The Mongoose Angiographic Catheter is intended for the delivery of contrast media to selected sites in the vascular system of adult and pediatric patients of all ages in conjunction with routine diagnostic procedures.
Device Story
Single-use, sterile, non-pyrogenic intravascular catheter; used in angiographic procedures to deliver contrast media to vascular sites. Available in 3.3F, 4.0F, and 5F sizes; lengths 35-135 cm; various distal tip shapes (pigtail, JB, JR, JL, MP, Cobra). Features hydrophilic coating, wire-braided reinforcement, and radiopaque barium sulfate for fluoroscopic visualization; optional side holes for contrast dispersion. Includes optional inserter accessory for pigtail configurations. Operated by physicians in clinical settings. Device facilitates diagnostic imaging; aids in visualization of vascular anatomy; supports clinical decision-making through improved contrast delivery.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including dimensional and functional performance testing (surface requirements, corrosion resistance, tensile strength, leakage, radiodetectability, and catheter fittings) and biocompatibility testing per ISO 10993.
Technological Characteristics
Single-lumen intravascular catheter; wire-braided reinforcement; hydrophilic coating; radiopaque barium sulfate for visualization. Sizes: 3.3F, 4.0F, 5F. Lengths: 35-135 cm. Sterilization: Ethylene oxide (ISO 11135-1:2007). Single-use, disposable.
Indications for Use
Indicated for delivery of contrast media to vascular sites in adult and pediatric patients of all ages during routine diagnostic procedures.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
{0}------------------------------------------------
# Section 5. 510(k) Summary
Submission Correspondent Emergo Group, Inc. www.emergogroup.com
# Address
611 West Fifth Street Third Floor Austin, TX 78701
Phone (512) 327-9997
Fax (512) 327-9998
Contact Caroline Tontini Project Manager project.management@emergogroup.com
## Submission Sponsor
PediaCath Inc 10000 Cedar Avenue #16 Cleveland, OH 44106 USA 888-691-6161 (Telephone)
Date Prepared 15 December 2011
Trade Name Mongoose Angiographic Catheter
Classification Name Diagnostic Intravascular Catheter
Regulation Number 870.1200
Product Code DQO
JUL 1 1 2012
{1}------------------------------------------------
K113819
p. 3 of 5
Classification Panel Cardiovascular
### Device Class
Class II
#### Predicate Devices
- 1. K943739: Softouch Diagnostic Intravascular Catheter (Merit Medical is currently responsible for marketing the device originally submitted under this premarket notification).
- 2. K060116: Micrus Courier Microcatheters
### Indications for Use
The Mongoose Angiographic Catheter is intended for the delivery of contrast media to selected sites in the vascular system of adult and pediatric patients of all ages in conjunction with routine diagnostic procedures. .
#### Device Description
The Mongoose Angiographic Catheter is a single-use, sterile, non-pyrogenic disposable intravascular catheter for use in angiographic procedures. The Mongoose Angiographic Catheter is intended for adult and pediatric patients of all ages.
The Mongoose Angiographic Catheter is available in 3.3 F, 4.0 F and 5 F sizes. These catheters are available in effective lengths ranging from 35-135 cm and a variety of distal end shapes. The distal tip, soft tube and shaft of Mongoose Angiographic Catheters are manufactured with radiopaque barium sulfate to aid in device visualization under fluoroscopy. Catheters have a hydrophilic coating and are reinforced with wire braiding. Side holes are available to disperse the contrast medium.
An inserter is made available as an accessory to the Mongoose Angiographic Catheters with a pigtail distal configuration. The inserter facilitates insertion of the angiographic catheter into an introducer.
### Technological Characteristics and Substantial Equivalence
The following table compares the Mongoose Angiographic Catheter to the predicate devices with respect to intended use and technological characteristics, providing more detailed information regarding the basis for the determination of substantial equivalence.
