K120392 · Berchtold GmbH & Co. KG · FSY · Feb 16, 2012 · General, Plastic Surgery
Device Facts
Record ID
K120392
Device Name
CHROMOPHARE F 628 AND F 528
Applicant
Berchtold GmbH & Co. KG
Product Code
FSY · General, Plastic Surgery
Decision Date
Feb 16, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4580
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The CHROMOPHARE® F 628 and F 528 are intended to be used to provide visible illumination to the surgical field of the patient. The surgical lights BERCHTOLD CHROMOPHARE® F 628 and F 528 are intended to illuminate locally the operating site on the patient's body with a high intensity, shadow free, "cold" light.
Device Story
Surgical lights (F 628, F 528) provide high-intensity, shadow-free, "cold" illumination to surgical fields. System utilizes LEDs combined with reflector technology to reduce heat radiation and pattern variation. Features include swivel arms, optional integrated in-light video cameras, and "EndoLite"/"GuideLite" modes for endoscopic procedures. Configurable for ceiling, wall, or floor mounting. Operated by surgical staff in clinical/OR settings. Output provides visual clarity for surgeons, facilitating surgical tasks and improving visibility of the operating site.
Clinical Evidence
Bench testing only. Device performance validated against IEC 60601-2-41:2001 specifications for surgical lamps. No clinical data utilized.
Technological Characteristics
LED-based surgical lighting with reflector technology. Composite frame construction. Features swivel arms, optional in-light video camera, and endoscopic lighting modes (EndoLite/GuideLite). Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and IEC 60601-2-41:2001 (performance). Control system via wall boxes. Sterilizable and disposable handles.
Indications for Use
Indicated for local illumination of the surgical operating site on a patient's body during surgical procedures.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{0}------------------------------------------------
K120392 162
## FEB 1 6 2012
-
: 上
# 510 (k) Summary
Submitter:
BERCHTOLD Holding GmbH Ludwigstaler Straße 125 D- 78532 Tuttlingen Germany
Contact Person:
Volker Hornscheidt Regulatory Affairs 0049 7461 181-374 Phone: 0049 7461 181-8374 Fax: Volker.Hornscheidt@BERCHTOLD.biz
Preparation Date:
November 18, 2011
Trade Name:
November 18, 2011
CHROMOPHARE® F 628 CHROMOPHARE® F 528
Surgical lamp Common Name:
Light, Surgical, Ceiling mounted Classification Name:
Predicate Device:
Device Description:
The new BERCHTOLD CHROMOPHARE® F 628 and F 528 surgical lights are suitable for all types of surgical procedures. With the use of Light Emitting Diodes (LEDs) in combination with reflector technology, BERCHTOLD realizes high illumination intensity with a lower heat radiation and less pattern variation compared to previous generations of BERCHTOLD products. The light features easy-to-operate swivel arms. Each light functions with an optional integrated in-light video camera and/or with upward "EndoLite" and/or with downward "GuideLite" for endoscopic procedures. The lights could be combined among each other or with other CHROMOPHARE devices, on ceiling mounted, wall mounted and floorstand configurations.
BERCHTOLD CHROMOPHARE® E 668 (K090378)
Intended Use of the device:
Indications for Use:
The CHROMOPHARE® F 628 and F 528 are intended to be used to provide visible illumination to the surgical field of the patient.
The surgical lights BERCHTOLD CHROMOPHARE® F 628 and F 528 are intended to illuminate locally the operating site on the patient's body with a high intensity, shadow free, "cold" light.
{1}------------------------------------------------
K120392 2 of 2
## Technological Characteristics Comparison Summary:
- The BERCHTOLD CHROMOPHARE® F 628 and F 528 have the same technological characteristics ﻣﻤﺴﺎ as the predicate BERCHTOLD CHROMOPHARE® E 668 (K090378). Following is a summary of the technological characteristics of the new device in comparison to those of the predicate device. The input power is the same; the protection against electrical shock is the same; the lamp technology is the same, the power consumption per sources is comparable; the color rendering index is slightly higher, the color temperature is the same, the focusing mechanism is the same; the use of sterilizable and disposable handles is the same; the control system via wall boxes is the same; and ombient illumination and in-light camera system capabilities are the same. The predicate device and the new F 628 and F 528 utilize differing materials in construction, such as a composite frame for the new device versus an all-aluminum housing for the predicate devices were subjected to testing in accordance with IEC 60601-1 and IEC 60601-1-2 for mechanical and electrical safety, just as the predicate device. Any technological differences between the F 628 and F 528, and the predicate device do not affect the safety and efficacy, or alter the intended use.
#### Performance Comparison Summary:
This device conforms to IEC 60601-2-41:2001 specifications for performance of surgical lamps. Testing was conducted in accordance with this standard to support substantial equivalence. The F rooting has conceed as did the predicate device, all requirements of the performance standard and achieved specified desired values per the product design specifications. Clinical testing was not required or utilized to support substantial equivalence.
## Substantial Equivalence Conclusion and Determination:
Based on the summarized technological characteristic similarities and differences, in conjunction with the testing conducted to support conformity to IEC 60601-2-41:2001 and performance data tabulated in detail within this submission per FDA "Guidance Document for Surgical Lamp 510(k)s", the CHROMOPHARE F 628 and F 528 are found to be substantially equivalent to the identified predicate device.
#### Conformity Summary:
This device conforms to IEC 60601-2-41:2001 specifications for performance of surgical lamps. This device conforms to IEC 60601-1 and IEC 60601-1-2 for medical device and electrical safety.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002
FEB 1 6 2012
Berchtold GmbH and Co. KG % Underwriters Laboratories, Inc. Mr. Jeff D. Rongero 12 Laboratory Drive Research Triangle Park, North Carolina 27709
Re: K120392
Trade/Device Name: Chromophare® F 628 Chromophare® F 628 Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulatory Class: Class II Product Code: FSY Dated: February 06, 2012 Received: February 08, 2012
Dear Mr. Rongero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2 - Mr. Jeff D. Rongero
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Eric L. Keith
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name:
## CHROMOPHARE® F 628 CHROMOPHARE® F 528
Indications for Use:
The surgical lights BERCHTOLD CHROMOPHARE® F 628 and F 528 are
intended to illuminate locally the operating site on the patient's body with
a high intensity, shadow free, "cold" light.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R.P. Ogden for MXA
(Division Sign-Off)
Division of Surgical. Orthopedic, and Restorative Devices
. 510(k) Number,
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.