K113767 · Essex Industries, Inc. · BYJ · Jun 29, 2012 · Anesthesiology
Device Facts
Record ID
K113767
Device Name
MOUNTED MEDICAL OXYGEN SYSTEM
Applicant
Essex Industries, Inc.
Product Code
BYJ · Anesthesiology
Decision Date
Jun 29, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5655
Device Class
Class 2
Attributes
Therapeutic, Expedited Review
Indications for Use
The MMOS is intended to convert liquid oxygen to gaseous oxygen for delivery to a patient and for delivery to rescue personnel to supplement environmental oxygen at high altitudes while mounted in rescue vehicles, e.g. HH-60 and HC-130 Guardian Angel (GA) rescue vehicles, at one-half (0.5) to fifteen (15) liters per minute (LPM) and fifty (50) pounds per square inch gauge (psig).
Device Story
Portable liquid oxygen (LOX) system; stores 4L LOX in thermally insulated cylinder; converts LOX to gaseous oxygen via heat exchanger. Used in rescue vehicles (HH-60, HC-130) by rescue personnel to treat 1-2 patients or provide supplemental oxygen during high-altitude/parachuting operations. Connects to masks, cannulas, or Bag Valve Masks (BVM). Features include pressure/quantity monitoring, visual low-quantity alarm (at 10%), and 9V lithium battery power. Provides gaseous oxygen at 50±5 psig; flow rate up to 22 LPM combined. Enhances patient care by providing portable, reliable oxygen supply in isolated/field environments.
Clinical Evidence
Bench testing only. Engineering personnel completed capability, performance, and environmental testing; results confirmed device meets specifications for safety and effectiveness.
Technological Characteristics
Thermally insulated container (49 CFR §178.57, Spec 4L); aluminum sheet metal enclosure; heat exchanger for LOX-to-gas conversion. 4L LOX capacity. 9V lithium battery. Dimensions: 10.63" x 17.00" x 8.75". Weight: 35 lbs (filled). Pressure: 50±5 psig. Monitoring: pressure and LOX quantity with visual low-quantity alarm.
Indications for Use
Indicated for patients requiring supplemental oxygen and rescue personnel requiring supplemental oxygen at high altitudes. Used in rescue vehicles (e.g., HH-60, HC-130). Delivers 0.5-15 LPM at 50 psig.
Regulatory Classification
Identification
A portable liquid oxygen unit is a portable, thermally insulated container of liquid oxygen that is intended to supplement gases to be inhaled by a patient, is sometimes accompanied by tubing and an oxygen mask. An empty portable liquid oxygen unit is a device, while the oxygen contained therein is a drug.
Predicate Devices
Portable Therapeutic Liquid Oxygen System (NPTLOX) (K033000)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a circular shape with a 'C' shape cut out from the center. The 'C' shape is not fully closed, leaving a small gap on the right side. The area between the outer circle and the 'C' shape is filled with a gray pattern, creating a visual contrast.
8007 Chivvis Drive · St. Louis. MO 63123-2395 tel: 314.832.8077 • fax: 314.832.8208 www.essexind.com • www.medlox.com
JUN 2 9 2012
#### K-113767 510(k) Summary
1. Owner Information
| Owner's Name: | Essex Cryogenics of Missouri, Inc. |
|-----------------|------------------------------------|
| Address: | 8007 Chivvis Drive |
| | St. Louis, Missouri 63123-2395 |
| Phone Number: | 314-832-8077 |
| Fax Number: | 314-832-8208 |
| Contact Person: | Kenneth L. Seise |
| Date: | December 16, 2011 |
2. Medical Device Information
| Trade Name: | Mounted Medical Oxygen System (MMOS) |
|----------------------|--------------------------------------|
| Common Name: | Portable Liquid Oxygen System |
| Classification Name: | Portable Liquid-Oxygen Unit |
- 3. Substantial Equivalence to Predicate Medical Device
The MMOS is substantially equivalent to the Portable Therapeutic Liquid Oxygen System (NPTLOX) (K033000).
