K120060 · Alpinion Medical Systems Co., Ltd. · IYN · Mar 16, 2012 · Radiology
Device Facts
Record ID
K120060
Device Name
E-CUBE 9
Applicant
Alpinion Medical Systems Co., Ltd.
Product Code
IYN · Radiology
Decision Date
Mar 16, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device is intended for use by a qualified physician for the evaluation of soft tissue and blood flow in the clinical applications; Fetal; Abdominal (renal & GYN/pelvic); Pediatric; Small Organ (breast, testes, thyroid); Trans-rectal(TR); Trans-vaginal(TV); Musculo-skeletal(Conventional); Musculo-skeletal (Superficial); Cardiac (adult & pediatric); Peripheral Vascular (PV); and Urology (including prostate).
Device Story
E-CUBE 9 is an ultrasonic pulsed Doppler imaging system; used in clinical settings (e.g., abdominal, cardiac, OB/GYN) by qualified physicians. Device inputs ultrasound signals via various transducers (e.g., SC1-6, L3-12H, C1-6); processes signals to generate diagnostic images (B-mode, M-mode, PWD, Color/Power Doppler, Tissue Harmonic Imaging, 3D/4D). System features wide flat panel display and ergonomic control panel for user interface. Output allows physicians to evaluate soft tissue and blood flow; aids in clinical diagnosis and patient management. No remote processing or cloud submission described.
Clinical Evidence
No clinical studies were required to support substantial equivalence. Evidence is based on bench testing, including biocompatibility, acoustic output, and safety compliance (thermal, electrical, electromagnetic, mechanical).
Technological Characteristics
Ultrasonic pulsed Doppler imaging system. Includes various transducers (e.g., SC1-6, L3-12H, C1-6, SP1-5, E3-10, L3-8, SP3-8). Features B-mode, M-mode, PWD, Color/Power Doppler, Tissue Harmonic Imaging, 3D/4D modes. Biocompatible patient-contact materials. Complies with medical device safety standards for electrical, thermal, and mechanical safety.
Indications for Use
Indicated for use by qualified physicians for diagnostic ultrasound imaging or fluid flow analysis of soft tissue in fetal, abdominal, pediatric, small organ, trans-rectal, trans-vaginal, musculo-skeletal, cardiac, peripheral vascular, and urological clinical applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
LOGIQ P5/A5 Diagnostic Ultrasound System (K060993)
Submission Summary (Full Text)
{0}------------------------------------------------
MAR 1 6 2012
(120060)
Page 1 at 2
510(K)-E-CUBE 9
**510(k) Summary**
In accordance with 21CFR807.92, the following summary of information is provided;
| <b>Date</b> | Jan 5 <sup>th</sup> 2012 |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Submitter:</b> | ALPINION MEDICAL SYSTEMS Co., Ltd.<br>Address: 1,6 and 7FL, 222-22, Guro-dong, Guro-gu, Seoul, 152-848, Korea Republic |
| <b>Primary Contact Person</b> | Donghwan Kim<br>QARA Manager<br>Address: 1,6 and 7FL, 222-22, Guro-dong, Guro-gu, Seoul, 152-848, Korea Republic<br>Phone: +82 70 7465 2068<br>Fax: +82 2 851 5594<br>Email: donghwan.kim@alpinion.com |
| <b>Secondary Contact Person</b> | Yuchi Chu<br>Address: Suite 229, 10604 NE 38th Place, Kirkland, WA 98033, United States<br>Phone: 425 949 4907<br>Fax: 425 949 4908<br>Email: ychu@alpinionus.com |
| <b>Device Trade Name:</b> | E-CUBE 9 |
| <b>Common/Usual Name:</b> | Ultrasonic Pulsed Doppler Imaging System |
| <b>Classification Names</b> | System, Imaging, Pulsed Doppler Ultrasonic |
| <b>Product Code:</b> | Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550 90-IYN<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO<br>Diagnostic Ultrasound Transducer, 21CFR 892.1570, 90-ITX |
| <b>Predicate Device(s)</b> | K060993 LOGIQ P5/A5 Diagnostic Ultrasound System |
ALPINION MEDICAL SYSTEMS Co., Ltd.
