K113702 · Fotona D.D. · PDZ · Mar 28, 2012 · General, Plastic Surgery
Device Facts
Record ID
K113702
Device Name
FOTONA XP ND: YAG LASER SYSTEM
Applicant
Fotona D.D.
Product Code
PDZ · General, Plastic Surgery
Decision Date
Mar 28, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Fotona XP Nd:YAG Laser System Family and its accessories are intended for use in the following procedures: Podiatry (ablation, vaporization, incision, excision, and coagulation of soft tissue) including: - Matrixectomy - - Radical nail excision - - Periungual and subungual warts - - Plantar warts - - Neuromas - - Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes - Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.)
Device Story
Fotona XP Nd:YAG Laser System is a flashlamp-pumped solid-state laser (1064 nm) for podiatric surgical procedures. System includes optical cavity with Nd:YAG crystal, red diode aiming beam, and fiber-optic delivery system with optical handpiece. User activates laser via footswitch. Device operates in pulsed or QCW modes with variable pulse durations (0.1-50 msec), repetition rates up to 100 Hz, and pulse energies up to 20 J. Clinicians use device in clinical settings to perform soft tissue procedures and treat onychomycosis. Output parameters are selected via keyboard/control panel. Laser energy facilitates tissue ablation, vaporization, incision, excision, and coagulation. Benefits include precise soft tissue management and temporary increase of clear nail in onychomycosis patients.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and performance comparisons to predicate devices.
Indicated for podiatry patients requiring soft tissue ablation, vaporization, incision, excision, or coagulation, including matrixectomy, radical nail excision, periungual/subungual warts, plantar warts, neuromas, and temporary increase of clear nail in patients with onychomycosis (dermatophytes or yeasts).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
K113702 Page w of a
# 5. 510(k) Summary
| Submitter's Name: | Fotona d.d.<br>Stegne 7<br>SI-1000 Ljubljana, Slovenia<br>Phone: +386 15009100<br>Fax: +386 5009 200 |
|-------------------|------------------------------------------------------------------------------------------------------|
| Contact Person: | Stojan Trošt, QA&RA Manager<br>Phone: +386 1 5009 299<br>E-mail: stojan.trost@fotona.com |
MAR 2 8 2012
Date: December 14, 2011
Fotona XP Nd:YAG Laser System Family Device Trade Name: Nd:YAG Surgical Laser Common Name: Instruments, Surgical, Powered, Laser Classification Name: 79-GEX
### DEVICE DESCRIPTION
The Fotona XP Nd:YAG Laser System Family is based on thè Nd:YAG (1064 nm) laser technology. There is one optical cavity containing the Nd: YAG crystal. The Nd: YAG laser is activated by means of the use of a flashlamp. After the cavity, a red diode aiming beam is reflected onto a coaxial bcam path using a beamsplitter assembly. The combined therapeutic and aiming beams are guided by an optical fiber delivery system to an optical handpiece. Optionally, the laser beams can be guided through fiber having SMA905 connector on the proximal end and bare fiber on distal end. The user activates laser emission by means of a footswitch.
The Nd: YAG (1064 nm) laser can be operated with variable pulse durations in the range of 0.1-50 msec, with repetition rates up to 100 Hz and laser pulse energies up to 20 J. Depending on the type of treatment, the user can choose between a PULSE mode user interface, and a QCW (Quasi-Continuous Wave) mode user interface. The laser operation is the same for both modes, the difference is only in the laser parameters that can be directly selected from the keyboard.
### INTENDED USE
Fotona XP Nd:YAG Laser System Family and its accessories are intended for use in the following procedures:
Podiatry (ablation, vaporization, incision, excision, and coagulation of soft tissue) including:
- Matrixectomy ー
- Radical nail excision -
- Periungual and subungual warts ー
- Plantar warts l
- Neuromas -
- Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes ー Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.)
