K120026 · Rand S.R.L. · LGZ · May 8, 2012 · General Hospital
Device Facts
Record ID
K120026
Device Name
PERFORMER HT/ HANG&GO HT BASIC
Applicant
Rand S.R.L.
Product Code
LGZ · General Hospital
Decision Date
May 8, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the PERFORMER HT is to provide isolated hyperthermic perfusion in the thoracic or peritoneal cavity by means of extra-corporeal circulation of warmed, physiologically compatible sterile solution, according to a protocol to be selected by the physician. The device must be used by a qualified medical professional who is experienced in the operation of this or similar equipment.
Device Story
PERFORMER HT is a prescription electromechanical device for isolated hyperthermic perfusion in thoracic or peritoneal cavities. It uses an extracorporeal circuit with a disposable kit (HANG&GO HT BASIC) to circulate and heat sterile solutions. Input data includes patient information, temperature probes, pressure sensors, and flow rates. The device features roller pumps to circulate fluid through a heater exchanger to reach target temperatures. A central information display (CID) allows clinicians to set treatment parameters, monitor real-time graphical trends, and manage alarm limits. The system automatically detects air in the circuit, monitors pressures/volumes, and triggers audible alarms or stops operation during hazardous conditions. Used in clinical settings by qualified medical professionals, the device enables controlled hyperthermic treatment, potentially benefiting patients by maintaining precise therapeutic temperatures during perfusion protocols.
Clinical Evidence
Bench testing only. Testing verified system ability to heat and pump fluids accurately, maintain temperature, control pressure/volume/flow, detect air, and trigger alarms for unsafe conditions. Human factors study conducted in a simulated environment to validate user interface and mitigate task-related risks.
Technological Characteristics
Electromechanical system with roller-type fluid pumps and touch-screen interface. Includes disposable sterile kit (HANG&GO HT BASIC). Monitors temperature, pressure, flow, and air detection. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), ISO 10993 (biocompatibility), and IEC 62304 (software lifecycle). Sterilization via ethylene oxide. Connectivity includes non-volatile memory for parameter storage.
Indications for Use
Indicated for patients requiring isolated hyperthermic perfusion in the thoracic or peritoneal cavity. Must be used by qualified medical professionals experienced in this equipment.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
K120-026
## 5. 510(k) Summarv
MAY - 8 2012
## 21 CFR 807.92(a)(1): Submitter and Manufacturing site information RanD S.r.l. via Statale 12. 62 41036 Medolla (MO) - ITAL Y Telephone: +39 0535 49283 Fax: +39 0535 660636 Contact person: Simone Ceretti
## Establishment registration number: 3003793891
### Date of the present summary preparation: 28th December, 2011
| 21 CFR 807.92(a)(2): Device information | | | |
|-----------------------------------------|------------------------------------|--|--|
| Trade name | PERFORMER HT / HANG&GO HT BASIC | | |
| Device Classification Name | Warmer, thermal, infusion fluid | | |
| Classification Review Panel | General Hospital | | |
| Classification Product Code | LGZ | | |
| Regulatory Class | Unclassified, pre-amendment device | | |
### 21 CFR 807.92(a)(3): Predicate devices
The PERFORMER HT is substantially equivalent to the Belmont Hyperthermia Pump (K070654, dtd June 8th, 2007).
## 21 CFR 807.92(a)(4): Description of the device
The PERFORMER HT is a prescription electromechanical device intended to provide isolated hyperthermic perfusion in the thoracic or peritoneal cavity by means of extracorporeal circulation of warmed. physiologically compatible sterile solution.
The PERFORMER HT equipment is designed to:
- operate the treatment execution by means of a color touch-screen display .
- . control the temperature of solution perfused to the Patient through the extracorporeal circuit
- control the flow rates of roller pumps .
- control the volume in the extracorporeal circuit and Patient's body cavity .
- monitor Patient's body cavity temperatures .
- monitor the extracorporeal circuit pressures .
- . detect air in the circuit
- change the fluid path in the extracorporeal circuit by means of 2 two-ways pinch valves .
The Central Information Display (CID) enables the User to interact with the PERFORMER HT equipment. The touch-screen display allows the User to:
- inout Patient information .
- operate the various treatment phases
- display treatment parameters (e.g, flows, pressures, temperatures) .
- set treatment parameters .
