← Product Code [CDT](/productcode/CDT) · K113830

# VANTERA CLINICAL ANALYZER (K113830)

_Liposcience · CDT · Aug 30, 2012 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K113830

## Device Facts

- **Applicant:** Liposcience
- **Product Code:** [CDT](/productcode/CDT.md)
- **Decision Date:** Aug 30, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1705
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K113830 · Aug 30, 2012 | VANTERA CLINICAL ANALYZER | Liposcience | Retrospective analysis of serum samples from a reference population | The sponsor used a retrospective cohort of 452 apparently healthy individuals to determine the distribution of LDL-P levels and establish reference ranges for the device. | Reference range; Retrospective cohort; Population distribution |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective analysis of a reference population | Apparently healthy men (n=158) and women (n=294); Sample Size: 452 | Not applicable for this study | Distribution of LDL-P levels by percentile |

## Indications for Use

The Vantera® Clinical Analyzer is an automated laboratory test analyzer which measures the 400 MHz proton nuclear magnetic resonance (NMR) spectrum of clinical samples to produce signal amplitudes, converting these signal amplitudes to analyte concentration. The device includes a 400 MHz NMR spectrometer and software to analyze digitized spectral data. This instrumentation is intended to be used with NMR based assays to detect multiple analytes from clinical samples. The NMR LipoProfile® test, when used with the Vantera® Clinical Analyzer, an automated NMR spectrometer, measures lipoprotein particles to quantify LDL particle number (LDL-P), HDL cholesterol (HDL-C), and triglycerides in human serum and plasma using nuclear magnetic resonance (NMR) spectroscopy. LDL-P and these NMR-derived concentrations of HDL-C and triglycerides are used in conjunction with other lipid measurements and clinical evaluation to aid in the management of lipoprotein disorders associated with cardiovascular disease.

## Device Story

Vantera Clinical Analyzer is an automated 400 MHz proton NMR spectrometer; processes human serum/plasma samples. System includes sample handling assembly, NMR subassembly, and enclosure. Input: clinical samples; system performs automated dilution and spectral measurement. Operation: NMR spectrum deconvolution at ~0.8 ppm (methyl group protons) to produce signal amplitudes for lipoprotein subclasses; conversion to analyte concentrations. Output: LDL-P, HDL-C, and triglyceride levels. Used in clinical laboratories; operated by lab personnel. Results used by clinicians alongside other lipid measurements for cardiovascular disease management. Benefits: provides lipoprotein particle quantification to aid in managing lipid disorders.

## Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP5-A2; linearity per EP6-A; interference per EP7-A2. Method comparison study (n=1482-1520 samples) against predicate NMR Profiler showed high correlation (r=0.978-0.998) and low bias (<5%) across medical decision limits. Matrix comparison (n=50) confirmed performance across LipoTube, plain serum, and EDTA plasma. Reference range study (n=452) established LDL-P distributions in healthy adults.

## Technological Characteristics

400 MHz proton NMR spectrometer; automated sample handling with onboard dilution. Materials: glass/plastic sample containers, aqueous NMR diluent, TMA calibrator. Connectivity: networked control system (Host, Process Control, NMR Control computers). Software: spectral deconvolution algorithm. Sterilization: N/A (clinical analyzer).

## Regulatory Identification

A triglyceride test system is a device intended to measure triglyceride (neutral fat) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders.

## Predicate Devices

- NMR Profiler and NMR Lipoprofile test (k111516)
- Luminex LX 100/200 Instrument (k073506)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

A. 510(k) Number:
k113830

B. Purpose for Submission:
New device

C. Measurand:
LDL-P (low density lipoprotein particle number), HDL cholesterol (HDL-C), and triglycerides

D. Type of Test:
Nuclear Magnetic Resonance (NMR) spectroscopy assay

E. Applicant:
LipoScience Inc.

F. Proprietary and Established Names:
NMR LipoProfile® test on Vantera® Clinical Analyzer

G. Regulatory Information:

|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  NSU | Class II | 21 CFR 862.2570, Instrumentation for clinical multiplex test system | Clinical Chemistry (75)  |
|  MRR | Class I, meets limitations per 21 CFR 862.9(c)(4) | 21 CFR 862.1475, Lipoprotein test system | Clinical Chemistry (75)  |
|  LBS |   | 21 CFR 862.1175, Cholesterol test system | Clinical Chemistry (75)  |
|  CDT |   | 21 CFR 862.1705, Triglyceride test system | Clinical Chemistry (75)  |

{1}

H. Intended Use:

1. Intended use(s):

See Indications for use below.

