K120271 · Medtronic Sofamor Danek USA, Inc. · JEY · May 30, 2012 · Dental
Device Facts
Record ID
K120271
Device Name
ARTISAN SPACE MAINTENANCE SYSTEM
Applicant
Medtronic Sofamor Danek USA, Inc.
Product Code
JEY · Dental
Decision Date
May 30, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4760
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ARTISAN™ Space Maintenance System is indicated for use as temporary implants to stabilize and support autograft, autograft extenders, allograft, and bone void fillers and/or fractured bone segments with or without bone plates or titanium mesh in bony defects of oral maxillofacial anatomy. The ARTISAN™ Space Maintenance System is intended for temporary use in oral maxillofacial surgical reconstruction and dental regeneration procedures for maintaining space during bone grafting procedures and to support soft tissue until bone formation.
Device Story
System of temporary implants for oral maxillofacial bone grafting; includes porous titanium mesh, fixation screws, tenting screws, and socket preservation screws. Used to stabilize/support bone graft materials and fractured bone segments; prevents soft-tissue collapse into graft sites. Operated by oral maxillofacial surgeons in clinical settings. Provides semi-protected space for bone regeneration. Subject of 510(k) to establish MR-Conditional labeling for 1.5T and 3T environments.
Clinical Evidence
No clinical data. Evidence based on bench performance testing including static axial pullout, subsidence, cantilever bending, and removal torque testing per ASTM F543-02, and MRI safety assessments (field interaction, torque, heating, and artifact) per ASTM F2052-06, F2213-06, F2182-11, and F2119-07.
Technological Characteristics
System components: porous titanium mesh, titanium alloy fixation/tenting/socket preservation screws. Materials: medical grade titanium, titanium alloy, cobalt-chromium-molybdenum alloy. Testing standards: ASTM F543-02 (mechanical), ASTM F2052-06, F2213-06, F2182-11, F2119-07 (MRI safety). MR-Conditional at 1.5T and 3T.
Indications for Use
Indicated for patients requiring temporary stabilization and support of bone grafts (autograft, allograft, bone void fillers) or fractured bone segments in oral maxillofacial bony defects. Used in surgical reconstruction and dental regeneration to maintain space and support soft tissue during bone formation.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
{0}------------------------------------------------
laazl
# ARTISAN™ Space Maintenance System 510(k) Summarv
MAY 3 0 2012
Below is the 510(k) summary submitted to the FDA containing information per Code of Federal Requlations, Title 21, Part 807.92: This 510(K) is submitted for one reason: to provide appropriate MRI labeling for the subject devices, while also providing MRI technologists with a method of concluding whether an MRI scan can be performed on the device and specific instructions on how to perform the scan.
- Submitter Information 1.
Medtronic Sofamor Danek Sponsor: 1800 Pyramid Place Memphis. TN 38132
Contact Person: Lin Wu, Regulatory Affairs Specialist lin.wu@medtronic.com (901) 396-3133
Date Prepared: April 18, 2012
- 11. Device Name
Proprietary Trade Name: ARTISAN™ Space Maintenance System Common/Usual Names:
Classification Name: Bone Plate (Regulation Number 21 CFR 872.4760) JEY, DZE Product Code:
#### lll. Predicate Devices
ARCHITEX™ Space Maintenance System (K100779 SE 12/13/2010, and K110259 SE 04/13/2011)
#### IV. Device Description
The ARTISAN™ Space Maintenance System is a comprehensive, allinclusive system which contains a variety of implants designed to temporarily reconstruct bony deficiencies common to the oral cavity that are not intrinsic to the stability of the bony structure. Bony deficiencies
{1}------------------------------------------------
may be naturally occurring osseous defects, surgically created osseous defects, or osseous defects created from traumatic injury to the bone.