- 4. Medical Device Description
#### Physical and performance characteristics
The MMOS, when filled with liquid oxygen, is used to provide medical treatment to injured isolated personnel and to supplement environmental oxygen during high altitude and parachuting operations. The MMOS is primarily employed on board the HH-60, HC-130, and Guardian Angel (GA) rescue Vehicle to treat 1-2 patients for durations of 4-8 hours. The following general requirements apply.
- The MMOS supplies 93% oxygen concentration when filled from an Oxygen a) Generator System (OGS).
- The MMOS is capable of delivering gaseous oxygen to two patients at a b) combined maximum system flow rate of 22 Liters-Per-Minute (LPM) ambient.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo that resembles the letter 'C' inside of a circle. The 'C' is not fully closed, and it has a small extension on the right side. The logo is composed of two concentric circles, with the 'C' shape formed by the inner circle and the extension. The lower right portion of the logo is shaded, creating a visual contrast.
8007 Chivvis Drive · St. Louis. MO 63123-2395 tel: 314.832.8077 • fax: 314.832.8208 www.essexind.com · www.medlox.com
- The MMOS is capable of delivering oxygen at flow rate of 7 LPM for a minimum c) duration of 8 hours at one outlet.
- d) The MMOS operating pressure is 50± 5 pounds per square inch gauge (psig).
- The liquid oxygen (LOX) capacity is 4 liters. e)
- The MMOS has two secondary accessory ports in addition to both patient ports. f)
- The MMOS operates up to 14,000 Ft. MSL. g)
- The oxygen delivery pressure is monitored and displayed. h)
- The liquid oxygen quantity is monitored and displayed. i)
- The MMOS has a visual low quantity alarm that triggers when the liquid capacity j) is at or below 10%.
- Power required by the MMOS is a self-contained 9-volt lithium battery. k)
- The MMOS has the capability to be filled with LOX by standard Department of 1) Defense (DoD) and North Atlantic Treaty Organization (NATO) servicing connectors.
- The overall system weight (when filled with 4.0 liters of liquid oxygen) is 35 lbs. m) Empty weight is 25 lbs.
- The MMOS is 10.63 inches wide x 17.00 inches long x 8.75 inches wide within a n) tolerance of .06 inches in all directions.
#### How the device functions
The MMOS provides for storing of 4 liters of liquid oxygen and converting this liquid into its gaseous state. The gaseous oxygen is capable of being delivered in controlled amounts to provide medical treatment to injured patients and uncontrolled amounts to drive respiratory medical devices or supplemental oxygen in high altitude and parachuting operations. The MMOS is capable of being filled with liquid oxygen from parachages Generator System (OGS) and with current liquid oxygen storage/ filling stations.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a circular shape with a gap in the lower right quadrant. Inside the larger circle is a smaller, incomplete circle that is connected to a smaller, filled circle on the left side. The overall design resembles a stylized letter 'C' within a circle, with the gap in the outer circle mirroring the opening of the 'C'.
1
8007 Chivvis Drive · St. Louis. MO 63123-2395 tel: 314.832.8077 • fax: 314.832.8208 www.essexind.com • www.medlox.com
MMOS contains liquid oxygen (LOX). The MMOS contains a thermally insulated container of liquid oxygen (LOX) that is intended to supplement gases to be inhaled by a patient. The MMOS supports medical devices provided by the user including masks, cannulas, and Bag Valve Mask (BVM) being attached to the flow control patient outlets. An empty portable liquid oxygen unit is a device, while the oxygen contained therein is a drug.
The MMOS is portable. The handle on top of the housing, as well as the handles on the left and right side of the front-angled user interface, make the MMOS portable. The MMOS can be carried onboard, tied down, transported and operationally perform on various aircraft and ground vehicles.
The MMOS converts liquid oxygen from its insulated container through its heat exchanger into gaseous oxygen and finally the gaseous oxygen is available for distribution from an outlet port on the user interface. After connecting a tube assembly connector of a mask, cannula, and Bag Valve Mask (BVM) to an outlet port, masks, cannulas, and BVMs or other similar medical device (none of these devices is included in the MMOS) the patient can inhale the gaseous oxygen.
