The intended use of the Ortho Solutions Trauma Plates for Osteosynthesis of fixation device(s) is to draw two or more aligned bone fragments together to facilitate healing. Ortho Solutions Bone Plates, Bone Screws and Washers System is indicated for use in fixation of various fractured bones. Mini plates and screws 3.5mm or smaller are intended for use in small bones in the hand or foot and include the metacarpal, metatarsal, and phalange bones. Small and large plates and screws 4.0mm and larger are intended for use in large and long bones which can include the clavicle, scapula, pelvis, calcaneus, talus and long bones (humerus, ulna, radius, femur, tibia, and fibula) in the skeletal anatomy. Ortho Solutions Compression Hip Screw CHS/SCS System is indicated for fracture fixation in the proximal and distal regions of the human anatomy skeletal femur. Specific indications include: intracapsular fractures of the femoral neck; trochaneric and subtrochanteric neck fractures; osteotomies for patients with diseases or deformities of the hip; hip arthrodesis; supracondylar fractures and distal femoral fractures. The Ortho Solutions Trauma Plates for Osteosynthesis is not intended for spinal use.
Device Story
Metallic bone fixation system; includes bone plates, screws, and washers; CHS/SCS system for femoral fractures. Used in surgical settings by orthopedic surgeons to stabilize bone fragments; facilitates healing by drawing fragments together. Components include mini, small, and large fragment plates; various cortical/cancellous screws; angled plates for hip fixation. Ancillary instrumentation (guide pins, drills) provided. Sterile, single-use devices. Provides mechanical stabilization of fractures; aids in bone alignment and healing.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Materials: 316LVM stainless steel per ASTM F138. Components: Bone plates, screws, washers, and CHS/SCS angled plates. Dimensions: Various hole lengths and screw diameters (1.5mm to 6.5mm). Energy source: None (mechanical). Connectivity: None. Sterilization: Provided sterile.
Indications for Use
Indicated for fracture fixation in patients requiring stabilization of small bones (hand/foot: metacarpal, metatarsal, phalange) or large/long bones (clavicle, scapula, pelvis, calcaneus, talus, humerus, ulna, radius, femur, tibia, fibula). CHS/SCS system indicated for femoral neck/trochanteric/subtrochanteric/supracondylar/distal femoral fractures, hip osteotomies, and hip arthrodesis. Contraindicated for spinal use.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Bone Plates, Screws & Washers aap Implantate AG (K072411)
Fixation Plate and Screw System Mahe Medical GmbH (K102845)
{0}------------------------------------------------
Page 1/3
# SAFE MEDICAL DEVICES ACT OF 1990 510(k) Summary
NAME OF FIRM:
Ortho Solutions Limited West Station Business Park Spital Road Maldon ESSEX. CM9 6FF United Kingdom
510(k) FIRM CONTACT:
Al Lippincott Engineering Consulting Services, Inc. 3150 E. 200th St. Prior Lake, MN 55372 e-mail: allippincott@msn.com
DATE:
January 30, 2012
TRADE NAME:
COMMON NAME:
CLASSIFICATION:
Compression Hip Screw System (CHS/SCS) Smooth or threaded metallic bone fixation fastener
Ortho Solutions Trauma Plates for Osteosynthesis
(per 21CFR888.3040) - Screw, Fixation, Bone.
Bone Plates, Bone Screws & Washers System;
Single/multiple component metallic bone fixation appliances and accessories (per 21CFR888.3030) - Plate, Fixation, Bone; Washer, Bolt Nut; Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component.
DEVICE PRODUCT CODE:
HRS
#### SUBSEQUENT PRODUCT CODE:
HWC, HTN, KTT
#### SUBSTANTIALLY EQUIVALENT DEVICES
'Bone Plates, Screws & Washers' aap Implantate AG (K072411) 'Fixation Plate and Screw System' Mahe Medical GmbH (K102845) 'CHS/SCS System' Smith & Nephew (K993289, K921786, K895241) 'CHS/SCS System' DePuy (K946156) 'CHS/SCS System' Howmedica (K781762) 'CHS/SCS System' aap Implantate AG (K071852)
{1}------------------------------------------------
<120360 Pege
Ortho Solutions Limited - 510(k) Summary - K120360:
#### DEVICE DESCRIPTION:
The Ortho Solutions Trauma Plates for Osteosynthesis consists of general predicate type trauma implant components commonly found with large companies with orthopedic markets in the United States. These 'general trauma implant devices' consist of the following categories:
### 1. Bone Plates, Bone Screws and Washers System
# 2. Adult Compression Hip Screw (CHS/SCS) System
A brief and concise description of each system is as follows:
1. Bone Plates, Bone Screws and Washers System: This implant system consists of a Mini Fragment System (MFS), a Small Fragment System (SFS), and a Large Fragment System (LFS).
The Mini Fragment System is composed of 2.7mm A.C.P Plates in various hole lengths. These mini plates are secured to bone using 1.5mm, 2.0mm, and 2.7mm Cortical Screws. All Mini Fragment System plates and screws are intended for use in small bone extremity skeletal anatomy. The Small Fragment System is composed of 1/3 Tubular, 3.5mm A.C.P., 3.5mm Reconstruction, 3.5mm Cloverleaf, and 3.5mm 'T' - Plates in various hole lengths, Left, Right, and Oblique configurations. These small fragment plates are secured to bone using 3.5mm and 4.0mm cortical and cancellous screws, selftapping, partial, and fully threaded. A 7 x 3.6mm Washer is also offered. All Bone Screws with or without the washer can be used separate from the plates. All Small Fragment plates and screws are intended for use in small bone skeletal anatomy. The Large Fragment System is composed of 4.5mm Narrow A.C.P, 4.5mm Broad A.C.P., 'T', 'T-Buttress', and 'L-Buttress' Left & Right Angled Plates in various hole length configurations. These large plates are secured to bone using 4.5mm, 4.0mm Malleolar, and 6.5mm cortical and cancellous screws, self-tapping, partial, and fully threaded. A washer is also offered. All bone screws with or without the washer can be used separate from the plates. All large fragment plates and screws are intended for use in large and long bone skeletal anatomy. All bone plates, bone screws, and washers are made of surgical grade 316LVM Stainless Steel to ASTM F138. All plates, screws, and washers are offered 'sterile' to the customer.
