The Dental Handpiece Cleaner and Lubricant is intended to be used during routine maintenance in order to lubricate and clean air-powered Dental Handpieces (including low speed and high speed) and Dental air motors after each patient use and prior to sterilization. The Dental Handpiece Lubricant is intended to be used during routine maintenance in order to lubricate air-powered Dental Handpieces (including low speed and high speed) and Dental air motors after each patient use and prior to sterilization.
Device Story
Aerosol and liquid maintenance products for air-powered dental handpieces; used in dental clinics by staff. Products contain mineral oil and isopropyl alcohol (aerosol versions) or refined mineral oil (lubricant version). Applied by spraying into air drive tube or dropping into tube; air pressure delivers lubricant to internal bearings, gears, and auto chucks. Removes contamination and reduces friction; ensures proper handpiece function; supports maintenance protocols prior to sterilization. Benefits include extended device life and consistent performance.
Clinical Evidence
No clinical data; substantial equivalence supported by bench testing comparing subject devices to predicate. Bench tests evaluated bur extraction force, concentricity, noise level, and torque, demonstrating performance equivalent to the predicate.
Technological Characteristics
Aerosol and liquid formulations. Aerosol contains mineral oil, isopropyl alcohol, and propellants (propane/n-butane). Lubricant contains 100% refined mineral oil. Insoluble in water. Applied via air drive tube. Class I device, 21 CFR 872.4200.
Indications for Use
Indicated for routine maintenance, cleaning, and lubrication of air-powered dental handpieces (low and high speed) and dental air motors after patient use and prior to sterilization.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
Predicate Devices
PANA SPRAY lubricant oil by NSK America Corp. (K052700)
Submission Summary (Full Text)
{0}------------------------------------------------
K113674
Image /page/0/Picture/2 description: The image shows the words "Handpiece Headquarters" in a bold, stylized font. Above the words is a line graph. The text is the main focus of the image.
A Henry Schein® COMPANY
### 510(K) SUMMARY:
- a- Submitted by: HANDPIECE HEADQUARTERS 620 S. Placentia Ave., Placentia CA 92870 Tel. 714-579-0175 Fax. 714-579-0186
- b- Contact person: Tina Steffanie-Oak, Tel. 717-335-7230, ext. 4150 Fax. 717-335-7240 Email: tina.steffanie-oak@henryschein.com
### c- Date summary prepared: 04/2/2012
### d- Device Name:
| Common Name: | Dental Handpiece and Accessories, Dental Handpiece Cleaner and<br>Lubricant. |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name: | Spray & Clean Handpiece Cleaner & Lubricant<br>Maxima Handpiece Cleaner & Lubricant<br>EZCARE Handpiece Cleaner & Lubricant<br>(3 trade names carry the same chemical formula ) |
| Common Name: | Dental Handpiece and Accessories, Dental Handpiece Lubricant. |
| Trade Name: | Maxima Handpiece Lubricant |
| Classification Name: | Class I Device, 21 CFR 872.4200, Handpiece, Air-power<br>Dental, EFB |
e- Substantial Equivalency is claimed against the following devices: PANA SPRAY lubricant oil by NSK America Corp. - K052700
f- Description of the device:
Dental Handpiece Cleaner and Lubricant (Spray & Clean Handpiece Cleaner & Lubricant, Maxima Handpiece Cleaner & Lubricant, and EZcare Handpiece Cleaner & Lubricant), This is an aerosol product that is used during routine maintenance. It contains two major components, which are Mineral oil and Isopropyl Alcohol. The product lubricates and cleans dental handpieces and air motors when sprayed directly into the air drive tube. It lubricates and cleans out contamination from all internal parts, such as the bearings, gears and auto chucks.
{1}------------------------------------------------
# Dental Handpiece Lubricant: Maxima Handpiece Lubricant
Maxima Handpiece Lubricant contains mineral oil that drops directly into the air drive tube of the dental handpiece. When the user operates the handpiece, the pressure of air in the air drive tube will deliver the oil into the handpiece to lubricate the bearings, gears, and auto chucks.
# g- Statement of Intended Use:
The Dental Handpiece Cleaner and Lubricant is intended to be used during routine maintenance in order to lubricate and clean air-powered Dental Handpieces (including low speed and high speed) and Dental air motors after each patient use and prior to sterilization.
The Dental Handpiece Lubricant is intended to be used during routine maintenance in order to lubricate air-powered Dental Handpieces (including low speed and high speed) and Dental air motors after each patient use and prior to sterilization.
