K120466 · Nanosphere, Inc. · NTI · Nov 6, 2012 · Clinical Toxicology
Device Facts
Record ID
K120466
Device Name
VERIGENE CYP2C19 NUCLEIC ACID TES (2C19)
Applicant
Nanosphere, Inc.
Product Code
NTI · Clinical Toxicology
Decision Date
Nov 6, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3360
Device Class
Class 2
Indications for Use
The Verigene® CYP2C19 Nucleic Acid Test (CYP2C19 Test), performed using the sample-to-result Verigene System, is a qualitative multiplexed in vitro diagnostic test for the simultaneous detection and identification of an individual's CYP450 2C19 genotype in genomic deoxyribonucleic acid (DNA) obtained from EDTA-anticoagulated whole blood samples. The Verigene CYP2C19 Nucleic Acid Test (CYP2C19 Test) is indicated for use in clinical laboratories upon prescription by the attending physician as an aid to clinicians in determining therapeutic strategy for therapeutics that are metabolized by the CYP450 2C19 gene product, specifically *2, *3, and *17. The Verigene CYP2C19 Nucleic Acid Test (CYP2C19 Test) is not indicated to be used to predict drug response or non-response.
Device Story
Verigene System is a bench-top molecular diagnostics workstation for automated genotyping. Input: EDTA-anticoagulated whole blood. Process: Automated DNA extraction using magnetic glass particles; genomic DNA fragmented/denatured; hybridized to capture oligonucleotides on glass slide; mediator oligonucleotides and gold nanoparticle probes added; optical detection of nanoparticle probes on microarray. Output: Genotype report listing detected alleles. Used in clinical laboratories by technicians/lab personnel. Healthcare providers use results to inform therapeutic strategies for drugs metabolized by CYP2C19 gene products. Benefits: Provides rapid, automated, multiplexed genotyping to assist clinical decision-making.
Clinical Evidence
Performance was validated via analytical studies and a method comparison. Precision/reproducibility studies (n=384 and n=480) showed 100% and 99.6% final agreement with bi-directional sequencing (BDS), respectively. A method comparison study (n=670) across 3 sites demonstrated 99.6% final agreement with BDS. Analytical specificity testing confirmed no interference from endogenous substances (albumin, bilirubin, triglycerides, cholesterol). No sample carry-over was observed. The study established a recommended blood input volume of 1.0 ± 0.05 mL.
Technological Characteristics
System includes disposable multi-chambered fluidic test cartridges and Verigene Processor SP/Reader. Uses magnetic glass particle DNA extraction, microarray hybridization, and gold nanoparticle probe optical detection. Standalone workstation. Software-based automated analysis.
Indications for Use
Indicated for clinical laboratory use, upon physician prescription, to identify CYP2C19 genotypes (*2, *3, *17) in EDTA-anticoagulated whole blood. Aids clinicians in determining therapeutic strategies for drugs metabolized by these alleles. Not for predicting drug response or non-response.
Regulatory Classification
Identification
A drug metabolizing enzyme genotyping system is a device intended for use in testing deoxyribonucleic acid (DNA) extracted from clinical samples to identify the presence or absence of human genotypic markers encoding a drug metabolizing enzyme. This device is used as an aid in determining treatment choice and individualizing treatment dose for therapeutics that are metabolized primarily by the specific enzyme about which the system provides genotypic information.
Special Controls
The special control is FDA's guidance document entitled "Class II Special Controls Guidance Document: Drug Metabolizing Enzyme Genotyping System."
*Classification.* Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Drug Metabolizing Enzyme Genotyping Test System.” See § 862.1(d) for the availability of this guidance document.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k120466
B. Purpose for Submission:
New device
C. Measurand:
Genotype of Cytochrome P450 2C19 (CYP2C19) variants
D. Type of Test:
Genotyping microarray
E. Applicant:
Nanosphere, Inc.
