Published clinical literature; Historical clinical case studies/histological evaluations
The sponsor utilized a literature review of historical clinical studies and decades of organizational experience with calcium hydroxide preparations to support the substantial equivalence of the device to predicate products.
Literature review; Historical clinical experience; Human pulp response; Dentinogenesis
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Literature review of historical clinical studies (1950s-1995); Retrospective literature review of clinical and histological studies; Study Period: 1950-1995
NuCal is used by the dental professional as a cavity liner and as a temporary root canal dressing.
Device Story
NuCal is a non-setting, pre-mixed, radiopaque calcium hydroxide paste; pH > 12. Device dispensed via 27-gauge applicator tip into dental cavities or root canals. Used by dental professionals in clinical settings. Paste consistency allows for easy placement and removal with water. Acts as a liner or temporary dressing to support dental procedures. Benefits include ease of handling and radiopacity for visualization.
Clinical Evidence
Bench testing only. Performance metrics include: appearance (thin white paste), consistency (smooth, homogenous), flow through 27-gauge tip, specific gravity (1.470 g/ml), pH (> 12), and radiopacity (equal to aluminum). Shelf life validated via accelerated testing at 37°C. No clinical trials were conducted; equivalence is supported by bench data and historical literature on calcium hydroxide preparations.
Technological Characteristics
Non-setting, pre-mixed, radiopaque paste. Composition: Calcium hydroxide, Barium sulfate, Polyethylene glycol, Water. pH > 12. Radiopacity equal to aluminum. Dispensed via 27-gauge applicator tip. Mechanical/chemical device; no software or energy source.
Indications for Use
Indicated for use by dental professionals as a cavity liner and temporary root canal dressing.
Regulatory Classification
Identification
A calcium hydroxide cavity liner is a device material intended to be applied to the interior of a prepared cavity before insertion of restorative material, such as amalgam, to protect the pulp of a tooth.
{0}------------------------------------------------
# PULPDENT CORPORATION
80 Oakland Street Watertown, MA 02472 USA
41200003
Tel: 617 926 6666 Fax: 617 926 6262 Email: Pulpdent@pulpdent.com
APR - 5 2012
# 510(k) SUMMARY
December 28, 2011
#### Contact:
Kenneth J. Berk 80 Oakland Street PO Box 780 Watertown, MA 02472 USA Telephone: Fax: Email:
617-926-6666 617-926-6262 ken@pulpdent.com
#### DEVICE:
Trade Name: NuCal Classification Name: Calcium hydroxide cavity liner. FDA Product Code: 76 EJK, 21 CFR Part 872.3250
#### PREDICATE DEVICES:
Pulpdent TempCanal Pulpdent Multi-Cal UltraDent UltraCal XS
INTENDED USE: NuCal is used by the dental professional as a cavity liner and as a temporary root canal dressing.
DESCRIPTION: NuCal is a non-setting, pre-mixed, radiopaque calcium hydroxide paste that has a pH > 12, is easily removed from the root canal with water, has been designed for ease of handling and will flow through a 27 gauge applicator tip.
#### COMPARISON WITH PREDICATE PRODUCTS:
NuCal is substantially equivalent in design, composition, performance and intended use to the predicate products:
| PRODUCT | DESCRIPTION | INTENDED USE | COMPOSITION |
|----------------------------------|------------------------------------------------------------------|----------------------------------------------------|---------------------------------------------------------------------|
| NuCal | Non-setting, pre-mixed,<br>radiopaque calcium hydroxide<br>paste | Cavity liner and temporary root<br>canal dressing. | Calcium hydroxide<br>Barium sulfate<br>Polyethylene glycol<br>Water |
| Pulpdent TempCanal<br>K944945 | Non-setting, pre-mixed,<br>radiopaque calcium hydroxide<br>paste | Cavity liner and temporary root<br>canal dressing. | Calcium hydroxide<br>Barium sulfate<br>Aqueous gel matrix |
| Pulpdent Multi-Cal<br>K944945 | Non-setting, pre-mixed,<br>radiopaque calcium hydroxide<br>paste | Cavity liner and temporary root<br>canal dressing. | Calcium hydroxide<br>Barium sulfate<br>Aqueous gel matrix |
| UltraDent UltraCal XS<br>K970114 | Non-setting, pre-mixed,<br>radiopaque calcium hydroxide<br>paste | Cavity liner and temporary root<br>canal dressing. | Calcium hydroxide<br>Barium sulfate<br>Aqueous matrix |
{1}------------------------------------------------
# PULPDENT CORPORATION
80 Oakland Street Watertown, MA 02472 USA
Tel: 617 926 6666 Fax: 617 926 6262 Email: Pulpdent@pulpdent.com
### SUMMARY OF PERFORMANCE TESTING - BENCH
The following test results demonstrate that NuCal performs as intended:
| Appearance | Thin white paste. | |
|----------------------|----------------------------------------------------------------------------|--|
| Consistency | Smooth, homogenous, thin paste. | |
| Paste characteristic | Dispenses easily through 27 gauge applicator tip without clogging or force | |
| Specific gravity | 1.470 g/ml | |
| pH | > 12 | |
| Radiopacity | Equal to the same thickness of aluminum | |
| Shelf life | Two years, based on accelerated testing at 37°C | |
#### CONCLUSION:
From the above comparisons, the bench testing and the decades of organizational experience with calcium hydroxide preparations, it can be concluded that NoCal is substantially equivalent in design, composition, performance and intended use to the predicate products listed above. The predicate products have been found substantially equivalent under the 510(k) Premarket Notification process as Class II Dental Devices under CFR 872.3250 and have been used by dental professionals for more than 20 years. A search of the relevant scientific literature shows that calcium hydroxide pastes used for cavity lining and root canal dressing have been studied since the 1950's. See References below.
