K120119 · Sunny Medical Device (Shenzhen) Co., Ltd. · DQX · May 11, 2012 · Cardiovascular
Device Facts
Record ID
K120119
Device Name
SUNMED GUIDE WIRE
Applicant
Sunny Medical Device (Shenzhen) Co., Ltd.
Product Code
DQX · Cardiovascular
Decision Date
May 11, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
Sunmed Guidewires are used for the introduction and placement of diagnostic or interventional devices in the coronary and peripheral vasculature and may be used to reach and cross a target lesion.
Device Story
Sunmed Guide Wires are single-use, sterile, radiodetectable guide wires used for coronary and peripheral vascular procedures. They facilitate the introduction and placement of diagnostic or interventional devices and assist in crossing target lesions. The device is operated by clinicians using the Seldinger technique; fluoroscopy is used to confirm wire position. Construction includes a safety wire, core wire, and wound spring for flexibility. Available in three versions: stainless steel, hydrophilic-coated, and PTFE (Teflon)-coated (some with Benzalkonium Heparin coating). Variations include fixed or moveable cores, straight or 'J' tip configurations, and various diameters and lengths. The device provides a track for interventional tools, potentially benefiting patients by enabling minimally invasive access to vascular lesions.
Indicated for introduction and placement of diagnostic or interventional devices in coronary and peripheral vasculature; may be used to reach and cross target lesions.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
# 5. 510(k) Summary
MAY 1 1 2012
K120119
# Sunny Medical Device (Shenzhen) Co., Ltd.
Suite 401 Dongpingxin Creative Science Park, Zhongxincheng Longcheng Street, Longgang District, Shenzhen, Guangdong Province P.R. China 518172
Tel: +86 755 3383-1052 Fax: +86 755 2893-8097, +852 3171-6516, +1 949 423-0168 Email: sunnymedical_cn@139.com, jamesqizhang@gmail.com Website: www.sunnymedical.com.cn
#### SUMMARY
| K Reference # | K120119 |
|-------------------|-------------------------------------------|
| Submitter's name: | Sunny Medical Device (Shenzhen) Co., Ltd. |
| Address: | Suite 401 Dongpingxin Creative Science |
Address. Park, Zhongxincheng Longcheng Street, Longgang District, Shenzhen, Guangdong Province, P.R. China 518172 Phone: +86-755 3383-1052 Fax number: +86 755 2893-8097/+1 949 423-0168
Name of contact person:
Greg Holland Regulatory Specialists, Inc. 3722 Ave. Sausalito Irvine, CA 92606 Phone: 949-262-0411 Fax: 949-552-2821
May 8, 2012 Date the summary was revised:
| Name of the device: | Sunmed Guide Wires |
|---------------------------|--------------------|
| Trade or proprietary name | Sunmed Guide Wires |
| Common or usual name | Guide wires |
| Product Code | Classification Regulation | Classification Name |
|--------------|---------------------------|---------------------|
| DQX | 870.1330 | Catheter guide wire |
#### The legally marketed devices to which we are claiming equivalence [807.92(a)(3)]:
| 510(k) | Trade or Proprietary or Model Name | Manufacturer |
|--------------|-------------------------------------|----------------|
| 1<br>K993000 | Bard Hydrophilic Coated Guide Wires | C.R. Bard Inc. |
| 2<br>K082094 | PTFE (Teflon) Coated Guide Wires | C.R. Bard Inc. |
{1}------------------------------------------------
#### Description of the device:
K120119 D. Doff
Sunmed Guidewires are used for the introduction and placement of diagnostic or interventional devices in the coronary and peripheral vasculature and mav be used to reach and cross a target lesion.
The Sunmed Guide Wires are available in three versions with a variety of tips:
Stainless steel guide wire Hydrophilic coated quide wire PTFE (Teflon®) coated guide wires
The Seldinger technique is used with these quidewires.
The stainless steel guide wires are standard wires with various tip configurations. The guide wire construction consists of a safety wire, a core wire, and a wound spring for flexibility. The features that distinguish individual wires consist of the core type (fixed or moveable), a straight or "J" tip configuration, diameters and lengths.
The Hydrophilic coated guide wires are constructed from the stainless steel core wire, with varying core lengths and diameters for each design.
The PTFE (Teflon®) coated guide wires are also manufactured from stainless steel wire, with PTFE (Teflon®) coating, or Benzalkonium Heparin (BH) coating applied over the PTFE (Teflon®) coating.