## h - Device and Predicate Comparison Table:
| Descriptive<br>Information | Subject Device: | Predicate: |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| | Spray & Clean Handpiece Cleaner &<br>Lubricant, Maxima Handpiece<br>Cleaner & Lubricant, and EZCARE<br>Handpiece Cleaner & Lubricant<br><br>Maxima Handpiece Lubricant | NSK PANA SPRAY<br>LUBRICANT OIL SUPERIOR<br>CLEANING &LUBRICANT<br><br>NSK PANA SPRAY<br>LUBRICANT OIL SUPERIOR<br>CLEANING &LUBRICANT |
| Indication of use | The Dental Handpiece Cleaner and<br>Lubricant is intended to be used<br>during routine maintenance in order to<br>lubricate and clean air-powered<br>Dental Handpieces (including low<br>speed and high speed) and Dental air<br>motors after each patient use and<br>prior to sterilization.<br><br>The Dental Handpiece Lubricant is<br>intended to be used during routine<br>maintenance in order to lubricate air-<br>powered Dental Handpieces<br>(including low speed and high speed)<br>and Dental air motors after each<br>patient use and prior to sterilization. | This new device is a lubricant<br>and is intended for use to clean<br>and lubricate the dental and<br>medical surgical handpieces. |
| Components<br>name/<br>Weight<br>percentage<br>(% w/w) | Spray & Clean Handpiece Cleaner &<br>Lubricant, Maxima Handpiece<br>Cleaner & Lubricant, and EZCARE<br>Handpiece Cleaner & Lubricant<br>(3 trade names carry the same<br>chemical formula )<br>Propellant: (84% w/w)<br>Propane 7.5%<br>n-Butane 17.5%<br>Isopropanol 59% | PANA SPRAY<br><br>Propellant: (63% w/w)<br>Propane<br>Isobutane<br>n-Butane |
{2}------------------------------------------------
| | Lubricant: (16% w/w)<br>Mineral oil 16%<br>Hydrocarbon Range C16-C28 0% | Lubricant: (16% w/w)<br>Petroleum Oil<br>Hydrocarbon Range C16-C28 |
|-----------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| | Maxima Handpiece Lubricant<br>Lubricant: (100% w/w)<br>Severely refined Mineral oil 100%<br>Hydrocarbon Range C16-C28 0% | Lubricant: (16% w/w)<br>Petroleum Oil<br>Hydrocarbon Range C16-C28 |
| Fragrance | N/A | N/A |
# Physical Property comparison table:
| Physical property | Predicate Device:<br>Pana Spray Lubricant Oil | Subject Device:<br>Spray & Clean Handpiece<br>Cleaner and Lubricant | Subject Device:<br>Maxima Handpiece<br>Lubricant |
|---------------------|-----------------------------------------------|---------------------------------------------------------------------|--------------------------------------------------|
| Physical State | Aerosol Product | Aerosol Product | Transparent Water-white Liquid |
| Color | clear | Clear | Clear |
| Solubility in Water | Insoluble | Insoluble | Insoluble |
| Specific Gravity | $0.63 \ (H_2O=1)$ | .70-.90 g/cc | $0.835 @ 15.6 \degree C \ (H_2O=1)$ |
| Boiling Point | -44 deg F-31 deg F | Propellant: 7.29 °F<br>Concentrate: 180 °F | 310 °C (590 °F) |
| Viscosity | n/a | n/a | n/a |
# i - Description of Non-Clinical Test Data used to support SE Decision:
The table below provides a general summary of the performance testing which was conducted on the subject devices and the predicate device. The testing results indicate that the subject devices meet their design performance requirements.
| Comparison | Test | Method | Conclusion |
|----------------------------------------------------------------------------|-------------------------------------|-----------------------------------------------------------------------------|-----------------------------------------------|
| Spray & Clean<br>Handpiece<br>Cleaner &<br>Lubricant vs.<br>NSK Pana Spray | Handpiece<br>Performance<br>Testing | Cycle<br>Bur extraction force,<br>concentricity, noise<br>level, and torque | Met handpiece<br>performance<br>specification |
| Maxima<br>Handpiece<br>Lubricant vs.<br>NSK Pana Spray | Handpiece<br>Performance<br>Testing | Cycle<br>Bur extraction force,<br>concentricity, noise<br>level, and torque | Met handpiece<br>performance<br>specification |
# j - Description of Clinical Test Data used to support SE Decision:
Clinical test data was not required to support our substantial equivalence determination.
# k - Conclusions Drawn from Non-Clinical Data:
Based on the information provided in this submission Handpiece Headquarters believes that the Dental Handpiece Cleaner & Lubricant and Dental Handpiece Lubricant are intended for the same use and are substantially equivalent to the predicate devices identified.
{3}------------------------------------------------
l – Biocompatibility: Biocompatibility testing has not been conducted on the subject devices, because the chemical components have already been incorporated in legally marketed devices with similar conditions of use and have a demonstrated history of biocompatibility.
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Tina Steffanie-Oak Consultant Handpiece Headquarters Inc, - HPR Inc. 620 S. Placentia Avenue Placentia, California 92870
APR - 5 2012
Re: K113674
Trade/Device Name: Dental Handpiece Cleaner and Lubricant, and a Dental Handpiece Lubricant (sold under these brand names Spray & Clean Handpiece Cleaner & Lubricant, Maxima Handpiece Cleaner & Lubricant, EZcare Handpiece Cleaner & Lubricant, and Maxima Handpiece Lubricant)
Regulation Number: 21 CFR 872.4200
Regulation Name: Dental Handpiece and Accessories
Regulatory Class: I
Product Code: EFB
Dated: April 2, 2012
Received: April 3, 2012
Dear Ms. Steffanie-Oak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
## Page 2 - Ms. Steffanie-Oak
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Nh for
Anthony D. Watson, B.S., M.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{6}------------------------------------------------
Indications for Use
113672
510(k) Number (if known):
Device Name: Dental Handpiece Cleaner and Lubricant, and a Dental Handpiece Lubricant (sold under these brand names Spray & Clean Handpiece Cleaner & Lubricant, Maxima Handpiece Cleaner & Lubricant, EZcare Handpiece Cleaner & Lubricant, and Maxima Handpiece Lubricant)
Indications for Use:
The Dental Handpiece Cleaner and Lubricant is intended to be used during routine maintenance in order to lubricate and clean air-powered Dental Handpieces (including low speed and high speed) and Dental air motors after each patient use and prior to sterilization.
The Dental Handpiece Lubricant is intended to be used during routine maintenance in order to lubricate air-powered Dental Handpieces (including low speed and high speed) and Dental air motors after each patient use and prior to sterilization.
Prescription Use X (Per 21 CFR 801 Subpart D) Subpart C)
AND/OR
Over-The-Counter Use
(Per 21 CFR 801
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Othuraltneen for MSR
Page 1 of 1
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.