#### REFERENCES
- American Dental Association. Accepted Dental Therapeutics, 39th edition. Calcium 1. hydroxide preparations. 1982 July; 300-301.
- 2. Behnen MJ, West LA, Liewehr FR, et al. Antimicrobial activity of several calcium hydroxide preparations in root canal dentin. J Endod 2001 Dec; 27(12):765-7.
- ന് Berk H. The effect of calcium hydroxide-methyl cellulose paste on the dental pulp. J Dent Children. 17(4) Oct. 1950.
- Berk H. Maintaining vitality of injured permanent anterior teeth. JADA 1954; 49:391-401 4.
- Berk H. Preservation of the dental pulp in deep seated cavities. J Amer Dent Assoc 1957; 5. 54:226-33.
- Berk H, Krakow AA. Efficient Vital Pulp Therapy. Dent Clin NA. 1965 July; 373-85. 6.
- Berk H, Stanley HR. Pulp healing following capping in human sound and carious teeth. J 7. Dent Res. 1958; 37(66) Abstract.
{2}------------------------------------------------
80 Oakland Street Watertown, MA 02472 USA
Tel: 617 926 6666 Fax: 617 926 6262 Email: Pulpdent@pulpdent.com
- ಹ Heithersay GS. Calcium hydroxide in the treatment of pulpless teeth with associated pathology. J Brit Endod Society, 1975;8(2), 74-93.
- 9. Kirk EEJ, Lim KC, Khan MOG. A comparison of dentinogenesis on pulp capping with calcium hydroxide in paste and cement form. Oral Surg Oral Med Oral Pathol 1989; 68(2):210-19.
- Kundsin RB, Perkins RE. Test of antibacterial activity of Pulpdent TempCanal (Calcium 10. hydroxide preparation). Kundsin Laboratory, Harvard Medical School, Brigham and Women's Hospital.
- 11. Phaneuf RA, Frankl SN, Ruben MP. A comparative histological evaluation of three calcium hydroxide preparations on the human primary dental pulp. J Dent for Children. 1968 January; 35:61-76.
- Subay RK, Suzuki S, et al. Human pulp response after partial pulpotomy with two calcium 12. hydroxide products. Oral Surg Oral Med Oral Pathol Oral Radiol Endo 1995; 80(3):330-7.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and in a bold, sans-serif font. The text is black and the background is white. The text is slightly tilted.
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Kenneth J. Berk Director Pulpdent Corporation 80 Oakland Street Watertown, Massachusetts 02472
APR - 5 2012
Re: K120003 Trade/Device Name: Nucal Regulation Number: 21 CFR 872.3250 Regulation Name: Calcium Hydroxide Cavity Liner Regulatory Class: II Product Code: EJK Dated: February 17, 2012 Received: February 21, 2012
Dear Mr. Berk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
# Page 2- Mr. Berk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffice s/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
# Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number: K120003
Device Name: Pulpdent NuCal
# Indications For Use:
NuCal is a non-setting, pre-mixed, radiopaque calcium hydroxide paste used as a cavity liner and as a temporary root canal dressing.
Prescription Use _X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Renner
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesiology, General Devices Division of Antrol, Dental Devices
510(k) Number: K120003
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.