← Product Code [NTN](/productcode/NTN) · K113679

# LUCEA LED SURGICAL LIGHT SYSTEM (K113679)

_Maquet S.A.S. · NTN · Jan 12, 2012 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K113679

## Device Facts

- **Applicant:** Maquet S.A.S.
- **Product Code:** [NTN](/productcode/NTN.md)
- **Decision Date:** Jan 12, 2012
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

MAQUET LUCEA LED® Surgical Light Systems are intended to be used to provide visible illumination of the surgical area or the patient during surgical operations, diagnostics and treatment.

## Device Story

LUCEA LED® Surgical Light System provides high-quality illumination for surgical suites, examining rooms, and doctor's surgeries. Device consists of LED lightheads (LUCEA 50 and 100 models) mounted via ceiling, wall, or mobile stands. Features include mechanical focusing via central handle, optional integrated zoom camera for image capture/sharing, and optional infrared remote control for intensity and camera adjustments. Operated by surgical staff to illuminate operating fields. System supports various configurations, including dual-head setups and triple-axis ceiling mounts for integration with other medical devices. Benefits include improved visualization of surgical area/patient during procedures, diagnostics, and treatments.

## Clinical Evidence

No clinical data is required for this device classification submission; bench testing only.

## Technological Characteristics

LED-based surgical illumination system. Configurations: ceiling, wall, or mobile mounted. Features mechanical focusing, optional zoom camera, and infrared remote control. Conforms to UL 60601-1, IEC 60601-2-41, IEC 60601-1, IEC 60601-1-2, and FCC Part 15 standards. Connectivity includes video pre-wiring for camera integration.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- MAQUET POWERLED™ Surgical Light System ([K070442](/device/K070442.md))
- MAQUET BLUE 30 & 80 Surgical Light System ([K954169](/device/K954169.md))
- MAQUET BLUE 80 Hospital Light System ([K970886](/device/K970886.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K 11 3679

JAN 1 2 2012

## 510(k) Summary As required by section 807.92

.)

## LUCEA LED® Surgical Light System

| Submitter's Name & Address      | MAQUET S.A.S.<br>Parc de Limère<br>Avenue de la Pomme de Pin<br>CS 10008 Ardon<br>45074 Orléans - Cedex 2<br>France                                                                                 |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter of this submission    | Ms. Marie-Françoise Cabel<br>Director, Quality and Regulatory Affairs<br>Maquet S.A.S<br>Phone: +33 238 25 88 72<br>Email: marie-francoise.cabel@maquet-sa.fr                                       |
| Applicant Correspondent         | Ms. Whitney Törning<br>Director, Regulatory Affairs<br>Maquet Inc.<br>45 Barbour Pond Drive<br>Wayne, NJ 07470<br>Phone: 973-709-7994<br>Fax: 973-973-807-9210<br>Email: whitney.torning@maquet.com |
| Date prepared                   | December, 9th 2011                                                                                                                                                                                  |
| Proprietary Name                | LUCEA LED® Surgical Light System                                                                                                                                                                    |
| Common Name                     | Surgical Light                                                                                                                                                                                      |
| Device product codes            | FTD (light, surgical) for ceiling and wall-mounted models<br>FSS (light, surgical, floor standing) for floor standing models                                                                        |
| Device classification           | Class II, acording to regulation number 21 CFR 878.4580                                                                                                                                             |
| Predicate Device identification |                                                                                                                                                                                                     |

# · MAQUET POWERLED™ Surgical Light System – 510(k) No. K070442

- · MAQUET BLUE 30 & 80 Surgical Light System (ceiling mounted) 510(k) No. K954169
- · MAQUET BLUE 80 Hospital Light System (floor standing) 510(k) No. K970886

{1}------------------------------------------------

MAQUET S.A.S. FDA 510(K) Summary
Device: LUCEA LED® Surgical Light System

113679

#### Device description:

MAQUET LUCEA LED® Surgical Light Systems have been developed in order to provide MAQUET EUGEAEED Surgied Lightnology. An innovative design combined with a functional shape offers an efficient product to the surgical staff.

functional shape oners an cindical provide hights provide high quality illumination Designed for minor ourgory, without any compromises on the major enhancements offered by MAQUET surgical lights.

