K120502 · Shenzhen Creative Industry Co., Ltd. · DQA · Dec 7, 2012 · Cardiovascular
Device Facts
Record ID
K120502
Device Name
PC-60NW FINGERTIP OXIMETER
Applicant
Shenzhen Creative Industry Co., Ltd.
Product Code
DQA · Cardiovascular
Decision Date
Dec 7, 2012
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The PC-60NW Fingertip Oximeter is intended for measuring the pulse rate and functional oxygen saturation (SpO2) through patient's finger, and indicating the pulse intensity by a bar-graph display. This device is powered by 2 AAA batteries. This device is applicable for spot-checking of SpO2 and pulse rate in home and clinic for adult and pediatric patients. This device is recommended for use on the index finger, for patients with fingers of 1.0 - 2.2 cm thick.
Device Story
Fingertip pulse oximeter; measures SpO2 and pulse rate via finger-mounted sensor; displays data on OLED color dot matrix screen including plethysmogram, bar graph, and perfusion index. Used in home or clinic settings by patients or clinicians for spot-checking. Features automatic start-up upon finger insertion; includes wireless module for data transmission to computer. Provides real-time physiological monitoring to assist in identifying hypoxemia; aids clinical decision-making by offering non-invasive, convenient alternative to blood gas analysis.
Clinical Evidence
Bench testing only. Device validated against system V&V plan, including electrical safety (IEC 60601-1-1), EMC (IEC 60601-1-2), biocompatibility (ISO 10993-1), and pulse oximeter performance standards (ISO 9919:2005, ISO 80610-2-61). No clinical trial data presented.
Technological Characteristics
OLED color dot matrix display; wireless communication module; powered by 2 AAA batteries. Biocompatible materials per ISO 10993-1. Standards: IEC 60601-1-1 (electrical safety), IEC 60601-1-2 (EMC), ISO 9919:2005 and ISO 80610-2-61 (SpO2 performance).
Indications for Use
Indicated for spot-checking pulse rate and functional oxygen saturation (SpO2) in adult and pediatric patients with finger thickness 1.0-2.2 cm. Recommended for use on index finger in home or clinical settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Shenzhen Creative Industry Co., Ltd, PC-60 Fingertip Oximeter (K063641)
Submission Summary (Full Text)
{0}------------------------------------------------
KI2OSO2
# 510(k) Summary
# DEC 0 7 2012
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
# Date: October 15, 2012
# 1. Company:
Name - Shenzhen Creative Industry Co., Ltd. Address Addr. 2/F, Block 3, Nanyou Tian'an Industry Town, Shenzhen, GD 518054, P.R. China Telephone - +86-755-26434955 Fax - +86-755-26434955 Contact - Mrs. Wang Jia Email charliemack@irc-us.com
## Correspondent:
Name- IRC Address- 77325 Joyce Way, Echo, Oregon 97826 Telephone- 931-625-4938 Fax- 541-376-5063 Contact- Charlie Mack Email- charliemack@irc-us.com
# 2. Device :
| Trade/proprietary name | PC-60NW Fingertip Oximeter |
|------------------------|----------------------------|
| Common Name | Oximeter |
| Classification Name | Oximeter |
3. Predicate Devices :
Shenzhen Creative Industry Co., Ltd, PC-60 Fingertip Oximeter K063641
This submission is an Abbreviated 510(k) submission, based upon standards and the predicate is simply the original basis of the submitted device.
4. Classifications Names & Citations :
21CFR 870.2700 Oximeter, DQA, Class2
{1}------------------------------------------------
Description :
General
The pulse oxygen saturation is the percentage of HbO2 in the total Hb in the blood, and the so-called O2 concentration in the blood. It is an important bio-parameter for the respiration. Many of the respiration disease will cause hypoxemia, even damage the patient's life. As a result, monitoring the SpO2 is indispensable in the clinical rescuing. The traditional monitoring method is to take the patient's blood sample and analyze the sample by the blood gas analyzer, measure out the O2 pressure and calculate the SpO2 finally. Traditionally SpO2 measure is based on the clinical help of sapling human blood, and getting O2 pressure by hemo-air analysis using hemo-air analyzer, then calculate the SpOz. This method is inconvenient and discontinuous. For the purpose of measuring the SpO2 more easily and accurately, our company developed the Fingertip Oximeter. The device can measure the body pulse simultaneously.