2. Indication(s) for use:

The Vantera® Clinical Analyzer is an automated laboratory test analyzer which measures the 400 MHz proton nuclear magnetic resonance (NMR) spectrum of clinical samples to produce signal amplitudes, converting these signal amplitudes to analyte concentration. The device includes a 400 MHz NMR spectrometer and software to analyze digitized spectral data. This instrumentation is intended to be used with NMR based assays to detect multiple analytes from clinical samples.

The NMR LipoProfile® test, when used with the Vantera® Clinical Analyzer, an automated NMR spectrometer, measures lipoprotein particles to quantify LDL particle number (LDL-P), HDL cholesterol (HDL-C), and triglycerides in human serum and plasma using nuclear magnetic resonance (NMR) spectroscopy. LDL-P and these NMR-derived concentrations of HDL-C and triglycerides are used in conjunction with other lipid measurements and clinical evaluation to aid in the management of lipoprotein disorders associated with cardiovascular disease.

3. Special conditions for use statement(s):

For in vitro diagnostic use only, for prescription use only.

4. Special instrument requirements:

All performance was evaluated on the The Vantera® Clinical Analyzer

I. Device Description:

The test system includes the following components:

- Diluent 1 (NMR LipoProfile® test) – 8 x 250 mL bottles of aqueous solution containing Na2EDTA (5.0 mM), CaCl2 (1.0 mM), KCL (120 mM), Na2HPO4 -7H2O (50 mM), pH 7.4, 6.0 M NaOH, 1.0 M HCl.
- WASH (NMR Fluidics System Solution) – single 2L bottle of Triton X-100-0.1% v/v, Liqui Nox 0.1% v/v in Type 2 water, pH 10.0, sodium bicarbonate (anhydrous), sodium carbonate (anhydrous), 6.0 M NaOH.
- NMR Reference Standard (calibrator) – 6 x 30 mL bottles of 0.2% w/v aqueous solution of Trimethyl Acetate (TMA) disodium salt (15.0 mM) containing Na2EDTA (5.0 mM), CaCl2 (3.0 mM), KCl (120 nM), D2O 10% v/v, 6.0 M NaOH, 1.0 M HCl. Each box of NMR

{2}

Reference Standard is supplied with specimen tube barcodes.

- QC Material – 2 levels of human based control material (6 x 3mL each) which include values for LDL-P, TG, and HDL-C as assigned at LipoScience.

The control materials contain human source material. Each donor unit is tested by FDA – approved methods and found non-reactive for hepatitis B surface antigen (HBsAg), antibody to hepatitis C, and antibody to HIV-1/2.

All products using human source material should be handled as potentially infectious, because no test method can offer complete assurance that infectious agents are absent.

- Vantera Clinical Analyzer - 400 MHz proton nuclear magnetic resonance spectrometer interfaced with sample handling assembly, deconvolution software and provided with a system User's Manual.

J. Substantial Equivalence Information:

1. Predicate device name(s):

1. NMR Profiler and NMR Lipoprofile test
2. Luminex LX 100/200 Instrument

2. Predicate 510(k) number(s):

1. k111516
2. k073506

3. Comparison with predicate:

Instruments:

|  Items | Vantera® Clinical Analyzer for use with NMR LipoProfile® test (Candidate Device) | Luminex LX 100/200 Instrument (Perdicate Device) k073506  |
| --- | --- | --- |
|  Similarities  |   |   |
|  Instrument Intended Use | Same | A clinical multiplex instrument intended to measure and sort multiple signals generated in an In Vitro diagnostic assay from a clinical sample. This instrumentation is used with a specific assay to measure multiple similar analytes that establish a single indicator to aid in diagnosis. The device  |

{3}

4

|   |  | includes a signal reader unit, raw data storage mechanisms, data acquisition software and software to process detected signals.  |
| --- | --- | --- |
|  Multi-Analyte | Same | Yes  |
|  System Fluidics | Same | Utilizes system fluidics to deliver sample to the site of sample analysis  |
|  System Calibration | Same | Calibration Required  |
|  Specimen Identification | Same | Barcode reader entry of sample ID  |
|  Data Acquisition Software | Same | Data acquisition software and software to process detected signals  |
|  Differences  |   |   |
|  Test Principle | Nuclear magnetic resonance – 400 MHz proton NMR | Bead based multiplexing - Fluorescence  |
|  Sample handling | Serum/Plasma Samples are diluted onboard system | Samples are manually prepared then presented to system.  |