{2}------------------------------------------------
| Parameter | PediaCath<br>Mongoose Angiographic<br>Catheter | Merit Medical Performa<br>Pediatric JL/JR and Angiography<br>Pigtail Catheters | Micrus Courier Microcatheters |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | To be assigned | K943739 | K060116 |
| Indications for Use | The Mongoose Angiographic<br>Catheter is intended for the<br>delivery of contrast media to<br>selected sites in the vascular<br>system of adult and pediatric<br>patients of all ages in<br>conjunction with routine<br>diagnostic procedures. | Angiographic catheters are<br>designed to be used for<br>delivering radiopaque media to<br>selected sites in the vascular<br>system in conjunction with<br>routine diagnostic procedures.<br>Angiographic catheters with<br>marker bands may also be used<br>for anatomical measurements. | The Micrus Courier<br>microcatheters are intended to<br>assist in the delivery of<br>diagnostic agents, such as<br>contrast media, and therapeutic<br>agents such as occlusion coils,<br>into peripheral, coronary, and<br>neuro vasculature. |
| Regulation Number | 870.1200 | 870.1200 | 870.1200 |
| FDA Product Code | DQO | DQO | DQO |
| Prescription/<br>OTC Use | Prescription | Prescription | Prescription |
| Single-Use/<br>Reusable | Single-Use | Single-Use | Single-Use |
| Dimensions | | | |
| French Sizes (available) | 3.3 F, 4 F, 5 F | 3 F, 4 F, 5 F, 6 F, 7 F | 2.3-2.4 F Proximal OD/<br>1.8-1.9 F Low Profile Distal OD |
| Total Length (cm) | 42 cm – 142 cm | 40 cm – 100 cm | 156 cm |
| Recommended Guide Wire | Varies by French Size and Distal<br>Tip Shape from 0.030" (0.76 mm)<br>to 0.038" (0.97 mm) | Varies by catheter French Size<br>from .021" (0.53 mm) to .038"<br>(0.97 mm) | .014" |
| Physical Characteristics | | | |
| Lumen Construction | Single | Single | Single |
| Coating | Hydrophilic | No | Hydrophilic |
| Parameter | PediaCath<br>Mongoose Angiographic<br>Catheter | Merit Medical Performa<br>Pediatric JL/JR and Angiography<br>Pigtail Catheters | Micrus Courier Microcatheters |
| Reinforcement Material | Wire braid design | Wire braid design | Wire braid design |
| Distal Tip Shapes | - Pigtail<br>- JB<br>- JR<br>- JL<br>- MP<br>- Cobra | - JL<br>- JR<br>- Pigtail | - Straight<br>- 45°<br>- 90°<br>- Custom tip shape accomplished using steam-shaping mandrel supplied with catheter |
| Side Holes | Yes | Yes | No |
| Sterilization and Shelf Life | | | |
| Sterilization Method | Ethylene oxide | Ethylene oxide | Ethylene oxide |
| Non-Pyrogenic | Yes | Yes | Yes |
| Sterile Package | Pouch | Pouch | Pouch |
| Shelf Life (Use By Date) | Three (3) years | Three (3) years (at minimum) | One (1) year |
mparison Table Davico C-1 oblo
KI13819
p. 5 of 5
{3}------------------------------------------------
.
.
K113819
p..4 of 5.
1. The state of the state
:
and the control control of the control of the control of
.
{4}------------------------------------------------
K11-3819
p. 5 of 5
#### Summary of Non-Clinical Data Submitted
Performance testing was conducted on the Mongoose Angiographic Catheter to establish substantial equivalence. Testing was conducted according to protocols based on international standards and in-house requirements, and included dimensional and functional testing. Functional performance testing included surface requirements, corrosion resistance, tensile strength, leakage tests, radiodetectability and tests specific to catheter fittings. The Angiographic Catheter was subjected to biocompatibility testing in accordance with ISO 10993-Additionally the Mongoose Angiographic Catheter was adopted into the existing 1:2009. ethylene oxide sterilization cycle which was validated in accordance with !SO 11135-1:2007.
#### Safety and Effectiveness
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device, or has the same intended use and different technological characteristics, and it can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
It has been shown in this 510(k) submission that the differences between the Mongoose Angiographic Catheter and predicate devices do not raise any questions regarding its safety and effectiveness. The Mongoose Angiographic Catheter, as designed and manufactured, therefore is determined to be substantially equivalent to the referenced predicate devices.
{5}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Pediacath Inc. c/o Ms. Caroline Tontini Emergo Group, Inc. 611 West 5th Street Third Floor Austin, TX 78701
Re: K113819
Trade/Device Name: Mongoose Angiographic Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: DQO Dated: June 21, 2012 Received: June 22, 2012
Dear Mr. Mr. Moran:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
JUL 1 1 2012
{6}------------------------------------------------
Page 2 - Ms. Timothy Moran
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set de riot the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you desire speethe ad ready Jor Jor J CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
M. G. Hilleben
C مر Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
| 510(k) Number (if known): | K113819 |
|---------------------------|---------|
|---------------------------|---------|
Device Name:
Mongoose Angiographic Catheter
Indications for Use:
The Mongoose Angiographic Catheter is intended for the delivery of contrast media to selected sites in the vascular system of adult and pediatric patients of all ages in conjunction with routine diagnostic procedures.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED T
Concurrence of CDRH, Office of Device Evaluation (ODE)
Cmsf. Ytllahan
Division Sian-Off) Division of Cardiovascular Devices Page 1 of 1
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.