F. Proprietary and Established Names:
Verigene® CYP2C19 Nucleic Acid Test (CYP2C19)
G. Regulatory Information:
1. Regulation section:
21 CFR § 862.3360; Drug Metabolizing Enzyme Genotyping Systems
21 CFR § 862.2570; Instrumentation for Clinical Multiplex Test Systems
2. Classification:
Class II
3. Product code:
NTI; Drug metabolizing enzyme genotyping systems
NSU; Instrumentation for Clinical Multiplex Test Systems
4. Panel:
Toxicology (91)
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
See Indications for use below
2. Indication(s) for use:
The Verigene® CYP2C19 Nucleic Acid Test (2C19), performed using the sample-to-result Verigene System, is a qualitative multiplexed in vitro diagnostic test for the simultaneous detection and identification of an individual’s CYP450 2C19 genotype in genomic deoxyribonucleic acid (DNA) obtained from EDTA-anticoagulated whole blood samples. The Verigene CYP2C19 Nucleic Acid Test
{1}
(2C19) is indicated for use in clinical laboratories upon prescription by the attending physician as an aid to clinicians in determining therapeutic strategy for therapeutics that are metabolized by the CYP450 2C19 gene product, specifically *2, *3, and *17. The Verigene CYP2C19 Nucleic Acid Test (2C19) is not indicated to be used to predict drug response or non-response.
3. Special conditions for use statement(s):
For prescription use only
The information provided from this test may supplement decision-making and should only be used in conjunction with routine monitoring by a physician. Because of the variability in the knowledge of clinical utility with specific drugs that are metabolized by CYP2C19, clinicians should use professional judgment in the interpretation of results from this test. Results from this type of assay should not be used in predicting a patient's response to drugs for which the drug metabolizing enzyme activity of the allele, or the drug metabolic pathway, has not been clearly established.
4. Special instrument requirements:
Verigene® System
I. Device Description:
The Verigene CYP2C19 Nucleic Acid Test (2C19) is a molecular assay which detects specific nucleic acid targets in a microarray format. For each of the nucleic acid sequences detected by the 2C19 test, Capture and Mediator oligonucleotides are utilized for gold nanoparticle probe-based endpoint detection. The Capture oligonucleotides bind to a specific portion of the nucleic acid target and are themselves bound onto a substrate in the microarray. The Mediator oligonucleotides bind to a different portion of the same nucleic acid target and also allow binding of a gold nanoparticle probe to a portion of the Mediator oligonucleotide that is complementary to the oligonucleotides attached to the gold nanoparticle probe. Specific silver enhancement of the bound gold nanoparticle probes at the capture sites results in gold-silver aggregates that scatter light with high efficiency. The 2C19 test is performed on the Verigene System, a "sample-to-result", fully automated, bench-top molecular diagnostics workstation. The Verigene System consists of 2 components: the Verigene Reader and the Verigene Processor SP. The 2C19 test utilizes single-use disposable test consumables and a self-contained Verigene Test Cartridge for each sample tested. For the 2C19 test, the Verigene System allows automated nucleic acid extraction from blood specimens and target detection of specific DNA.
The Reader is the Verigene System's user interface, which serves as the central control unit for all aspects of test processing and results generation. The Reader's graphical user interface guides the user through test processing and test results using a barcode scanner. The user inserts the Test Cartridge into the Verigene Processor
2
{2}
SP, which executes the test procedure, automating the steps of (1) sample preparation; i.e.; cell lysis and magnetic bead-based DNA isolation from EDTA-anticoagulated whole blood samples and (2) hybridization; i.e.; detection and identification of DNA in a microarray format using gold nanoparticle probe-based technology.
After test processing is complete, the user removes the Test Cartridge from the Processor SP, removes the reagent pack from the substrate holder, and inserts the substrate holder into the Reader for analysis. Light scatter from the capture spots is imaged by the Reader. Presence of the gold nanoparticle probes at a particular location on the substrate is assessed optically. The relative density of probes adhered at each set of capture locations on the substrate is then translated into a signal intensity which is used to assess the relative level of captured DNA target to determine a genotype.