- . set alarm limits
- display graphical trending of temperature probes connected to the HTS module
{1}------------------------------------------------
- · read alarm and alert messages
The PERFORMER HT is intended to be used with the HANG&GO HT BASIC, a disposable, single use kit intended to circulate and filtrate the clear solutions in the thoracic or abdominal cavity.
The HANG&GO HT BASIC includes all necessary lines and accessories for the treatment execution:
- · circulation line
- ·. heater bag
- · reservoir
- · table pack (Patient inlet and outlet lines) with thermal protections
The HANG&GO HT BASIC is supplied sterile, packaged in a proprietary-design container, and is intended to be used only with the PERFORMER HT.
### 21 CFR 807.92(a)(5): Intended use
The intended use of the PERFORMER HT is to provide isolated hyperthermic perfusion in the thoracic or peritoneal cavity by means of extra-corporeal circulation of warmed. physiologically compatible sterile solution, according to a protocol to be selected by the physician.
The device must be used by a qualified medical professional who is experienced in the operation of this or similar equipment.
### Predicate device intended use
The intended use of the Belmont Hyperthermia Pump is to raise the temperature of the thoracic or peritoneal cavity to the desired target temperature by continuously lavaging the cavity with circulating warmed sterile solution, according to a protocol to be selected by the physician.
#### 21 CFR 807.92(a)(6): Comparison of technological characteristics
The PERFORMER HT has the same operating principle, energy type, environmental specifications, and performance specifications of the Belmont device. It uses roller-type fluid pumps, touch screen to direct the User through set-up preparation and use, control and monitoring of fluid temperatures, and a proprietary-design disposable set including large fluid reservoir to circulate sterile fluid into and out of the body cavity. Flow from the reservoir is delivered by action of the roller pump to the heater exchanger to be heated to the desired target temperature, and then to the Patient, From the Patient, the fluid is pumped from the return line back into the reservoir. The PERFORMER HT monitors circulating flow, temperatures, pressures and volumes, detects air in the circuit and automatically responds to ineffective or unsafe operating conditions by sounding audible alarms, stopping heating and pumping in case of hazardous conditions for User/Patient.
{2}------------------------------------------------
# 21 CFR 807.92(b)(1): Discussion of the non-clinical tests
The PERFORMER HT is designed to, complies with, and has been tested to, the FDA Recognized Consensus Standards listed in the table below, as applicable to the equipment, the disposable lines set and the components:
| Area | US Recognized standard # | EU Recognized standard # | Title |
|---------------------------|---------------------------------------------|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General | ISO 14971:2007 | EN ISO 13485:2003 / AC:2007 | Medical devices -- Quality management systems -- Requirements for regulatory purposes |
| | | EN ISO 14971:2007 | Medical devices - Application of risk management to medical devices |
| | | | |
| Packaging and<br>labeling | | EN 980:2008 | Symbols for use in the labelling of medical devices |
| | | EN 1041:2008 | Information supplied by the manufacturer with medical devices |
| | | ASTM D4169-04a | Standard Practice for Performance Testing of Shipping Containers and Systems |
| | | | |
| Biocompatibility | AAMI ANSI ISO 10993-1:2009 | EN ISO 10993-1:2009 | Biological evaluation of medical devices - Part 1: Evaluation and testing |
| | | EN ISO 10993-4:2002 | Biological evaluation of medical devices - Part 4: Selection of tests for interaction with blood |
| | AAMI ANSI ISO 10993-5:2009 | EN ISO 10993-5:2009 | Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity |
| | AAMI ANSI ISO 10993-7:2008 | EN ISO 10993-7:2009 | Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals |
| | ISO 10993-10:2002/Amd<br>1:2006 | EN ISO 10993-<br>10:2002/Amd 1:2006 | Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity |
| | AAMI ANSI ISO 10993-<br>11:2006 | EN ISO 10993-11:2006 | Biological evaluation of medical devices - Part 11: Tests for systemic toxicity |
| | IEC 60601-1:1988/Amd<br>1:1991 & Amd 2:1995 | EN 60601-1:1990/Amd<br>1:1993 & Amd 2:1995 | Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995 |
| | | | |
| | | | |
| Electrical safety | IEC 60601-1-1:2000 | EN 60601-1-1:2001 | Medical electrical equipment - Part 1-1: General requirements for safety -<br>Collateral standard: Safety requirements for medical electrical systems |
| | AAMI ANSI IEC 60601-<br>1-2:2001 | EN 60601-1-2:2001 | Medical Electrical Equipment - Part 1-2: General Requirements for Safety -<br>Collateral Standard: Electromagnetic Compatibility - Requirements and Tests |
| | | | |
