← Product Code [LGZ](/productcode/LGZ) · K120026

# PERFORMER HT/ HANG&GO HT BASIC (K120026)

_Rand S.R.L. · LGZ · May 8, 2012 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K120026

## Device Facts

- **Applicant:** Rand S.R.L.
- **Product Code:** [LGZ](/productcode/LGZ.md)
- **Decision Date:** May 8, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5725
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The intended use of the PERFORMER HT is to provide isolated hyperthermic perfusion in the thoracic or peritoneal cavity by means of extra-corporeal circulation of warmed, physiologically compatible sterile solution, according to a protocol to be selected by the physician. The device must be used by a qualified medical professional who is experienced in the operation of this or similar equipment.

## Device Story

PERFORMER HT is a prescription electromechanical device for isolated hyperthermic perfusion in thoracic or peritoneal cavities. It uses an extracorporeal circuit with a disposable kit (HANG&GO HT BASIC) to circulate and heat sterile solutions. Input data includes patient information, temperature probes, pressure sensors, and flow rates. The device features roller pumps to circulate fluid through a heater exchanger to reach target temperatures. A central information display (CID) allows clinicians to set treatment parameters, monitor real-time graphical trends, and manage alarm limits. The system automatically detects air in the circuit, monitors pressures/volumes, and triggers audible alarms or stops operation during hazardous conditions. Used in clinical settings by qualified medical professionals, the device enables controlled hyperthermic treatment, potentially benefiting patients by maintaining precise therapeutic temperatures during perfusion protocols.

## Clinical Evidence

Bench testing only. Testing verified system ability to heat and pump fluids accurately, maintain temperature, control pressure/volume/flow, detect air, and trigger alarms for unsafe conditions. Human factors study conducted in a simulated environment to validate user interface and mitigate task-related risks.

## Technological Characteristics

Electromechanical system with roller-type fluid pumps and touch-screen interface. Includes disposable sterile kit (HANG&GO HT BASIC). Monitors temperature, pressure, flow, and air detection. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), ISO 10993 (biocompatibility), and IEC 62304 (software lifecycle). Sterilization via ethylene oxide. Connectivity includes non-volatile memory for parameter storage.

## Regulatory Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

## Predicate Devices

- Belmont Hyperthermia Pump ([K070654](/device/K070654.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K120-026

## 5. 510(k) Summarv

MAY - 8 2012

## 21 CFR 807.92(a)(1): Submitter and Manufacturing site information RanD S.r.l. via Statale 12. 62 41036 Medolla (MO) - ITAL Y Telephone: +39 0535 49283 Fax: +39 0535 660636 Contact person: Simone Ceretti

## Establishment registration number: 3003793891

### Date of the present summary preparation: 28th December, 2011

| 21 CFR 807.92(a)(2): Device information |                                    |  |  |
|-----------------------------------------|------------------------------------|--|--|
| Trade name                              | PERFORMER HT / HANG&GO HT BASIC    |  |  |
| Device Classification Name              | Warmer, thermal, infusion fluid    |  |  |
| Classification Review Panel             | General Hospital                   |  |  |
| Classification Product Code             | LGZ                                |  |  |
| Regulatory Class                        | Unclassified, pre-amendment device |  |  |

### 21 CFR 807.92(a)(3): Predicate devices

The PERFORMER HT is substantially equivalent to the Belmont Hyperthermia Pump (K070654, dtd June 8th, 2007).

## 21 CFR 807.92(a)(4): Description of the device

The PERFORMER HT is a prescription electromechanical device intended to provide isolated hyperthermic perfusion in the thoracic or peritoneal cavity by means of extracorporeal circulation of warmed. physiologically compatible sterile solution.

The PERFORMER HT equipment is designed to:

- operate the treatment execution by means of a color touch-screen display .
- . control the temperature of solution perfused to the Patient through the extracorporeal circuit
- control the flow rates of roller pumps .
- control the volume in the extracorporeal circuit and Patient's body cavity .
- monitor Patient's body cavity temperatures .
- monitor the extracorporeal circuit pressures .
- . detect air in the circuit
- change the fluid path in the extracorporeal circuit by means of 2 two-ways pinch valves .

The Central Information Display (CID) enables the User to interact with the PERFORMER HT equipment. The touch-screen display allows the User to:

- inout Patient information .
- operate the various treatment phases
- display treatment parameters (e.g, flows, pressures, temperatures) .
- set treatment parameters .
- . set alarm limits
- display graphical trending of temperature probes connected to the HTS module

{1}------------------------------------------------

- · read alarm and alert messages
The PERFORMER HT is intended to be used with the HANG&GO HT BASIC, a disposable, single use kit intended to circulate and filtrate the clear solutions in the thoracic or abdominal cavity.

The HANG&GO HT BASIC includes all necessary lines and accessories for the treatment execution:

- · circulation line
- ·. heater bag
- · reservoir
- · table pack (Patient inlet and outlet lines) with thermal protections

The HANG&GO HT BASIC is supplied sterile, packaged in a proprietary-design container, and is intended to be used only with the PERFORMER HT.