J. Substantial Equivalence Information:
1. Predicate device name(s): INFINITI CYP2C19 Assay
2. Predicate K number(s): k101683
3
{3}
3. Comparison with predicate:
| Item | Verigene CYP2C19 Test | INFINITI CYP2C19 Assay (k101683) |
| --- | --- | --- |
| Similarities | | |
| Intended Use/Indications for Use | In vitro diagnostic test for the identification of a patient's CYP450 2C19 genotype in genomic DNA obtained from whole blood samples | Same |
| Warnings and Precautions | For use in clinical laboratories upon prescription by the attending physician | Same |
| Contraindication (s) | Assay is not intended to be used to predict drug response or non-response | Same |
| Test Cartridge | Disposable single-use, multi-chambered fluidic cartridge. | Same |
| Quality control | Internal procedural/instrument quality controls; Internal Negative Control, Sample processing control, external positive and negative assay controls | Same |
| Target Mutations | *2, *3, and *17 genotype of CYP 2C19 | Same |
| Type of Test | Genotyping microarray | Same |
| Interference | None observed for bilirubin, triglyceride, cholesterol, and albumin | Same |
| Differences | | |
| Detection Method | Gold/Ag nanoparticle probe detection of DNA on complementary oligo-microarray | Measures fluorescent signals of labeled DNA target hybridized to the microarray |
| Sample Type used by the test system | EDTA-anticoagulated whole blood | Genomic DNA extracted from EDTA-anticoagulated whole blood |
| DNA Extraction Method | Automated DNA extraction with “blood sample to result system” | DNA extraction using method validated by the laboratory |
K. Standard/Guidance Document Referenced (if applicable):
None
L. Test Principle:
The Verigene® System is comprised of test consumables and shared instrumentation. All Verigene tests are formatted in self-contained test specific Verigene Test Cartridges which serve to analyze a nucleic acid sample that is presented to them. Nucleic acids are prepared directly from a whole blood specimen using magnetic glass particles in an extraction tray and input automatically into a Test Cartridge inside the Verigene Processor SP. Test progress is tracked and directed by the
{4}
Verigene Reader instrument, which serves as a central control unit for each Verigene System.
## Test Procedure
Genomic DNA is extracted from the white blood cells in a whole blood specimen, fragmented and denatured. This fragmented, single-stranded genomic DNA hybridizes to complementary sequence-specific DNA oligonucleotides, known as capture oligonucleotides, arrayed on the surface of a substrate (glass slide). A second DNA oligonucleotide is then hybridized to the captured genomic DNA that was captured initially. This oligonucleotide is known as a mediator oligonucleotide containing 2 sequence domains: 1 domain is complementary to the genomic DNA target and a second domain is complementary to a common oligonucleotide attached to a signal generating gold nanoparticle probe. After washing away any DNA not affixed to the captures, the probe is exposed to the captured mediator/target compound where it hybridizes to any captured mediators. Presence of the gold nanoparticle probes at a particular location on the substrate is assessed optically. The relative density of probes at each set of captures on the substrate is then translated into a signal intensity which is used to assess relative levels of captured DNA target to determine a genotype.
## Result Interpretation
A genotype result is calculated using a normalized ratio of the signal of wild type capture locations on the microarray to the mutant capture locations on the microarray.
Ratio = (wild type signal - mutant signal) / (wild type signal + mutant signal) When the ratio is near 1.0, the wild type signal is higher and the genotype result is wild type. When the ratio is near 0, the wild type signal is equivalent to the mutant signal and the genotype result is heterozygous. When the ratio is near -1.0, the mutant signal is higher and the genotype result is mutant.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
The precision study was conducted at 1 testing site (Nanosphere) and the reproducibility study was conducted at 3 sites (2 external and 1 internal sites). All 3 sites tested the same 8 member sample panel consisting of 8 whole blood specimens containing genotypes *1/*1, *1/*2, *2/*2, *2/*17, *1/*3 (2 tested), *1/*17, and *17/*17. Precision and reproducibility results generated by the 2C19 test were compared to results generated by bi-directional sequencing (BDS), confirming the genotype of the panel member before prior to testing.
