The ACE Axcel Clinical Chemistry System is an automated, discrete, bench-top, random access analyzer that is intended for in vitro diagnostic use in the quantitative determination of constituents in blood and other fluids. The ACE Alkaline Phosphatase Reagent is intended for the quantitative determination of alkaline phosphatase activity in serum using the ACE Axcel Clinical Chemistry System. Measurements of alkaline phosphatase are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. The ACE Amylase Reagent is intended for the quantitative determination of α-amylase activity in serum using the ACE Axcel Clinical Chemistry System. Amylase measurements are used primarily for the diagnosis and treatment of pancreatitis (inflammation of the pancreas). This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. The ACE LDH-L Reagent is intended for the quantitative determination of lactate dehydrogenase activity in serum using the ACE Axcel Clinical Chemistry System. Lactate dehydrogenase measurements are used in the diagnosis and treatment of liver diseases such as acute viral hepatitis, cirrhosis, and metastatic carcinoma of the liver, cardiac diseases such as myocardial infarction and tumors of the lung or kidneys. This test is intended for use in physician office laboratories or clinical laboratories. For in vitro diagnostic use only.
Device Story
Reagents for quantitative enzymatic photometric determination of ALP, Amylase, and LDH activity in serum samples; used on ACE Axcel Clinical Chemistry System. System measures rate of absorbance change at specific wavelengths (ALP: 408 nm; Amylase: 408 nm; LDH: 340 nm) to calculate enzyme activity. Used in clinical laboratories or physician office laboratories by trained operators. Output provides quantitative enzyme activity levels (U/L) to assist clinicians in diagnosing/monitoring liver, bone, pancreatic, and cardiac conditions. Benefits include rapid, automated diagnostic testing for patient management.
Clinical Evidence
Bench testing only. Precision studies conducted over 22 days and at three Physician Office Laboratory (POL) sites. Accuracy evaluated via correlation studies comparing ACE Axcel (y) to predicate ACE system (x) using 111-121 patient samples per analyte. Correlation coefficients (r) ranged from 0.9957 to 1.0000. Detection limits: 1.3 U/L (ALP), 8.5 U/L (Amylase), 8.3 U/L (LDH).
Technological Characteristics
Enzymatic photometric assays. Reagents contain substrates (p-nitrophenyl phosphate, 2-chloro-p-nitrophenyl-α-D-maltotrioside, L-lactate/NAD), buffers, and preservatives. Kinetic reaction measured at 37°C. Connectivity: Integrated with ACE Axcel Clinical Chemistry System. Software: Embedded firmware for automated spectrophotometric analysis.
Indications for Use
Indicated for quantitative determination of alkaline phosphatase, α-amylase, and lactate dehydrogenase activity in serum for diagnosis and treatment of liver, bone, parathyroid, intestinal, pancreatic, and cardiac diseases. Intended for use in clinical laboratories or physician office laboratories.
Regulatory Classification
Identification
An alkaline phosphatase or isoenzymes test system is a device intended to measure alkaline phosphatase or its isoenzymes (a group of enzymes with similar biological activity) in serum or plasma. Measurements of alkaline phosphatase or its isoenzymes are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.
Predicate Devices
Alfa Wassermann ACE Clinical Chemistry System and ACE Reagents (K931786)
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k113436
B. Purpose for Submission:
New device
C. Measurand:
Alkaline Phosphatase, Amylase, and Lactate Dehydrogenase
D. Type of Test:
Quantitative, enzymatic activity
E. Applicant:
Alfa Wassermann Diagnostic Technologies, LLC
F. Proprietary and Established Names:
ACE Alkaline Phosphatase Reagent
Amylase Reagent
ACE LDH-L Reagent
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| CJE | II | 862.1050, Alkaline phosphatase or isoenzymes test system | 75-Chemistry |
| CIJ | II | 862.1070, Amylase test system | 75-Chemistry |
| CFJ | II, exempt, meets limitations of exemption. 21 CFR 862.9 (c) (4) and (9) | 862.1440, Lactate dehydrogenase test system | 75-Chemistry |
{1}
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
The ACE Alkaline Phosphatase Reagent is intended for the quantitative determination of alkaline phosphatase activity in serum using the ACE Axcel Clinical Chemistry System. Measurements of alkaline phosphatase are used in the diagnosis and treatment of liver, bone, parathyroid and intestinal diseases. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
The ACE Amylase Reagent is intended for the quantitative determination α-amylase activity in serum using the ACE Axcel Clinical Chemistry System. Amylase measurements are used primarily for the diagnosis and treatment of pancreatitis (inflammation of the pancreas). This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
The ACE LDH-L Reagent is intended for the quantitative determination of lactate dehydrogenase activity in serum using the ACE Axcel Clinical Chemistry System. Lactate dehydrogenase measurements are used in the diagnosis and treatment of liver diseases such as acute viral hepatitis, cirrhosis, and metastatic carcinoma of the liver, cardiac diseases such as myocardial infarction and tumors of the lung and kidneys. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
3. Special conditions for use statement(s):
For in vitro diagnostic use only. For prescription and point-of-care use.
