K113597 · Minilap Technologies · GEI · Jun 3, 2012 · General, Plastic Surgery
Device Facts
Record ID
K113597
Device Name
MINI-LAP BIPOLAR ELECTROCAUTERY DEVICES
Applicant
Minilap Technologies
Product Code
GEI · General, Plastic Surgery
Decision Date
Jun 3, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MiniLap Technologies family of BiPolar instruments with bipolar cautery have applications in a variety of general, thoracic, gynecologic (except for use in female sterilization), urologic, laparoscopic and endoscopic procedures for manipulation and coagulation of tissue.
Device Story
Disposable bipolar electrosurgical grasper; used in general, thoracic, gynecologic, urologic, laparoscopic, and endoscopic procedures. Device features 2.4mm insulated needle/cannula shaft with retractable grasper; percutaneously inserted to surgical site without traditional conduit. Connects to standard bipolar cables and generators for tissue cauterization. Operated by surgeons in clinical settings. Provides precise tissue manipulation and coagulation; benefits include reduced invasiveness via percutaneous access.
Clinical Evidence
Bench testing only. Mechanical and electrical performance evaluations conducted. Biocompatibility of materials confirmed via history of use in medical industry.
Indicated for patients undergoing general, thoracic, gynecologic (excluding female sterilization), urologic, laparoscopic, and endoscopic procedures requiring tissue manipulation and coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Wolf Medical Instruments Bipolar Forceps (K023813)
{0}------------------------------------------------
# 510(k) Summary
# JUN - 3 2012
### Date Prepared [21 CFR 807.92(a)(1)] November 1, 2011, Updated May 31, 2012
### Submitter's Information [21 CFR 807.92(a)(1)]
Regulatory Contact Allan Alward 88 Ashford Avenue Dobbs Ferry, NY 10522
Sponsor/Manufacturer Mini Lap Technologies, Inc. 88 Ashford Avenue Dobbs Ferry, NY 10522 Contact: Dr. S. Ravikumar Tel: 914 591 8400
FDA Establishment Registration Number is 3007123990
### Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)|
Trade Names
- MINI LAP Bipolar Electrocautery Devices ◆
Device Common, Usual or Classification Names
Laparoscopic Instruments, Graspers, Cannula, Trocar, Manual Surgical Instruments, Bipolar Instruments
#### Classification Panel
Classification of this device would fall under the responsibility of the General & Plastic Surgery panel.
#### Class
Class 2 device under the following product codes/regulations:
- KOG, 21 CFR 876.1500, Endoscope Accessories
- 트 FBQ, 21 CFR 878.5090, Trocar
- . GEI, 21 CFR 878.4400, Electrosurgical cutting and coagulation accessories
{1}------------------------------------------------
#### Predicate Device [21 CFR 807.92(a)(3)]
The following devices have been identified as predicate devices:
- . Wolf Medical Instruments Bipolar Forceps - K023813
- MINI LAP Instruments K070686 .
### Description of the Device [21 CFR 807.92(a)(4)]
The MiniLap Bipolar Electrocautery devices are a family of disposable bipolar electrosurgical graspers that directly penetrate soft tissue to access certain areas of the human anatomy without the need for a traditional insertion conduit. These disposable instruments are used for electrosurgical cauterization during general laparoscopic procedures.
Fabricated from a stainless steel alloy, MiniLap Bipolar instruments consist of an integrated 2.4mm (+/- .1mm) insulated needle/cannula shaft that houses a retractable grasper instrument. The shaft can be introduced percutaneously to the surgical site, after which the working portion of the instrument can be deployed to approximate and cauterize soft tissue. Each BiPolar instrument has a male connection provided as an integral part of the instrument that may be utilized for the bipolar connection when attached to standard two plug bipolar cables and their generators. The bipolar connection is compatible with manufacturers of standard dual female bipolar plug cables.
The instruments come in multiple lengths from 150mm to 300mm.
