K120274 · Codman and Shurtleff, Inc. · HCG · Mar 2, 2012 · Neurology
Device Facts
Record ID
K120274
Device Name
DELTAMAXX 18 MICROCOIL SYSTEM
Applicant
Codman and Shurtleff, Inc.
Product Code
HCG · Neurology
Decision Date
Mar 2, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5950
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DELTAMAXX™ Microcoil Delivery System is intended for endovascular embolization of intracranial aneurysms, other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae, and is also intended for arterial and venous embolizations in the peripheral vasculature.
Device Story
DELTAMAXX™ 18 Microcoil System is an implantable neurovascular embolization device used to pack aneurysm lumens, isolating them from arterial pressure to promote thrombosis. The system consists of a platinum-tungsten alloy microcoil (available in Stretch Resistant or Cerecyte® models) attached to a Device Positioning Unit (DPU). It is deployed via a Micrus Detachment Control System through compatible 2-tip marker microcatheters (0.0165"–0.019" ID). Used in clinical settings by physicians, the device is advanced into the target vessel; once positioned, the coil is detached from the DPU. The resulting coil mass covers the aneurysm neck, re-establishing parent artery flow and inducing solidification of trapped blood into scar tissue. This line extension of the DELTAPAQ® system offers larger diameters (3–24 mm) and longer lengths (up to 60 cm) to treat varied vascular geometries.
Clinical Evidence
Bench testing only. No clinical data was required to establish substantial equivalence. Verification and validation testing included delivery force, microcatheter compatibility, secondary shape retention, spring constant (stiffness), detachment zone tensile strength, and durability/detachability (cyclic advancement/retrieval).
Technological Characteristics
Implant: Platinum-tungsten alloy wire wound into primary coil (triangular profile) and secondary helical shape. Internal suture: Polypropylene (SR) or absorbable polymer (Cerecyte®). Dimensions: 3-24 mm diameter, up to 60 cm length. Delivery: DPU compatible with 0.0165"-0.019" ID microcatheters. Detachment: Electrical via Micrus Detachment Control System. Sterilization: E-beam radiation. Biocompatibility: ISO 10993 compliant.
Indications for Use
Indicated for endovascular embolization of intracranial aneurysms, neurovascular abnormalities (arteriovenous malformations, arteriovenous fistulae), and arterial/venous embolizations in peripheral vasculature.
Regulatory Classification
Identification
A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).
Predicate Devices
Micrus Microcoil System DELTAPAQ® 10 Stretch Resistant (K080379)
Micrus Microcoil System DELTAPAQ® 10 Cerecyte® (K080437)
{0}------------------------------------------------
Codman & Shurtleff, Inc.
MAR - 2 2012
. Special 510(k) DELTAMAXX™ 18 Microcoil System
.
## 510(k) Summary
(As Required By 21 CFR 807.92(a))
#### A. Submitter Information
| Submitter's name: | Codman & Shurtleff, Inc. |
|-----------------------|-----------------------------------|
| Address: | 325 Paramount Drive |
| | Raynham, MA 02767 |
| Telephone: | 408-433-1420 |
| Fax: | 408-433-1585 |
| Contact Person: | Kim Fonda |
| Date of Submission: | January 27, 2012 |
| B. Trade/Device Name: | DELTAMAXX™ 18 Microcoil System |
| Common Name: | Artificial embolization device |
| Classification Name: | Neurovascular Embolization Device |
| Regulation Number: | Class II per 21 CFR 882 5950 |
# C. Predicate Devices:
| Device | Company | 510(k) Number/<br>Concurrence<br>Date | Product Code | Predicate For: |
|-----------------------------------------------------------------------------|-------------------------------------------------------------------------|---------------------------------------|--------------|-----------------------------------------------------------------------|
| Primary: Micrus<br>Microcoil<br>System<br>DELTAPAQ® 10<br>Stretch Resistant | Codman &<br>Shurtleff, Inc<br>(Previously Micrus<br>Endovascular, Corp) | K080379<br>6/30/2008 | HCG | Intended Use<br>Design<br>Materials<br>Manufacturing<br>Sterilization |
| Primary: Micrus<br>Microcoil<br>System,<br>DELTAPAQ® 10<br>Cerecyte® | Codman &<br>Shurtleff, Inc<br>(Previously Micrus<br>Endovascular, Corp) | K080437<br>5/8/2008 | HCG | Intended Use<br>Design<br>Materials<br>Manufacturing<br>Sterilization |
| Micrus Microcoil<br>System<br>PRESIDIO® 18 | Codman &<br>Shurtleff, Inc<br>(Previously Micrus<br>Endovascular, Corp) | K062036<br>8/25/2006 | HCG | Design |
{1}------------------------------------------------
#### D. Device Description:
The intent of endovascular treatment using embolic microcoils is to pack the lumen of the aneurysm with the microcoil mass, thereby eliminating the influx of fresh blood into the aneurysm. The aneurysm neck will be covered with loops of coils to help eliminate the influx of fresh blood, a blood flow pathway is then re-established down the lumen of the parent artery, and the weakened wall of the aneurysm is isolated from arterial pressures. The existing blood and soft clot trapped within the lumen of the aneurysm begins to solidify, ultimately becoming hardened clot and scar tissue. The process will re-establish blood flow down the lumen of the parent artery and thromboses the aneurysm.
