K120075 · Advanced Instrumentations, Inc. · BSZ · Jan 26, 2012 · Anesthesiology
Device Facts
Record ID
K120075
Device Name
AM-6000 ANESTHESIA MACHINE
Applicant
Advanced Instrumentations, Inc.
Product Code
BSZ · Anesthesiology
Decision Date
Jan 26, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5160
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The AM-6000 Anesthesia machine is intended for general anesthesia use. The AM-6000 Anesthesia machine will delivery operator set concentrations of oxygen and anesthesia gases as well as deliver controlled breaths to the patient with either a constant or a deceleration flow pattern. AM-6000 Anesthesia machine is also intended to allow for the provision of manual ventilation. The AM-6000 Anesthesia machine is intended for use by Healthcare professionals who are trained in the administration of anesthesia. The AM-6000 Anesthesia machine is intended for use on the neonatal to adult patient populations in all ventilation modes. The AM-6000 Anesthesia machine is intended to be used in the environments where anesthesia is to be administered by Healthcare professionals trained in administering anesthesia.
Device Story
Continuous flow anesthesia machine; provides manual or automatic ventilation, fresh gas delivery, and anesthetic agent delivery. Inputs: operator-set concentrations of oxygen and anesthesia gases. Operation: electronic PEEP, dual flow tubes for gas monitoring, two vaporizers, and heated breathing system to minimize condensation. Outputs: controlled breaths (constant or deceleration flow patterns) and manual ventilation. Used in clinical settings by trained healthcare professionals. Features battery backup for power outages. Benefits: provides controlled anesthesia delivery and ventilation support across patient populations while reducing risk of hypoxic mixtures and inadvertent flow control adjustments.
Clinical Evidence
No clinical data. Bench testing only. Device software validated per FDA guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
Technological Characteristics
Continuous flow anesthesia machine; dual flow tubes; two vaporizers; heated breathing system. Ventilation modes: CMV, PCV, SIMV, PSV with electronic PEEP. Battery backup included. Software validated per FDA guidance.
Indications for Use
Indicated for general anesthesia administration in neonatal to adult patient populations. Intended for use by trained healthcare professionals in clinical environments where anesthesia is administered. Contraindications not specified.
Regulatory Classification
Identification
A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).
{0}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
120075 The assigned 510(k) number is:
Submitter
Advanced Instrumentations, Inc. 6800 N.W. 77" Court Miami, FI 33166 305-477-6331 Telephone: 305-477-5351 Fax:
Registration # 1066270
### Official correspondent :
Jorge Millan, PhD Email: jmillan@hiatec.org 601 West 20 St Hialeah, FL 33010 Phone : (305) 925-1260
Date Prepared:
January 6, 2012
Device name and classification:
- AM-6000 Anesthesia Machine . Device Name:
- Classification Name: 868.5160 System, Gas-Machine, Anesthesia ● Product code: BSZ
- II Regulatory Class: ●
{1}------------------------------------------------
AM-6000 Anesthesia Machine.
### Predicate Device:
Navigator Anesthesia Delivery System. K080175 Manufacturer: AS3000 Datascope Corp.
### Device Description:
The AM-6000 Anesthesia Machine is a continuous flow Anesthesia Machine which offers manual or automatic vehtilation, easily adjustable fresh gas delivery, anesthetic agent delivery, ventilation monitoring, convenient ergonomics, and state-of-the-art safety systems. The AM-6000 is designed to decrease the risk of hypoxic mixtures and the inadvertent movement of the air flow control knobs. Additionally, the AM-6000 provides battery power in the event of an AC power outage.
Multiple ventilation modes, i.e., CMV, PCV, SIMV and PSV, are offered by the AM-6000 with electronic PEEP available in each of the modes. The fresh gas dosing subsystem offers features of a traditional Anesthesia Machine along with dual flow tubes which display the gas flows at all times. The AM-6000 contains two vaporizers and a heated breathing system to minimize condensation and return moisture to the patient.