Device:

|  Items | Vantera® Clinical Analyzer for use with NMR LipoProfile® test (Candidate Device) | LipoScience NMR LipoProfile® test and NMR Profiler (Predicate Device) k111516  |
| --- | --- | --- |
|  Similarities  |   |   |
|  Device Intended Use | Same | For the measurement of lipoprotein particles to quantify LDL particle number (LDL-P), HDL cholesterol (HDL-C), and triglycerides in human serum and plasma using nuclear magnetic resonance (NMR) spectroscopy. LDL-P and these NMR-derived concentrations of HDL-C and triglycerides are used in conjunction with other lipid measurements and clinical evaluation to aid in the management of lipoprotein disorders associated with cardiovascular disease  |

{4}

|  Test Principle | Same | Nuclear magnetic resonance – 400 MHz proton NMR  |
| --- | --- | --- |
|  Multi-Analyte | Same | Yes  |
|  System Fluidics | Same | Manual  |
|  Sample Type | Same | Human Serum and Plasma  |
|  System Calibration | Same | Calibration Required  |
|  Specimen Identification | Same | Barcode reader entry of sample ID  |
|  Data Acquisition Software | Same | Data acquisition software and software to process detected signals  |
|  Spectral Deconvolution | Same | Linear least-squares with singular value decomposition of the spectra from each specimen.  |
|  Measuring Range LDL-P | Same | 300-3500 nmol/L  |
|  Measuring Range TG | Same | 5-1100 mg/dL  |
|  Measuring Range HDL-C | Same | 7-140 mg/dL  |
|  Differences  |   |   |
|  Sample handling | Serum/Plasma Samples are diluted onboard system | Samples are manually prepared then presented to system.  |
|  System Fluidics | Utilizes system fluidics to deliver sample to the site of sample analysis | Manual  |

# K. Standard/Guidance Document Referenced (if applicable):

- CLSI Guideline EP05-A2: Evaluation of Precision Performance of Qualitative Measurement Methods
- CLSI Guideline EP06-A: Evaluation of the Linearity of Qualitative Measurement Methods
- CLSI Guideline EP07-A2: Interference Testing in Clinical Chemistry
- CLSI Guideline EP09-A2: Method Comparison and Bias Estimation Using Patient Samples
- CLSI Guideline EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation
- CLSI Guideline EP14-A2: Evaluation of Matrix Effects
- CLSI Guideline C28-A3: Defining Establishing, and Verifying Reference Intervals in the Clinical Laboratory
- IEC 610 I 0-1:200 I, 2nd Edition; Safety requirements for electrical equipment for measurement, control and laboratory use. Part I

{5}

L. Test Principle:

Vantera Clinical Analyzer

The Vantera Clinical Analyzer is a clinical laboratory analyzer that employs nuclear magnetic resonance spectroscopic detection to quantify multiple analytes in biological fluid specimens, specifically blood plasma and serum.

The Vantera Clinical Analyzer system design is divided into 3 major subassemblies: a sample handling assembly, an NMR subassembly, and an enclosure. The Vantera Clinical Analyzer control system is distributed across three separate computers:

- The Host (1U) controls user interface, data handling, results calculation, system startup and shutdown.
- The Process Control (4U) schedules and manages all activities required to process a sample, controls all hardware in the sample handling subsystem, and manages remote access to the system.
- The NMR Control Computer controls all magnet operations.

Two of these computers are contained within the Sample Handling Subassembly (1U and 4U) and one in the NMR Subassembly (NMR Console).

NMR LipoProfile test

The NMR LipoProfile test involves measurement of the 400 MHz proton NMR spectrum of a plasma/serum sample, deconvolution of the composite signal at approximately 0.8 ppm to produce signal amplitudes of the lipoprotein subclasses that contribute to the composite plasma/serum signal, and conversion of these subclass signal amplitudes to lipoprotein subclass concentrations. The ~0.8 ppm plasma NMR signal arises from the methyl group protons of the lipids carried in the LDL, HDL and VLDL subclasses of varying diameters. The NMR signals from the various lipoprotein subclasses have unique and distinctive frequencies and line shapes, each of which is accounted for in the deconvolution analysis model. Each subclass signal amplitude is proportional to the number of subclass particles emitting the signal, which enables subclass particle concentrations to be calculated from the subclass signal amplitudes derived from the spectral deconvolution analysis. LDL subclass particle concentrations, in units of nanomoles of particles per liter (nmol/L), are summed to give the reported total LDL particle concentration (LDL-P). By employing conversion factors assuming that the various lipoprotein subclass particles have cholesterol and triglyceride contents characteristic of normolipidemic individuals, HDL cholesterol and triglyceride concentrations are also derived.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

{6}

Within-run and within-laboratory precision were evaluated in accordance with the methods defined in the CLSI Guideline EP5-A, “Evaluation of Precision Performance of Clinical Chemistry Devices.”

Three serum pools with different analyte concentrations were tested in multiple runs over multiple days. For the Within-Lab, multiple instruments, operators and reagent lots were incorporated into the testing and a variance component analysis was conducted to estimate the individual sources of the total system variability.