Intra-laboratory precision was evaluated by testing the 8 member panel in duplicate twice daily by 2 operators for 12 non-consecutive days at 1 testing site, generating a total of 48 replicates per specimen and an overall total of
5
{5}
384 data points. Three lots of test cartridges and 3 lots of extraction trays were used for the study. Of the 384 samples tested, there was 100% agreement for all panel members as compared to BDS. There were 9 “No Calls” in the study for an initial call rate of 97.7% (375/384). A final call rate of 100% (384/384) was achieved upon repeat testing of all 9 samples. There were 2 Pre-Analysis Errors (Pre-AE) (1 tip test error and 1 tip seal error) for a 0.6% error rate. The precision study results are summarized below:
| Genotype | Number Tested | Reps/Sample | Initial Results | | | | | Final Results | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | No. of Genotype Calls | | No Calls | Agreement | 95% One-sided CI LL | No. of Genotype Calls | | No Calls | Agreement | 95% One-sided CI LL |
| | | | Correct | Incorrect | | | | Correct | Incorrect | | | |
| *1/*1 | 1 | 48 | 47 | 0 | 1 | 97.9 | 90.5 | 48 | 0 | 0 | 100 | 94.0 |
| *1/*2 | 1 | 48 | 47 | 0 | 1 | 97.9 | 90.5 | 48 | 0 | 0 | 100 | 94.0 |
| *1/*3 | 1 | 48 | 46 | 0 | 2 | 95.8 | 87.5 | 48 | 0 | 0 | 100 | 94.0 |
| *1/*3 | 1 | 48 | 46 | 0 | 2 | 95.8 | 87.5 | 48 | 0 | 0 | 100 | 94.0 |
| *1/*17 | 1 | 48 | 47 | 0 | 1 | 97.9 | 90.5 | 48 | 0 | 0 | 100 | 94.0 |
| *2/*2 | 1 | 48 | 48 | 0 | 0 | 100 | 94.0 | 48 | 0 | 0 | 100 | 94.0 |
| *2/*17 | 1 | 48 | 47 | 0 | 1 | 97.9 | 90.5 | 48 | 0 | 0 | 100 | 94.0 |
| *17/*1 7 | 1 | 48 | 47 | 0 | 1 | 97.9 | 90.5 | 48 | 0 | 0 | 100 | 94.0 |
| ALL | 8 | 384 | 375 | 0 | 9 | 97.7 | 96.0 | 384 | 0 | 0 | 100 | 99.2 |
Reproducibility was evaluated with the 8 member panel tested in duplicate twice daily by 2 operators for 5 non-consecutive days. The testing regimen generated 60 replicates (20 replicates per site) per specimen for a total of 480 data points. Four lots of test cartridges and 4 lots of extraction trays were used for the study. There were 15 initial "No Calls" in the study for an initial call rate of 96.9% (465/480). A final call rate of 99.6% (478/480) was achieved upon successful repeat testing of 13 samples, where 2 samples did not resolve after re-testing. Of the 480 samples tested, with 2 un-resolved No Call samples, the percent agreement for all panel members compared to BDS was 99.6% (478/480). There were 6 Pre-AEs in the study for a 1.4% error rate. The data is summarized below:
{6}
| Genotype | Number Tested | Reps/Sample | Initial Results | | | | | Final Results | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | No. of Genotype Calls | | No Calls | Agreement | 95% One-sided CI LL | No. of Genotype Calls | | No Calls | Agreement | 95% One-sided CI LL |
| | | | Correct | Incorrect | | | | Correct | Incorrect | | | |
| *1/*1 | 1 | 60 | 59 | 0 | 1 | 98.3 | 92.3 | 60 | 0 | 0 | 100 | 95.1 |
| *1/*2 | 1 | 60 | 59 | 0 | 1 | 98.3 | 92.3 | 60 | 0 | 0 | 100 | 95.1 |
| *1/*3 | 1 | 60 | 56 | 0 | 4 | 93.3 | 85.4 | 60 | 0 | 0 | 100 | 95.1 |
| *1/*3 | 1 | 60 | 57 | 0 | 3 | 95.0 | 87.6 | 59 | 0 | 1 | 98.3 | 92.3 |
| *1/*17 | 1 | 60 | 59 | 0 | 1 | 98.3 | 92.3 | 60 | 0 | 0 | 100 | 95.1 |
| *2/*2 | 1 | 60 | 59 | 0 | 1 | 98.3 | 92.3 | 60 | 0 | 0 | 100 | 95.1 |
| *2/*17 | 1 | 60 | 57 | 0 | 3 | 95.0 | 87.6 | 59 | 0 | 1 | 98.3 | 92.3 |
| *17/*17 | 1 | 60 | 59 | 0 | 1 | 98.3 | 92.3 | 60 | 0 | 0 | 100 | 95.1 |
| ALL | 8 | 480 | 465 | 0 | 15 | 96.9 | 95.2 | 478 | 0 | 2 | 99.6 | 98.7 |
b. Linearity/assay reportable range: Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Controls: Positive or negative DNA assay quality controls are not included as part of the assay. When external quality control is needed, the sponsor recommends that previously-characterized clinical specimens (positive control samples to test each genotype and at least 1 negative control wild type) be used with the assay during instrument installation, test validation, verification of the performance of a new lot of consumables, or when the integrity of storage conditions is in question. The package insert outlines QC options for evaluating (1) DNA extraction and hybridization steps together and (2) DNA extraction and hybridization steps separately. For (1), combined QC steps (DNA extraction and hybridization together), fresh samples of known genotype stored at $2 - 8^{\circ}\mathrm{C}$ be used within 10 days of collection. In the event that insufficient volume of whole blood for certain genotypes cannot be obtained, QC option (2) can be used where the extraction and hybridization control steps are divided. For the extraction control step, whole blood samples (stored at $2 - 8^{\circ}\mathrm{C}$ for up to 10 days after collection) should be used where the following results are obtained: DNA concentration range within $160\mathrm{ng} / \mu \mathrm{L}$ and $400\mathrm{ng} / \mu \mathrm{L}$ and absorbance $\mathrm{A}_{260} / \mathrm{A}_{280}$ ratio range within 1.7-2.2. To QC the hybridization control step, previously characterized genomic DNA (stored at $-20^{\circ}\mathrm{C}$ for up to 9 months) with a concentration range of $160\mathrm{ng} / \mu \mathrm{L}$ to $400\mathrm{ng} / \mu \mathrm{L}$ and absorbance $\mathrm{A}_{260} / \mathrm{A}_{280}$ ratio range from 1.7-2.2 should be used and the test results should match the expected genotype result for the sample.
