ELEMENT BLOOD GLUCOSE MONITORING SYSTEM
Device Facts
| Record ID | K113670 |
|---|---|
| Device Name | ELEMENT BLOOD GLUCOSE MONITORING SYSTEM |
| Applicant | Infopia Co, Ltd. |
| Product Code | CGA · Clinical Chemistry |
| Decision Date | May 8, 2012 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 862.1345 |
| Device Class | Class 2 |
Indications for Use
The Element™ Blood Glucose Monitoring System is for the quantitative measurement of glucose in capillary whole blood taken from the fingertip, ventral palm, dorsal hand, upper arm, forearm, calf and/or thigh. Element The Blood Glucose Testing System is for in vitro diagnostic use and is not to be used for the diagnosis of or screening for diabetes or for neonatal use. Element ™ Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared. The Element ™ Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid in the management of diabetes. Alternative site Testing (ventral palm, dorsal hand, upper arm, calf and/or thigh) should be done during steady- state times when glucose is not changing rapidly. The Element ™ Test Strips are for use with the Element™ Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip, ventral palm, dorsal hand, upper arm, forearm, calf and/or thigh, The Element TM control solutions are for use with the Element ™ Blood Glucose Monitoring system to check that the meter and the test strips are working together properly and the test is performing correctly.
Device Story
System measures glucose concentration in capillary whole blood via electrical current generated in test strips; meter processes current to provide quantitative glucose reading. Used at home by patients with diabetes for self-monitoring; aids in diabetes management. Includes meter, test strips, and control solutions for system verification. Patient applies blood sample to strip; meter displays result. Supports alternative site testing during steady-state glucose levels. Device modifications from predicate include battery type, display icon location, LCD material, meter dimensions, and removal of voice function.
Clinical Evidence
Clinical performance evaluation conducted to validate consumer use and professional accuracy. Results demonstrated substantial equivalence to the predicate device. Non-clinical testing included verification, validation, and reliability testing, all meeting pre-determined pass/fail criteria. Disinfection study validated CaviWipes (EPA 46781-8) for 1,095 cycles (simulating 3 years) with no performance degradation.
Technological Characteristics
In vitro diagnostic glucose test system. Uses electrochemical sensing principle. Components: meter, test strips, control solutions. Modifications include updated battery, LCD material, and physical dimensions. Disinfection validated for 1,095 cycles. No specific software algorithm class or connectivity details provided.
Indications for Use
Indicated for the quantitative measurement of glucose in fresh capillary whole blood from the fingertip, palm, or forearm by people with diabetes at home or by healthcare professionals in clinical settings as an aid to monitor the effectiveness of diabetes control programs.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
- Element Plus Blood Glucose Monitoring System (k103021)