#### Intended Use [21 CFR 807.92(a)(5)]
The MiniLap Technologies family of BiPolar instruments with bipolar cautery have applications in a variety of general, thoracic, gynecologic (except for use in female sterilization), urologic, laparoscopic and endoscopic procedures for manipulation and coagulation of tissue.
#### Technological Characteristics [21 CFR 807.92(a)(6)]
We believe the MINI LAP Bipolar Electrocautery Devices are substantially equivalent to the predicate devices. see chart
{2}------------------------------------------------
| Characteristic | Mini Lap Technologies | Richard Wolf Medical<br>Instruments Corp. | Mini Lap Technologies |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | BiPolar Electrocautery Devices | | MINI LAP Instruments |
| | | Bipolar Forceps | |
| | | | |
| 510(k) | K113597 | K023813 | K070686 |
| Indications for Use | The MiniLap Technologies family<br>of BiPolar instruments with<br>bipolar cautery have applications<br>in a variety of general, thoracic,<br>gynecologic (except for use in<br>female sterilization), urologic,<br>laparoscopic and endoscopic<br>procedures for manipulation and<br>coagulation of tissue. | For use in open and laparoscopic<br>surgery where grasping,<br>coagulating, and transecting<br>tissue is indicated. | Penetrate soft tissue to access<br>certain areas of the human<br>anatomy. The devices are used to<br>grasp, hold and manipulate other<br>soft internal tissues as well as<br>items such as hernia mesh. |
| Needle diameter | 2.4mm ± .1mm | 3.5 mm, 5 mm | 2.4 mm |
| Device length | 150-300 mm | 320 mm, 450 mm | 150-300 mm |
| Material Composition | Medical Grade Stainless Steel<br>17-7 SS, 300 SS | Medical Grade Stainless Steel | Medical Grade Stainless Steel<br>470 SS, 300 SS |
| Sterilization | Sterile, single use | Non sterile, reusable | Sterile, single use |
| Power Source | Bipolar Electrosurgical Generator | Bipolar Electrosurgical Generator | N/A |
| Power Ranges | 70 watts maximum, 20-35 watts<br>normal operating range | 50 watts maximum, 20-35 watts<br>normal operating range | |
| Safety | Safety Interlock | Safety Interlock is not applicable | Safety Interlock |
| | Insulated shaft and jaws | Insulated jaws | |
| | Compliant to applicable IEC<br>electrical safety standards | Compliant to applicable IEC<br>electrical safety standards | |
| Rotation | 175 degree shaft rotation, each<br>direction | | 360 degree shaft rotation |
# Technological Characteristics Comparison Chart
### Performance Data [21 CFR 807.92(b)(1)]
The subject device has been subjected to and passed a variety of mechanical and electrical tests and evaluations. Additionally, the device is composed of biocompatible materials with a history of usage in the medical device industry.
# Conclusion [21 CFR 807.92(b)(3)]
We believe the changes are minimal and conclude that the subject device is as safe and effective as the predicate device.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wing-like shapes, symbolizing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MiniLap Technologies % Mr. Allan Alward Vice President Research and Development 88 Ashford Ave. Dobbs Ferry. NY 10522
JUN. - 3 2012
Re: K113597
Trade/Device Name: Mini-Lan Bipolar Electrocautery Devices Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: May 15, 2012 Received: May 16, 2012
Dear Mr. Alward:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
# Page 2 - Mr. Allan Alward
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical, Orthopedic & Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K113597
Device Name: Mini-Lap Bipolar Electrocautery Devices
Indications for Use:
The MiniLap Technologies family of BiPolar instruments with bipolar cautery have applications in a variety of general, thoracic, gynecologic (except for use in female sterilization), urologic, laparoscopic and endoscopic procedures for manipulation and coagulation of tissue.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sign-Off
(Division Sign-Off) (Division Sign-Off)
Division of Surgical, Orthopedic.
Division of Surgical, Orthopedic. Division of Surgices and Restorative Devices
K113597
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.