K120420 · Shanghai Double Dove Industrial Co., Ltd. · FMI · Oct 18, 2012 · General Hospital
Device Facts
Record ID
K120420
Device Name
STERILE BLUNT FOR SINGLE USE
Applicant
Shanghai Double Dove Industrial Co., Ltd.
Product Code
FMI · General Hospital
Decision Date
Oct 18, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
The Sterile Blunt for Single Use includes two models, which are blunt fill needle and blunt plastic cannula. The blunt fill needle is used in conjunction with a syringe to penetrate the vial/ampoules stopper intended for injection or aspiration of fluid. The blunt plastic cannula is used in conjunction with a syringe to access the pre-slit septum covering injection sites on I.V. System intended for injection or aspiration of fluid.
Device Story
Sterile Blunt for Single Use consists of two models: blunt fill needle (stainless steel tube, hub, sheath) and blunt plastic cannula (molded plastic blunt tip, hub, sheath). Devices attach to syringes to facilitate fluid aspiration or injection. Blunt fill needle penetrates vial/ampoule stoppers; blunt plastic cannula accesses pre-slit septa on I.V. systems. Designed to replace sharp hypodermic needles to prevent accidental needle-stick injuries. Used in clinical settings by healthcare providers. Sterilized via EO or radiation to SAL 10^-6. Benefits include reduced risk of sharps injury during fluid handling.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility (cytotoxicity, irritation, sensitization, systemic toxicity), sterility validation (ISO 11135-1, ISO 11137-2), package integrity (ASTM F1140-07, ASTM F88-09), and physical performance (ISO 9626, ISO 7864, ISO 594-1).
Technological Characteristics
Materials: Stainless steel (needle), molded plastic (cannula). Connectivity: None. Energy: None. Sterilization: EO or Radiation (SAL 10^-6). Standards: ISO 9626, ISO 7864, ISO 594-1, ISO 10993 series, ISO 11135-1, ISO 11137-2, USP 34, ASTM F1140-07, ASTM F88-09.
Indications for Use
Indicated for use with a syringe for fluid injection or aspiration. Blunt fill needle penetrates vial/ampoule stoppers; blunt plastic cannula accesses pre-slit septa on I.V. systems. Intended for single use.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
OCT 1 8 2012
# Attachment III 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
KIZOYNO The assigned 510(k) Number:
- Date of Submission: 17 OCT 2012 1 .
- 2. Sponsor
Shanghai Double-Dove Industry Co, Ltd. No.1888 Huhang Road, Fengxian Economic Zone Shanghai, 201400, China
Establishment Registration No .: 3004416208
Contact Person Mr. Sanba Yang Vice General Manager Tel: +86-21-67104888 Fax: +86-21-67104666 Email: Yangsbmaster@hotmail.com
- 3. Submission Correspondent Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 237-023, Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net
### 4. Proposed Device Identification
Proposed Device Name: Sterile Blunt for Single Use Proposed Device Model: Blunt Fill Needle and Blunt Plastic Cannula
Classification: II Product Code: FMI Regulation Number: 21 CFR 880.5570 Review Panel: General Hospital
l l l - 1
ﻴﻨﻴ
{1}------------------------------------------------
Intended Use Statement:
The Sterile Blunt for Single Use includes two models, which are blunt fill needle and blunt plastic cannula.
The blunt fill needle is used in conjunction with a syringe to penetrate the vial/ampoules stopper intended for injection or aspiration of fluid.
The blunt plastic cannula is used in conjunction with a syringe to access the pre-slit septum covering injection sites on I.V. System intended for injection or aspiration of fluid.
5. Predicate Device Identification
> 510(k) Number:K 102328 Product Name: Blunts Manufacturer: Shan Dong Wei Gao Group Medical Polymer Co., LTD
### 6. Device Description
The proposed device, Sterile Blunt for Single Use, includes two models, which are blunt fill needle and blunt plastic cannula.
The blunt fill needle consists of a stainless steel needle tube, a hub and a sheath. It is used to penetrate the vial and/or ampoules stopper intended for injection or drawl fluid. It is not intended to access the injection site on I.V. System.
The blunt plastic cannula consists of molded plastic blunt tipped cannula, a hub and a sheath. It is used to replace hypodermic needles to prevent from the unintended injury by metal needle.