The ARTISAN™ Space Maintenance System includes a selection of Socket Preservation screws which are designed to aid in extraction socket grafting while simultaneously supporting the original gingival margins and papilla. The system contains a porous titanium mesh (with only one size of 76mm x 44mm x 0.2mm), titanium allov mesh fixation screws (with sizes of 1.2mm or major diameters x 4mm or 6mm or 10mm or 10mm or 12mm or 14mm lengths, and with sizes of 1.4mm maior diameters x 3mm or 5mm or 7mm lengths), titanium allov tenting screws (with sizes of 1.2mm or 1.4mm major diameters x 8mm or 10mm or 12mm or 14mm lengths x 5mm head diameter), and titanium allov socket preservation screws (with sizes of 1.2mm or 1.4mm major diameters x 14mm or 18mm or 20mm lengths x 3/4mm or 4/5mm or 5/6mm head diameters). When used as indicated, this system provides a semi-protected space to stabilize, support, and protect bone graft (autograff, autograft extenders, allograft, and bone void fillers) by minimizing soft-tissue collapse into the graft recipient site.
## Never use stainless steel and titanium implant components in the same construct.
Medical grade titanium, titanium alloy, and medical grade cobaltchromium-molybdenum allov may be used together.
# Technological Characteristics
The ARTISAN™ Space Maintenance System has the same design, same materials, and physical properties as their respective predicate devices, and is substantially equivalent to their predicate devices.
#### V. Indications for Use
The ARTISAN™ Space Maintenance System is indicated for use as temporary implants to stabilize and support autograft, autograft extenders, allograft, and bone void fillers and/or fractured bone segments with or without bone plates or titanium mesh in bony defects of oral maxillofacial anatomy.
The ARTISAN™ Space Maintenance System is intended for temporary use in oral maxillofacial surqical reconstruction and dental reqeneration procedures for maintaining space during bone grafting procedures and to support soft tissue until bone formation.
The ARTISAN™ Space Maintenance System has the same indications for use as their respective predicate devices.
{2}------------------------------------------------
#### VI. General discussion of the non-clinical tests submitted
For a determination of substantial equivalence, the following analyses and bench performance tests were performed on Subject Devices and Predicate Devices: h
- > Analysis of Dimensional and Material Features
- Static Axial Pullout Testing in accordance with ASTM F543-02: A
- Static Subsidence Testing (to date there are no industry standards); >
- A Static Cantilever Bending Testing (to date there are no industry standards):
- Static "Removal Torque" and "Torque to Failure" testing in accordance A with ASTM F543-02
For a determination of MRI labeling and safety information, the following analyses and bench tests were performed on the Subject Devices:
- > Magnetic Field Interaction Evaluation at 3-Tesla (ASTM F2052-06)
- Qualitative Torque Assessment (ASTM F2213-06) >
- > MRI related Heating Assessment at 1.5-Resla and 3-Tesla (ASTM F2182-11)
- > Artifact Assessment at 3-Tesla (ASTM F2119-07)
#### VII. Conclusions drawn from the non-clinical Tests
Results of the MRI testing have demonstrated that the ARTISAN™ Space Maintenance System can be classified as MR-Conditional in 1.5 and 3 Tesla environments.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all capital letters. The text is black and appears to be printed on a white background. The text is centered horizontally in the image.
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three human profiles incorporated into its body. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Lin Wu Regulatory Affairs Specialist Medtronic Sofamor Danek USA, Inc. 1800 Pyramid Place Memphis, Tennessee 38132
MAY 3 0 2012
Re: K120271
Trade/Device Name: ARTISAN™ SPACE MAINTENANCE SYSTEM Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: JEY, DZE Dated: May 24, 2012 Received: May 24, 2012
Dear Mr. Wu: -
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 - Mr. Wu
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Owens
Anthony D. Watson, B.S., M.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### 510(k) Number (if known):
# Device Name: ARTISAN™ SPACE MAINTENANCE SYSTEM
### INDICATIONS FOR USE:
The ARTISAN™ Space Maintenance System is indicated for use as temporary implants to stabilize and support autograft extenders, allograft, and bone void fillers and/or fractured bone segments with or without bone plates or titanium mesh in bony defects of oral maxillofacial anatomy.
The ARTISAN™ Space Maintenance System is intended for temporary use in oral maxillofacial surgical reconstruction and dental regeneration procedures for maintaining space during bone grafting procedures and to support soft tissue until bone formation.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susa Russer
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K120627
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.