K120131 · Konica Minolta Medical & Graphic, Inc. · MQB · Feb 10, 2012 · Radiology
Device Facts
Record ID
K120131
Device Name
REGIUS SIGMA2
Applicant
Konica Minolta Medical & Graphic, Inc.
Product Code
MQB · Radiology
Decision Date
Feb 10, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The Direct Digitizer, REGIUS SIGMA2 is an X-ray image reader which uses a stimulable phosphor plate (Plate) as X-ray detector installed in a separate cassette. It reads the image recorded on the Plate and transfers the image data to an externally connected device such as a host computer, an order input device, an image display device, a printer, an image data filing device, and other image reproduction devices. It is designed and intended to be used by trained medical personnel in a clinic, a radiology department in a hospital and in other medical facilities. This device is not intended for use in mammography.
Device Story
Compact X-ray image reader; utilizes stimulable phosphor plate (Plate) in cassette as X-ray detector. Device accepts cassette; performs laser scan and photoelectric conversion of latent X-ray images into digital data. Output transferred to external host computer, display, printer, or filing device. Operated by trained medical personnel in clinics or radiology departments. Modification from predicate (REGIUS SIGMA) involves firmware update to increase processing capacity via optimized feed sequence and faster erasing LED. Device facilitates digital X-ray image reproduction for clinical decision-making.
Clinical Evidence
Bench testing only. Performance data compared against predicate device (REGIUS SIGMA) to verify that modifications to processing speed did not impact safety or efficacy. No clinical data presented.
Indicated for use by trained medical personnel in clinical, hospital, or medical facility settings to read X-ray images from stimulable phosphor plates. Not intended for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K120131
# 510(k) Summary
FEB 1 0 2012
#### As required by 807.92
## 1. Company Identification
Konica Minolta Medical & Graphic, Inc. No.1 Sakura-machi, Hino-shi, Tokyo 191-8511, Japan Establishment Registration Number: 3004485675
## 2. Submitter's Name and Address
Shigeyuki Kojima Manager Regulations and Standards Section, Quality Assurance Center No. 1 Sakura-machi, Hino-shi, Tokyo 191-8511, Japan Telephone: 81-42-589-8429 Fax: 81-42-589-8053
## 3. Date of Submission
January 13, 2012
### 4. Device Trade Name
Direct Digitizer, REGIUS SIGMA2
#### 5. Classification
Class II , 90 MQB, 21 CFR 892.1650.
## 6. Predicate Device
Direct Digitizer, REGIUS SIGMA, 510(k) number: K103703
#### 7. Indications for Use
The Direct Digitizer, REGIUS SIGMA2 is an X-ray image reader which uses a stimulable phosphor plate (Plate) as X-ray detector installed in a separate cassette. It reads the image recorded on the Plate and transfers the image data to an externally connected device such as a host computer, an order input device, an image display device, a printer, an image data filing device, and other image reproduction devices. It is designed and intended to be used by trained medical personnel in a clinic, a radiology department in a hospital and in other medical facilities.
This device is not intended for use in mammography.
#### 8. Device Description
The Direct Digitizer, REGIUS SIGMA2 is a compact X-ray image reader which uses a stimulable phosphor plate (Plate) as X-ray detector installed in a
5-1
{1}------------------------------------------------
cassette, and reads the image recorded on the Plate by inserting a cassette in the entrance slot of this device. By means of laser scan and photoelectric method, this device reads the X-ray image data created in form of a latent image on the Plate exposed by an external X-ray generating device, and converts the read data into digital.
The device is comprised of a reading unit with cassette containing Plate.
Plates and cassette remain unchanged from the predicate device, REGIUS SIGMA.
The image data transfer to an externally connected device such as a host computer, an order input device, an image display device, a printer, an image data filing device, and other image reproduction devices.
The basic operations of REGIUS SIGMA2 and the predicate device, REGIUS SIGMA, such as a starting, a shut down, a registration-of-patient, a setting of a various condition are operated with the Medical Image Processing Workstation, ImagePilot (operator console) which is cleared 510(k), K071436.
The modification was made from the REGIUS SIGMA to the REGIUS SIGMA2 to increase processing capacity. To increase the processing capacity, the firmware (device mechanical control software) of the reading unit is modified. The outline is as follows.
The processing capacity is increased by controlling the feed sequence.
(1) Increasing the speed of removal of Plate from cassette and the sequence of storage into cassette
(2) Increasing the erasing speed by changing the erasing LED
The feed sequence related to Image Quality (Reading speed, Open/Close timing of sub-scanning nip) is not changed.
#### 9. Performance Testing
Performance data from non-clinical testing of the REGIUS SIGMA2 is compared with data from the predicate device, REGIUS SIGMA. This comparison showed that the REGIUS SIGMA2 performed as well as the predicate device.
#### 10. Substantial Equivalence to Predicate Device
The predicate device and the REGIUS SIGMA2 are the X-ray image reader which use the same stimulable phosphor plate (Plate) as X-ray detector installed in a separate cassette. The principals of operation and technological characteristics of the predicate device and the REGIUS SIGMA2 is the same. The performance test results show that there is no new safety and efficacy issue of the REGIUS SIGMA. The indications for use of REGIUS SIGMA remain unchanged from the predicate device. For details, please refer to Section 8, Device description.
$$\mathfrak{k}\cdot\mathfrak{k}$$
{2}------------------------------------------------
## 11. Safety Information
The REGIUS SIGMA2 has been tested and shown to meet the requirements of the following standards.
: IEC60601-1 Ed.2(1988)+ A1(1991)+A2(1995) Safety standard Electromagnetic Compatibility : IEC60601-1-2 Ed.3(2007) Radiation safety: 21 CFR 1040.10, IEC60825-1(1993)+A1(1997)+A2:2001 The Risk Analysis for the REGIUS SIGMA2 was conducted on the basis of ISO14971, "Medical devices - Application of risk management to medical devices". As a result of risk control measures, the risk associated with all of the identified hazards was reduced to an acceptable level.
## 12. Conclusion
Comprehensively, we judged that the REGIUS SIGMA2 has the same technological characteristics as the predicate devices. This 510(k) has demonstrated substantial equivalence as the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle with its wings outstretched.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
KONICA MINOLTA MEDICAL & GRAPHIC, INC Mr. Russell D. Munves Consultant Storch Amini & Munves, P.C. Two Grand Central Tower, 25th Floor 140 East 45th Street NEW YORK NY 10017
FEB 1 0 2012
Re: K120131
Trade/Device Name: Direct Digitizer, REGIUS SIGMA2 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: MOB Dated: January 13, 2012 Received: January 17, 2012
#### Dear Mr. Munves:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirences as be form inrketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nredicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary Pastel
Mary S. Pastel. Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name Direct Digitizer, REGIUS SIGMA2
Indications for Use:
The Direct Digitizer, REGIUS SIGMA2 is an X-ray image reader which uses a stimulable phosphor plate (Plate) as X-ray detector installed in a separate cassette. It reads the image recorded on the Plate and transfers the image data to an externally connected device such as a host computer, an order input device, an image display device, a printer, an image data filing device, and other image reproduction devices. It is designed and intended to be used by trained medical personnel in a clinic, a radiology department in a hospital and in other medical facilities.
This device is not intended for use in mammography.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Muthal D.K.
Division Sign-off Office of In Vitro Diagnostic Devices Evaluation and Safety
510(k) hla0131
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.