K113472 · CAO Group, Inc. · GEX · Aug 30, 2012 · General, Plastic Surgery
Device Facts
Record ID
K113472
Device Name
PRECISE SHP DIODE LASER
Applicant
CAO Group, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 30, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Precise SHP Diode Laser is indicated for the removal of lesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation on oral soft tissue for the specific dental and oral surgery procedures of gingivectomy, freeectomy, operculectomy, contouring, biopsy, troughing, ulcer care, abscess care, sulcular debridement, soft tissue curettage, and removal of inflamed edematous tissue.
Device Story
Precise SHP Diode Laser is a powered surgical instrument for dental soft tissue procedures. Device uses solid-state diodes to generate 810nm (± 20nm) laser energy; maximum 3W output. Energy delivered to surgical site via optical fiber system housed in metal handpiece with disposable single-use tips. Features include 630nm aiming beam; adjustable power settings; continuous or pulsed delivery modes. Operated by dentist in clinical setting via foot-actuated switch. Laser energy performs excision, incision, ablation, and hemostasis. Benefits include precise tissue removal and controlled coagulation during oral surgery.
Clinical Evidence
Bench testing only. Evaluation confirmed device met design criteria for essential performance, output requirements per 21 CFR 1010/1040, and correct operation of all safety features.
Technological Characteristics
Solid-state diode laser; 810nm ± 20nm wavelength; 3W max output. 630nm aiming beam. Fiber optic delivery system with disposable single-use tips. Foot-switch activation. Complies with 21 CFR 1040.10/1040.11, IEC 60601-1 (3rd Ed), IEC 60601-1-2, IEC 60601-1-4, IEC 60601-1-6, IEC 60601-2-22, and IEC 60825-1.
Indications for Use
Indicated for patients requiring oral soft tissue surgery, including gingivectomy, frenectomy, operculectomy, contouring, biopsy, troughing, ulcer/abscess care, sulcular debridement, soft tissue curettage, and removal of inflamed edematous tissue.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
K113472
# 510(k) Summary of Safety and Effectiveness
CAO Group, Inc. 4628 West Skyhawk Drive West Jordan, UT 84084 Tel: 801-256-9282 Fax: 801-256-9287
Prepared By: Robert K. Larsen, Preparation Date: November 15, 2011
#### Device Name:
| Trade Name: | Precise SHP Diode Laser |
|-------------------------|-----------------------------------|
| Common Name: | Soft Tissue Diode Laser |
| Product Classification: | Powered Laser Surgical Instrument |
## Legally Marketed Predicate Devices for Substantial Equivalence:
Odyssey Navigator Diode Laser, manufactured by Ivoclar Vivadent, Inc. (K062258)
DenLaser 800 Plus, manufactured by CAO Group, Inc. (K062619)
#### Rationale for Substantial Equivalence:
The aforementioned devices share similar indications for use with the present device for excision, incision, ablation, and photocoagulation on soft tissue for a variety of procedures in dentistry. The predicate devices and submitted device share similar design features including wavelength, operating controls, and laser delivery method. The devices share similar methods of control systems, safety features, and performance monitoring. The devices share similar performance specifications including power output and energy type.
#### Description of Submitted Device:
The Precise SHP Diode Laser is a device for delivering laser energy to human soft tissue for a variety of surgical procedures and treatments. This energy is generated by solid-state diodes, which provide a consistent and reliable generation of laser energy at 810 ± 20nm for a maximum of 3 watts of energy output. The laser energy is delivered to surgical site by means of an optical fiber system, which allows for the safe transmission of laser energy to the site without creating undue risk to the patient or operatory staff by errant or collateral laser emissions. The device features some user definable settings, including a switchable 630nm aiming beam,
{1}------------------------------------------------
adjustable power output for both the working beam and aiming beam, and continuous delivery or pulse delivery options.
The working end of the delivery fiber is contained within a metal handpiece with a disposable single-use tip. This handpiece system is provided with the device. The activation of the working beam diodes is completed by use of a foot-actuated switch.
#### Intended Uses of the Submitted Device:
The Precise SHP Diode Laser is indicated for the removal of lesions, excision, vaporization, ablation, hemostasis, and photocoagulation on oral soft tissue for the specific dental and oral surgery procedures of gingivectomy, operculectomy, contouring, biopsy, troughing, ulcer care, abscess care, sulcular debridement, soft tissue curettage, and removal of inflamed edematous tissue.
#### Technological Characteristics and Substantial Equivalence:
The Odyssey Navigator Diode Lasers uses solid state diodes to generate laser energy in the 810nm range. This system uses a fiber delivery system to transmit laser energy to the surgical site. The system also features a 630nm aiming beam and features controls that allow for adjusting the output of the working beam, and switching between a continuous or pulsed-mode laser emissions. The maximum output of the working beam is 3 watts.
The DenLaser 800 Plus uses solid state diodes to generate laser energy in the 810nm range. This system uses a fiber delivery system to transmit laser energy to the surgical site. The system also features a 630nm aiming beam and features controls that allow for adjusting the output of the working beam, and switching between a continuous or pulsed-mode laser emissions. The maximum output of the working beam is 5 watts.
#### Conformity to Standards:
The Precise SHP Diode Laser is designed to comply with the performance requirements . of 21 CFR 1040.10 and 1040.11, with permissible deviations relative to Laser Notice 50, dated June 24, 2007. The device also complies with the recognized standards of IEC 60601-2-22 Edition 3 and IEC 60825-1 Edition 2. The device also designed in compliance to the entirety of 1EC 60601-1: 3nd Edition, 1EC 60601-1-2, IEC 60601-1-4, and IEC 60601-1-6.
#### Performance Data
Bench testing on an evaluation sample of the current device revealed that the device met the design criteria for essential performance, and satisfied the performance requirements
{2}------------------------------------------------
indicated in 21 CFR 1010 and 21 CFR 1040. Device outputs were within performance requirements and all safety features and functions were operating correctly.
### Conclusion
The Precise SHP Diode Laser is substantially equivalent to the listed predicate devices without raising any new issues of safety or effectiveness. This device shares similar intended uses, operating principles, design features, and similar functional and performance abaracteristics. The device is designed to comply with relevant federal and international safety and performance standards.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
AUG 30 2012
CAO Group, Incorporated % Mr. Robert K. Larsen Regulatory Affairs Manager 4628 West Skyhawk Drive West Jordan, Utah 84084
Re: K113472
Trade/Device Name: Precise SHP Diode Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX
Dated: August 20, 2012 Received: August 22, 2012
Dear Mr. Larsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I Drines interes and regulations administered by other Federal agencies. You must
{4}------------------------------------------------
#### Page 2 - Mr. Robert K. Larsen
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
FOR
Mark N. Melkerson
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name:
Indications For Use:
The Precise SHP Diode Laser is indicated for the removal of lesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation on oral soft tissue for the specific dental and oral surgery procedures of gingivectomy, freeectomy, operculectomy, contouring, biopsy, troughing, ulcer care, abscess care, sulcular debridement, soft tissue curettage, and removal of inflamed edematous tissue.
Prescription Use X (Per 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Melh R. Sella Lis mkn
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K113472
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.