K113713 · Free Wheelchair Mission · IOR · May 7, 2012 · Physical Medicine
Device Facts
Record ID
K113713
Device Name
GEN_2 MECHANICAL WHEELCHAIR
Applicant
Free Wheelchair Mission
Product Code
IOR · Physical Medicine
Decision Date
May 7, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3850
Device Class
Class 1
Indications for Use
The GEN 2 Mechanical Wheelchair is intended to provide mobility to persons restricted to a seated position.
Device Story
The GEN 2 Mechanical Wheelchair is a manually operated, all-purpose mobility device designed for use in developing countries. It features a powder-coated steel frame, adjustable seating, large 8-inch castor wheels, 26-inch spoke wheels with pneumatic tires, and an extra-thick cushion. The device is designed to provide a personalized fit for users, with adjustable footrests and back upholstery. It is intended for use by individuals restricted to a seated position to improve their mobility and quality of life. The device is operated by the user or a caregiver. It is a non-powered, mechanical device that does not require complex clinical workflows or electronic processing.
Clinical Evidence
Bench testing only. The device underwent mechanical performance testing according to international standards, including ISO 7176-1 (static stability), ISO 7176-3 (effectiveness of brakes), ISO 7176-5 (dimensions/mass), ISO 7176-7 (seating dimensions), ISO 7176-8 (static, impact, and fatigue strength), ISO 7176-15 (documentation), and ISO 7176-16 (resistance to ignition). No clinical data was required or provided.
Technological Characteristics
Powder-coated steel frame (ASTM A 53/A 53M-2005, JIS G 3444-2004); manual operation; 26-inch spoke wheels with pneumatic tires; 8-inch polypropylene castor wheels; polypropylene footplates; adjustable back upholstery. Standards applied: ISO 7176-1, 3, 5, 7, 8, 15, 16. Weight limit: 220 lbs. No electronic components or software.
Indications for Use
Indicated for persons restricted to a seated position requiring mobility assistance.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the "Free Wheelchair Mission". The logo consists of a drawing of two wheels, one larger than the other, with the words "Free Wheelchair Mission" written below. Underneath the name is the text "TRANSFORMING LIVES THROUGH THE GIFT OF MOBILITY".
KII3713 PAGE 1 OF 4
MAY - 7 2012
# 510(K) SUMMARY (21 CFR 807.92)
## GEN 2 MECHANICAL WHEELCHAIR
Free Wheelchair Mission 510(k) Owner: 15279 Alton Parkway, Suite 300 Irvine, CA 92618 Tel: 949-273-0858 Fax: 949-273-8471
Contact Person: Sharon Rockwell Tel: 714-695-9269 E-mail: srockwell@writeme.com
December, 2011 Date Prepared:
GEN 2 Mechanical Wheelchair Trade Name:
Mechanical Wheelchair Common Name:
Classification Name: Mechanical wheelchair per 21 CFR 890.3850, IOR
Ki Mobility Catalyst Wheelchair, K062660 Predicate Devices:
The GEN 2 mechanical wheelchair is a highly adaptable all-Device Description: purpose wheelchair created for use in developing countries, while allowing for a healthy, personalized fit. The wheelchair features adjustable seating, large castor wheels, an extra-thick cushion and adjustable footrests.
The GEN_2 is a manually operated device intended to provide Intended Use: mobility for persons restricted to a sitting position.
> The Indications for Use are identical to those of the predicate devices.