All quality control requirements and testing should be performed in
{7}
conformance with local, state and/or federal regulations or accreditation requirements.
**Imaging Controls**: A set of oligonucleotide spots on the substrate is used as Imaging Controls. The imaging controls act as a quality check for proper fluid control and movement between the Test Cartridge and the instrument. If the imaging controls are absent, then a “No Call” is an output and the user is informed that the reason is due to an absence of imaging controls. The imaging control signal is due to signal probe hybridization to an oligonucleotide on the substrate. Signal will only be present if the probe reagent and the signal enhancement reagents are pumped into and held in the hybridization chamber for the specified time period. A Negative Control sequence is included as a control for non-specific binding within the test owing to factors including inadequate hybridization stringency, fluidic failures resulting in high non-specific signals, and other user- or instrument-related failures. A set of background spots on the substrate is used as a background control. The user is informed if the “No Call” result is due to the background signal being too high. Each target oligonucleotide and background location has 6 or more separate spots in the array. The array spots are grouped in at least 3 different locations on the substrate surface. Mathematical algorithms, called “spot filters”, compare the “like” array spots as a set. If individual spots in the set differ, then they are discarded. If the total number of spots in a set becomes less than 4 due to the spot filters, then a “No Call” results. The spot comparison algorithms protect against some hybridization failures; they also protect against fluidics failures such as the hybridization chamber being partially filled.
**Reagent Stability**: Stability testing of the CYP2C19 test involved taking individual test kit units arranged in the final packaging configuration and placing them in qualified stability chambers in a temperature controlled environment and testing at defined time intervals. For shipping stability, a subset of 2C19 test kit lots were shipped in their intended outer packaging to several sites and returned to Nanosphere and tested for performance. Successful product testing for a minimum of twelve weeks nominal exposure is required prior to marketing of the product.
**Sample Stability**: The sponsor has demonstrated that specimens (EDTA anti-coagulated whole blood samples) can be kept refrigerated (2 to 8°C) 10 days after specimen collection and used with the CYP2C19 test. Previously characterized genomic DNA (160 ng/μL to 400 ng/μL and A₂₆₀/A₂₈₀ from 1.7-2.2) can be stored at -20°C for up to 9 months and used with the CYP2C19 test.