These guide wires are radiodetectable and fluoroscopy may be used to confirm position.
These medical devices are single use and supplied sterile by EQ.
{2}------------------------------------------------
# List of devices and predicate K reference numbers:
# 120119
| DIAMETER | LENGTH | TIP | COATING | SUNNY CATALOG # | BARD K # |
|----------|--------|----------|------------------------|-------------------|-------------------|
| 0.035 | 45 | 3 mm J | PTFE | SM-GW-3J35045P-TP | K082094 |
| 0.035 | 50 | 1.5 mm J | PTFE | SM-GW-1J35050P-TP | K082094 |
| 0.025 | 80 | Straight | PTFE | SM-GW-3S25080P-TP | K082094 |
| 0.032 | 80 | Straight | PTFE | SM-GW-3S32080P-TP | K082094 |
| 0.035 | 80 | Straight | PTFE | SM-GW-3S35080P-TP | K082094 |
| 0.035 | 80 | 3 mm J | PTFE | SM-GW-3J35080P-TP | K082094 |
| 0.038 | 80 | Straight | PTFE | SM-GW-3S38080P-TP | K082094 |
| 0.025 | 100 | Straight | PTFE | SM-GW-3S25100P-TP | K082094 |
| 0.025 | 100 | 3 mm J | PTFE | SM-GW-3J25100P-TP | K082094 |
| 0.035 | 100 | Straight | PTFE | SM-GW-3S35100P-TP | K082094 |
| 0.035 | 100 | 3 mm J | PTFE | SM-GW-3J35100P-TP | K082094 |
| 0.038 | 100 | Straight | PTFE | SM-GW-3S38100P-TP | K082094 |
| 0.038 | 100 | 3 mm J | PTFE | SM-GW-3J38100P-TP | K082094 |
| 0.025 | 120 | Straight | PTFE | SM-GW-3S25120P-TP | K082094 |
| 0.025 | 120 | 3mm J | PTFE | SM-GW-3J25120P-TP | K082094 |
| 0.032 | 120 | Straight | PTFE | SM-GW-3S32120P-TP | K082094 |
| 0.032 | 120 | 3 mm J | PTFE | SM-GW-3J32120P-TP | K082094 |
| 0.035 | 120 | Straight | PTFE | SM-GW-3S35120P-TP | K082094 |
| 0.035 | 120 | 1.5 mm J | PTFE | SM-GW-1J35120P-TP | K082094 |
| 0.035 | 120 | 3 mm J | PTFE | SM-GW-3J35120P-TP | K082094 |
| 0.035 | 120 | 6 mm J | PTFE | SM-GW-6J35120P-TP | K082094 |
| 0.038 | 120 | Straight | PTFE | SM-GW-3S38120P-TP | K082094 |
| 0.038 | 120 | 3 mm J | PTFE | SM-GW-3J38120P-TP | K082094 |
| 0.045 | 120 | Straight | PTFE | SM-GW-3S45120P-TP | K082094 |
| 0.018 | 145 | Straight | PTFE | SM-GW-3S18145P-TP | K082094 |
| 0.024 | 145 | 3 mm J | PTFE | SM-GW-3J24145P-TP | K082094 |
| 0.025 | 145 | Straight | PTFE | SM-GW-3S25145P-TP | K082094 |
| 0.025 | 145 | 1.5 mm J | PTFE | SM-GW-1J25145P-TP | K082094 |
| 0.028 | 145 | Straight | PTFE | SM-GW-3S28145P-TP | K082094 |
| 0.028 | 145 | 1.5 mm J | PTFE | SM-GW-1J28145P-TP | K082094 |
| 0.028 | 145 | 3 mm J | PTFE | SM-GW-3J28145P-TP | K082094 |
| 0.032 | 145 | Straight | PTFE | SM-GW-3S32145P-TP | K082094 |
| 0.032 | 145 | 1.5 mm J | PTFE | SM-GW-1J32145P-TP | K082094 |
| 0.032 | 145 | 3 mm J | PTFE | SM-GW-3J32145P-TP | K082094 |
| 0.035 | 145 | Straight | PTFE | SM-GW-3S35145P-TP | K082094 |
| 0.035 | 145 | 1.5 mm J | PTFE | SM-GW-1J35145P-TP | K082094 |
| 0.035 | 145 | 3 mm J | Heparin | SM-GW-3J35145E-TP | K082094 |
| 0.035 | 145 | 3 mm J | PTFE | SM-GW-3J35145P-TP | K082094 |
| 0.035 | 145 | 6 mm J | PTFE | SM-GW-6J35145P-TP | K082094 |
| 0.038 | 145 | Straight | PTFE | SM-GW-3S38145P-TP | K082094 |
| 0.038 | 145 | 3 mm J | Heparin | SM-GW-3J38145E-TP | K082094 |
| P. 4 of | | | | | |
| DIAMETER | LENGTH | TIP | COATING | SUNNY CATALOG # | BARD K # |
| 0.038 | 145 | 3 mm J | PTFE | SM-GW-3J38145P-TP | K082094 |
| 0.038 | 145 | 6 mm J | PTFE | SM-GW-6J38145P-TP | K082094 |