{1}------------------------------------------------
### Technological Characteristics:
Technological Characteristics: The GEN_2 mechanical wheelchair is substantially equivalent in design, materials, and intended use to the Ki Mobility Catalyst mechanical wheelchair as described in the table below:
| Feature | Catalyst | GEN 2 |
|-----------------------|---------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Intended Use | As a means of mobility for<br>persons restricted to a sitting<br>position. | To provide mobility to persons<br>restricted to a seated position. |
| Frame material | 6061-T6 Aluminum | Powder coated steel frame that<br>meets ASTM A 53/A 53M-2005<br>and JIS G 3444-2004 standards |
| Frame widths | 14-20" | 13.5-19.5" |
| Overall widths | 20.5-26.5" | 29" |
| Seat depths | 14-20" | 11.5-17" |
| Back heights | 8.5-19" | 11.8-19" |
| Weight limit | 220 lbs | 220 lbs |
| Chair weight | 23 lbs (without footrests) | 36 lbs (with footrests) |
| Armrests | Flip back height adjustable desk<br>and full length arm pads.<br>Tubular swing-away. | Fixed in place to serve as<br>armrests but not to restrict<br>transfers. |
| Front end type | Swing-away, non-swing-away | Swing-away |
| Back type | Standard | Standard |
| Footrest hangers | 70°, elevating leg rest | NA |
| Footplates | Composite, foam, angle<br>adjustable | Polypropylene, 280 mm<br>adjustable range with angles<br>adjustable to 0, ± 7°, ± 14°, and ±<br>21°. |
| Extension tubes | Extra short, short, med, long | NA |
| Back upholstery | Low, med, tall, adjustable | Adjustable to 4 different heights |
| Axle plates | Standard, curved, amputee,<br>offset | NA |
| Wheel sizes | 22, 24, 26 | 26" |
| Wheel types | Spoke, composite mag, Octopus<br>(performance spoke) | Spoke |
| Tire types | Pneumatic, full profile<br>polyurethane, iron cap<br>(puncture resistant), high<br>pressure | Pneumatic |
| Handrims | Aluminum, plastic coated<br>projections | Powder coated steel |
| Feature | Catalyst | GEN 2 |
| Caster sizes | 4, 5, 6, 8" | 8" |
| Caster types | Poly, pneumatic, pneumatic<br>with airless insert | Polypropylene |
| Forks sizes | 4, 5, 6, 7" frog legs | 5" |
| Wheel locks | Push to lock, pull to lock, low<br>profile | Pull to lock |
| Anti tips tubes | Yes | No |
| Standards applied | Static Stability and Fatigue<br>Strength, ANSI/RESNA std<br>sections 1, 5, 7, 8, 16 and 93 | ISO 7176-1, 3, 5, 7, 8, 15, 16 |
| Tubing wall thickness | 0.070" on frame except front<br>vertical tube which is 0.1". X-<br>tube is 0.070" | 0.079" throughout |
| Tube properties | 6061-T4 prior to welding.<br>After welding, the component is<br>treated up to T-6. | ASTM A 53/A 53M-2005<br>JIS G 3444-2004 |
{2}------------------------------------------------
The indications for use for the GEN 2 and the predicate device are identical. The size ranges in terms of frame width, seat depth, back depths, chair weight and weight limit overlap those of the predicate device. Most other features are very similar. The fundamental technology, welded tube frame, fire retardant upholstery, and wheel locks are the same. The information provided supports a substantial equivalence decision. Any differences in features to the predicate do not affect the performance of the device and do not raise new types of safety or effectiveness questions.
- Non-Clinical
Non-clinical testing has been performed to mechanical wheelchair Performance Data: international standards including:
1) ISO 7176-1:1999 Determination of static stability
2) ISO 7176-3:2006 Determination of effectiveness of breaks
3) ISO 7176-5:2008 Determination of dimensions, mass and manoeuvring space
4) ISO 7176-7:1998 Measurement of seating and wheel dimensions
5) ISO 7176-8:1998 Requirements and test methods for static, impact and fatigue
6) ISO 7176-15:1996 Requirements for information disclosure, documentation and labeling
7) ISO 7176-16: 1997 Resistance to ignition of upholstered parts --Requirements and test methods
{3}------------------------------------------------
The ISO tests were conducted with methods that met the following standards:
1) ISO 7176-11:1992 Test dummies
2) ISO 7176-13: 1989 Determination of coefficient of friction of test surfaces
The non-clinical test results demonstrate the GEN 2 Mechanical Conclusions: Wheelchair does not raise any issues regarding safety and effectiveness. The testing supports a determination of substantial equivalence to mechanical wheelchairs previously cleared by FDA.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of an eagle or bird-like figure with three wing-like shapes extending upwards and to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Free Wheelchair Mission % Ms. Sharon Rockwell Consultant 5582 Chalon Road Yorba Linda, California 92886
MAY - 7 2012
、Re: K113713
Trade/Device Name: Free Wheelchair Mission GEN 2 Mechanical Wheelchair Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical wheelchair Regulatory Class: Class I Product Code: IOR Dated: March 26, 2012 Received: March 27, 2012
Dear Ms. Rockwell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
#### Page 2 - Ms. Sharon Rockwell
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Eunel Keith
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Free Wheelchair Mission GEN_2 Mechanical Wheelchair
Indications for Use:
The GEN 2 Mechanical Wheelchair is intended to provide mobility to persons restricted to a seated position.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
Division Sign-Off
Page ( of )
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K113713
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.