K120152 · Riemser Arzneimittel AG · NPM · Aug 13, 2012 · Dental
Device Facts
Record ID
K120152
Device Name
CERASORB PLUS
Applicant
Riemser Arzneimittel AG
Product Code
NPM · Dental
Decision Date
Aug 13, 2012
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
CERASORB® Plus is recommended for: - · Defects after extirpation of bone cysts - · Augmentation of the atrophied jaw ridge - · Sinus lift or sinus base elevation respectively - · Filling alveolar defects after tooth extraction for jaw ridge support - · Filling extraction defects to create an implant site - · Filling of two or multi-walled infrabony pockets, and bifurcation and trifurcation defects - · Defects after operative removal of retained teeth or corrective osteotomes
Device Story
CERASORB® Plus is a sterile, synthetic, porous bone void filler composed of pure-phase Beta-Tricalcium Phosphate (β-TCP). Provided as roundish and polygonal crushed morsels, the material features an interconnected pore network with porosities of 35% and 65%. Used by dental surgeons in clinical settings, the granules are applied directly to bony defects. Upon contact with vital bone, the material acts as an osteoconductive scaffold, gradually resorbing and being replaced by new host bone. The device is intended for single-use and is supplied in double sterile packaging. It provides a biocompatible, non-toxic alternative to autograft or allograft materials, facilitating bone regeneration in various dental and oral surgical applications.
Clinical Evidence
No clinical data provided in this submission. Safety and effectiveness are supported by historical animal and human studies on Beta-Tricalcium Phosphate and the performance of predicate CERASORB® devices.
Technological Characteristics
Material: Pure-phase Beta-Tricalcium Phosphate (β-TCP) compliant with ASTM F 1088-04. Form: Sterile, porous, roundish/polygonal crushed morsels. Porosity: Interconnected, 35% and 65%. Mechanism: Osteoconductive, resorbable. Sterilization: Gamma irradiation. Single-use.
Indications for Use
Indicated for patients requiring bone void filling in oral/maxillofacial and dental procedures, including bone cyst removal, jaw ridge augmentation, sinus lift, alveolar defect filling, implant site preparation, and treatment of infrabony, bifurcation, or trifurcation defects.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
K 120152
CERASORB® PLUS
Abbreviated 510 (K) Summary
**RIEMSER**
ITEM
## Abbreviated 510 (k) Summary:
CERASORB® Plus Date Prepared: 1 December 2011
AUG 1 3 2012
10,3
#### SUBMISSION INFORMATION 1.
| Name and Address of the Sponsor: | Riemser Arzneimittel AG<br>An der Wiek 7<br>17493 Greifswald<br>Germany |
|----------------------------------|-------------------------------------------------------------------------|
|----------------------------------|-------------------------------------------------------------------------|
Contact person:
Christoph Wahl Regulatory affairs dental medical devices/ Product Group Responsible Tel : +49 - 38351-76667 Fax: +49 - 38351-7688694 E - Mail: wahl@RIEMSER.de
Registered U. S. Agent:
James M. Clinton 5105 Fairoaks Rd. Durham, NC 27712, USA Phone: 919 - 247-0479, fax: 919 - 287-2551 E - Mail: clintonim(wearthiink.net
