K120415 · Myoscience, Inc. · GXH · Jun 22, 2012 · Neurology
Device Facts
Record ID
K120415
Device Name
CRYO-TOUCH III
Applicant
Myoscience, Inc.
Product Code
GXH · Neurology
Decision Date
Jun 22, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Myoscience Cryo-Touch III is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. The Cryo-Touch III is not indicated for treatment of central nervous system tissue.
Device Story
Portable cryogenic surgical device; utilizes liquid nitrous oxide (N2O) to cool a needle probe 'Tip' via Joule-Thomson effect and/or Latent Heat of Vaporization. Used to destroy tissue or create peripheral nerve lesions for pain management. Comprised of re-usable control unit, re-usable handpiece, and single-patient-use tip. May be used with standard off-the-shelf nerve stimulators for precise nerve localization. Operated by clinicians in surgical settings. Output is localized extreme cold applied to target tissue. Benefits include precise, controlled tissue destruction or nerve block.
Clinical Evidence
No clinical data. Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, ISO 10993-1, ISO 11137-1, ISO 11607, and ASTM F882-84.
Technological Characteristics
Portable cryogenic surgical unit. Components: re-usable control unit, re-usable handpiece, single-use tip. Cooling mechanism: Joule-Thomson effect/Latent Heat of Vaporization using N2O. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10993-1, ISO 11137-1, ISO 11607, ASTM F882-84. Sterilization: ISO 11137-1 compliant.
Indications for Use
Indicated for surgical tissue destruction and peripheral nerve lesioning for pain blocking. Contraindicated for central nervous system tissue.
Regulatory Classification
Identification
A cryogenic surgical device is a device used to destroy nervous tissue or produce lesions in nervous tissue by the application of extreme cold to the selected site.
{0}------------------------------------------------
## myoscience
# Cryo-Touch III
### Premarket Notification
### Section 5: 510(k) Summary
#### Device Information:
| Category | Comments |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor / Submitter: | Myoscience, Inc<br>525 Chesapeake Drive<br>Redwood City, CA 94063<br>(650) 474-2600<br>(650) 474-2700 |
| Correspondent Contact<br>Information: | Tracey Henry<br>Vice President RAQA, Clinical Compliance<br>525 Chesapeake Drive<br>Redwood City, CA 94063<br>(650) 474-2600<br>(650) 474-2900 |
| Device Common Name: | Cryogenic Surgical device |
| Device Classification & Code: | Class II, GXH |
| Device Classification Name: | Cryosurgical unit and accessories (21 CFR 882.4250) |
| Device Proprietary Name: | Myoscience Cryo-Touch III |
#### a. Predicate Device Information:
The Cryo-Touch III is substantially equivalent to the following currently legally marketed devices:
| 510(k) Number | Product | Sponsor |
|---------------|---------------|-----------------|
| K083493 | Cryo-Touch | Myoscience, Inc |
| K102021 | Cryo-Touch II | Myoscience, Inc |
#### b. Date Summary Prepared
February 8, 2012
#### c. Description of Device
The Myoscience Cryo-Touch III is a portable cryogenic surgical device used to destroy tissue and/or produce lesions in nervous tissue through application of extreme cold to the selected site. The device is based on introduction of a needle probe, "Tip" cooled by the cryogenic fluid (liguid nitrous oxide, N₂O) to a selected area. The Tip is cooled by the Joule-Thomson effect and/or Latent Heat of Vaporization. The Cryo-Touch III may be used in conjunction with a standard off-the-shelf nerve stimulator device in applications where precise nerve location is desired.
#### Device Design
The device is comprised of four main components:
- 1. a re-usable Control Unit,
- 2. a re-usable Handpiece,
- a single patient use Tip, and 3.
Page
{1}------------------------------------------------
# myoscience
| Test Performed | Result |
|-----------------------------------------|---------------------------------------------|
| Temperature<br>reproducibility | PASS, Substantially equivalent to predicate |
| Mechanical<br>Integrity for system | PASS, Substantially equivalent to predicate |
| Disable valve<br>reliability | PASS, Substantially equivalent to predicate |
| Thermal fuse<br>testing | PASS, Substantially equivalent to predicate |
| Nitrous exposure | PASS, Substantially equivalent to predicate |
| Cryozone size | PASS, Substantially equivalent to predicate |
| Needle Integrity | PASS, Substantially equivalent to predicate |
| Device Durability | PASS, Substantially equivalent to predicate |
| Sterilization and<br>Shelf Life Testing | PASS, Substantially equivalent to predicate |
| Electrical Safety<br>Testing | PASS, Substantially equivalent to predicate |
| Software Testing | PASS, Substantially equivalent to predicate |
| Safety Testing | PASS, Substantially equivalent to predicate |
| Biocompatibility<br>Testing | PASS, Substantially equivalent to predicate |
This performance testing demonstrated that the device is in compliance with pertinent standards (IEC 60601-1, IEC 60601-1-2, ISO 10993-1, ISO 11137-1, ISO 11607, and ASTM F882-84), the product labeling, and is substantially equivalent to the predicates.
Clinical Testing Submitted: None
#### g. Conclusion
Myoscience concludes that the Cryo-Touch III described in this submission is substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol, with three overlapping, curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MyoScience, Incorporated % Ms. Tracey Henry Sr. Director, Regulatory Affairs and Quality Assurance 525 Chesapeake Drive Redwood City, CA 94063
JUN 2 2 2012
Re: K120415
Trade/Device Name: Myoscience Cryo-touch III Regulation Number: 21 CFR 882.4250 Regulation Name: Cryogenic Surgical Device Regulatory Class: II Product Code: GXH Dated: June 13, 2012 Received: June 15, 2012
Dear Ms. Henry:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for answer to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability additeration. Thease now u, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean 1 toure be actived a determination that your device complies with other requirements of the Act that I DA has made a actoring administered by other Federal agencies. You must
{3}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n you attp://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 1000 the regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Kesia Alexander
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices Office of Device Evaluation Health Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Cryo-Touch III Premarket Notification
### Section 4: Indications for Use Statement
510(k) Number:
Device Name: Myoscience Cryo-Touch III
#### Indications for Use:
The Myoscience Cryo-Touch III is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. The Cryo-Touch III is not indicated for treatment of central nervous system tissue.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-the-Counter Use (21 CFR 807 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of
Daryl Kausin
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear. Nose and Throat Devices
K120415 510(k) Number.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.