### 5. Intended Use of Medical Device
The MMOS is intended to convert liquid oxygen to gaseous oxygen for delivery to a patient and for delivery to rescue personnel to supplement environmental oxygen at high altitudes while mounted in rescue vehicles, e.g. HH-60 and HC-130 Guardian Angel (GA) rescue vehicles, at one-half (0.5) to fifteen (15) liters per minute (LPM) and fifty (50) pounds per square inch gauge (psig).
6. Comparison to Predicate Medical Device
The NPTLOX is the predicate medical device to which MMOS is compared. Both are portable. Personnel can move an NPTLOX device by picking up the unit by its two side handles and walking with it while personnel can move an MMOS device by picking up the unit by its two handles on the front-angled user interface or single top handle. Both the NPTLOX and MMOS devices are similar in that they include a thermally insulated container. Both containers are designed built, and tested per 49 CFR §178.57, Specification 4L welded insulated cylinders, and are housed in aluminum sheet metal enclosures that contain heat exchangers that convert the liquid oxygen to gaseous oxygen. Both devices have ports that can be connected to tubing intended for connection to an
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a white circle with a shape inside that resembles the letter 'C'. The 'C' shape is also white and is surrounded by a thicker, gray ring. The 'C' shape has a small, rounded extension on its top end, and the bottom end is open, creating a gap in the circle.
8007 Chiwis Drive · St. Louis. MO 63123-2395 tel: 314.832.8077 · fax: 314.832.8208 www.essexind.com · www.medlox.com
oxygen mask or other similar medical device to ultimately supplement gases to be inhaled by a patient.
Both devices can be monitored for pressure and liquid oxygen quantity.
The differences in the NPTLOX and MMOS are in the container size and LOX capacity. The NPTLOX container is a sphere that is approximately 13.5 inches in diameter with a LOX capacity of 25 liters while the MMOS container is a cylinder that is approximately 11.3 inches long and 6.1 inches in diameter with a LOX capacity of 4 liters. Thus both the NPTLOX and MODS are portable liquid-oxygen units that function the same but are simply different sizes.
7. Medical Device Tests
Essex Cryogenics of Missouri, Inc. Engineering personnel completed extensive MMOS capability, performance, and environmental testing with no issues arising regarding its safety and efficiency. The combined testing and analysis of results provides assurance that the device meets it specifications and is safe and effective for its intended use.
8. Conclusions
Based on review of the design and test results, Essex Cryogenics of Missouri, Inc. believes that no significant differences exist between this medical device, MMOS, and the predicate medical device, NPTLOX, and therefore the MMOS as safe, as effective and performs as well as or better than the NPTLOX.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three lines representing the staff and two snakes intertwined around it. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Essex Industries, Incorporated Mr. Ken Seise Quality Assurance Manager Essex Cryogenics of Missouri, Incorporated 8007 Chivvis Drive Saint Louis, Missouri 63123
Re: K113767
Trade/Device Name: Mounted Medical Oxygen System Regulation Number: 21 CFR 686.5655 Regulation Name: Portable Liquid Oxygen Unit Regulatory Class: II Product Code: BYJ Dated: June 14, 2012 Received: June 18, 2012
#### Dear Mr. Seise:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
JUN 2 9 2012
{5}------------------------------------------------
#### Page 2- Mr. Seise
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
th for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a circular design with a smaller circle connected to the larger circle by a horizontal line. The larger circle has a gap in the lower right quadrant. The design appears to be a stylized representation of a letter or symbol.
8007 Chiwis Drive · St. Louis. MO 63123-2395 tel: 314.832.8077 • fax: 314.832.8208 www.essexind.com • www.medlox.com
### Indications for Use Statement
510(k) Number (if known):
Device Name: Mounted Medical Oxygen System
Indications for Use: The MMOS is intended to convert liquid oxygen to gaseous oxygen for delivery to a patient and for delivery to rescue personnel to supplement environmental oxygen at high altitudes while mounted in rescue vehicles, e.g. HH-60 and HC-130 Guardian Angel (GA) rescue vehicles, at one-half (0.5) to fifteen (15) liters per minute (LPM) and fifty (50) pounds per square inch gauge (psig).
Prescription Use (CFR Title 21, Part 801, Subpart D) AND/OR
Over-The-Counter Use (CFR Title 21, Part 801, Subpart C)
(PLEASE DON NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schullhm
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
4-1
510/k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.