.
.
{1}------------------------------------------------
K120060
Page 2 of 2
E-CUBE 9 product is an ultrasound imaging system for medical Device Description: diagnosis. This product can be used for the applications of abdominal, obstetrics, gynecology, small parts, cardiology, vascular, etc.
> The system platform provides optimal patient diagnosis workflow with the wide flat panel display, ergonomic control panel with easy user interface, optimal image quality
Indications For Use:
The device is intended for use by a qualified physician for the evaluation of soft tissue and blood flow in the clinical applications; Fetal; Abdominal (renal & GYN/pelvic); Pediatric; Small Organ (breast, testes, thyroid); Trans-rectal(TR); Trans-vaginal(TV); Musculo-skeletal(Conventional); Musculo-skeletal (Superficial); Cardiac (adult & pediatric); Peripheral Vascular (PV); and Urology (including prostate).
Technology:
Determination of Substantial Equivalence:
E-CUBE 9 employs the same fundamental scientific technology as its predicate device.
#### Summary of Non-Clinical Tests:
E-CUBE 9 has been evaluated for biocompatibility, acoustic output as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform to applicable medical device safety standards. E-CUBE 9 and its application comply with voluntary standards as detailed in this premarket submission. The following quality management system measures were applied to the development of E-CUBE 9:
- Medical Device Risk Management �
- Requirements Reviews �
- Design Reviews �
- Component Verification �
- Integration Review (System Verification) �
- Performance Testing (System Verification) �
- Safety Testing (Compliance Test) �
- � Design Validation
Transducer materials and other patient contact materials are biocompatible.
#### Summary of Clinical Tests:
The subject of this premarket submission, E-CUBE 9, did not require clinical studies to support substantial equivalence.
Alpinion Medical Systems Co., Ltd. Considers E-CUBE 9 to be as Conclusion: safe, as effective, and performance is substantially equivalent to the predicate device.
ALPINION MEDICAL SYSTEMS Co., Ltd. will update and include in this summary any other information deemed reasonably necessary by the FDA or the requirements will be published in guidance documents.
ALPINION MEDICAL SYSTEMS Co., Ltd.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" written around it.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Alpinion Medical Systems Co., Ltd. % Mr. Yuchi Chu Correspondent AUS Co., Inc. 10604 NE 38th Place, Suite 229 KIRKLAND WA 98033
Re: K120060
Trade/Device Name: E-CUBE 9 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: January 9, 2012 Received: January 9, 2012
Dear Mr. Chu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the E-CUBE 9, as described in your premarket notification:
#### Transducer Model Number
| SC1-6 | L3-12H |
|--------|--------|
| C1-6 | SP1-5 |
| SVC1-6 | E3-10 |
| VC1-6 | L3-8 |
| L3-12 | SP3-8 |
MAR. 1 6 2012
{3}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Joshua Nipper at (301) 796-6524.
Sincerely Yours,
Janine M. Morris
Acting Directd Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K111864
Device Name: E-CUBE 9
Indications for Use:
The device is intended for use by a qualified physician for the evaluation of soft tissue and blood flow in the clinical applications; Fetal; Abdominal (renal & GYN/pelvic); Pediatric; Small Organ (breast, testes, thyroid); Trans-rectal(TR); Trans-vaginal(TV); Musculo-skeletal(Conventional); Musculo-skeletal (Superficial); Cardiac (adult & pediatric); Peripheral Vascular (PV); and Urology (including prostate).
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
(Division Sign-Off)
Division Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K. K120060
ALPINION MEDICAL SYSTEMS Co., Ltd.