**d. Detection limit**:
The upper and lower limits of detection of the assay were assessed by analysis of 7 whole blood samples (*1/*1, *1/*2, *1/*3, *1/*17, *2/*2, *2/*17 and
8
{8}
*17/*17 genotypes) with 5 input volumes (between $800~\mu \mathrm{L}$ and $1200~\mu \mathrm{L}$ in $100~\mu \mathrm{L}$ increments in replicates of 40) to determine the lowest and highest input volumes that would consistently give an initial call rate of $\geq 90\%$ and $100\%$ accuracy (vs. BDS). The blood samples were then tested with the assay over a period of 4 days by 2 different users. No repeat testing was performed for samples yielding a "no call" result. A total of 1,400 tests were completed. No incorrect calls were observed for the 1,348 valid test results generated (i.e., $>4\%$ no-calls; $n = 52$ ). Overall, the CYP2C19 test detected each genotype consistently ( $\geq 90.0\%$ ) and accurately ( $100\%$ vs. BDS) for all genotypes tested between the specimen volume ranges of $1000 \pm 200\mu \mathrm{L}$ . There were no incorrect calls at any of the input levels (data summarized below). The Sponsor states that a significant degradation in test performance was not observed when testing the blood input volumes at either end of the $20\%$ range around the recommended input value of $1\mathrm{ml}$ (i.e., $1000 \pm 200~\mu \mathrm{L}$ ). Therefore, $1.0 \pm 0.05\mathrm{mL}$ of whole blood is the recommended sample volume tolerance for the assay. Furthermore, an additional study evaluating the impact of test performance at blood input volumes of 0, 500 and $750~\mu \mathrm{L}$ (in replicates of 40 for the $*1/*2$ specimen) demonstrated initial call rates of 0, 67.5 (27/40) and $97.5\%$ (39/40) for all 3 additional input volumes tested, respectively. Initial call rate for blood input volumes greater than $800~\mu \mathrm{L}$ are summarized in the table below:
| Blood Input (μL) | | Initial Call Rate and 95% Confidence Interval(1) for Each Genotypea | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | *1/*1 | *1/*2 | *1/*3 | *1/*17 | *2/*2 | *2/*17 | *17/*17 | All |
| 800 | 97.5% (39/40) | 95.0% (38/40) | 100% (40/40) | 97.5% (39/40) | 100% (40/40) | 97.5% (39/40) | 92.5% (37/40) | 97.1% (272/280) |
| | 86.8-99.9 | 83.1-99.4 | 91.2-100 | 86.8-99.9 | 91.2-100 | 86.8-99.9 | 79.6-98.4 | 94.5-98.8 |
| 900 | 100% (40/40) | 95.0% (38/40) | 95.0% (38/40) | 95.0% (38/40) | 100% (40/40) | 95.0% (38/40) | 97.5% (39/40) | 96.8% (271/280) |
| | 91.2-100 | 83.1-99.4 | 83.1-99.4 | 83.1-99.4 | 91.2-100 | 83.1-99.4 | 86.8-99.9 | 94.0-98.5 |
| 1000 | 97.5% (39/40) | 100% (40/40) | 97.5% (39/40) | 95.0% (38/40) | 100% (40/40) | 97.5% (39/40) | 100% (40/40) | 98.2% (275/280) |
| | 86.8-99.9 | 91.2-100 | 86.8-99.9 | 83.1-99.4 | 91.2-100 | 86.8-99.9 | 91.2-100 | 95.9-99.4 |
| 1100 | 95.0% (38/40) | 97.5% (39/40) | 92.5% (37/40) | 97.5% (39/40) | 92.5% (37/40) | 92.5% (37/40) | 95.0% (38/40) | 94.6% (265/280) |
| | 83.1- | 86.8- | 79.6- | 86.8- | 79.6- | 79.6- | 83.1- | 91.3-97.0 |
{9}
| | 99.4 | 99.9 | 98.4 | 99.9 | 98.4 | 98.4 | 99.4 | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| 1200 | 95.0%
(38/40)
83.1-
99.4 | 92.5%
(37/40)
79.6-
98.4 | 97.5%
(39/40)
86.8-
99.9 | 97.5%
(39/40)
86.8-
99.9 | 90.0%
(36/40)
75.1-
99.8 | 92.5%
(37/40)
79.6-
98.4 | 97.5%
(39/40)
86.8-
99.9 | 94.6%
(265/280)
91.3-97.0 |
| Total By
Genotype | 97.0%
(194/20
0)
93.6-
98.9 | 96.0%
(192/20
0)
92.3-
98.3 | 96.5%
(193/20
0)
92.9-
98.6 | 96.5%
(193/20
0)
92.9-
98.6 | 96.5%
(193/20
0)
92.9-
98.6 | 95.0%
(190/20
0)
91.0-
97.6 | 96.5%
(193/20
0)
92.9-
98.6 | 96.3%
(1348/14
00)
95.2-97.2 |
1) %CI values were computed using the SAS 9.1.3, exact binomial method.
a) Genotype determined by bi-directional sequencing
The Sponsor has noted in their labeling that the ability of the CYP2C19 test to consistently detect the genotype of a specimen will be negatively impacted at whole blood sample input volumes less than 750 µL, based on the call rates in the study results cited above.
e. Analytical specificity:
Interference from potential interfering substances was evaluated to assess the genotyping assay performance of the CYP2C19 test in the presence of excess amounts of endogenous substances. Seven EDTA anticoagulated whole blood control samples (each containing individually containing genotypes *1/*1, *1/*2, *1/*3, *1/*17, *2/*2, *2/*17, and *17/*17) were processed in replicates of 30 using 1.0 mL of blood for each replicate. These samples were used as the base matrix into which the 5 interfering substances were spiked and compared to unspiked control samples. The following compounds at the following concentrations were tested: albumin (6000 mg/dL), bilirubin (conjugated and unconjugated, 20 mg/dL), triglycerides (3000 mg/dL) and cholesterol (500 mg/dL). The concentrations of these endogenous substances were at higher levels for each of the substances than is normally observed in human blood. Sample genotypes were verified by BDS.