### 21 CFR 807.92(a)(5): Intended use

The intended use of the PERFORMER HT is to provide isolated hyperthermic perfusion in the thoracic or peritoneal cavity by means of extra-corporeal circulation of warmed. physiologically compatible sterile solution, according to a protocol to be selected by the physician.

The device must be used by a qualified medical professional who is experienced in the operation of this or similar equipment.

### Predicate device intended use

The intended use of the Belmont Hyperthermia Pump is to raise the temperature of the thoracic or peritoneal cavity to the desired target temperature by continuously lavaging the cavity with circulating warmed sterile solution, according to a protocol to be selected by the physician.

#### 21 CFR 807.92(a)(6): Comparison of technological characteristics

The PERFORMER HT has the same operating principle, energy type, environmental specifications, and performance specifications of the Belmont device. It uses roller-type fluid pumps, touch screen to direct the User through set-up preparation and use, control and monitoring of fluid temperatures, and a proprietary-design disposable set including large fluid reservoir to circulate sterile fluid into and out of the body cavity. Flow from the reservoir is delivered by action of the roller pump to the heater exchanger to be heated to the desired target temperature, and then to the Patient, From the Patient, the fluid is pumped from the return line back into the reservoir. The PERFORMER HT monitors circulating flow, temperatures, pressures and volumes, detects air in the circuit and automatically responds to ineffective or unsafe operating conditions by sounding audible alarms, stopping heating and pumping in case of hazardous conditions for User/Patient.

{2}------------------------------------------------

# 21 CFR 807.92(b)(1): Discussion of the non-clinical tests

The PERFORMER HT is designed to, complies with, and has been tested to, the FDA Recognized Consensus Standards listed in the table below, as applicable to the equipment, the disposable lines set and the components:

| Area                      | US Recognized standard #                    | EU Recognized standard #                   | Title                                                                                                                                                                                 |
|---------------------------|---------------------------------------------|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General                   | ISO 14971:2007                              | EN ISO 13485:2003 / AC:2007                | Medical devices -- Quality management systems -- Requirements for regulatory purposes                                                                                                 |
|                           |                                             | EN ISO 14971:2007                          | Medical devices - Application of risk management to medical devices                                                                                                                   |
|                           |                                             |                                            |                                                                                                                                                                                       |
| Packaging and<br>labeling |                                             | EN 980:2008                                | Symbols for use in the labelling of medical devices                                                                                                                                   |
|                           |                                             | EN 1041:2008                               | Information supplied by the manufacturer with medical devices                                                                                                                         |
|                           |                                             | ASTM D4169-04a                             | Standard Practice for Performance Testing of Shipping Containers and Systems                                                                                                          |
|                           |                                             |                                            |                                                                                                                                                                                       |
| Biocompatibility          | AAMI ANSI ISO 10993-1:2009                  | EN ISO 10993-1:2009                        | Biological evaluation of medical devices - Part 1: Evaluation and testing                                                                                                             |
|                           |                                             | EN ISO 10993-4:2002                        | Biological evaluation of medical devices - Part 4: Selection of tests for interaction with blood                                                                                      |
|                           | AAMI ANSI ISO 10993-5:2009                  | EN ISO 10993-5:2009                        | Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity                                                                                                    |
|                           | AAMI ANSI ISO 10993-7:2008                  | EN ISO 10993-7:2009                        | Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals                                                                                             |
|                           | ISO 10993-10:2002/Amd<br>1:2006             | EN ISO 10993-<br>10:2002/Amd 1:2006        | Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity                                                                            |
|                           | AAMI ANSI ISO 10993-<br>11:2006             | EN ISO 10993-11:2006                       | Biological evaluation of medical devices - Part 11: Tests for systemic toxicity                                                                                                       |
|                           | IEC 60601-1:1988/Amd<br>1:1991 & Amd 2:1995 | EN 60601-1:1990/Amd<br>1:1993 & Amd 2:1995 | Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995                                                                 |
|                           |                                             |                                            |                                                                                                                                                                                       |
|                           |                                             |                                            |                                                                                                                                                                                       |
| Electrical safety         | IEC 60601-1-1:2000                          | EN 60601-1-1:2001                          | Medical electrical equipment - Part 1-1: General requirements for safety -<br>Collateral standard: Safety requirements for medical electrical systems                                 |
|                           | AAMI ANSI IEC 60601-<br>1-2:2001            | EN 60601-1-2:2001                          | Medical Electrical Equipment - Part 1-2: General Requirements for Safety -<br>Collateral Standard: Electromagnetic Compatibility - Requirements and Tests                             |
|                           |                                             |                                            |                                                                                                                                                                                       |
|                           | IEC 60601-1-4:2000                          | EN 60601-1-<br>4:1996/Amd 1:1999           | Medical electrical equipment - Part 1-4:<br>General requirements for safety -<br>Collateral standard: Programmable electrical medical systems                                         |
|                           |                                             |                                            |                                                                                                                                                                                       |
|                           |                                             |                                            |                                                                                                                                                                                       |
|                           |                                             | EN 60601-1-6:2004                          | Medical electrical equipment -- Part 1-6:<br>General requirements for basic safety and essential performance - Collateral Standard: Usability                                         |
|                           |                                             |                                            |                                                                                                                                                                                       |
|                           |                                             |                                            |                                                                                                                                                                                       |
|                           |                                             |                                            |                                                                                                                                                                                       |
|                           |                                             |                                            |                                                                                                                                                                                       |
| Sterility                 | AAMI ANSI ISO 11135-<br>1:2007              | EN ISO 11135-1:2007                        | Sterilization of health care products -<br>Ethylene oxide - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices. |
|                           |                                             |                                            |                                                                                                                                                                                       |
|                           |                                             |                                            |                                                                                                                                                                                       |
|                           |                                             |                                            |                                                                                                                                                                                       |
|                           | AAMI ANSI ISO 11607-<br>1:2006              | EN ISO 11607-1:2009                        | Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems                                               |
|                           |                                             |                                            |                                                                                                                                                                                       |
|                           |                                             |                                            |                                                                                                                                                                                       |
|                           |                                             |                                            |                                                                                                                                                                                       |
|                           |                                             |                                            |                                                                                                                                                                                       |
|                           | AAMI ANSI ISO 11607-<br>2:2006              | EN ISO 11607-2:2006                        | Packaging for terminally sterilized<br>medical devices - Part 2: Validation<br>requirements for forming, sealing and<br>assembly processes                                            |
|                           | ASTM F1980-07                               |                                            | Standard Guide for Accelerated Aging of<br>Sterile Barrier Systems for Medical<br>Devices.                                                                                            |
|                           | AAMI ANSI ISO 11737-<br>1:2006              | EN ISO 11737-1:2006                        | Sterilization of medical devices-<br>Microbiological methods-Part 1:<br>Determination of the population of<br>microorganisms on product                                               |
| Human factors             | IEC 62366:2007                              | EN 62366:2008                              | Medical devices - Application of usability<br>engineering to medical devices                                                                                                          |
| Software                  | IEC 62304:2006                              | EN 62304:2006                              | Medical device software - Software life<br>cycle processes                                                                                                                            |