#### 2. DEVICE IDENTIFICATION
| Proprietary Name: | CERASORB® Plus |
|-----------------------------------|-----------------------------------|
| Common Name: | Bone Void Filler, Synthetic |
| Classification Name: | Bone Grafting Material, Synthetic |
| Classification: | Class II, Special Controls. |
| Classification regulation Number: | 21CFR 872.3930 |
| Product Code: | LYC |
## PREDICATE DEVICES 3.
CERASORB® DENTAL: PMA800035/ K051443
Revised 20 Jun 2012
{1}------------------------------------------------
# CERASORB® PLUS
# Abbreviated 510 (K) Summary
TEM 6
## CERASORB® M DENTAL: K051443
#### INTENDED USE 4.
CERASORB® Plus is recommended for:
- · Defects after extirpation of bone cysts
- · Augmentation of the atrophied jaw ridge
- · Sinus lift or sinus base elevation respectively
- · Filling alveolar defects after tooth extraction for jaw ridge support
- · Filling extraction defects to create an implant site
- · Filling of two or multi-walled infrabony pockets, and bifurcation and trifurcation defects
- · Defects after operative removal of retained teeth or corrective osteotomes
#### 5. DESCRIPTION OF THE DEVICE
CERASORB® Plus is a sterile, synthetic, porous and biocompatible ceramic matrix roundish and polygonal crushed shaped morsels. This modification consists of purephase Beta-Tricalcium Phosphate with a phase purity of ≥ 99% and complies with the ASTM F 1088-04. The devices, when applied to a bony defect, create a network of large, smoothly interconnected pores providing different porosities (CERASORB® Plus mixture of granules with porosity of 35% and of 65%).
CERASORB® Plus is manufactured by a validated manufacturing process which guarantees batch to batch conformity and reproducibility. Due to their synthetic nature CERASORB® Plus does not pose any risk of potential allergic reactions and are neither locally nor systemically toxic.
In contact with vital bone the CERASORB® Plus granules, morsels or granulate is reabsorbed and gradually replaced by new bone.
CERASORB® Plus is provided in double sterile packages (sterilization via gamma irradiation) and are for single-use only.
The use of Beta-Tricalcium Phosphate in vivo is well known and has been studied and evaluated in numerous animal and human studies over the vears. These studies indicate overall excellent performance, biocompatibility, resorption and bone regeneration with CERASORB® derived thereof the performance of CERASORB® Plus in the areas of its intended use and have proven that the material is non-toxic, biocompatible, resorbable while in contact with vital bone over time is replaced by new bone at the defect size.
{2}------------------------------------------------
# CERASORB® PLUS
# Abbreviated 510 (K) Summary
TEM 6
## BASIS FOR SUBSTANTIAL EQUIVALENCE 6.
CERASORB® Plus is are substantially equivalent to a number of currently marketed and approved/cleared bone void fillers for defects in the oral/maxillofacial and dental region, such as CERASORB® DENTAL (formally regulated as a PMA800035/ K051443) and CERASORB® M DENTAL (K051443)
Although the manufacturing process of the material is different (Beta-Tricalcium Phosphate in different porosities and morsel sizes), the intended use, recommended indications for use, target population, anatomical site, and performance data for CERASORB® Plus and the predicate devices are essentially similar. Also, all materials are resorbable and biocompatible. In contact with vital bone any of the bone grafting materials is reabsorbed and gradually replaced by new bone.
Information provided in this submission proves the effectiveness and safety of CERASORB® Plus design.
## STATEMENT OF TECHNOLOGICAL COMPARISON 7.
The CERASORB® Plus modification consist of pure phase Beta-Tricalcium Phosphate ceramic material according to ASTM F 1088-04. The material is of interconnecting porosity, osteoconductive and resorbable.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three overlapping wings, symbolizing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Riemser Arzneimittel AG C/O Mr. James Clinton Quality And Regulatory Consulting, Limited Liability Company 5105 Fairoaks Road Durham, North Carolina 27712
AUG 1 3 2012
Re: K120152
Trade/Device Name: CERASORB® Plus Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: NPM Dated: June 26, 2012 Received: July 02, 2012
Dear Mr. Clinton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{4}------------------------------------------------
Page 2- Ms. Clinton
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Rumm
Anthony D. Watson, B.S., M.S., M.B.A. Director Director of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
K120752
## Indications for Use
510(k) Number: K120152
CERASORB® Plus Device Name:
Indications for Use:
CERASORB® Plus is recommended for
- · Defects after extirpation of bone cysts
- · Augmentation of the atrophied jaw ridge
- · Sinus lift or sinus base elevation respectively
- · Filling alveolar defects after tooth extraction for jaw ridge support
- · Filling extraction defects to create an implant site
- · Filling of two or multi-walled infrabony pockets, and bifurcation and trifurcation defects
- · Defects after operative removal of retained teeth or corrective osteotomes
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susem Reuryer
، Division Sign-Off) ivision of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K-20-152
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.