In the presence of these substances, each genotype was detected consistently with a final call rate of 100% (except for *1/*3 in the presence of albumin, *2/*17 in the presence of conjugated bilirubin and *17/*17 in the presence of cholesterol that yielded final call rates of 29/30) and 100% accuracy vs. BDS. The sponsor concluded that the compounds do not interfere with the performance of the assay.
Sample carry-over:
To evaluate possible cross-contamination and carry-over between test runs,
{10}
samples of alternating genotypes were tested consecutively on the same device. Demonstration of a $\geq 90\%$ no call rate and $\geq 99\%$ concordance with BDS was the criteria by which to establish that there is no sample carry-over and cross-over contamination between procedures on the Verigene Processor $SP$ using different sample cartridges. Whole blood with $*1/*2$ genotype was tested on 10 Verigene instruments, followed immediately by whole blood with $*1/*1$ genotype. Next, whole blood with $*1/*17$ genotype was tested on the same 10 instruments, followed by whole blood with $*1/*1$ genotype. This was repeated 3 times, such that 30 replicates were generated for $*1/*2$, 30 replicates for $*1/*17$, and 60 replicates for $*1/*1$ on a total of 10 instruments. A total of 120 test runs were generated. All samples were concordance with BDS and there were a total of 4 "no calls" during the study as summarized below:
| Genotype | Total # Replicates | Calls | No Calls | Total Call Rate (%) |
| --- | --- | --- | --- | --- |
| *1/*1 | 60 | 58 | 2 | 96.7 |
| *1/*2 | 30 | 30 | 0 | 100 |
| *1/*17 | 30 | 28 | 2 | 93.3 |
The sponsor concluded that the no calls observed in the study reflected the typical performance of the CYP2C19 assay and that no sample carry-over was illustrated between the samples.
f. Assay cut-off: Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Method comparison studies were conducted at 3 external clinical study sites using according to CLSI EP9-A2 guidelines. Each site utilized 1 Verigene Reader and up to 12 Verigene Processor SP instruments, depending on each site's specimen enrollment and testing volume. Clinical samples, consisting of EDTA anti-coagulated whole blood samples from the general population of known Asian donors, were collected by 2 third party facilities and shipped to each study site. Specimen tests resulting in a "No Call" or a pre-analysis error (pre-AE) were repeat tested once. Residual DNA from each clinical specimen after testing was sequenced using BDS. A total of 670 samples (125-281 samples per site) were tested. The table below shows the number of samples tested, the initial and final call rates at each of the sites and the combined statistics for each. There were 35 initial no calls of which 34 were successfully repeated. The initial call rate was $94.8\%$ (635/670) and the final call rate was $99.9\%$ (669/670), as summarized below:
{11}
12
| Genotype (a) | No. Tested | Reps per Sample | No. Correct Genotype Calls | No. Incorrect Calls | No Calls | Agreement | 95% Two-sided Confidence Interval Lower Limit |
| --- | --- | --- | --- | --- | --- | --- | --- |
| *1 / *1 (b) | 260 | 1 | 260 | 0 | 0 | 100% | 98.6% |
| *1 / *2 | 177 | 1 | 176 | 0 | 1(c) | 99.4% | 96.9% |
| *1 / *3 | 24 | 1 | 23 | 1 | 0 | 95.8% | 78.9% |
| *1 / *17 | 114 | 1 | 113 | 1 | 0 | 99.1% | 95.2% |
| *2 / *2 | 32 | 1 | 32 | 0 | 0 | 100% | 89.1% |
| *2 / *3 | 13 | 1 | 13 | 0 | 0 | 100% | 75.3% |
| *2 / *17 | 30 | 1 | 30 | 0 | 0 | 100% | 88.4% |
| *3 / *17 | 1 | 1 | 1 | 0 | 0 | 100% | 2.5% |
| *3 / *3 | 1 | 1 | 1 | 0 | 0 | 100% | 2.5% |
| *17 / *17 | 18 | 1 | 18 | 0 | 0 | 100% | 81.5% |
| Total | 670 | 1 | 667 | 2(c) | 1 | 99.6% | 98.7% |
a) Genotype determined by bi-directional sequencing
b) *1/*1 samples are inferred if they are wild-type for *2, *3 and *17
c) CYP2C19 result was no-call for both initial and final test runs; BDS result *1/*2.