{1}------------------------------------------------
. K113702 Page (2) of (2)
### PREDICATE DEVICES
- Cutera GenesisPlus (K103626) l
- PinPointe FootLaser (K093547) l
## SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
The Fotona XP Laser System Family has identical technological and similar design characteristics as the predicate devices. The output characteristics are for the intended use the same as those of the predicate devices.
A comparison of the technical characteristics for the intended use of the Fotona XP Laser System Family with those of the predicate devices is provided in the Table 1 below.
| Characteristic | Fotona XP Nd:YAG<br>Laser System Family | PinPointe Footlaser<br>K093547 | Cutera GenesisPlus Laser<br>K103626 |
|------------------|-------------------------------------------------------------|---------------------------------------------------------------|-----------------------------------------------------------|
| Laser source | Nd:YAG | Nd:YAG | Nd:YAG |
| Wavelength | 1064 nm | 1064 nm | 1064 nm |
| Laser media | Flashlamp pumped solid<br>state rod | Flashlamp pumped solid<br>state rod | Flashlamp pumped solid<br>state rod |
| Aiming beam | 650 nm (≤ 1 mW) | 630-680 nm (≤ 2,5 mW) | 630-680 nm (≤ 2,5 mW) |
| Output mode | Pulsed, Multimode | Pulsed, Multimode | Pulsed, Multimode |
| Energy per pulse | up to 3,5 J (podiatry)<br>up to 20 J (other<br>indications) | up to 3,5 J | up to 3,5 J |
| Pulsewidth | 0,1 - 2 ms (podiatry)<br>0,1 - 50 ms (other<br>indications) | 0,1-3 ms | 0,1-3 ms<br>6-20 ms (other<br>indications) |
| Repetition rate | 1 - 100 Hz | 5 - 100 Hz | 5 - 100 Hz |
| Output power | up to 30 W | up to 100 W | up to 100 W |
| Beam delivery | Fiber | Fiber | Fiber |
| Spot sizes | 1 -1.5 mm (podiatry)<br>0,2 - 10 mm (other<br>indications) | 1 -1.5 mm (podiatry)<br>Other spot sizes are not<br>published | 1-1.5 mm (podiatry)<br>up to 13 mm (other<br>indications) |
| User interface | Push-button control panel | LCD color touchscreen or<br>push-button control panel | LCD color touchscreen |
| Laser activation | Footswitch | Footswitch | Footswitch |
| Input power | 230V, 17A 50/60 Hz | 200-240V, 50/60 Hz | NA |
| Dimensions | 60x33x82 cm | 33x36x81 cm | NA |
| Weight | 78,5 kg (maximum power<br>configuration | 17,2 kg | NA |
Table 1: Comparison table between Fotona XP Nd:YAG Laser System Family and predicate device.s
## STATEMENT OF SUBSTANTIAL EQUIVALENCE
Fotona XP Nd:Y AG Laser System Family is substantially equivalent to the predicate devices Cutera GenesisPlus (K103626) and PinPointe FootLaser (K093547) in terms of indications for use and technology based on technical and functional characteristics.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, with stylized lines representing its wings and body.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 13, 2013
Fotona D.D. % Mr. Stojan Trošt QA/RA Manager Stegne 7 SI-1000 Ljubljana, Slovenia
Re: K113702
Trade/Device Name: Fotona XP Nd:YAG Laser System Family Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: PDZ, GEX Dated: February 27, 2012 Received: February 29, 2012
Dear Mr. Trošt:
This letter corrects our substantially equivalent letter of March 28, 2012.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Stojan Trošt
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
FOR
Peter D. Rumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K!!3702
## 4. Indications for Use Statement
510(k) Number (if known):
#### Device Name: Fotona XP Nd: YAG Laser System Family
Fotona XP Nd:YAG Laser System Family is indicated for podiatry (ablation, vaporization, incision, excision, and coagulation of soft tissue) including:
- Matrixectomy -
- Radical nail excision -
- Periungual and subungual warts ・
- Plantar warts -
- Neuromas -
- Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes -Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.)
Prescription Use: __ X (21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use: (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R.P. Dale for mxm
Division Sign-Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K113702
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.