Within-run precision - A single run of 20 replicates for the low, medium and high pool was conducted on one NMR instrument. A single operator conducted all three runs.

|   | Pool #1 |   |   | Pool #2 |   |   | Pool #3  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Mean | SD | %CV | Mean | SD | %CV | Mean | SD | %CV  |
|  LDL-P, nmol/L | 842.6 | 48.5 | 5.8 | 1309.5 | 39.1 | 3.0 | 1837.7 | 50.3 | 2.7  |
|  HDL-C, mg/dL | 29.1 | 1.17 | 4.0 | 51.1 | 1.43 | 2.8 | 86.9 | 2.29 | 2.6  |
|  Triglycerides, mg/dL | 70.1 | 1.6 | 2.3 | 169.2 | 3.5 | 2.1 | 356.1 | 4.2 | 1.2  |

Within-laboratory precision - Two runs per day with two replicates per run for three pools were tested on three instruments for a total of 20 testing days.

|   | Pool #1 |   |   | Pool #2 |   |   | Pool #3  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Mean | SD | %CV | Mean | SD | %CV | Mean | SD | %CV  |
|  LDL-P, nmol/L | 988.6 | 48.84 | 5.3 | 1266.7 | 32.57 | 4.0 | 1943.5 | 63.42 | 3.9  |
|  HDL-C, mg/dL | 28.9 | 0.80 | 2.8 | 50.7 | 1.02 | 2.0 | 85.2 | 1.51 | 1.8  |
|  Triglycerides mg/dL | 68.8 | 1.59 | 2.3 | 166.3 | 3.92 | 2.4 | 352.2 | 9.36 | 2.7  |

Reproducibility - A reproducibility study was conducted in accordance to EP5-A2 at 3 sites incorporating five levels of serum panels at or around the medical decision limits. The panels were tested for 5 days, 6 runs per day, 2 replicates per run. The overall precision estimates are described below.

{7}

|   | Pool #1 |   |   | Pool #2 |   |   | Pool #3  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Mean | SD | %CV | Mean | SD | %CV | Mean | SD | %CV  |
|  LDL-P, nmol/L | 520.2 | 52.94 | 10.2 | 1182.1 | 82.19 | 7.0 | 1343.4 | 97.37 | 7.2  |
|  HDL-C, mg/dL | 20.1 | 1.26 | 6.3 | 30.3 | 2.60 | 8.6 | 52.0 | 2.45 | 4.7  |
|  Triglycerides, mg/dL | 67.6 | 2.76 | 4.1 | 71.8 | 3.21 | 4.5 | 136.4 | 5.41 | 4.0  |
|   | Pool #4 |   |   | Pool #5  |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  Mean | SD | %CV | Mean | SD | %CV  |
|  LDL-P, nmol/L | 1996.2 | 96.39 | 4.8 | 3214.8 | 165.44 | 5.1  |
|  HDL-C, mg/dL | 75.8 | 4.56 | 6.0 | 88.2 | 3.91 | 4.4  |
|  Triglycerides, mg/dL | 161.4 | 8.66 | 5.4 | 348.0 | 14.99 | 4.3  |

## b. Linearity/assay reportable range:

Three serum pools were prepared from patient specimens with low, medium and high values of LDL-P, HDL-C and Triglycerides (TG) as determined by NMR LipoProfile test. The pools were mixed in different proportions to produce eleven (for LDL-P) or Twelve (12) (TG and HDL-C) different samples with widely varying target concentrations. Mean values from analysis of four replicates of each pool were compared to the expected target values to determine the percent bias for each sample. The serum pools were analyzed according to EP6-A. Tables and regression plots of the linearity data for LDL-P, HDL-P and Triglycerides are given below:

LDL-P Linear Regression Analysis:

$$
y = 1.0193x + 7.8226
$$

$$
R^2 = 0.9949
$$

Sample range tested: 248.8 – 4442.8

HDL-C Linear Regression Analysis:

$$
y = 1.0486x - 0.3459
$$

$$
R^2 = 0.9961
$$

Sample range tested: 5.0 – 151.5

TG Linear Regression Analysis:

$$
y = 1.008x - 0.3979
$$

$$
R^2 = 0.9999
$$

Sample range tested: 5.5 – 1356.3

{8}

9

Claimed Reportable Range for each analyte:

|  LDL-P | 300 – 3500 nmol/L  |
| --- | --- |
|  HDL-C | 7 – 140 mg/dL  |
|  TG | 5 – 1100 mg/dL  |

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

The sponsor states in the labeling that this device has not been certified or tested by the Cholesterol Reference Method Laboratory Network.