| | IEC 60601-1-4:2000 | EN 60601-1-<br>4:1996/Amd 1:1999 | Medical electrical equipment - Part 1-4:<br>General requirements for safety -<br>Collateral standard: Programmable electrical medical systems |
| | | | |
| | | | |
| | | EN 60601-1-6:2004 | Medical electrical equipment -- Part 1-6:<br>General requirements for basic safety and essential performance - Collateral Standard: Usability |
| | | | |
| | | | |
| | | | |
| | | | |
| Sterility | AAMI ANSI ISO 11135-<br>1:2007 | EN ISO 11135-1:2007 | Sterilization of health care products -<br>Ethylene oxide - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices. |
| | | | |
| | | | |
| | | | |
| | AAMI ANSI ISO 11607-<br>1:2006 | EN ISO 11607-1:2009 | Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems |
| | | | |
| | | | |
| | | | |
| | | | |
| | AAMI ANSI ISO 11607-<br>2:2006 | EN ISO 11607-2:2006 | Packaging for terminally sterilized<br>medical devices - Part 2: Validation<br>requirements for forming, sealing and<br>assembly processes |
| | ASTM F1980-07 | | Standard Guide for Accelerated Aging of<br>Sterile Barrier Systems for Medical<br>Devices. |
| | AAMI ANSI ISO 11737-<br>1:2006 | EN ISO 11737-1:2006 | Sterilization of medical devices-<br>Microbiological methods-Part 1:<br>Determination of the population of<br>microorganisms on product |
| Human factors | IEC 62366:2007 | EN 62366:2008 | Medical devices - Application of usability<br>engineering to medical devices |
| Software | IEC 62304:2006 | EN 62304:2006 | Medical device software - Software life<br>cycle processes |
## Table 1: Applicable Reference Standards list
{3}------------------------------------------------
Tests were carried out on the PERFORMER HT system (equipment and disposable) to assess:
- . The ability of the system to heat and pump fluids accurately over the full range of flow rate and operating conditions;
- The ability of the system to maintain fluid temperature over the full range of fluid . flow rates:
- . The ability of the system to control the pressures, volumes and flow rate accuracy over the full range of fluid flow rate within the operating temperature and pressure range;
- . The ability of the system to detect and alarm at unsafe or ineffective operating conditions including operator errors, the failure of the system sensors, and other internal system malfunctions;
- The ability of the system to mitigate against known or predictable operator errors; .
- The ability of the system to store parameters and data in non-volatile memory. .
The PERFORMER HT performed within the specifications in all of the above tests.
## Software verification and validation
PERFORMER HT software ability to meet performance specifications has been verified through the tests described above. Software control of all functions, actuators and parameters has been verified at unit, integration and system level.
## Human Factors
A human factors non-clinical study was conducted in a simulated use environment with the purpose to optimize the system's user interface through human factors analysis, testing and validation. The study evaluated user- and patient-safety risks associated with the design of the PERFORMER HT system and demonstrated that the task-risks and errors that occur during use of the device have been identified, addressed and either eliminated or reduced.
## 21 CFR 807.92(b) (3): Conclusions
Based upon safety and performance testing and compliance with voluntary standards, RanD believes that the PERFORMER HT is substantially equivalent to the predicate device, and does not raise any new questions of safety or effectiveness.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is oriented towards the upper right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Simone Ceretti Regulatory Affairs RanD S.r.l. Via Statle 12, 62 MEDOLLA ITALY 41036
MAY - 8 2012
Re: K120026
Trade/Device Name: PERFORMER HT System Regulation Number: Unclassified Regulation Name: Thermal Fluid Warmer Regulatory Class: II Product Code: LGZ Dated: April 24, 2012 Received: April 26, 2012
Dear Ms. Ceretti:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your (1 vir 1), it may be our of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2 - Ms. Ceretti
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony O.m
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{6}------------------------------------------------
# 4. Indications for Use Statement
510(k) Number (if known): none
Device Name: PERFORMER HT system
Indications for Use:
The intended use of the PERFORMER HT is to provide isolated hyperthermic perfusion in the thoracic or peritoneal cavity by means of extra-corporeal circulation of warmed, physiologically compatible sterile solution, according to a protocol to be selected by the physician.
The device must be used by a qualified medical professional who is experienced in the operation of this or similar equipment.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kld C. Ch 5/8/12
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K120026
4 - 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.