. Hayes Medical changed its name to Consensus Orthopedics in 2008. This system consists of three primary components: the neck, the body, and the stem. This submission is to addition of a set of modified ("Plus") stems.
These minor modifications in body thickness are designed to allow for better contact between the implant and the inner surface of the medullary canal during revision surgery. The sizes of stems offered with the modified design are within the ranges previously cleared for the UniSyn system. The Plus stems are offered in neutral or +1 mm styles, short or long, in 10 to 20 mm diameters, and lengths from 110 to 210 mm. There is no change to the manufacturing, packaging, or sterilization processes. This change will not affect the indications for use nor any of the labeling.
{1}------------------------------------------------
# 120595
#### Indications for Use:
A) Significantly impaired joints resulting from rheumatoid, osteo, and posttraumatic arthritis.
B) Revision of failed femoral head replacement, hip arthroplasty or other hip procedures.
C) Proximal femoral fractures.
D) Avascular necrosis of the femoral head.
E) Non-union of proximal femoral neck fractures.
F) Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, pseudarthrosis conversion, and structural abnormalities.
UNISYN stems used with roughened and plasma coated bodies are intended for cemented or uncemented use. UNISYN stems used with plasma/HA or HA coated bodies are intended for uncemented use only.
### Substantial Equivalence:
### Technological Characteristics/Substantial Equivalence:
The new "Plus" stems for the UniSyn hip systems employ the same materials, basic design features, manufacturing processes, packaging, and indications as the predicate stems for the UHS. Therefore, the new Plus stems for use with the UNISYN Hip System are substantially equivalent to legally marketed predicate devices (Table 2.1).
| 510(k)<br>Number | Trade Name | 510(k) holder | 510(k)<br>Release Date |
|------------------|------------------------------|---------------------|------------------------|
| K003649 | Triton Hip System | Hayes Medical, Inc. | 02/23/2001 |
| K062383 | UniSyn HA Plasma Modular Hip | Hayes Medical, Inc. | 12/21/2006 |
Table 2.1: Legally marketed devices to which substantial equivalence is claimed:
#### Non-Clinical Performance Data:
The UNISYN Hip System (UHS) implant components use the identical materials, design construct features of the predicate components. The UHS implant components were evaluated using a Failure Modes and Effects Analysis (FMEA). Distal fatigue performance of the UniSyn Plus grit blasted stem was evaluated per ISO 7206-4 on a worst case example. No other nonclinical bench testing was deemed necessary over that used in support of the predicates based on identical design and materials.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
OCT 12 2012
Consensus Orthopedics, Incorporated % Mr. Matthew M. Hull, RAC Director, QS and RA 1115 Windfield Way, Suite 100 El Dorado Hills, California 95762
Re: K120595
Trade/Device Name: UNISYN Hip System Regulation Number: 21 CFR 888.3360 Regulation Name: Hip joint femoral (hemi hip) metallic cemented or uncemented prosthesis Regulatory Class: Class II Product Code: LWJ, JDI, LZO Dated: October 4, 2012 Received: October 5, 2012
Dear Mr. Hull:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
#### Page 2 - Mr. Matthew M. Hull, RAC
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
For
J. Melkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## 12-0595
#### 1. INDICATIONS FOR USE STATEMENT
510(k) Number: K120595
Device Name: UNISYN Hip System
#### Indications for Use:
Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis. Revision of failed femoral head replacement, hip arthroplasty or other hip procedures.
- A. Proximal femoral fractures.
- B. Avascular necrosis of the femoral head.
- C. Non-union of proximal femoral neck fractures.
- D. Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, pseudarthrosis conversion, and structural abnormalities.
UNIS YNTM assemblies with roughened and plasma coated bodies are intended for cemented or uncemented use. UNISYNTM assemblies with plasma/HA or HA coated bodies are intended for uncemented use only.
Prescription Use X (21 CFR Part 801 Subpart D) AND/OR
Over the Counter Use · (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
K. Harrel.
Post
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K120595
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.