Traceability for calibrator – TMA (Trimethylacetic acid, Sodium salt, from Trimethyl acetate hydrate CAS-No 143174-36-1, and Ethylenediamine Tetraacetic acid CAS-No 139-33-3) is used as the NMR calibrator for the NMR clinical analyzers. TMA is used routinely as a calibrator once daily during instrument startup to establish daily normalization factors. It also serves as a quality assessment tool to ensure quality NMR spectra are produced by the NMR analyzer. New calibrator material is run in parallel with the existing calibrator in five separate runs. The TMA calibrator is characterized by the manufacturer using infrared spectrum (chemical structure), elemental analysis, and titration.

Traceability for control – Controls are included with the NMR LipoProfile® test. Bio-Rad Liquichek controls (k012513) are obtained by LipoScience and are assigned target values and expiration dates in house by LipoScience prior to providing the control sets to the users. The LDL-P specific value assignment occurs via in-house created master calibrators which are traceable to a purchased source material characterized by lipoprotein metabolism profiling using a combination of electrophoresis, ultracentrifugation, and automated enzymatic quantification of cholesterol and triglycerides (fractions HDL, LDL, VLDL, and Lp[a]). Each lot of control set is value assigned from the master calibrators through evaluation of 5 replicate measurements per run using 3 instruments analyzing 2 runs per day for 2 days.

Stability – The stability protocol of the TMA calibrator material and the control sets was reviewed and found to be adequate. Stability studies support a stability claim of TMA calibrator stability for 18 months either refrigerated or at room temperature. Stability for the quality controls is designated by the manufacturer and confirmed using stability studies in house. The stability protocols and acceptance criteria were reviewed and found to be adequate, with a specific stability claim for LDL-P of 6 months frozen at -20°C

Sample Stability – The sample stability protocol and acceptance criteria were reviewed and found to be adequate. The stability study supports a claim of the following:

{9}

10

|  Tube | Processing | Storage | LDL-P Stability | TG Stability | HDL-C Stability  |
| --- | --- | --- | --- | --- | --- |
|  LipoTube (serum) | Normal | Refrigerated | 8 days | 12 days | 7 days  |
|   |  Delayed Centrifugation | Refrigerated | 3 days | 12 days | 4 days  |
|  EDTA (plasma) | Normal | Refrigerated | 12 days | 12 days | 8 days  |

d. Detection limit:

**Limit of the Blank** – Five delipidated serum pools containing very low concentration were tested in replicates of 4 for 3 days. The Limit of Blank (LoB) was calculated non-parametrically for each analyte and determined to be the following:

LDL-P: 0.0 nmol/L

HDL-C: 2.7 mg/dL

TG: 1.1 mg/dL

**Limit of Detection** – Non-lipoprotein specimens were analyzed 60 consecutive times for 3 days. The Limit of Detection (LoD) was calculated parametrically for each analyte and supported the claimed measuring range of the assay.

LDL-P: 40.7 nmol/L

HDL-C: 3.5 mg/dL

TG: 2.5 mg/dL

**Limit of Quantitation** – Five serum pools containing very low concentrations were tested in replicates of 4 for 3 days. The Limit of Quantification (LoQ) was mathematically calculated for each analyte by plotting the %CV on the Y-axis against low concentration pools and determined to be the following:

LDL-P: 132 nmol/L

HDL-C: 4.0 mg/dL

TG: 4.0 mg/dL

The LoQ values listed above support the sponsors claimed measuring ranges of:

|  LDL-P | 300 – 3500 nmol/L  |
| --- | --- |
|  HDL-C | 7 – 140 mg/dL  |
|  TG | 5 – 1100 mg/dL  |

e. Analytical specificity:

Endogenous substances normally found in blood and exogenous substances (common and prescription drugs) were evaluated for potential interference with the NMR LipoProfile® test by LipoScience. Each potential interferent was diluted in the appropriate solvent and analyzed on the NMR Profiler to determine if it demonstrated

{10}

peak(s) in the  $0.7 - 1.0$  ppm region. Five endogenous agents and twenty three drugs were screened for potential interfering effects to NMR LipoProfile test using concentrations in accordance to CLSI EP7-A2 guidelines.

