← Product Code [LZC](/productcode/LZC) · K120336

# NITRILE POWDER FREE EXAMINATION GLOVES (VIO) TESTED FOR USE WITH CHENITRILE POWDER FREE EXAMINATION (ABLU) TESTED FOR U (K120336)

_Hartalega Sdn Bhd · LZC · Jun 25, 2012 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K120336

## Device Facts

- **Applicant:** Hartalega Sdn Bhd
- **Product Code:** [LZC](/productcode/LZC.md)
- **Decision Date:** Jun 25, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The Nitrile Powder Free Examination Gloves (White) Tested for Use with Chemotherapy Drugs is a non sterile disposable device intended for medical and dental purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs. The Nitrile Powder Free Examination Gloves (VBLU - Violet Blue) Tested for Use with Chemotherapy Drugs is a non - sterile disposable device intended for medical and dental purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs. The Nitrile Powder Free Examination Gloves (ABLU - Aqua Blue) Tested for Use with Chemotherapy Drugs is a non - sterile disposable device intended for medical and dental purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs. The Nitrile Powder Free Examination Gloves (VIO - Violet) Tested for Use with Chemotherapy Drugs is a non - sterile disposable device intended for medical and dental purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.

## Device Story

Nitrile powder-free examination gloves; non-sterile; disposable. Worn by medical/dental professionals to provide barrier protection against patient-to-examiner contamination. Tested for permeation resistance against 12 specific chemotherapy drugs (e.g., Carmustine, Cisplatin, Cyclophosphamide, etc.). Breakthrough times vary by drug and glove color variant. Used in clinical/dental settings. Provides protection during chemotherapy drug handling; minimizes exposure risk to healthcare personnel.

## Clinical Evidence

Bench testing only. Permeation testing performed for 12 chemotherapy drugs per ASTM standards to determine minimum breakthrough times. No clinical data provided.

## Technological Characteristics

Nitrile rubber material; powder-free; non-sterile; disposable; multiple color variants (White, Violet Blue, Aqua Blue, Violet).

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUN 2 5 2012

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Nurul Aisyah Kong Quality Assurance Manager Hartalega, Sdn. Bhd. No. 7. Kawasan Perusahaan Suria Bestari Jaya Malaysia 45600

Re: K120336

Trade/Device Name: Nitrile Powder-Free Examination Gloves Tested for Use with Chemotherapy Drugs; White, Violet Blue, Aqua Blue and Violet

Regulation Number: 21 CFR §880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZC Dated: June 15, 2012 Received: June 20, 2012

Dear Ms. Kong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.

{1}------------------------------------------------

#### Page 2- Ms. Kong

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirement's as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

For

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

#### Enclosure

{2}------------------------------------------------

## KI20336

## INDICATION FOR USE

510(k) Number (if known): K120336

Device Name:

Nitrile Powder Free Examination Gloves (White) Tested for Use with Chemotherapy Drugs

Indication for Use:

The Nitrile Powder Free Examination Gloves (White) Tested for Use with Chemotherapy Drugs is a non sterile disposable device intended for medical and dental purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs. The list of Chemotherapy Drugs tested (with breakthrough times) as per below:

| Chemotherapy Drugs                         | Minimum Breakthrough Time (Minutes) |
|--------------------------------------------|-------------------------------------|
| 1. Carmustine (3.3 mg/ml)                  | 30.2                                |
| 2. Cisplatin (1.0 mg/ml)                   | > 240                               |
| 3. Cyclophosphamide (Cytoxan) (20.0 mg/ml) | > 240                               |
| 4. Dacarbazine (10.0 mg/ml)                | > 240                               |
| 5. Doxorubixin Hydrochloride (2.0 mg/ml)   | > 240                               |
| 6. Etoposide (Toposar) (20.0 mg/ml)        | > 240                               |
| 7. Fluorouracil (50.0 mg/ml)               | > 240                               |
| 8. Methotrexate (25.0 mg/ml)               | > 240                               |
| 9. Mitomycin C (0.5 mg/ml)                 | > 240                               |
| 10. Paclitaxel (6.0 mg/ml)                 | > 240                               |
| 11. Thiotepa (10.0 mg/ml)                  | > 240                               |
| 12. Vincristine Sulfate (1.0 mg/ml)        | > 240                               |

Please note that the following drugs have extremely low permeation times of 30.2 minutes with Carmustine (3.3 mg/ml)

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over - The - Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Oft

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page 1 of

510(k) Number: K120336

{3}------------------------------------------------

### INDICATION FOR USE

510(k) Number (if known): K120336

Device Name:

Nitrile Powder Free Examination Gloves (Violet Blue) Tested for Use with Chemotherapy Drugs

Indication for Use:

