K120005 · Isopure, Corp. · KPO · Mar 29, 2012 · Gastroenterology, Urology
Device Facts
Record ID
K120005
Device Name
ACDS
Applicant
Isopure, Corp.
Product Code
KPO · Gastroenterology, Urology
Decision Date
Mar 29, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.5820
Device Class
Class 2
Indications for Use
The Isopure ACDS with remote fill is intended to be used in Hemodialysis facilities for the storage and distribution of acid concentrate to be used in the treatment of Hemodialysis patients.
Device Story
Isopure ACDS stores and distributes pre-mixed acidified concentrate in hemodialysis facilities. System comprises linear polyethylene bulk storage tanks, 115 VAC controller, distribution pump, piping, and optional remote fill station. Controller manages pump operation and remote filling via electrical receptacles. Level floats monitor tank volume, triggering low-level alarms and enabling remote fill power. System delivers concentrate through distribution loops to dialysis machines or via ball valves to fill transportable jugs. Pressure regulators maintain output < 5 psig at stations. Used by facility staff to ensure consistent concentrate availability for patient treatment. Benefits include centralized storage, automated distribution, and reduced manual handling of concentrate.
Clinical Evidence
Bench testing only. Validation protocol VP12-002 evaluated operation of the ACDS controller, distribution pump, piping, remote fill station, and remote alarm. Testing verified leak-free assembly, proper function of level floats for alarm/fill activation, pump flow rates, and pressure regulation.
Technological Characteristics
Materials: Linear virgin polyethylene tanks, Schedule 80 PVC piping, polyethylene flex tubing. Energy: 115 VAC electrical controller. Components: Distribution pump, level floats, pressure regulators (0-25 psi), autolock fittings. Connectivity: Standalone distribution loop. Sterilization: Not specified.
Indications for Use
Indicated for use in hemodialysis facilities for storage and distribution of acid concentrate for the treatment of hemodialysis patients.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Predicate Devices
Pure Water Inc's Acidified Storage and Distribution System (K993058)
Submission Summary (Full Text)
{0}------------------------------------------------
## MAR 2 9 2012
OF 3
#### L. 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
Isopure Corp 1. Submitter's Name:
> 141 Citizens Blvd. Simpsonville, KY 40067 Telephone: (502) 722-1000 Kevin C. Gillespie
Contact person: . December 29, 2011 Date of Summary:
- 2. Device Name: Isopure ACDS (Acidified Concentrate Distribution and Storage) Device Classification Name: Dialysate concentrate for Hemodialysis (liquid or powder) (21 CFR \$876.5820, KPO).
3. Legally Marketed Devices to which Equivalence is Claimed: Isopure Corp purchased the Pure Water Inc's Acidified Storage and Distribution System with Optional Remote Fill (k993058) from Pure Water Inc. Isopure intends to manufacture and market the device at its Simpsonville K Y location. The intended use, functionally of neither the device nor any of the contacting components for the device has changed.
Device Description: The Isopure ACDS (Acidified Concentrate Distribution and Storage) is 4. designed for storage of pre-mixed acidified concentrate in a tank for transfer through a loop to the individual dialysis machines or to fill jugs which then can be transported to the individual dialysis machines when no distribution loop exists.
ISOPURE ACDS (Acidified Concentrate Distribution and Storage) with optional remote fill uses "one-piece molded seamless tanks constructed of linear polyethylene" for bulk storage of acidified concentrate.
The optional remote fill allows the storage tanks to be filled by the vendor from outside the building. The Isopure ACDS (Acidified Concentrate Distribution and Storage) with optional remote fill meets or exceeds all Association for the Advancement of Medical Instrumentation (AAMI) National Standards for Hemodialysis.
Indications for use: The Isopure ACDS (Acidified Concentrate Distribution and Storage) with 5. remote fill is intended to be used in Hemodialysis facilities for the storage and distribution of acid concentrate to be used in the treatment of Hemodialysis patients.
Descriptive Summary of Technological Characteristics and Those of Predicate Devices: The 6. technological characteristics of the device are the same as the original submitted device under 501(k) K993058, Pure Water, Inc.'s Acidified Storage and Distribution System cleared January 24, 2000.