If a potentially interfering substance was suspected to have significant interference defined as difference from control greater than  $10\%$ , a spiking study was completed where the substance was added to sample pools containing two different levels of LDL-P, HDL-C and triglycerides for a paired difference test.

|  Endogenous |   | Exogenous (OTC drugs, etc.)  |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Potential Interferent | Test Concentration | Potential Interferent | Test Concentration | Potential Interferent | Test Concentration  |
|  Hemoglobin | 0.5 g/dL | Acetaminophen | 1324 μmol/L | Metformin Hydrochloride | 3.62 mmol/L  |
|  Bilirubin, unconj. | 342 μmol/L 20 mg/dL | Acetylsalicylic acid | 3.62 mmol/L | Metoprolol tartrate | 18.7 μmol/L  |
|  Creatinine | 442 μmol/L 5 mg/dL | Atorvastatin | 600 μg Eq/L | Naproxen Sodium | 2170 μmol/L  |
|  Urea | 42.9 mmol/L 260 mg/dL | Clopidogrel hydrogensulfate** | 95.7 μmol/L | Nicotinic Acid Sodium salt | 8.28 mmol/L  |
|  Uric acid | 1.4 mmol/L 23.5 mg/dL | Enalaprilat Dihydrate | 0.86 μmol/L | Nifedipine | 1156 nmol/L  |
|  Protein (albumin) | 6 g/dL, 60g/L | Fenofibrate | 125 μmol/L | Pioglitazone hydrochloride | 152.7μmol/L  |
|  Bilirubin, conj | 342 μmol/L 28.9 mg/dL | Furosemide | 181 μmol/L | Piroxicam | 181 μmol/L  |
|   |  | Glipizide | 4.48 μmol/L | Pravastatin | 107.5 μmol/L  |
|   |  | Hydralazine hydrochloride | 915.4 μmol/L | Salicylic Acid* | 1.3 mmol/L  |
|   |  | Heparin | 3000U/L | Simvastatin | 114.7 μmol/L  |

{11}

Ibuprofen Sodium salt 2425 μmol/L

Isosorbide dinitrate 636 nmol/L

Menhaden oil (Fish Oil) 2.4 mg/mL

*Salicyclic acid at ≥ 1.3 mmol/L was determined to interfere with LDL-P

**Clopidogrel hydrogensulfate at ≥ 95.7 μmol/L was determined to interfere with LDL-P

f. Assay cut-off:

Not applicable

2. Comparison studies:

a. Method comparison with predicate device:

LDL-P: The comparison was conducted in agreement with CLSI EP9-A2: Method Comparison and Bias Estimation Using Patient Samples. The study involved testing singlicates followed by statistical analysis of 1482 freshly-collected clinical de-identified specimens across the range of the LDL-P test on the Vantera Clinical Analyzer (candidate device) vs the NMR Profiler (predicate device). Bias was calculated for LDL-P from the linear regression line at the medical decision limits of 1000, 1300 and 1600 nmol/L as well. The range of the samples was 303 to 3479 nmol/L.

Linear regression/least squares analysis:

$$
y = 1.03x - 36.60
$$

$$
r = 0.978
$$

Observed LDL-P Bias per linear regression at Medical Decision Limits:

|  LDL-P (nmol/L) | Absolute Bias | Percent Bias (%)  |
| --- | --- | --- |
|  1000 | -10.0 | -1.0  |
|  1300 | -2.0 | -0.2  |
|  1600 | 6.0 | 0.4  |

{12}

In addition to conducting a method comparison for the overall data set, the data were classified into the three following segments based on LDL-P values; (1) results below the medical decision limits of 1000 nmol/L, (2) results between 1000 and 1600 nmol/L and (3) results above 1600 nmol/L. For each segment the bias was determined. The determined % Bias for each segment was less than 5% for each level).

**HDL-C**: The comparison was conducted in agreement with CLSI EP9-A2: Method Comparison and Bias Estimation Using Patient Samples. The study involved testing singlicates followed by statistical analysis of 1518 freshly-collected clinical de-identified specimens across the range of the HDL-C test on the Vantera Clinical Analyzer (candidate device) vs the NMR Profiler (predicate device). Bias was calculated for HDL-C from the linear regression line at the medical decision limits of 40 and 60 mg/dL as well. The range of the samples was 7.0 to 132 mg/dL.

**Linear regression/least squares analysis:**
$$
y = 1.04x - 1.20
$$
$$
r = 0.989
$$

**Observed HDL-C Bias per linear regression at Medical Decision Limits:**

|  HDL-C (mg/dL) | Absolute Bias | Percent Bias (%)  |
| --- | --- | --- |
|  40 | 0.3 | 0.8  |
|  60 | 1.1 | 1.8  |

In addition to conducting a method comparison for the overall data set, the data were classified into the three following segments based on HDL-C values; (1) results below the medical decision limits of 40 mg/dL, (2) results between 40 and 60 mg/dL and (3) results above 60 mg/dL. For each segment the bias was determined. The determined % Bias for each segment was less than 3% for each level.