## Table 1: Applicable Reference Standards list

{3}------------------------------------------------

Tests were carried out on the PERFORMER HT system (equipment and disposable) to assess:

- . The ability of the system to heat and pump fluids accurately over the full range of flow rate and operating conditions;
- The ability of the system to maintain fluid temperature over the full range of fluid . flow rates:
- . The ability of the system to control the pressures, volumes and flow rate accuracy over the full range of fluid flow rate within the operating temperature and pressure range;
- . The ability of the system to detect and alarm at unsafe or ineffective operating conditions including operator errors, the failure of the system sensors, and other internal system malfunctions;
- The ability of the system to mitigate against known or predictable operator errors; .
- The ability of the system to store parameters and data in non-volatile memory. .

The PERFORMER HT performed within the specifications in all of the above tests.

## Software verification and validation

PERFORMER HT software ability to meet performance specifications has been verified through the tests described above. Software control of all functions, actuators and parameters has been verified at unit, integration and system level.

## Human Factors

A human factors non-clinical study was conducted in a simulated use environment with the purpose to optimize the system's user interface through human factors analysis, testing and validation. The study evaluated user- and patient-safety risks associated with the design of the PERFORMER HT system and demonstrated that the task-risks and errors that occur during use of the device have been identified, addressed and either eliminated or reduced.

## 21 CFR 807.92(b) (3): Conclusions

Based upon safety and performance testing and compliance with voluntary standards, RanD believes that the PERFORMER HT is substantially equivalent to the predicate device, and does not raise any new questions of safety or effectiveness.

{4}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is oriented towards the upper right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Simone Ceretti Regulatory Affairs RanD S.r.l. Via Statle 12, 62 MEDOLLA ITALY 41036

MAY - 8 2012

Re: K120026

Trade/Device Name: PERFORMER HT System Regulation Number: Unclassified Regulation Name: Thermal Fluid Warmer Regulatory Class: II Product Code: LGZ Dated: April 24, 2012 Received: April 26, 2012

Dear Ms. Ceretti:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your (1 vir 1), it may be our of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

#### Page 2 - Ms. Ceretti

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony O.m

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

#### Enclosure

{6}------------------------------------------------

# 4. Indications for Use Statement

510(k) Number (if known): none

Device Name: PERFORMER HT system

Indications for Use:

The intended use of the PERFORMER HT is to provide isolated hyperthermic perfusion in the thoracic or peritoneal cavity by means of extra-corporeal circulation of warmed, physiologically compatible sterile solution, according to a protocol to be selected by the physician.

The device must be used by a qualified medical professional who is experienced in the operation of this or similar equipment.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Kld C. Ch 5/8/12
(Division Sign-Off)

Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K120026

4 - 1

---

**Source:** [https://fda-staging.innolitics.com/device/K120026](https://fda-staging.innolitics.com/device/K120026)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