{5}------------------------------------------------
Diagnostic Ultrasound Indications for Use Form
## E-CUBE 9 Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | |
|------------------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| | B | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic | | | | | | | | | |
| Fetal | P | P | P | | P | P | | P | P |
| Abdominal | P | P | P | | P | P | P | P | P |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | N | N | N | | N | N | N | N | N |
| Small Organ<br>(breast, testes, thyroid) | P | P | P | | P | P | N | P | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | P | P | P | | P | P | | P | |
| Trans-vaginal | P | P | P | | P | P | | P | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | P | P | P | | P | P | N | P | |
| Musculo-skeletal<br>(Superficial) | P | P | P | | P | P | N | P | |
| Intravascular | | | | | | | | | |
| Cardiac Adult | P | P | P | N | P | P | P | P | |
| Cardiac Pediatric | N | N | N | N | N | N | N | N | |
| Intravascular (Cardiac) | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Intra-cardiac | | | | | | | | | |
| Peripheral vessel | P | P | P | | P | P | N | P | |
| Urology (including prostate) | N | N | N | | N | N | N | N | |
N = new indication; P = previously cleared by FDA; E = added under appendix
* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
E-2
Prescription User (Per 21 CFR 801.109)
ALPINION MEDICAL SYSTEMS Co., Ltd.
sion Sign Division of Radiologica! Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510K =
{6}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
## E-CUBE 9 with SC1-6 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | Combined*<br>(Specify) | Other**<br>(Specify) |
|------------------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| | B | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | | |
| Ophthalmic | | | | | | | | | |
| Fetal | P | P | P | | P | P | P | P | |
| Abdominal | P | P | P | | P | P | P | P | |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | N | N | N | | N | N | N | N | |
| Small Organ<br>(breast, testes, thyroid) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | | | | | | | | | |
| Musculo-skeletal<br>(Superficial) | | | | | | | | | |
| Intravascular | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Intra-cardiac | | | | | | | | | |
| Peripheral vessel | | | | | | | | | |
| Urology (including prostate) | N | N | N | | N | N | N | N | |
N = new indication; P = previously cleared by FDA; E = added under appendix
* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
| Prescription User (Per 21 CFR 801.109) |
|------------------------------------------------------------|
| ALPINION MEDICAL SYSTEMS Co., Ltd. |
| (Division Sign-Off) |
| Division of Radiological Devices |
| Office of In Vitro Diagnostic Device Evaluation and Safety |
| 510K |
| K120060 |
{7}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
## E-CUBE 9 with C1-6 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | |
|------------------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| | B | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic | | | | | | | | | |
| Fetal | P | P | P | | P | P | P | P | |
| Abdominal | P | P | P | | P | P | P | P | |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | N | N | N | | N | N | N | N | |
| Small Organ<br>(breast, testes, thyroid) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | | | | | | | | | |
| Musculo-skeletal<br>(Superficial) | | | | | | | | | |
| Intravascular | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Intra-cardiac | | | | | | | | | |
| Peripheral vessel | | | | | | | | | |
| Urology (including prostate) | N | N | N | | N | N | N | N | |
N = new indication; P = previously cleared by FDA; E = added under appendix
* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
## The even do not write below this line-continue on another bage in meeded) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
| Prescription User (Per 21 CFR 801.109) | |
|------------------------------------------------------------------------------------------------|---------|
| ALPINION MEDICAL SYSTEMS Co., Ltd. | |
| (Division Sign-Off) | |
| Division of Radiological Devices<br>Office of In Vitro Diagnostic Device Evaluation and Safety | |
| 510K | K120868 |
{8}------------------------------------------------
Diagnostic Ultrasound Indications for Use Form
## E-CUBE 9 with SVC1-6 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|------------------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|--|
| | B | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) | |
| Ophthalmic | | | | | | | | | | |
| Fetal | P | P | P | | P | P | P | P | P | |
| Abdominal | P | P | P | | P | P | P | P | P | |
| Intra-operative (Specify) | | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Pediatric | N | N | N | | N | N | N | N | N | |
| Small Organ<br>(breast, testes, thyroid) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | | | | | | | | | | |
| Musculo-skeletal<br>(Superficial) | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Urology (including prostate) | N | N | N | | N | N | N | N | N | |
N = new indication; P = previously cleared by FDA; E = added under appendix
* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
E-2
(Prescription User (Per 21 CFR 801.109)
ALPINION MEDICAL SYSTEMS Co., Ltd.