The LUCEA LED® Surgical Lights are well-suited for installation in surgical suites, examining rooms, doctor's surgeries and external consultations.

The LUCEA LED® product family is available in the following versions:

| Device name              | Device Description                                                                                                                                                                                     |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| WALL MOUNTED VERSIONS    |                                                                                                                                                                                                        |
| LCA 50 WALL              | LUCEA LED 50 SURGICAL LIGHT, WALL MOUNTED VERSION WHICH INCLUDES A LUCEA 50<br>LIGHTHEAD, A WALL ERGODISC SUSPENSION AND A POWER SUPPLY KIT                                                            |
| LCA 100 WALL             | LUCEA LED 100 SURGICAL LIGHT, WALL MOUNTED VERSION WHICH INCLUDES A LUCEA 100<br>LIGHTHEAD, A WALL ERGODISC SUSPENSION AND A POWER SUPPLY KIT                                                          |
| CEILING MOUNTED VERSIONS |                                                                                                                                                                                                        |
| LCA DUO 50               | LUCEA LED DUO 50 SURGICAL LIGHT, CEILING MOUNTED, WHICH INCLUDES TWO LUCEA 50<br>LIGHTHEADS - UNLIMITED ROTATION ON CENTRAL AXIS, ERGODISC SUSPENSION                                                  |
| LCA 50                   | LUCEA LED 50 SURGICAL LIGHT, CEILING MOUNTED, WHICH INCLUDES A LUCEA 50 LIGHTHEAD -<br>UNLIMITED ROTATION ON CENTRAL AXIS, ERGODISC SUSPENSION                                                         |
| LCA 100 V                | LUCEA LED 100 SURGICAL LIGHT, CEILING MOUNTED, WHICH INCLUDES A LUCEA 100<br>LIGHTHEAD - UNLIMITED ROTATION ON CENTRAL AXIS, ERGODISC SUSPENSION, VIDEO<br>PREWIRING                                   |
| LCA DUO 100 V            | LUCEA LED DUO 100 SURGICAL LIGHT, CEILING MOUNTED, WHICH INCLUDES TWO LUCEA 100<br>LIGHTHEADS - UNLIMITED ROTATION ON CENTRAL AXIS, ERGODISC SUSPENSION, VIDEO<br>PREWIRING                            |
| LCA DUO 100+50 V         | LUCEA LED 100+50 V SURGICAL LIGHT, CEILING MOUNTED, WHICH INCLUDES A LUCEA 100<br>LIGHTHEAD WITH VIDEO PREWIRING AND A SATELLITE LUCEA 50 - UNLIMITED ROTATION ON<br>CENTRAL AXIS, ERGODISC SUSPENSION |
| MOBILE VERSIONS          |                                                                                                                                                                                                        |
| LCA 50 MOBILE            | LUCEA LED 50 SURGICAL LIGHT, MOBILE VERSION WHICH INCLUDES A LUCEA 50 LIGHTHEAD, A<br>MOBILE STAND INCLUDING THE POWER SUPPLY                                                                          |
| LCA 50 MOBILE B          | LUCEA LED 50 SURGICAL LIGHT, MOBILE VERSION WHICH INCLUDES A LUCEA 50 LIGHTHEAD, A<br>MOBILE STAND INCLUDING THE POWER SUPPLY AND A BATTERY PACK                                                       |
| LCA 100 MOBILE           | LUCEA LED 100 SURGICAL LIGHT, MOBILE VERSION WHICH INCLUDES A LUCEA 100 LIGHTHEAD,<br>A MOBILE STAND INCLUDING THE POWER SUPPLY                                                                        |
| LCA 100 MOBILE B         | LUCEA LED 100 SURGICAL LIGHT, MOBILE VERSION WHICH INCLUDES A LUCEA 100 LIGHTHEAD,<br>A MOBILE STAND INCLUDING THE POWER SUPPLY AND A BATTERY PACK                                                     |

{2}------------------------------------------------

MAQUET S.A.S. FDA 510(k) Summary FDA 510(K) Surfimaly
Device: LUCEA LED® Surgical Light System

K 113679

#### Intended Use:

MAQUET LUCEA LED® Surgical Light Systems are intended to be used to provide visible illumination of the surgical area or the patient during surgical operations, diagnostics and treatment.