The DELTAMAXX " 18 Microcoil System consists of an embolic coil ("Microcoil") attached to a Device Positioning Unit (DPU), and is compatible with commercially available 2-tip marker microcatheters which have internal lumen diameters between 0.0165 inches and 0.019 inches. The DELTAMAXX ™ 18 Microcoils are provided in Stretch Resistant (SR) or Cerecyte® models and will be offered in lengths up to 60 cm and diameters ranging from 3 mm to 24 mm.
The DELTAMAXX™ 18 Microcoils are fabricated from the same platinum alloy wire used in the DELTAPAQ® Microcoils. The wire is wound into a primary coil (containing either a polypropylene (SR) or absorbable polymer suture (Cerecyte®) inside the wind and then formed into a secondary helical coil shape. The Microcoil is packaged in an introducer sheath designed to protect the coil in the packaging dispenser and to provide support for introducing the coil into the infusion catheter. The coil is the implantable segment of the device, and is detached from the device positioning unit by the Micrus Detachment Control System.
The DELTAMAXX™ 18 Microcoil System is a line extension to the currently marketed DELTAPAQ® 10 Microcoil System, as submitted in premarket notification numbers K080379 and K080437. The modifications include minor design changes required to expand the product line to include larger diameters and longer lengths as well as a minor manufacturing process change.
{2}------------------------------------------------
#### E. Intended Use:
The DELTAMAXX™ Microcoil Delivery System is intended for endovascular embolization of intracranial aneurysms, other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae, and is also intended for arterial and venous embolizations in the peripheral vasculature.
# F. Summary of technological characteristics of the proposed to the predicate device:
The DELTAMAXX™18 Microcoil System (including SR and Cerecyte® models) is substantially equivalent to the Micrus DELTAPAQ® 10 Microcoil System (including SR and Cerecyte® models). This modification expands the product line using existing technology and no new technological characteristics are being introduced with the proposed device.
The DELTAMAXX"18 Microcoil System has the same intended use as the DELTAPAQ® 10 Microcoil System and is similar with regard to design, manufacturing, sterilization process, and materials.
The DELTAMAXX™ 18 Microcoil System is also similar to the PRESIDIO® 18 Microcoil System with regard to the larger diameters and longer lengths.
A summary table including specifications of the proposed device compared with those of the predicate devices follows.
{3}------------------------------------------------
Codman & Shurtleff, Inc
ELTAMAXX™ 18 Microcoil System
# rative Informatio
| Comparative Information<br>Characteristic | DELTAMAXX™ | DELTAPAQ® 10<br>(Primary Predicate) | PRESIDIO® 18<br>(Predicate) |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The DELTAMAXX Microcoil<br>Delivery System is intended for<br>endovascular embolization of<br>intracranial aneurysms, other<br>neurovascular abnormalities such as<br>arteriovenous malformations and<br>arteriovenous fistulae, and is also<br>intended for arterial and venous<br>embolizations in the peripheral<br>vasculature. | The Micrus Microcoil Delivery System<br>is intended for endovascular<br>embolization of<br>intracranial aneurysms, other<br>neurovascular abnormalities such as<br>arteriovenous malformations and<br>arteriovenous fistulae, and are also<br>intended for arterial and venous<br>embolizations in the peripheral<br>vasculature. | The Micrus Microcoil Delivery System<br>is intended for endovascular<br>embolization of<br>intracranial aneurysms, other<br>neurovascular abnormalities such as<br>arteriovenous malformations and<br>arteriovenous fistulae, and are also<br>intended for arterial and venous<br>embolizations in the peripheral<br>vasculature. |
| Component and or<br>material | | | |
| Implant material | Platinum Tungsten Alloy Wire | Platinum Tungsten Alloy Wire | Platinum Tungsten Alloy Wire |
| Internal suture<br>material | SR - Polypropylene<br>Cerecyte | SR - Polypropylene<br>Cerecyte | Cerecyte |
| Primary wind OD | 0.015" | 0.0105" | 0.015" |
| Profile of primary wind | Triangular | Triangular | Circular |
| Primary wind pattern<br>(Deltawind<br>Technology) | Rotated in a repeating pattern | Rotated in a repeating pattern | Cylindrical |
| Secondary Shape Coil<br>OD | 3mm – 24mm | 1.5mm to 10mm | 8mm – 20mm |
| Secondary shape | Helical | Helical | Spherical |
| Coil length | 12 cm minimum<br>60cm maximum | 2cm minimum<br>25cm maximum | 30cm minimum<br>50cm maximum |
| Characteristic | DELTAMAXX™ | DELTAPAQ® 10<br>(Primary Predicate) | PRESIDIO® 18<br>(Predicate) |
| DPU Delivery System<br>Length | 150 cm | 150 cm | 150 cm |
| Microcatheter<br>compatibility | Compatible with 2-tip marker micro<br>catheters of ID from 0.0165" to 0.019" | Compatible with 2-tip marker micro<br>catheters of ID from 0.014" to 0.017" | Compatible with 2-tip marker micro<br>catheters of ID from 0.017" to 0.021" |
| Detachment System | Uses Micrus Connecting Cable | Uses Micrus Connecting Cable | Uses Micrus Connecting Cable |
| | Uses Micrus Detachment Control Box | Uses Micrus Detachment Control Box | Uses Micrus Detachment Control Box |
| Sterilization | E-Beam Radiation | E-Beam Radiation | E-Beam Radiation |
| Packaging | Packaged in a plastic hoop and enclosed<br>in a pouch. | Packaged in a plastic hoop and enclosed<br>in a pouch. | Packaged in a plastic hoop and enclosed<br>in a pouch. |
{4}------------------------------------------------
Codman & Shurtleff, In
omparative Information (con
Note: The DELTAMAXX" 18 Microcoil System is also similar to the MicroPlex Coil System- Cosmos " (K093358) and the
{5}------------------------------------------------
# G. Testing Summary:
Bench testing data demonstrates that the device performs according to its description and intended use, and establishes the performance characteristics of the modifications to this device. The DELTAMAXX™18 Microcoil System passed equivalent bench testing as compared to the DELTAPAQ 10 System and PRESIDIO 18 System predicate devices. Clinical testing was not required to establish substantial equivalence.