The Nitrile Powder Free Examination Gloves (VBLU - Violet Blue) Tested for Use with Chemotherapy Drugs is a non - sterile disposable device intended for medical and dental purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs. The list of Chemotherapy Drugs tested (with breakthrough times) as per below:

| Chemotherapy Drugs                         | Minimum Breakthrough Time (Minutes) |
|--------------------------------------------|-------------------------------------|
| 1. Carmustine (3.3 mg/ml)                  | 30.3                                |
| 2. Cisplatin (1.0 mg/ml)                   | > 240                               |
| 3. Cyclophosphamide (Cytoxan) (20.0 mg/ml) | > 240                               |
| 4. Dacarbazine (10.0 mg/ml)                | > 240                               |
| 5. Doxorubixin Hydrochloride (2.0 mg/ml)   | > 240                               |
| 6. Etoposide (Toposar) (20.0 mg/ml)        | > 240                               |
| 7. Fluorouracil (50.0 mg/ml)               | > 240                               |
| 8. Methotrexate (25.0 mg/ml)               | > 240                               |
| 9. Mitomycin C (0.5 mg/ml)                 | > 240                               |
| 10. Paclitaxel (6.0 mg/ml)                 | > 240                               |
| 11. Thiotepa (10.0 mg/ml)                  | > 240                               |
| 12. Vincristine Sulfate (1.0 mg/ml)        | > 240                               |

Please not e that the following drugs have extremely low permeation times of 30.3 minutes with Carmustine (3.3 mg/ml)

Prescription Use (Part 21 CFR 801 Subpart D)

510(k) Number:

AND/OR

Over - The - Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page 1 of

**P. 2 of 4**

{4}------------------------------------------------

K120336

### INDICATION FOR USE

510(k) Number (if known): K120336

Device Name:

Nitrile Powder Free Examination Gloves (Aqua Blue) Tested for Use with Chemotherapy Drugs

Indication for Use:

The Nitrile Powder Free Examination Gloves (ABLU - Aqua Blue) Tested for Use with Chemotherapy Drugs is a non - sterile disposable device intended for medical and dental purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs. The list of Chemotherapy Drugs tested (with breakthrough times) as per below:

| Chemotherapy Drugs                         | Minimum Breakthrough Time (Minutes) |
|--------------------------------------------|-------------------------------------|
| 1. Carmustine (3.3 mg/ml)                  | 30.4                                |
| 2. Cisplatin (1.0 mg/ml)                   | > 240                               |
| 3. Cyclophosphamide (Cytoxan) (20.0 mg/ml) | > 240                               |
| 4. Dacarbazine (10.0 mg/ml)                | > 240                               |
| 5. Doxorubixin Hydrochloride (2.0 mg/ml)   | > 240                               |
| 6. Etoposide (Toposar) (20.0 mg/ml)        | > 240                               |
| 7. Fluorouracil (50.0 mg/ml)               | > 240                               |
| 8. Methotrexate (25.0 mg/ml)               | > 240                               |
| 9. Mitomycin C (0.5 mg/ml)                 | > 240                               |
| 10. Paclitaxel (6.0 mg/ml)                 | > 240                               |
| 11. Thiotepa (10.0 mg/ml)                  | 15.6                                |
| 12. Vincristine Sulfate (1.0 mg/ml)        | > 240                               |

Please note that the following drugs have extremely low permeation times of 30.4 minutes with Carmustine (3.3 mg/ml) and 15.6 minutes with Thiotepa (10.0 mg/ml)

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over – The – Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

signature

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

P. 3 of 4

Page 1 of

510(k) Number:

{5}------------------------------------------------

# K120336

### INDICATION FOR USE

510(k) Number (if known): K120336

Device Name:

Nitrile Powder Free Examination Gloves (Violet) Tested for Use with Chemotherapy Drugs

Indication for Use:

The Nitrile Powder Free Examination Gloves (VIO - Violet) Tested for Use with Chemotherapy Drugs is a non - sterile disposable device intended for medical and dental purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs. The list of Chemotherapy Drugs tested (with breakthrough times) as per below:

| Chemotherapy Drugs                         | Minimum Breakthrough Time (Minutes) |
|--------------------------------------------|-------------------------------------|
| 1. Carmustine (3.3 mg/ml)                  | 0.6                                 |
| 2. Cisplatin (1.0 mg/ml)                   | > 240                               |
| 3. Cyclophosphamide (Cytoxan) (20.0 mg/ml) | > 240                               |
| 4. Dacarbazine (10.0 mg/ml)                | > 240                               |
| 5. Doxorubixin Hydrochloride (2.0 mg/ml)   | > 240                               |
| 6. Etoposide (Toposar) (20.0 mg/ml)        | > 240                               |
| 7. Fluorouracil (50.0 mg/ml)               | > 240                               |
| 8. Methotrexate (25.0 mg/ml)               | > 240                               |
| 9. Mitomycin C (0.5 mg/ml)                 | > 240                               |
| 10. Paclitaxel (6.0 mg/ml)                 | > 240                               |
| 11. Thiotepa (10.0 mg/ml)                  | 46.7                                |
| 12. Vincristine Sulfate (1.0 mg/ml)        | > 240                               |

Please not e that the following drugs have extremely low permation times of 0.6 minutes with Carmustine (3.3mg/ml) and 46.7 minutes with Thiotepa (10.0 mg/ml)

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over - The - Counter Use ਮ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

E.P.S.

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page 1 of

P. 4 of 4

510(k) Number: K120336

---

**Source:** [https://fda-staging.innolitics.com/device/K120336](https://fda-staging.innolitics.com/device/K120336)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).