| 0.045 | 145 | Straight | PTFE | SM-GW-3S45145P-TP | K082094 |
| 0.065 | 145 | Straight | PTFE | SM-GW-3S65145P-TP | K082094 |
| 0.038 | 150 | Straight | PTFE | SM-GW-3S38150P-TP | K082094 |
| 0.038 | 150 | 3 mm J | PTFE | SM-GW-3J38150P-TP | K082094 |
| 0.035 | 175 | 3 mm J | PTFE | SM-GW-3J35175P-TP | K082094 |
| 0.025 | 200 | 1.5 mm J | PTFE | SM-GW-1J25200P-TP | K082094 |
| 0.032 | 200 | Straight | PTFE | SM-GW-3S32200P-TP | K082094 |
| 0.032 | 200 | 1.5 mm J | PTFE | SM-GW-1J32200P-TP | K082094 |
| 0.035 | 200 | Straight | PTFE | SM-GW-3S35200P-TP | K082094 |
| 0.035 | 200 | 1.5 mm J | PTFE | SM-GW-1J35200P-TP | K082094 |
| 0.035 | 200 | 3 mm J | PTFE | SM-GW-3J35200P-TP | K082094 |
| 0.038 | 200 | Straight | PTFE | SM-GW-3S38200P-TP | K082094 |
| 0.025 | 220 | Straight | PTFE | SM-GW-3S25220P-TP | K082094 |
| 0.035 | 220 | Straight | PTFE | SM-GW-3S35220P-TP | K082094 |
| 0.035 | 220 | 3 mm J | PTFE | SM-GW-3J35220P-TP | K082094 |
| 0.038 | 220 | Straight | PTFE | SM-GW-3S38220P-TP | K082094 |
| 0.035 | 260 | Straight | PTFE | SM-GW-3S35260P-TP | K082094 |
| 0.035 | 260 | 3 mm J | PTFE | SM-GW-3J35260P-TP | K082094 |
| 0.038 | 260 | Straight | PTFE | SM-GW-3S38260P-TP | K082094 |
| 0.035 | 300 | Straight | PTFE | SM-GW-3S35300P-TP | K082094 |
| 0.035 | 400 | Straight | PTFE | SM-GW-3S35400P-TP | K082094 |
| 0.025 | 145 | Straight | Hydrophilic | SM-GW-3S25145H-TP | K993000 |
| 0.035 | 145 | Straight | Hydrophilic | SM-GW-3S35145H-TP | K993000 |
| 0.038 | 145 | Straight | Hydrophilic | SM-GW-3S38145H-TP | K993000 |
| 0.032 | 40 | Straight | 302 Stainless<br>Steel | SM-GW-3S32040N-TP | Pre-amendment |
| 0.025 | 45 | 3 mm J | 302 Stainless<br>Steel | SM-GW-3J25045N-TP | Pre-amendment |
| 0.035 | 45 | Straight | 302 Stainless<br>Steel | SM-GW-3S35045N-TP | Pre-amendment |
| 0.035 | 45 | 3 mm J | 302 Stainless<br>Steel | SM-GW-3J35045N-TP | Pre-amendment |
| 0.038 | 45 | Straight | 302 Stainless<br>Steel | SM-GW-3S38045N-TP | Pre-amendment |
| 0.038 | 50 | 3 mm J | 302 Stainless<br>Steel | SM-GW-3J38050N-TP | Pre-amendment |
| 0.032 | 80 | Straight | 302 Stainless<br>Steel | SM-GW-3S32080N-TP | Pre-amendment |
| 0.035 | 80 | 3 mm J | 302 Stainless<br>Steel | SM-GW-3J35080N-TP | Pre-amendment |
| 0.035 | 100 | Straight | 302 Stainless<br>Steel | SM-GW-3S35100N-TP | Pre-amendment |
| DIAMETER | LENGTH | TIP | COATING | SUNNY CATALOG # | BARD K # |
| 0.035 | 100 | 3 mm J | 302 Stainless<br>Steel | SM-GW-3J35100N-TP | Pre-<br>amendment |
| 0.038 | 100 | Straight | 302 Stainless<br>Steel | SM-GW-3S38100N-TP | Pre-<br>amendment |
| 0.038 | 100 | 3 mm J | 302 Stainless<br>Steel | SM-GW-3J38100N-TP | Pre-<br>amendment |
| 0.045 | 100 | Straight | 302 Stainless<br>Steel | SM-GW-3S45100N-TP | Pre-<br>amendment |
| 0.035 | 120 | Straight | 302 Stainless<br>Steel | SM-GW-3S35120N-TP | Pre-<br>amendment |
| 0.035 | 120 | 3 mm J | 302 Stainless<br>Steel | SM-GW-3J35120N-TP | Pre-<br>amendment |
| 0.038 | 120 | Straight | 302 Stainless<br>Steel | SM-GW-3S38120N-TP | Pre-<br>amendment |
| 0.038 | 120 | 3 mm J | 302 Stainless<br>Steel | SM-GW-3J38120N-TP | Pre-<br>amendment |
| 0.045 | 120 | Straight | 302 Stainless<br>Steel | SM-GW-3S45120N-TP | Pre-<br>amendment |