#### Indications for Use:
The AM-6000 Anesthesia machine is intended for general anesthesia use. The AM-6000 Anesthesia machine will delivery operator set concentrations of oxygen and anesthesia gases as well as deliver controlled breaths to the patient with either a constant or a deceleration flow pattern. AM-6000 Anesthesia machine is also intended to allow for the provision of manual ventilation.
Intended operator:
The AM-6000 Anesthesia machine is intended for use by Healthcare professionals who are trained in the administration of anesthesia.
Intended Patient Populations:
The AM-6000 Anesthesia machine is intended for use on the neonatal to adult patient populations in all ventilation modes.
Intended Use Environment:
The AM-6000 Anesthesia machine is intended to be used in the environments where anesthesia is to be administered by Healthcare professionals trained in administering anesthesia.
{2}------------------------------------------------
# Effectiveness and Safety Contraindications:
### Clinical Test
Clinical testing is not required
### Non-clinical test:
Non-Childer, Color Anesthesia Machine has been tested and found to be in The AM-6000 Anesthesia Machine has been lested and found to and to and and compatibility standards.
compansion in the been developed to identify potential hazards and document
A risk and the becards. The device's software has been validated in A risk analysis has been developed to ldentify potchtian has been validated in
the mitigation of the hazards. The device's software has been validated in A the mitigation of the hazards. The devices sontware for the Content
accordance with the requirements set forth in the FDA Guidanes for the Content the more with the requirements set form in the PDA Guidanee Contained in Medical Devices (May 11, 2017) 2005).
### Comparison to the predicate device:
on has similar technology characteristics and has the same
The subject device to tarial components, same manufacturing process, same The subject device has similar technology charactering process, same
intended use, same material components, same measurement mode and intended use, same material classification, same measurement mode and
design principle, same electrication, same measurement mode and
intended use, same electrical classifica design principle; sun predicate device.
# Substantially Equivalent Determination:
Verification and validation testing was done on the AM-6000 Anesthesia Verification and validation testing was done on "the "Ath" on the Press.
Machine. This premarket notification demonstrates that AM-6000
Machine. This premains that starticles Machine. This premarket notification Submission Submittee device.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three stripes forming its wing and tail feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Advanced Instrumentations, Incorporated C/O Dr. Jorge Millan Official Correspondent Hialeah Technology Center, Incorporated West 20th Street Hialeah, Florida 33010
JAN 2 6 2012
Re: K120075
Trade/Device Name: AM-6000 Anesthesia Machine Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine for Anesthesia or Analgesia Regulatory Class: II Product Code: BSZ Dated: January 6, 2012 Received: January 10, 2012
Dear Dr. Millan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{4}------------------------------------------------
#### Page 2 - Dr. Millan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Susan Runoie
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
510(k) Number (if known):
Device Name:
AM-6000 Anesthesia Machine
Indications for Use:
Indications for of on the is intended for general anesthesia use. The The AM-6000 Anesthesia machine is intended for generations of oxygen
AM-6000 Anesthesia machine will delivery operator set concentrations of oxygen He All-6000 Anesthesia machine will deliver controlled breaths to the path with with with with with with with with AM-6000 And either a constant of a deceleration now pation of manual ventilation.
also intended to allow for the provision of manual ventilation.
Intended operator:
Intended operator:
The AM-6000 Anesthesia machine is intended for use by Healthcare
the AM-6000 Annest sinced in the administration of anesthesia. The AM-6000 Anesthesia machine Is Internation of anesthesia.
professionals who are trained in the administration of anesthesia.
Intended Patient Populations:
Intended Patient Populations
The AM-6000 Anesthesia machine is intended for use on the neonatal to adult
in the AM-6000 Anaily of vestigation modes The AM-6000 Ancothoold The All ventilation modes.
Intended Use Environment:1
Intended Use Environment.
The AM-6000 Anesthesia machine is intended to be used in the environments
in The AM-6000 Anesthesia machine is intended to be used in the ender the comments.
where anesthesia is to be administered by Healthcare professionals trained in administering anesthesia.
inistering anesthesia.
Prescription Use _X_ | resoription Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schulten
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of __ 1__
510(k) Number: K120075
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.