← Product Code [MLO](/productcode/MLO) · K120462

# AUDICOR CPAM WITH SDB (CARDIOPULMONARY AMBULATORY MONITOR WITH SLEEP DISORDERED BREATHING DETECTION) (K120462)

_Inovise Medical, Inc. · MLO · Sep 20, 2012 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K120462

## Device Facts

- **Applicant:** Inovise Medical, Inc.
- **Product Code:** [MLO](/productcode/MLO.md)
- **Decision Date:** Sep 20, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2800
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The AUDICOR System when used with AUDICOR Sensors on the chest wall and properly attached Holter unit, is intended for use on adults 18 years of age and older in acquiring, analyzing and reporting ECG, heart sound, sleep disordered breathing, snoring detection and activity level data and to provide interpretation of the data in an integrated report for consideration by physicians. The sleep disordered breathing analysis and reporting is intended for use on adult patients only as a screening device to determine the need for evaluation by polysomnography based on the patient's score. Subjects screened for sleep disordered breathing should have periods of sleep of at least 4 hours duration during which the ECG is predominantly sinus rhythm in nature. The AUDICOR recording may be obtained at any location specified by a physician including home, hospital or clinic. The device is intended for use only under the direct supervision of a physician.

## Device Story

Audicor CPAM with SDB is a wearable, battery-powered ambulatory monitor; captures ECG, heart sounds, snoring, and activity level via chest-wall sensors. Device records up to 48 hours of continuous data or 10-second snapshots. Data downloaded to analysis module; software trends parameters, detects events (bradycardia, tachycardia, atrial fibrillation, apnea/hypopnea), and generates integrated reports. Clinician reviews detected events and measurements; modifies interpretative statements. Used in home, hospital, or clinic under physician supervision. Benefits include automated screening for sleep-disordered breathing and cardiac rhythm monitoring to assist clinical decision-making for patients with suspected heart failure or acute coronary syndrome.

## Clinical Evidence

Bench testing only. Compliance with EN 60601-1 (safety), EN 60601-1-2 (EMC), ANSI/AAMI EC38 (ambulatory ECG), and ANSI/AAMI EC57 (cardiac rhythm/ST-segment measurement algorithms) demonstrated.

## Technological Characteristics

Wearable, battery-powered device. Sensors: chest-wall ECG and acoustic sensors, accelerometer. Connectivity: non-volatile memory for data download. Standards: ANSI/AAMI EC38, ANSI/AAMI EC57, EN 60601-1, EN 60601-1-2.

## Regulatory Identification

A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.

## Predicate Devices

- Audicor CPAM ([K110569](/device/K110569.md))
- Spacelabs Healthcare Pathfinder SL Holter Analyzer ([K110001](/device/K110001.md))
- Nox Medical Nox T3 Sleep Recorder ([K082113](/device/K082113.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K120462

17 of 124

| Inovise Medical, Inc. |
|-----------------------|
|-----------------------|

510(k) Notification: Modification to AUDICOR System - CPAM with SDB

டு/OKSummary of Safety and Effectiveness As required by 21 CFR, part 807 92 Submitted By: Inovise Medical, Inc. 8770 SW Nimbus Ave. Suite D Portland, OR 97008-7196 Phone 503-313-0149 Fax 503-431-3801 Contact: Earl Anderson Director, Quality and Regulatory Date Prepared: February, 08 2012 Proprietary Modification to AUDICOR CPAM (Cardiopulmonary Ambulatory Monitor) Name: Common/ Usual Ambulatory Monitor / Acoustic Cardiograph Name: Classification: 870.2800, MLO, class II, Electrocardiograph, Ambulatory (with analysis program) 868.2375, MNR, Ventilatory Effort Recorder Performance AAMI EC38 and AAMI EC57 Standards: The AUDICOR System when used with AUDICOR Sensors on the chest wall and Intended Use: properly attached Holter unit, is intended for use on adults 18 years of age and older in acquiring, analyzing and reporting ECG, heart sound, sleep disordered breathing, snoring detection and activity level data and to provide interpretation of the data in an integrated report for consideration by physicians. The sleep disordered breathing analysis and reporting is intended for use on adult patients only as a screening device to determine the need for evaluation by polysomnography based on the patient's score. Subjects screened for sleep disordered breathing should have periods of sleep of at least 4 hours duration during which the ECG is predominantly sinus rhythm in nature. The AUDICOR recording may be obtained at any location specified by a physician including home, hospital or clinic. The device is intended for use only under the direct supervision of a physician. The Audicor CPAM with SDB (Cardiopulmonary Holter with Sleep Disordered Device Description: Breathing detection) is an ambulatory device that can be used to capture 10second snapshots or up to 48 hours of continuous data from ECG, heart sounds, sleep disordered breathing and snoring detection, and activity level, particularly in patients suspected of heart failure or acute coronary syndrome. The Audicor CPAM with SDB includes software to display, analyze and provide a summary of patient data over time in a trended format. Notable events are detected and displayed for review by the clinician. The Audicor System analyzes and reports the following parameters: Heart rate including bradycardia and tachycardia events . Atrial fibrillation . ECG beat classification and morphology grouping with user-editing . . Heart rate variability . Snoring detection Sleep disordered breathing (apnea/hypopnea) events .