Results of verification and validation conducted on the DELTAMAXX™18 Microcoil System demonstrates that the system · performs as designed, is suitable for the intended · use, is substantially equivalent to the predicate device and therefore, does not raise any different questions of safety and effectiveness.
The following tests were conducted to verify the modified design for both Stretch Resistant and Cerecyte® models:
- . Delivery Force Test
- System Delivery Force
- Microcatheter compatibility .
- Secondary Shape Retention Test ♥
- Coil Secondary Shape OD .
- Spring Constant Test .
- Primary Wind Coil Stiffness .
- Detachment Zone Tensile Test .
- Coil Socket Ring Tensile Strength .
- Detachment Fiber Tensile Strength
- Stretch resistance of suture/Tensile Strength l
- Durability and Detachability Test .
- Durability-Cyclic advancement and retrieval of the microcoil system 트
- Detachment of the microcoil from the delivery system .
- Coil damage after microcoil system delivery .
- Microcatheter compatibility ■
{6}------------------------------------------------
#### Codman & Shurtleff, Inc.
The proposed DELTAMAXX™18 Microcoil System is similar to the currently marketed DELTAPAO 8 10 Microcoil System. The modifications include only minor design changes required to expand the product line to include larger diameters and longer lengths as well as a minor manufacturing process change.
The materials, suppliers, manufacturing process, packaging, and sterilization used in the DELTAMAXX" 18 Microcoil System are identical to the material, suppliers, manufacturing process, packaging, and sterilization used in the current DELTAPAQ® 10 Microcoil System with the exception of a minor manufacturing process change to the distal PET ring. The DELTAPAQ® 10 Microcoil System was previously demonstrated to be biocompatible and the data to support the biocompatibility was provided to FDA in the predicate device premarket notifications K080379 (6/30/08) and K080437 (5/8/2008).
The DELTAMAXX"18 Microcoil System continues to meet all the same biocompatibility requirements as the predicate device and as specified by the ISO 10993 Part I, and the General Program Memorandum # G95-1 on Biological Evaluation of Medical Devices.
Based upon the design, materials, function, intended use, comparison with currently marketed devices and the non-clinical testing performed by Codman & Shurtleff, Inc., it is concluded that the DELTAMAXX™18 Microcoil System is substantially equivalent to the DELTAPAQ ® 10 Microcoil System, and therefore does not raise any new questions of safety and effectiveness.
{7}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus, featuring three curved lines that converge at the bottom.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAR - 2 2012
Codman and Shurtleff, Inc. c/o Ms. Kim Fonda Manager Regulatory Affairs 325 Paramount Drive Raynham, MA 02767-0350
Re: K120274
Trade/Device Name: DELTAMAXX™ 18 Microcoil System Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG, KRD Dated: January 27, 2012 Received: January 30, 2012
Dear Ms. Fonda:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becared the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass sured in the exercised of the enactment date of the Medical Device Amendments, or to conninered pror to May 20, 1978, the encordance with the provisions of the Federal Food, Drug, de nees that have been roomston of a premarket approval of a premarket approval application (PMA). and Costience Fee (11et) that as novice, subject to the general controls provisions of the Act. The Tou may, incretore, manxel the Act include requirements for annual registration, listing of general controls proficious practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability adulteration. Thease noter u, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (boo are t). Existing major regulations affecting your device can be may be subject to additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
Page 2 - Ms. Kim Fonda
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
signature
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
Codman & Shurtleff, Inc.
#### Indications for Use Form
510(k) Number (if known): K120274
Device Name: DELTAMAXX™ 18 Microcoil System
The DELTAMAXX™ Microcoil Delivery System is intended for endovascular embolization of intracranial aneurysms, other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae, and is also intended for arterial and venous embolizations in the peripheral vasculature.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Jeffrey Toy
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K12-0274
14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.