(Division Sign-Off)
Division of Radiological Devices
ામદિક વા on and Satery
ETOK
{9}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
# E-CUBE 9 with VC1-6 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | |
|------------------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| | B | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic | | | | | | | | | |
| Fetal | P | P | P | | P | P | P | P | P |
| Abdominal | P | P | P | | P | P | P | P | P |
| Intra-operative (Specify) | . | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | N | N | N | . | N | N | N | N | N |
| Small Organ<br>(breast, testes, thyroid) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | | | | | | | | | |
| Musculo-skeletal<br>(Superficial) | | | | | | | | | |
| Intravascular | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Intra-cardiac | | | | | | | | | |
| Peripheral vessel | | | | | | | | | |
| Urology (including prostate) | N | N | N | | N | N | N | N | |
N = new indication; P = previously cleared by FDA; E = added under appendix
* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
## (Please do not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
| Prescription User (Per 21 CFR 801.109) |
|----------------------------------------|
|----------------------------------------|
ALPINION MEDICAL SYSTEMS Co., Ltd.
| (Division Sign-Off) |
|------------------------------------------------------------|
| Division of Radiological Devices |
| Office of Na Vitro Diagnostic Device Evaluation and Safety |
510K. K120060
{10}------------------------------------------------
Diagnostic Ultrasound Indications for Use Form
## E-CUBE 9 with L3-12 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | |
|------------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| | B | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic | | | | | | | | | |
| Fetal | | | | | | | | | |
| Abdominal | | | | | | | | | |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | N | N | N | N | N | N | N | | |
| Small Organ | P | P | P | | P | P | P | P | |
| (breast, testes, thyroid) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | P | P | P | | P | P | N | P | |
| Musculo-skeletal<br>(Superficial) | P | P | P | | P | P | N | P | |
| Intravascular | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Intra-cardiac | | | | | | | | | |
| Peripheral vessel | P | P | P | | P | P | N | P | |
| Urology (including prostate) | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under appendix
* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
| | Prescription User (Per 21 CFR 801.109) |
|------|------------------------------------------------------------|
| | ALPINION MEDICAL SYSTEMS Co., Ltd. |
| | (Division Sign-Off) |
| | Division of Radiological Devices |
| | Office of In Vitro Diagnostic Device Evaluation and Safety |
| 510K | |
{11}------------------------------------------------
# Diagnostic Ultrasound Indications for Use Form
## E-CUBE 9 with L3-12H Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) | |
|------------------------------------------|-------------------|---|-----|-----|------------------|-------------------------------|------------------------|----------------------|------------------|
| | B | M | PWD | CWD | Color<br>Doppler | | | | Power<br>Doppler |
| Ophthalmic | | | | | | | | | |
| Fetal | | | | | | | | | |
| Abdominal | | | | | | | | | |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | N | N | N | | N | N | | N | |
| Small Organ<br>(breast, testes, thyroid) | P | P | P | | P | P | | P | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | P | P | P | | P | P | | P | |
| Musculo-skeletal<br>(Superficial) | P | P | P | | P | P | | P | |
| Intravascular | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Intra-cardiac | | | | | | | | | |
| Peripheral vessel | P | P | P | | P | P | | P | |
| Urology (including prostate) | | | | | | | | | |
N = new indication, P = previously cleared by FDA; E = added under appendix
* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF MEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
ALPINION MEDICAL SYSTEMS Co., Ltd.
(Division Sign-OH)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
610K.