**TG**: The comparison was conducted in agreement with CLSI EP9-A2: Method Comparison and Bias Estimation Using Patient Samples. The study involved testing singlicates followed by statistical analysis of 1520 freshly-collected clinical de-identified specimens across the range of the TG test on the Vantera Clinical Analyzer (candidate device) vs the NMR Profiler (predicate device). Bias was calculated for TG from the linear regression line at the medical decision limits of 150, 200 and 500 mg/dL as well. The range of the samples was 18.0 to 1095 mg/dL.

**Linear regression/least squares analysis:**
$$
y = 1.00x + 0.92
$$
$$
r = 0.998
$$

{13}

Observed TG Bias per linear regression at Medical Decision Limits:

|  TG (mg/dL) | Absolute Bias | Percent Bias (%)  |
| --- | --- | --- |
|  150 | 1.2 | 0.8  |
|  200 | 1.3 | 0.7  |
|  500 | 1.8 | 0.4  |

In addition to conducting a method comparison for the overall data set, the data were classified into the three following segments based on TG values; (1) results below the medical decision limits of  $150\mathrm{mg / dL}$ , (2) results between 150 and  $200\mathrm{mg / dL}$  and (3) results above  $200\mathrm{mg / dL}$ . For each segment the bias was determined. The determined  $\%$  Bias for each segment was less than  $3\%$  for each level.

# b. Matrix comparison:

Samples from 50 subjects (46 native and 4 contrived) were analyzed for each analyte by collecting the patient specimen into each of the claimed tube types. Comparative analysis of the LipoTube vs plain serum and EDTA plasma resulted in the following:

LipoTube vs Plain Serum

|   | LDL-P (nmol/L) | TG (mg/dL) | HDL-C (mg/dL)  |
| --- | --- | --- | --- |
|  Slope | 0.9678 | 1.002 | 0.9739  |
|  Y-Int | 26.4 | -0.417 | 1.589  |
|  R sq | 0.96 | 0.99 | 0.99  |

LipoTube vs EDTA Plasma

|   | LDL-P (nmol/L) | TG (mg/dL) | HDL-C (mg/dL)  |
| --- | --- | --- | --- |
|  Slope | 0.9442 | 0.9212 | 0.9339  |
|  Y-Int | -13.9 | 2.414 | 2.079  |
|  R sq | 0.97 | 1.00 | 0.99  |

# 3. Clinical studies:

a. Clinical Sensitivity:

Not applicable

b. Clinical specificity:

Not applicable

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable

{14}

4. Clinical cut-off:

See expected values below

5. Expected values/Reference range:

The sponsor conducted a reference range study in order to determine the normal distribution of LDL-P levels expected in a representative sampling of the general population. Healthy adult patient plasma samples were obtained (n=452) and analyzed (age ranging from 18 to 84 years). Samples or patients with specific history of diabetes or cardiac, renal or lipid disease were excluded, as well as any patients who were pregnant or being treated for cancer. Sample results were analyzed together, and also were partitioned into percentiles and separated based upon patient sex and age. The results of the reference range study are listed in the tables below:

Distribution of LDL-P Observed in a Reference Population

|   | All (n=452) | Men (n=158) | Women (n=294) | All (n=452) | Men (n=158) | Women (n=294)  |
| --- | --- | --- | --- | --- | --- | --- |
|  Percentile | LDL-P (nmol/L) | LDL-P (nmol/L) | LDL-P (nmol/L) | LDL-C (mg/dL) | LDL-C (mg/dL) | LDL-C (mg/dL)  |
|  5 | 539 | 528 | 542 | 63 | 62 | 65  |
|  10 | 643 | 713 | 638 | 75 | 76 | 75  |
|  20 | 784 | 883 | 749 | 84 | 90 | 83  |
|  30 | 909 | 1004 | 863 | 94 | 100 | 91  |
|  40 | 1009 | 1087 | 970 | 102 | 107 | 98  |
|  50 | 1127 | 1241 | 1070 | 109 | 113 | 109  |
|  60 | 1248 | 1366 | 1202 | 118 | 128 | 115  |
|  70 | 1396 | 1505 | 1322 | 129 | 137 | 124  |
|  80 | 1572 | 1676 | 1482 | 140 | 147 | 136  |
|  90 | 1894 | 1941 | 1818 | 157 | 161 | 151  |
|  95 | 2047 | 2169 | 1986 | 169 | 171 | 169  |