{1}------------------------------------------------

- . Sleep disordered breathing score
- . Activity level
- Heart sound and combined ECG/heart sound measurements
- Heart rate distributions of heart sound parameters ●

The clinician can review automatically detected events and measurements, and modify them as well as modify the interpretative statements generated.

## Test Summary & Conclusion:

The Audicor System has been tested to the applicable requirements of the following standards, and shown to comply.

- EN 60601-1 Standard for Medical Electrical Equipment: General Requirements . for Safety
- . EN 60601-1-2 Standard for Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -Requirements and Tests
- ANSI/AAMI EC38:Ambulatory Electrocardiographs ●
- ANSI/AAMI EC57: Test and Reporting Performance Results of Cardiac . Rhythm and ST-segment Measurement Algorithms

The added algorithms for reporting of sleep disordered breathing, snoring and Substantial Equivalence: activity are substantially equivalent to:

- 1. Audicor CPAM (510(k) K110569), and
- 2. Spacelabs Healthcare Pathfinder SL Holter Analyzer (510(k) K110001)
- Nox Medical Nox T3 Sleep Recorder (510(k) K082113)

The Audicor System and the predicate devices are technologically equivalent in Technological that the devices: Characteristics:

- Are small, wearable, battery-powered devices .
- Acquire ECG data from patients using limited ECG leads ●
- Analyze the ECG signal to detect periods of sleep disordered breathing .
- . Acquire a sound signal for detection of snoring
- . Analyze the signals from an accelerometer to derive activity level
- Use non-volatile memory to store patient data for download to the analysis . module of the system

The Audicor System is different from the predicates in the following ways:

- Does not acquire EEG, EMG or EOG signals, or nasal pressure .

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized representation of a human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem. The emblem is positioned to the right of the text.

## Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Inovise Medical, Incorporated Mr. Earl Anderson Director, Quality and Regulatory 8770 SW Nimbus Avenue, Suite D Beaverton, Oregon 97008

SEP 20 2012

Re: K120462

Trade/Device Name: Audicor® CPAM with SDB (Cardiopulmonary Ambulatory Monitor with Sleep Disordered Breathing) Regulation Number: 21 CFR 870.2800

21 CFR 868.2375

Regulation Name: Medical Magnetic Tape Recorder, Breathing Frequency Monitor Regulatory Class: II Product Code: MLO, MNR Dated: September 6, 2012

Received: September 10, 2012

Dear Mr. Anderson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

## Page 2- Mr. Anderson

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Wh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known):

Device Name: Audicor System (CPAM with SDB)

Indications For Use:

The AUDICOR System when used with AUDICOR Sensors on the chest wall and properly attached Holter unit, is intended for use on adults 18 years of age and older in acquiring, analyzing and reporting ECG, heart sound, sleep disordered breathing and snoring detection and to provide interpretation of the data in an integrated report for consideration by physicians. The sleep disordered breathing analysis and reporting is intended for use on adult patients only as a screening device for obstructive or mixed apnea to determine the need for evaluation by polysomnography based on the patient's score.

Subjects screened for sleep disordered breathing should have periods of sleep of at least 4 hours duration during which the ECG is predominantly sinus rhythm in nature. The AUDICOR recording may be obtained at any location specified by a physician including home, hospital or clinic.

The device is intended for use only under the direct supervision of a physician.

Prescription Use XXX (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

L. Schuler

Division Slon Off Division of Anesthesiology, General Hospital Infection Control. Dental Dev

510(k) Number: K120462

---

**Source:** [https://fda-staging.innolitics.com/device/K120462](https://fda-staging.innolitics.com/device/K120462)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