| Pure Water, Inc K993058 | Isopure Corp |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The Acidified Tank(s) are designed to store acidified concentrate. The tank(s) are constructed of one piece linear virgin polyethylene. Tanks feature a vented sealed lid to reduce particle contamination. (Pure water acidified storage & distribution 510(k) submission 9/10/99) | The Acidified Tank(s) are designed to store acidified concentrate. The tank(s) are constructed of one piece linear virgin polyethylene. Tanks feature a vented sealed lid to reduce particle contamination. |
## Summary of Comparisons of Components
{1}------------------------------------------------
# OF 3
| Acid Fill Controller is a 115 VAC controller that is<br>used to operate the distribution pump when the<br>concentrate is distributed through a loop to the<br>dialysis stations. The controller will also operate<br>the remote filling station by providing power to the<br>receptacle located at the fill station. (Pure water<br>acidified storage & distribution 510(k) submission<br>9/10/99) | Acid Fill Controller is a 115 VAC controller that is<br>used to operate the distribution pump when the<br>concentrate is distributed through a loop to the<br>dialysis stations. The controller will also operate<br>the remote filling station by providing power to the<br>receptacle located at the fill station. |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The filter assembly filters our impurities that may<br>be present in the Acidified concentrate. The filter<br>assembly is comprised of a clear filter housing and<br>a 5 micron 10 inch filter. (Pure water acidified<br>storage & distribution 510(k) submission 9/10/99) | This filter was eliminated because a filter is located<br>on the acidified concentrate mixer for this purpose.<br>Additional filters would serve no purpose and only<br>offers additional opportunities for leakage. |
| Auxiliary Components A ball valve is placed in the<br>distribution line to fill jugs of concentrate if<br>necessary. (Pure water acidified<br>storage &<br>distribution 510(k) submission 9/10/99) | Auxiliary Components A ball valve is placed in the<br>supply line or the side of the storage tank to fill<br>jugs of concentrate if necessary. |
| Schedule 80 PVC pipe and fittings: Schedule 80<br>PVC pipe and fittings are used to direct concentrate<br>flow. (Pure water acidified storage & distribution<br>510(k) submission 9/10/99) | Schedule 80 PVC pipe and fittings: Schedule 80<br>PVC pipe and fittings are used to direct concentrate<br>flow. |
| Polyethylene Flex Tubing: Polyethylene Flex<br>Tubing is used to direct concentrate flow. (Pure<br>water acidified storage & distribution 510(k)<br>submission 9/10/99) | Polyethylene Flex Tubing: Polyethylene Flex<br>Tubing is used to direct concentrate flow. |
| Autolock Fittings Autolock fittings are used to<br>connect tubing. (Pure water acidified storage &<br>distribution 510(k) submission 9/10/99) | Autolock Fittings Autolock fittings are used to<br>connect tubing. |
Performance Data: The Isopure ACDS ( Acidified Concentrate Distribution and Storage) is the same device that was manufactured by Pure Water, Inc. under 510(k) 993058. The applicant is notifying the FDA that the device DMR and 510(k) was sold to Isopure who will begin manufacturing the device at its Simpsonville KY location. The only change to the device is the name and manufacturing location. All performance data provided in the original 510(k) remains the same.
7. Non Clinical Testing: A validation protocol was performed VP12-002 The following validation test protocol addresses the operation of the ACDS controller, distribution pump, distribution piping, remote fill station, and remote alarm.
## ACDS Storage tank:
A 250 gallon storage tank will be assembled with the trim kit to test the operation of all assemblies of the trim kit. This will also test for leaks in the trim kit assemblies and connections to the storage tank. When connected to the bottom connection should provide acidified concentrate to the distribution pump. The valve will also allow flow to be terminated to the distribution pump when closed.
{2}------------------------------------------------
Level floats: The level floats 2 for the remote fill station the low float will have a dual purpose; I- for the low tank warning on the remote alarm, and 2- to turn power on for the optional remote fill station. The floats will be tested to ensure proper operation of the remote alarm low level indicator and for the optional remote fill station.
ACDS Control Box: The ACDS control box will be connected to the storage tank and also to the remote fill box to test the operation of the control box. Power will be connected to the control box and the distribution pump will be tested as well as the low level alarm and the optional remote fill station
Distribution Pump: The distribution pump will be connected to a 500 ft ½" polyethylene line to test the distribution of the acidified concentrate through the distribution line. A pressure gauge and a flow meter will also be connected to this line to measure the amount of flow as well as the pressure on the distribution line. The flow meter will be used to verify the operation of the flow switch on the ACDS distribution pump.
Remote Fill Station: A remote fill station will be connected to the storage tank and also to the controller. This will be used to test the on/off electrical connections of the controller in relation to the fill station.
Regulators: In the event that the pressure is excessive (> 5 psig) at the distribution box, a 0-25 regulator is supplied to regulate the pressure at each individual station.
## 8. Not Applicable to this application
Conclusion: Upon instituting the above changes, performance testing was undertaken to verify 9. that the device is as safe, as effective and performs as well as or better than the earlier versions of the Pure Water Inc's Acidified Concentrate Distribution and Storage system (K993058) cleared on January 24, 2000.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Kevin Gillespie President/CEO Isopure Corp 141 Citizens Boulevard SIMPSONVILLE KY 40067
MAR 2 9 2012
Re: K120005
Trade/Device Name: Isopure ACDS (Acidified Concentrate Distribution and Storage) Regulation Number: 21 CFR§ 876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: KPO Dated: February 24, 2012 Received: March 1, 2012
Dear Mr. Gillespie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 -
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Benjamin K. Eicher
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Statement
December 29, 2011
Page 1 of 1
## 510(k) Number: K120005
Device Name: Isopure ACDS (Acidified Concentrate Distribution and Storage)
## Indications for Use:
The Isopure ACDS with remote fill is intended to be used in Hemodialysis facilities for the storage and distribution of acid concentrate to be used in the treatment of Hemodialysis patients.
(Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
| Di sion of Reproductive, Gastro-Renal, and<br>Urological Devices | |
|------------------------------------------------------------------|---------|
| 510(k) Number | K120005 |
| Prescription Use | X | OR | Over-the-Counter Use |
|------------------|---|----|----------------------|
|------------------|---|----|----------------------|
(Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.