{12}------------------------------------------------
Diagnostic Ultrasound Indications for Use Form
## E-CUBE 9 with SP1-5 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | |
|------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| | B | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic | | | | | | | | | |
| Fetal | | | | | | | | | |
| Abdominal | P | P | P | | P | P | P | P | |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | N | N | N | | N | N | N | N | |
| Small Organ | | | | | | | | | |
| (breast, testes, thyroid) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethrai. | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | |
| (Conventional) | | | | | | | | | |
| Musculo-skeietal | | | | | | | | | |
| (Superficial) | | | | | | | | | |
| Intravascular | | | | | | | | | |
| Cardiac Adult | P | P | P | N | P | P | P | P | |
| Cardiac Pediatric | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Intra-cardiac | | | | | | | | | |
| Peripheral vessel | | | | | | | | | |
| Urology (including prostate) | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under appendix
* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANQTHER PAGE IF NEEDED) Concurrence of CDRH; Office of In-Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
ALPINION MEDICAL SYSTEMS Co., Ltd.
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K. K120068
E-d
{13}------------------------------------------------
Diagnostic Ultrasound Indications for Use Form
## E-CUBE 9 with E3-10 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | |
|------------------------------------------|-------------------|---|-----|-----|---------------|---------------|-------------------------|---------------------|-------------------|
| | B | M | PWD | CWD | Color Doppler | Power Doppler | Tissue Harmonic Imaging | Combined* (Specify) | Other** (Specify) |
| Ophthalmic | | | | | | | | | |
| Fetal | | | | | | | | | |
| Abdominal | | | | | | | | | |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | | | | | | | | | |
| Small Organ<br>(breast, testes, thyroid) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | P | P | P | | P | P | | P | |
| Trans-vaginal | P | P | P | | P | P | | P | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | | | | | | | | | |
| Musculo-skeletal<br>(Superficial) | | | | | | | | | |
| Intravascular | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Intra-cardiac | | | | | | | | | |
| Peripheral vessel | | | | | | | | | |
| Urology (including prostate) | N | N | N | | N | N | | N | |
N = new indication; P = previously cleared by FDA; E = added under appendix
* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IE NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
ALPINION MEDICAL SYSTEMS Co., Ltd.
Division of Radiological Devices Office of In Vitro Diagnostic luation and Safety
510K Kianna66
{14}------------------------------------------------
Diagnostic Ultrasound Indications for Use Form
## E-CUBE 9 with L3-8 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | |
|------------------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| | B | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic | | | | | | | | | |
| Fetal | | | | | | | | | |
| Abdominal | | | | | | | | | |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | N | N | N | | N | N | | N | |
| Small Organ<br>(breast, testes, thyroid) | N | N | N | | N | N | | N | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | N | N | N | | N | N | | N | |
| Musculo-skeletal<br>(Superficial) | N | N | N | | N | N | | N | |
| Intravascular | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Intra-cardiac | | | | | | | | | |
| Peripheral vessel | N | N | N | | N | N | | N | |
| Urology (including prostate) | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under appendix
* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IE NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
| | ALPINION MEDICAL SYSTEMS Co., Ltd. |
|------|------------------------------------------------------------|
| | (Division Sign-Off) |
| | Division of Radiological Devices |
| 510K | Office of in Vitro Diagnostic Device Evaluation and Safety |
| | K120068 |
{15}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
## E-CUBE 9 with SP3-8 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | |
|------------------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| | B | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic | | | | | | | | | |
| Fetal | | | | | | | | | |
| Abdominal | N | N | N | | N | N | N | N | |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | N | N | N | | N | N | N | N | |
| Small Organ<br>(breast, testes, thyroid) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | | | | | | | | | |
| Musculo-skeletal<br>(Superficial) | | | | | | | | | |
| Intravascular | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | N | N | N | | N | N | N | N | |
| Intravascular (Cardiac) | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Intra-cardiac | | | | | | | | | |
| Peripheral vessel | | | | | | | | | |
| Urology (including prostate) | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under appendix
* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
## (Please do not write below this line-continue on another Page if needed) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
ALPINION MEDICAL SYSTEMS Co., Ltd.
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K K120068
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.