The PC-60NW Fingertip Oximeter is intended for measuring the pulse rate and functional oxygen saturation (SpO2) through patient's finger, and indicating the pulse intensity by a bar-graph display. This device is powered by 2 AAA batteries, it is small in size, convenient to use, and easy to carry. This device is applicable for spot-checking of SpO2 and pulse rate in home and clinic. The index finger is the recommended site. It is intended for spotchecking adult and pediatric patients on fingers between 1.0 -2.2 . cm thick.
This oximeter has the function of wireless data transmission. The user can effectively transmit the data to computer through the wireless communication module.
{2}------------------------------------------------
### 5. Indication for use :
The PC-60NW Fingertip Oximeter is intended for measuring the pulse rate and functional oxygen saturation (SpO2) through patient's fing er, and indicating the pulse intensity by a bar-graph display.
This device is powered by 2 AAA batteries. This device is applicable for spot-checking of SpO2 and pulse rate in home and clinic for adult and pediatric patients. This device is recommended for use o n the index finger, for patients with fingers of 1.0 - 2.2 cm thick.
### 6. Comparison with predicate device :
Shenzhen Creative Industry Co., Ltd, PC-60 Fingertip Oximeter K063641
This submission is an Abbreviated 510(k) submission, based upon standards and the predicate is simply the original basis of the submitted device.
7. Safety and Performance Data :
# Summary of Performance Testing
The Model PC-60 Fingertip Oximeter substantially has been tested in accordance with the system V & V plan and summary included with the submission using production equivalent units prior to release to market. A risk analysis identifying potential hazards and documenting mitigation of the hazards has been developed and applied as part of design control procedure. Shenzhen Creative Industry, LTD quality system confirms to 21 CFR820, ISO 9001, ISO13485 and CMDCAS ISO 13485. Testing performed to Electrical Safety IEC 60601-1-1:1988, EMC Requirements IEC 60601-1-2, Biocompatibility IEC 10993-1, SPO2 Basic Safety and performance ISO 9919:2005, Basic Electrical safety and performance for Pulse Oximeters, ISO 80610-2-61 and pertinent FDA quidance documents, complied with the specific requirements for safety and performance.
{3}------------------------------------------------
- 8. Differences between Predicate and Submitted PC60-NW:
The only difference between the submitted device and the predicate is in the display. The predicate uses a seven segment LCD to display the SPO2 and Pulse rate. Also the predicate displays the pulse amplitude with a 16 segment bar graph. The submitted device uses an OLED color dot matrix to display SPO2, Pulse Rate/perfusion index and plethysmogram and bar graph. The submitted device has two new features of automatic start-up two seconds after the finger is inserted and the new device has wireless PC connection capabilities.
#### 9. Conclusions:
The performance data cited in the non- clinical testing performed is sufficient to demonstrate substantial equivalence to the predicate device. In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Shenzhen Creative Industry Co., Ltd. concludes that the PC-60NW Fingertip Oximeter is safe and effective and complies with the acceptance criteria in the SP02 standards and FDA guidance documents as described herein.
END
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the bird symbol. The text is in all caps and appears to be in a sans-serif font.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Shenzhen Creative Industry Company, Limited C/O IRC Mr. Charles Mack Principal Engineer 77325 Joyce Way ECHO, OR 97826
#### Re: K120502
Trade/Device Name: PC-60NW Fingertip Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: November 28, 2012 Received: December 04, 2012
## Dear Mr. Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
DEC 0 7 2012
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
**Kwame** O. Ulmer
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Indications for Use
# Indications for Use
510(k) Number (if known): K / 20562
Device Name: PC-60NW Fingertip Oximeter
Indications For Use:
- The PC-60NW Fingertip Oximeter is intended for measuring the pulse rate and functional oxygen saturation (SpO2) through patient's finger, and indicating the pulse intensity by a bar-graph display.
This device is powered by 2 AAA batteries. This device is applicable for spotchecking of SpO2 and pulse rate in home and clinic for adult and pediatric patients. This device is recommended for use on the index finger, for patients with fingers of 1.0 - 2.2 cm thick.
AND/OR Prescription Use _ (Part 21 CFR 801 Subpart D)
.. . . . . . . . . . . . .
Over-The-Counter Use __ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) raval Roaphs
Concurrence of CDRH, Office of Device Evaluation (ODE)
Lester W. Schultheis Jr 2012.12.07 11:31:42 -05'00'
(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital Division Control, Dental Devices
K120502 510(k) Number;
15
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.