| 0.035 | 145 | Straight | 302 Stainless<br>Steel | SM-GW-3S35145N-TP | Pre-<br>amendment |
| 0.035 | 145 | 3 mm J | 302 Stainless<br>Steel | SM-GW-3J35145N-TP | Pre-<br>amendment |
| 0.038 | 145 | Straight | 302 Stainless<br>Steel | SM-GW-3S38145N-TP | Pre-<br>amendment |
| 0.038 | 145 | 3 mm J | 302 Stainless<br>Steel | SM-GW-3J38145N-TP | Pre-<br>amendment |
| 0.035 | 250 | Straight | 302 Stainless<br>Steel | SM-GW-3S35250N-TP | Pre-<br>amendment |
| 0.038 | 260 | Straight | 302 Stainless<br>Steel | SM-GW-3S38260N-TP | Pre-<br>amendment |
| 0.038 | 300 | Straight | 302 Stainless<br>Steel | SM-GW-3S38300N-TP | Pre-<br>amendment |
| 0.035 | 400 | Straight | 302 Stainless<br>Steel | SM-GW-3S35400N-TP | Pre-<br>amendment |
{3}------------------------------------------------
K120119 14
{4}------------------------------------------------
20119 CC
### Package:
Material No.1: DuPont Tyvek package for medical use. Material No.2: Multiple-layer compound film made from of PE and PA.
# Indications:
Sunmed Guidewires are used for the introduction and placement of diagnostic or interventional devices in the coronary and peripheral vasculature and may be used to reach and cross a target lesion.
{5}------------------------------------------------
# Performance Testing:
K120119 P- 6 of 6
The following tests were performed on the Sunmed Guide Wires:
Degradation Test on Disposable Guide Wire Physical Performance Appearance Radiodetectability Eroding resistance Breakage resistance Deformation resistance Connection strength Package Performance Sterility ETO Residue ECH Residue
#### Technological Characteristics:
Summary of the technological characteristics of our device compared to the predicate device:
The predicates were compared in the following areas and found to have identical technological characteristics and to be equivalent because they are the identical products.
- Indications for Use Target Population Design Technique Sterility Biocompatibility Anatomical Sites
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the symbol. The caduceus is depicted in black, and the text is also in black, creating a monochrome design.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAY 1 1 . 2012
Sunny Medical Device (Shenzhen) Co., Ltd. c/o Mr. Greg Holland Regulatory Specialists, Inc. 3722 Ave. Sausalito Irvine, CA 92606
Re: K120119
> Sunmed Guide Wire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: April 20, 2012 Received: May 10, 2012
Dear Mr. Holland:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
#### Page 2 – Mr. Greg Holland
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
W. G. Hillen
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# 4. Indications for Use Statement
# Indications for Use
510(k) Number (if known): K120119
Device Name: Sunmed Guide Wires
Indications for Use:
Sunmed Guidewires are used for the introduction and placement of diagnostic or interventional devices in the coronary and peripheral vasculature and may be used to reach and cross a target lesion.
Prescription Use >>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>> (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
M. A. Kildeer
diovascular Devices
510(k) Number K120119
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.