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
The following table lists the alleles recognized by the device, the single nucleotide polymorphisms (SNPs) recognized by the device for each allele, enzyme activity, and references:
| Allele | Recognized Polymorphism | Polymorphisms Associated with Allele^{a} | Enzyme Activity | References |
| --- | --- | --- | --- | --- |
| *2^{(b)} | 19154G>A | -98T>C; 99C>T; 12122G>A; 12460G>C; 12662A>G; 12834G>C; 19154G>A; 19520A>G; 57740C>G; 79936T>A; 80160C>T; | None | de Morais et al, 1994^{2}
Ibeanu et al, 1998^{3}
Fukushima- |
{12}
| | | 80161A>G; 87275G>A | | Useaka et al, 2005^{4} |
| --- | --- | --- | --- | --- |
| *3 | 17948G>A | -889T>G; 12031T>G; 12122G>A; 12306G>A; 13166T>C; 17948G>A; 18911A>G; 80161A>G; 80248G>A; 87313A>C | None | de Morais et al, 1994^{2} Ibeanu et al, 1998^{3} Fukushima-Useaka et al, 2005^{4} |
| *17 | -806C>T | -3402C>T; -806C>T; 99C>T; 80161A>G | Increased | Sim et al, 2006^{5} Rudberg et al, 2008^{1} |
a) SNPs in **bold** are the major SNPs/alterations responsible for the phenotype of the corresponding allele and are unique to the mutation
b) The *10 mutation is adjacent to the *2 mutation, therefore the CYP2C19 Test contains two capture probes for *10. These probes are used to ensure the accuracy of *2 allele detection; however, the *10 mutations are not reported by the assay.
1 Rudberg, I; et al. Impact of the ultrarapid CYP2C19*17 allele on serum concentration of escitalopram in psychiatric patients. Clin Pharmacol Ther. 2008 Feb;83(2):322-7
2 de Morais SM et al. The major genetic defect responsible for the polymorphism of S-mephenytoin metabolism in humans. J Biol Chem. 1994 Jun 3; 269(22):15419-22
3 Ibeanu GC et al. Identification of new human CYP2C19 alleles(CYP2C19*6 and CYP2C19*2B) in a Caucasian poor metabolizer of mephenytoin. J Pharmacol Exp Ther. 1998 Sep;286(3):1490-5.
4 Fukushima-Uesaka H et al. Genetic variations and haplotypes of CYP2C19 in a Japanese population. Drug Metab Pharmacokinet. 2005 Aug;20(4):300-7
5 Sim SC et al. A common novel CYP2C19 gene variant causes ultra-rapid drug metabolism relevant for the drug response to proton pump inhibitors and anti depressants. Clin Pharmacol Ther. 2006; 79:103-113
4. Clinical cut-off: Not applicable
5. Expected values/Reference range:
The package insert states:
CYP2C19*2 arises from a G→A transition in exon 5 at position 681 of CYP2C19, creating an aberrant splice site. The CYP2C19 *2 polymorphism is seen in approximately 15% of Caucasians, 15% of African-Americans, and 30% of Asians. CYP2C19*3 arises from a G→A transition at position 636 in exon 4 of CYP2C19, which produces a truncated protein. The CYP2C19 *3 polymorphism is seen in approximately <1% of Caucasians, <1% of African-Americans, and 8%
{13}
of Asians¹. A poor metabolizer phenotype (caused by 2 non-functional CYP2C19 alleles is present in 4% of Caucasians, 5% of African-Americans, and up to 25% of Asians¹.
The CYP2C19*17 polymorphism is seen in approximately 18% of Swedes, 18% of Ethiopians, and more than 4% of Chinese².
¹ARUP Laboratories. Test Sheet: Cytochrome P450 2C19 (CYP2C19) 10 Mutations – for Detection of CYP2C19 Mutations Affecting Drug Metabolism. August 2010.
²Rudberg, I; et al. Impact of the ultrarapid CYP2C19*17 allele on serum concentration of escitalopram in psychiatric patients. Clin Pharmacol Ther. 2008 Feb;83(2):322-7.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.