Distribution of LDL-P Observed in a Reference Population of Healthy Subjects Ages 18 - 44

|   | All (n=329) | Men (n=115) | Women (n=214) | All (n=329) | Men (n=115) | Women (n=214)  |
| --- | --- | --- | --- | --- | --- | --- |
|  Percentile | LDL-P (nmol/L) | LDL-P (nmol/L) | LDL-P (nmol/L) | LDL-C (mg/dL) | LDL-C (mg/dL) | LDL-C (mg/dL)  |
|  5 | 532 | 513 | 528 | 62 | 61 | 64  |
|  10 | 634 | 654 | 630 | 72 | 72 | 72  |
|  20 | 753 | 862 | 726 | 82 | 89 | 81  |
|  30 | 885 | 961 | 820 | 90 | 99 | 87  |
|  40 | 966 | 1041 | 930 | 98 | 106 | 94  |
|  50 | 1045 | 1149 | 1008 | 107 | 110 | 101  |

{15}

16

|  60 | 1191 | 1322 | 1115 | 113 | 126 | 109  |
| --- | --- | --- | --- | --- | --- | --- |
|  70 | 1328 | 1466 | 1251 | 122 | 135 | 118  |
|  80 | 1492 | 1639 | 1406 | 136 | 145 | 127  |
|  90 | 1807 | 1945 | 1666 | 150 | 161 | 146  |
|  95 | 2028 | 2241 | 1976 | 165 | 171 | 156  |

Distribution of LDL-P Observed in a Reference Population of Healthy Subjects ≥Age 45

|   | All (n=123) | Men (n=43) | Women (n=80) | All (n=123) | Men (n=43) | Women (n=80)  |
| --- | --- | --- | --- | --- | --- | --- |
|  Percentile | LDL-P (nmol/L) | LDL-P (nmol/L) | LDL-P (nmol/L) | LDL-C (mg/dL) | LDL-C (mg/dL) | LDL-C (mg/dL)  |
|  5 | 590 | 559 | 586 | 71 | 56 | 71  |
|  10 | 733 | 772 | 726 | 83 | 78 | 84  |
|  20 | 871 | 951 | 842 | 94 | 93 | 94  |
|  30 | 1025 | 1186 | 998 | 106 | 103 | 109  |
|  40 | 1188 | 1290 | 1158 | 113 | 110 | 115  |
|  50 | 1295 | 1364 | 1224 | 123 | 119 | 124  |
|  60 | 1402 | 1478 | 1368 | 134 | 133 | 134  |
|  70 | 1545 | 1643 | 1504 | 140 | 141 | 140  |
|  80 | 1752 | 1731 | 1796 | 150 | 153 | 146  |
|  90 | 1949 | 1934 | 1949 | 164 | 162 | 169  |
|  95 | 2075 | 2562 | 2016 | 186 | 181 | 188  |

Labeling for expected values states that each laboratory should verify the validity of the stated reference range values for the population the laboratory serves.

HDL concentration and triglycerides: The proposed device has similar reference values as the predicate. The reference values for patient classification have been recommended by the NCEP and taken from literature* for HDL cholesterol and triglycerides for the assessment and management of CVD risk. The labeling states that each laboratory should verify the validity of these reference values for the population it serves.

|  HDL Cholesterol, mg/dL Classification  |   |
| --- | --- |
|  Low | High  |
|  < 40 | ≥ 60  |
|  Triglycerides, mg/dL Classification  |   |   |   |
| --- | --- | --- | --- |
|  Normal | Borderline High | High | Very High  |
|  < 150 | 150 - 199 | 200 - 499 | > 500  |

*Expert Panel on Detection, Evaluation and Treatment of High Cholesterol in Adults (Adult Treatment Panel III), May (2001).
NIH Publication No. 01 3305, ATP III Guidelines At-A-Glance, Quick Desk Reference, May (2001).
NIH Publication No. 01 3670, Third Report of National Cholesterol Education Program (NCEP)

{16}

N. Instrument Name:

Vantera Clinical Analyzer

O. System Descriptions:

1. Modes of Operation:

Not applicable

2. Software:

FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:

Yes ☐ X ☐ or No ☐

3. Specimen Identification:

Internal barcode scanner identifies the specimens and automatically loads identification into the software.

4. Specimen Sampling and Handling:

Samples are loaded onto the instrument manually and samples are scanned and diluted / prepared automatically by the system.

5. Calibration:

Instrument calibration occurs daily (once every eight-hour shift) with the provided TMA calibrator (trimethyl acetic acid) to ensure the homogeneity of the magnetic field (line-shape specifications), to check the temperature of the NMR system, and to calculate the daily normalization factor. Calibration procedures are described in the labeling, and include a manual loading of the calibrant and automatic calibration by the instrument software.

6. Quality Control:

Quality controls are treated as specimens with specific control barcode labels that are provided with the system. The operator’s manual recommends running the quality controls once daily or in accordance with the laboratory practice.

P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

Not applicable

17

{17}

Q. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

R. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

18

---

**Source:** [https://fda-staging.innolitics.com/device/K113830](https://fda-staging.innolitics.com/device/K113830)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
