The AMPLATZER TorqVue 45°x45°, LA1, and LA2 Delivery Sheaths are intended to provide a pathway through which devices are introduced within the chambers of the heart.
Device Story
Delivery sheath system designed to access and deliver devices to heart chambers. Components include sheath, dilator, and flush adaptor (for 10-14Fr sizes). Sheaths feature radiopaque markers for fluoroscopic visualization. TV45x45 model has 3D geometry with ~45° curves in two dimensions; TVLA1 has single 45° distal curve; TVLA2 has 45° distal curve plus 20-30° superior curve at distal tip. Available in 9Fr, 10Fr, 12Fr, 13Fr, and 14Fr sizes; lengths of 80 cm (TVLA) or 100 cm (TV45x45). Used by physicians in clinical settings to facilitate percutaneous delivery of cardiac devices. System enables device advancement, tracking, and recapture within the vasculature and heart chambers.
Clinical Evidence
Bench and laboratory testing performed, including dimensional, trackability, kink, flushability, handoff/advancement force, recapture force, tensile, torque, and leak testing. Design validation included an animal study. No human clinical data provided.
Technological Characteristics
Percutaneous delivery catheter system. Components: sheath, dilator, flush adaptor. Radiopaque materials used for fluoroscopic visibility. Sizes: 9Fr-14Fr. Lengths: 80 cm or 100 cm. Geometry: 45°x45° curve, single 45° curve, or 45° plus 20-30° superior curve. Non-powered, mechanical device.
Indications for Use
Indicated for providing a pathway for the introduction of devices into the chambers of the heart. Intended for prescription use.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
K120120
P. 1 of 4
## 510(K) SUMMARY
MAR 1 5 2012
| Submitter: | AGA Medical Corporation<br>5050 Nathan Lane<br>Plymouth, MN |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Rashmi Bhushan, PhD<br>Principal Regulatory Affairs Specialist<br>Phone: 651-756-5767<br>Fax: 763-647-5932<br>E-mail: rbhushan@sjm.com |
| Date Prepared: | January 13, 2012 |
| Trade Name: | AMPLATZER® TorqVue® 45°x45° Delivery Sheath<br>AMPLATZER® TorqVue® LA1 Delivery Sheath<br>AMPLATZER® TorqVue® LA2 Delivery Sheath |
| Common Name: | Catheter Delivery System |
| Classification: | Class II, 21 CFR 870.1250<br>Catheter, Percutaneous |
| Product Code: | DQY |
| Predicate Device(s): | AMPLATZER TorqVue 45°x45° Delivery Sheath (K083214) |
{1}------------------------------------------------
Device Description: The AMPLATZER TorqVue LA and AMPLATZER TorqVue 45x45 Delivery Sheaths are general purpose delivery sheaths designed to access and deliver a device to the chambers of the heart. The TV45x45 Delivery Sheath is curved approximately 45° in two dimensions resulting in a three-dimensional geometry. The TVLA1 provides a single 45° curve on the distal end of the sheath. The TVLA2 provides the same 45° curve as the TVLA1 with a second superior curve at the distal tip of 20-30°.
> The sheaths are offered in five sizes each (9Fr, 10Fr, 12Fr, 13FR, and 14Fr). The TVLA sheaths and are 80 cm in length and the TV45x45 sheath is 100 cm in length.
The devices include a sheath to enable device delivery, a dilator to ease entrance into and through the vasculature, and a flush adaptor for the 10, 12, 13, and 14Fr sizes to enable connection with syringes for flushing the sheath lumen. The sheaths and dilators are radiopaque for visibility under fluoroscopy. For added visibility, the sheaths also have a radiopaque marker band near the distal tip.
#### Intended Use: The AMPLATZER TorqVue 45°x45°, LA1, and LA2 Delivery Sheaths are intended to provide a pathway through which devices are introduced within the chambers of the heart.
{2}------------------------------------------------
:
.
| Comparison to<br>predicate: | The predicate device, the TV45x45, is currently marketed under<br>K083214 clearance for the 9, 10, 12, and 13Fr sizes. Design<br>changes were made to the predicate, which included: |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <b>14Fr TV45x45 Delivery Sheath</b> Addition of larger sheath size Change to flush adapter (14Fr devices only) Addition of additional backing card to packaging (14Fr devices only) <b>TVLA Delivery Sheaths</b> Addition of larger sheath size Change to device length Change to sheath distal curve geometry Change to flush adapter (14Fr devices only) Addition of additional backing card to packaging (14Fr devices only) |
| Functional and<br>Safety Testing: | Bench and laboratory testing was performed to support a<br>determination of substantial equivalence to the predicate device.<br>Results from the testing show the proposed device conforms to<br>the requirements for its intended use. This included the<br>following testing: Dimensional Tests Trackability Test Kink Test Flushability Test Handoff / Advancement Force Test Recapture Force Test Visual Inspection Post-Interaction Testing Sheath Tip Tensile Test Sheath Transition Tensile Test Sheath Surface Inspection Sheath torque to failure Dilator Hub Tensile Test Sheath Hub Tensile Test Sheath Luer Leak Test (water and air) Design validation testing (animal study) |
{3}------------------------------------------------
AGA Medical Corporation considers the 14Fr AMPLATZER Conclusion: TorqVue 45°x45° Delivery Sheath, 9-14Fr AMPLATZER TorqVue LA1 Delivery Sheaths, and 9-14Fr AMPLATZER TorqVue LA2 Delivery Sheaths to be substantially equivalent to the predicate device listed above. This conclusion is based upon the devices' similarities in principles of operation, technology, materials and indications for use.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's name in a circular arrangement around a central symbol. The symbol consists of three stylized shapes that resemble a person with outstretched arms, representing health and human services. The seal is black and white.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAR 1 5 2012
AGA Medical Corporation c/o Rashmi Bhushan, Ph.D. Principal Regulatory Affairs Specialist 5050 Nathan Lane Plymouth, MN 55433
Re: K120120
Trade/Device Name: AMPLATZER® TorqVue® 45°x45° Delivery Sheath AMPLATZER® TorqVue® LA1 Delivery Sheath AMPLATZER® TorqVue® LA2Delivery Sheath Regulation Number: 21 CFR 870.1250 Regulation Name: Catheter, Percutaneous Regulatory Class: Class II Product Code: DQY Dated: February 22, 2012 Received: February 23, 2012
Dear Dr. Bhushan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
Page 2 – Dr. Rashmi Bhushan
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing
practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram Ø. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use Statement
510(k) Number:
### K 120120 --
Device Name:
AMPLATZER TorqVue 45°x45° Delivery Sheath AMPLATZER TorqVue LA1 Delivery Sheath AMPLATZER TorqVue LA2 Delivery Sheath
Indications for Use:
The AMPLATZER TorqVue 45°x45°, LA1, and LA2 Delivery Sheaths are intended to provide a pathway through which devices are introduced within the chambers of the heart.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)

2 Devices
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.