2. Adult Compression Hip Screw (CHS/SCS) System: This implant system consists of various angled (950 for SCS, 130°, 135°, and 150° for CHS) high strength forged plates in multiple hole lengths, various Lag Screw lengths with various length Compression Screws, and various length 4.5mm Cortical Screws for securing the plate to femoral bone. Associated guide pin, drills, and ancillary instrumentation is available. The system is intended for use in repair of fractures of the femoral hip neck/head and supra-condylar area of the skeletal femur. All components are manufactured of surgical grade 316LVM Stainless Steel to ASTM F138. All plates and screws are offered 'sterile' to the customer.
Section XII
{2}------------------------------------------------
**Ortho Solutions Limited - 510(k) Summary – K120360:**
INTENDED USE:
The intended use of the Ortho Solutions Trauma Plates for Osteosynthesis of fixation device(s) is to draw two or more aligned bone fragments together to facilitate healing.
Ortho Solutions Bone Plates, Bone Screws and Washers System is indicated for use in fixation of various fractured bones. Mini plates and screws 3.5mm or smaller are intended for use in small bones in the hand or foot and include the metacarpal, metatarsal, and phalange bones. Small and large plates and screws 4.0mm and larger are intended for use in large and long bones which can include the clavicle, scapula, pelvis, calcaneus, talus and long bones (humerus, ulna, radius, femur, tibia, and fibula) in the skeletal anatomy.
Ortho Solutions Compression Hip Screw CHS/SCS System is indicated for fracture fixation in the proximal and distal regions of the human anatomy skeletal femur. Specific indications include: intracapsular fractures of the femoral neck; trochaneric and subtrochanteric neck fractures; osteotomies for patients with diseases or deformities of the hip; hip arthrodesis; supracondylar fractures and distal femoral fractures.
The Ortho Solutions Trauma Plates for Osteosynthesis is not intended for spinal use.
EQUIVALENCE:
The Ortho Solutions Trauma Plates for Osteosynthesis is substantially equivalent to predicate systems from many orthopedic companies (as listed).
#### SUMMARY OF TECH-NOLOGICAL CHAR-ACTERISTICS
The Ortho Solutions Trauma Plates for Osteosynthesis is Similar in Material. Geometry Design/Markings, and Indications to many predicate systems currently sold in the U.S. market.
#### SUMMARY OF SAFETY AND EFFECTIVENESS:
The Ortho Solutions Trauma Plates for Osteosynthesis is shown to be safe and effective for use as 'sterile' and for single-use in a surgical setting.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ortho Solutions Limited % Mr. Al Lippincott Engineering Consulting Services, Inc. 3150 E. 200th St. Prior Lake, MN 55372
MAY 2 9 2012
Re: K120360
Trade/Device Name: Ortho Solutions Trauma Plates for Osteosynthesis Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Product Code: HRS, HWC, HTN, KTT Dated: April 26, 2012 Received: April 30, 2012
Dear Mr. Lippincott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{4}------------------------------------------------
## Page 2 - Mr. Al Lippincott
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "ORTHO SOLUTIONS". The word "ORTHO" is in all caps and is to the left of a large, bold "S" that is enclosed in a circle. The word "SOLUTIONS" is in all caps and is to the right of the "S".
West Station Business Park, Spital Road, Maldon, Essex CM9 6FF Tel: 0870 7777515 Fax: 0870 7777525 Email: sales@orthosol.com www.orthosolutions.com
# Indications for Use
# 510(k) NUMBER: K120360 DEVICE NAME: Ortho Solutions Trauma Plates for Osteosynthesis INDICATIONS FOR USE:
The intended use of the Ortho Solutions Trauma Plates for Osteosynthesis of fixation device(s) is to draw two or more aligned bone fragments together to facilitate healing.
Ortho Solutions Bone Plates, Bone Screws, and Washers System is indicated for use in fixation of various fractured bones. Mini plates and screws 3.5mm or smaller are intended for use in small bones in the hand or foot and include the metacarpal, metatarsal, and phalange bones. Small and large plates and screws 4.0mm and larger are intended for use in large and long bones which can include the clavicle, scapula, pelvis, calcaneus, talus and long bones (humerus, ulna, radius, femur, tibia, and fibula) in the skeletal anatomy.
Ortho Solutions Compression Hip Screw CHS/SCS System is indicated for fracture fixation in the proximal and distal regions of the human anatomy skeletal femur. Specific indications include: intracapsular fractures of the femoral neck; trochaneric and subtrochanteric neck fractures; osteotomies for patients with diseases or deformities of the hip; hip arthrodesis; supracondylar fractures and distal femoral fractures.
The Ortho Solutions Trauma Plates for Osteosynthesis is not intended for
spinal use.(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K120360
Prescription Use X Over-The-Counter-Use_
(Per 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
· AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Registered Office: West Station Business Park, Spital Road, Maldon, Essex CM9 6FF Registered in England No: 4311933
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.