Both two models are available in either EO sterilized to achieve a Sterilized to achieve a Sterility Assurance Level (SAL) of 106. The proposed device is sealed in a sterility maintenance package to maintain its sterility during its shelf life of three (3) years.
### 7. Non-Clinical Test Conclusion
Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
ISO 9626:1991/Amendment 1:2001, Stainless steel needle tubing for the manufacture of medical devices.
{2}------------------------------------------------
ISO 7864:1993, Sterile hypodermic needles for single use.
ISO 594-1:1986, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirements.
ISO 10993-1: 2009, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.
ISO 10993- 4:2002/And 1:2006, Biological Evaluation of Medical Device. Part 4: Selection of rest for Interactions with Blood.
ISO 10993-5: 2009, Biological evaluation of medical devices - Part 5: Tests for In Viro cytotoxicity. 2-153
ISO 10993-7:2008, Biological evaluation of medical devices -Part 7: Ethylene oxide sterilization residuals
ISO 10993-10: 2002/Amd. 1:2006(E), Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity AMENDMENT I
ISO 10993-11: 2006. Biological evaluation of medical devices -- Part H : Tests for systemic toxicity. ISO 11135-1: 2007 Sterilization of health care products -- Ethylene oxide -- Part 1: Requirements Tor development, validation and routine control of a sterifization process for medical devices.
USP 34:2011, -: 85> Bacterial Endotoxins Test
ASTM F1140-07, Standard Test Methods for Internal Pressurization Failure Resisunce of Unrestrained Packages.
ASTM F88-09, Standard Test Method for Seal Strength of Flexible Barrier Materials.
ISO 11137-2: 2006. Sterifization of healtheare products-Radistion-Part2: Establishing ប្រើ sterifization dose.
#### Substantially Equivalent Conclusion 8.
The proposed device, Sterile Blunt for Single Use, has same intended use, configuration, operation mode with those of the predicate device, Blunts (K 102328). The materials are different, the biocompatibility test and leachable substance analysis demonstrated that the materails used for manufacturing of proposed dveice are safe and accpetable. The bevel of the proposed device is identical to that of the predicate device, while the gauges of proposed and predicate device is different. The physical performance test results submitted demonstrated the proposed device complied with the referenced standards with different gauge of the predicate device. The proposed device is available in EO sterilization or Radiation sterilization, the predicate device is only avaiable in radiation sterilization. Various tests, including biocompatibility, perfromance, package, endotoxin tests demonstrated that the both EO sterilization sterilization sterilization proposed device complied with acceptance criteria. In addition, both of the proposed and predicate device could achieve same Sterility Assurance Level (SAL)
Therefore, the proposed device, Sterile Blunt for Single Use, is determined to be Substantially Equivalent (SE) to the predicate device, Blunts (K102328), in respect of safety and effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is in all capital letters and is evenly spaced around the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Shanghai Double-Dove Industry Company, Limited C/O Ms. Diana Hong General Manager Mid-Link Consulting Company, Limited P.O. Box 237-023 Shanghai, CHINA 200237
OCT 1 8 2012
Re: K120420
Trade/Device Name: Sterile Blunt for Single Use Model: Blunt Fill Needle, Blunt Plastic Cannula Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: August 30, 2012 Received: September 7, 2012
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{4}------------------------------------------------
## Page 2 - Ms. Hong
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthem um
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
arket Notification 510(k) Submission
K 12O H2O
# Section II Indications for Use
510(k) Number:
Device Name: Sterile Blunt for Single Use Model: Blunt Fill Needle, Blunt Plastic Cannula
Indications for Use:
Sterile Blunt for Single Use includes two models, which are blunt fill needle and blunt plastic cannula.
The blunt fill needle is used in conjunction with a syringe to penetrate the vial/ampoules stopper intended for injection or aspiration of fluid.
The blunt plastic cannula is used in conjunction with a syringe to access the pre-slit septum covering injection sites on I.V. System intended for injection or aspiration of fluid.
OPRESCRIPTION USE (Part 21 CFR 801 Subpart D) ■OVER-THE-COUNTER USE (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
RC Chap 10/12/12 Page 1 of 1
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesional Devices. Division of AneSthoolors.
Infection Control, Dental Devices.
510(k) Number: K120920
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.