### Nonclinical Comparisons to Predicate Device

The LUCEA LED® Surgical Light (subject device) is similar to the predicate devices with the following modifications:

- Modified light head design, updating its appearance: One primary lighthead (type . LUCEA 100) and one secondary lighthead (type LUCEA 50).
- Added a mechanical focusing function in LUCEA 100 lightheads: Rotating the . central handle makes it possible to focus the light depending on the distance from the operating field
- Added an optional zoom camera accessory integrated in LUCEA 100 lightheads to . / Rodod an optional be images, which may then be shared, saved or broadcast.
- Added an optional infrared remote control accessory to control intensity of LUCEA . 50 and LUCEA 100 lightheads, as well as an optional zoom camera.
- Added an optional triple-axis ceiling down tube which enable combinations of . LUCEA 50-100 surgical light systems with 2 other medical devices.

{3}------------------------------------------------

MAQUET S.A.S. FDA 510(k) Summary Device: LUCEA LED® Surgical Light System

K113679

#### Test Data:

Test data support conformance to:

- · UL 60601-1, 1st Edition, 2006-04-26 (Medical Electrical Equipment, Part 1: General Requirements for Safety, includes National Differences for USA)
- · IEC 60601-2-41:2000, Medical electrical equipment Part 2-41: Particular IEC 00001-2-41:2000, Modioa. Shownaires and luminaires for diagnostics
- IEC 60601-1:1988 + A1:1991 + A2:1995, Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- · IEC 60601-1-2:2007: General requirements for basic safety and essential ILC - 00001 - 2.2001. Standard: Electromagnetic compatibility - Requirements and tests
- · FCC Part 15 (10) Code of Federal Regulations, Title 47 Telecommunication, 1 - Federal Communications Commission, Part 15 - Radio frequency Ghapter T = T ederar Oommanisations, limits and methods of measurement of radio disturbance characteristics of information technology equipment

#### Clinical Data:

No clinical data is required for this device classification submission.

#### Conclusion:

The modifications incorporated into the MAQUET LUCEA LED® Surgical Light System The mountcations incorporated into the the new MAQUET POWERLED™, BLUE 30 & 80, designs use those desired doolgh toatures . Based upon the information provided herein
and BLUE 80 Hospital Surgical Light Systems. Based upon the information in this bestems and BLOE ou Hospital Surgical Eight Oystems: Babea aportal Light Systems
this 510(k) Premarket Notification, we conclude that LUCEA LED® Surgical Light Systems this 310(K) Fremanel Notification, wo conside that as a is safe and effective when used as intended.

{4}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The bird is oriented towards the right. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

MAQUET S.A.S. % Ms. Whitney Torning Director, Regulatory Affairs 45 Barbour Pond Drive Wayne. New Jersey 07470

JAN 1 2 2012

Re: K113679

Trade/Device Name: MAQUET LUCEA LED® Surgical Light System Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulatory Class: Class II Product Code: NTN, FTD, FSS Dated: January 6, 2012 Received: January 9, 2012

Dear Ms. Torning:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{5}------------------------------------------------

Page 2 - Ms. Whitney Torning

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson

Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

## Indications for Use

510(k) Number (if known):_K // 36 79

## Device Name: MAQUET LUCEA LED® Surgical Light System

Indications for Use:

MAQUET LUCEA LED® Surgical Light Systems are intended to be used to provide visible illumination of the surgical area or the patient during surgical operations, diagnostics and treatment.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil R. Oyler Sor nxm

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K113679

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K113679](https://fda-staging.innolitics.com/device/K113679)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).