← Product Code [LLZ](/productcode/LLZ) · K120484

# VISIA ONCOLOGY (K120484)

_MeVis Medical Solutions AG · LLZ · Mar 27, 2012 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K120484

## Device Facts

- **Applicant:** MeVis Medical Solutions AG
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Mar 27, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

Visia™ Oncology is a medical software application intended for the visualization of images from a variety of image devices. The system provides viewing, quantification, communication, and printing of medical images. Visia™ Oncology is a noninvasive image analysis software package designed to support the physician in routine diagnostic oncology, staging and follow-up. Flexible layouts and automated image registration facilitate the synchronous display and navigation of multiple datasets for viewing data and easy follow-up comparison. The application provides a range of interactive tools specifically designed for segmentation and volumetric analysis of findings. The integrated reporting helps the user to track findings and note changes, such as shape or size, over time.

## Device Story

Visia™ Oncology is noninvasive medical image processing software; inputs medical images from CT systems or archives; operates on standard PC as standalone workstation or distributed server-client network. Designed for use by physicians (radiologists, clinicians) in clinical oncology settings for routine diagnostics, staging, and follow-up. Features include automated image registration for synchronous display/navigation of multiple datasets; interactive tools for segmentation and volumetric analysis of findings; integrated reporting for tracking changes in findings over time. Output is displayed/printed images and quantitative data; supports physician decision-making by facilitating comparison and monitoring of oncology findings. Benefits include improved workflow efficiency and longitudinal tracking of patient disease state.

## Clinical Evidence

No clinical data. Bench testing only; validation testing evaluated all input/output functions and operational modes to ensure compliance with risk management procedures and predetermined acceptance criteria.

## Technological Characteristics

Software-only medical device; no patient contact; no life-sustaining control. Operates on standard off-the-shelf PC. Supports network connectivity (server-client). Features automated image registration, segmentation, and volumetric analysis tools. No specific hardware materials or sterilization required.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- syngo CT Oncology ([K071310](/device/K071310.md))
- Vitrea® ([K071331](/device/K071331.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 1120484

# 510(k) Summary

Submitted By:

MeVis Medical Solutions AG Universitaetsallee 29 28359 Bremen, Germany

Establishment Name:

Establishment Registration Number:

Contact Person:

MeVis Medical Solutions AG

10039520

Thomas E. Tynes President & CEO MeVis Medical Solutions, Inc. N27 W24075 Paul Court - Suite 100 Pewaukee, WI 53072 USA Phone: +1-262-691-9530 Fax: +1-262-691-9531

| Date Prepared:                       | 2/14/2012                         |
|--------------------------------------|-----------------------------------|
| Trade Name:                          | Visia™ Oncology                   |
| Common Name:                         | Medical Image Processing Software |
| Classification Name:                 | Image Processing System           |
| Classification<br>Regulation Number: | 892.2050                          |
| Class:                               | II                                |
| Panel:                               | Radiology                         |
| Product Code:                        | LLZ                               |

Visia™ Oncology 510(k) 510(k) Summary - 2/14/2012 Page 1 of 3

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for MeVis MEDICAL SOLUTIONS. The logo is in black and white, with the words "MeVis" in large, bold letters. Below the word "MeVis" is the phrase "MEDICAL SOLUTIONS" in smaller letters. The logo appears to be for a medical company.

# Device Description

Visia™ Oncology is a noninvasive medical image processing software application intended for the visualization of images from various sources such as Computed Tomography systems or from image archives. The system provides viewing, quantification, manipulation, and printing of medical images.

Visia™ Oncology integrates within typical clinical workflow patterns through receiving and transferring medical images over a computer network. The software can be loaded on a standard off-the-shelf personal computer (PC) and can operate as a stand-alone workstation or in a distributed server-client configuration across a computer network.

Visia™ Oncology is designed to support the physician in routine diagnostic oncology, staging and follow-up. Flexible layouts and automated image registration facilitate the synchronous display and navigation of multiple datasets for viewing data and easy follow-up comparison. The application provides a range of interactive tools specifically designed for segmentation and volumetric analysis of findings. The integrated reporting helps the user to track findings and note changes, such as shape or size, over time.

## Intended Use

Visia™ Oncology is a medical software application intended for the visualization of images from a variety of image devices. The system provides viewing, quantification, communication, and printing of medical images. Visia™ Oncology is a noninvasive image analysis software package designed to support the physician in routine diagnostic oncology, staging and follow-up. Flexible layouts and automated image registration facilitate the synchronous display and navigation of multiple datasets for viewing data and easy follow-up comparison. The application provides a range of interactive tools specifically designed for segmentation and volumetric analysis of findings. The integrated reporting helps the user to track findings and note changes, such as shape or size, over time.

## Predicate Device Information

| Product         | Predicate Device Name | Predicate 510(k)<br>Submission Reference |
|-----------------|-----------------------|------------------------------------------|
| Visia™ Oncology | syngo CT Oncology     | K071310                                  |
| Visia™ Oncology | Vitrea®               | K071331                                  |

## Safety and Effectiveness

The Visia™ Oncology labeling contains instructions for use and necessary cautions, warnings and notes to provide for safe and effective use of the device. Risk Management is ensured via MeVis Medical Solution AG's Risk Management procedure, which is used to identify potential hazards. These potential hazards are controlled via software development and verification & validation testing.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for MeVis MEDICAL SOLUTIONS. The logo is in black and white, with the word "MeVis" in large, bold letters. Below the word "MeVis" is the phrase "MEDICAL SOLUTIONS" in smaller letters. The logo is simple and professional, and it is likely used to represent the company's brand.

## Nonclinical Testing and Performance Information

The complete system configuration has been assessed and tested at the manufacturer's facility and has passed all in-house testing criteria. The Validation Test Plan was designed to evaluate all input functions, output functions, and actions performed by the software in each operational mode and followed the process documented in the Validation Test Plan.

Nonclinical testing results are provided in the 510(k). Validation testing indicated that as required by the risk analysis, designated individuals performed all verification and validation activities and that the results demonstrated that the predetermined acceptance criteria were met.

#### Technological Characteristics

Visia™ Oncology is a software device that does not contact the patient, nor does it control any life sustaining devices. Diagnosis is not performed by the software but by Radiologists, Clinicians and referring Physicians.

A physician, providing ample opportunity for competent human interprets images and information being displayed and printed.

The new device and predicate devices are substantially equivalent in the areas of technical characteristics, general function, and intended use. The new device does not raise any new potential safety risks and is equivalent in performance to the existing legally marketed devices.

#### Conclusion

The 510(k) Pre-Market Notification for Visia™ Oncology contains adequate information, data, and nonclinical test results to enable FDA - CDRH to determine substantial equivalence to the oredicate device. MeVis Medical Solutions has determined that its device, Visia™, is substantially equivalent to the identified predicate device listed above. A comparison with the legally marketed predicate device indicates that it is substantially equivalent to this device, and that it does not raise any new safety or efficacy concerns. Nonclinical tests demonstrate that the device is safe, effective, and is substantially equivalent to the predicate device.

> Visia™ Oncology 510(k) 510(k) Summary - 2/14/2012 Page 3 of 3

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

MeVis Medical Solutions AG % Mr. Thomas E. Tynes President & CEO MeVis Medical Solutions, Inc. N27 W24075 Paul Court, Suite 100 PEWAUKEE WI 53072

Re: K120484

Trade/Device Name: Visia™ Oncology Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 14, 2012 Received: February 17, 2012

#### Dear Mr. Tynes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such 11 your w 100 to 100 to 100 regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

MAR 2 7 2012

{4}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for MeVis Medical Solutions. The logo is in black and white, with the word "MeVis" in large, bold letters. Below the word "MeVis" are the words "MEDICAL SOLUTIONS" in smaller letters. The logo is simple and professional, and it is likely used to represent the company's brand.

# Indications for Use Statement

510(k) Number (if known): _ |ん | 20484

Device Name: Visia™ Oncology

#### Indications for Use:

Visia Oncology is a medical software application intended for the visualization of images from a variety of image devices. The system provides viewing, quantification, manipulation, and printing of medical images. Visia Oncology is a noninvasive image analysis software package designed to support the physician in routine diagnostic oncology, staging and follow-up. Flexible layouts and automated image registration facilitate the synchronous display and navigation of multiple datasets for viewing data and easy follow-up comparison. The application provides a range of interactive tools specifically designed for segmentation and volumetric analysis of findings. The integrated reporting helps the user to track findings and note changes, such as shape or size, over time.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off Office of In Vitro Diaghostic Device Evaluation and Safety

510(k)

Visia™ Oncology 510(k) Indications for Use Statement - 2/14/2012 Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K120484](https://fda-staging.innolitics.com/device/K120484)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).