4. Special instrument requirements:
ACE Axcel Clinical Chemistry System
I. Device Description:
The ACE Alkaline Phosphatase Reagent and ACE LDH-L Reagent for the Axcel Clinical Chemistry System each come in a kit containing 6 liquid ready-to-use bottles; three bottles containing 30 mL of R1 (reagent 1), and three bottles containing 12 mL of R2 or (reagent 2). The ACE Amylase Reagent for the Axcel Clinical Chemistry System comes in a kit containing 3 liquid ready-to-use bottles containing 30 mL of R1 (reagent 1). Buffers contain preservatives (sodium azide) and stabilizers.
{2}
J. Substantial Equivalence Information:
1. Predicate device name(s):
ACE Clinical Chemistry System, ACE Alkaline Phosphatase Reagent
ACE Clinical Chemistry System, ACE Amylase Reagent
ACE Clinical Chemistry System, ACE LDH-L Reagent
2. Predicate 510(k) number(s):
k931786
3. Comparison to predicate
Comparison for Alkaline Phosphatase (ALP):
| | Candidate Device | Predicate Device |
| --- | --- | --- |
| 510(k) # | k113436 | k931786 |
| Similarities | | |
| Intended Use/Indications for Use | Same | ACE Alkaline Phosphatase Reagent is intended for the quantitative determination of alkaline phosphatase activity in serum. |
| Calibration | Same | Enzyme activity is directly determined by multiplying the change in absorbance per minute of the unknown samples by a constant factor based on the molar absorptivity of p-nitrophenoxide |
| Method Traceability | Same | Bowers, G.N. Jr. and McComb, R.B., Clin. Chem 12, 70 (1966); Tietz, N.W. et al., Clin. Chem. 29, 751 (1983). |
| Use of Controls | Same | Two levels of control per day |
| Basic Principle | Same | Enzymatic assay for alkaline phosphatase |
| Measurement Type | Same | Reaction of alkaline phosphatase with colorless substrate (p-nitrophenylphosphate) in alkaline solution results in formation of p-nitrophenol and inorganic phosphate. measured spectrophotometrically at 408/486 nm |
| Reactive Ingredients | Same | p-Nitrophenyl phosphate, Magnesium salt, AMP buffer (pH 10.45) |
| Non-reactive Ingredients | Same | Preservatives and activators |
| Dimensions | Same | Bottles with total volumes of 12 and 30 mL of reagent |
| Analysis | Same | 37°C |
{3}
| | Candidate Device | Predicate Device |
| --- | --- | --- |
| 510(k) # | k113436 | k931786 |
| Temperature | | |
| Reaction Type | Same | Kinetic |
| Sample Type | Same | Serum |
| Sample Volume | Same | 4 μL |
| Reaction Volume (total) | Same | 169 μL |
| Differences | | |
| Instrument Platforms | ACE Axcel Clinical Chemistry System | ACE and ACE Alera® Clinical Chemistry Systems |
| Detection Limit | 1.3 U/L | 2 U/L |
| Reportable Range | 9 to 1400 U/L | 2 to 1400 U/L |
| Endogenous Interferences | Bilirubin: For the low pool, no significant interference occurred below 27.95 mg/dL. Positive interference (19%) occurred at 55.9 mg/dL. For the high pool, no significant interference occurred.
Hemolysis: For the low pool, no significant interference occurred below 62.5 mg/dL. Negative interference (≥16%) occurred at ≥125 mg/dL. For the high pool, no significant interference occurred below 500 mg/dL. A MXINIT flag occurred at 1000 mg/dL for both pools.
Lipemia (Intralipid): For low and high pools, no significant interference occurred below 1000 mg/dL. A MXINIT flag occurred at 2000 mg/dL.
Ascorbic Acid: No significant interference. | Bilirubin: No significant interference below 12.5 mg/dL. Concentrations greater than 12.5 may cause interference. Samples with 25 mg/dL were found to positively interfere.
Hemolysis: No significant interference below 125 mg/dL. Concentrations greater than 125 may cause interference. Samples with 250 mg/dL were found to positively interfere.
Lipemia (Intralipid): No significant interference. |
| Precision (U/L) | Within run:
Sample A: Mean 50.0, SD 1.6, CV 3.2%
Sample B: Mean 690.0, SD 9.8, | Within run:
Sample A: Mean 42, SD 1.0, CV 2.4%
Sample B: Mean 161, SD 2.5, CV 1.6%
Sample C: Mean 332, SD 5.7, CV 1.7% |
{4}
Comparison for Amylase:
| 510(k) # | k113436 | k931786 |
| --- | --- | --- |
| | CV 1.4%
Sample C: Mean 1020.6, SD 13.2, CV 1.3%
Sample D: Mean 60.9, SD 1.6, CV 2.6%
**Total:**
Sample A: Mean 50.0, SD 2.2, CV 4.4%
Sample B: Mean 690.0, SD 19.9, CV 2.9%
Sample C: Mean 1020.6, SD 28.3, CV 2.8%
Sample D: Mean 60.9, SD 2.9, CV 4.7% | **Total:**
Sample A: Mean 42, SD 1.5, CV 3.6%
Sample B: Mean 161, SD 4.3, CV 2.7%
Sample C: Mean 332, SD 9.1, CV 2.7% |
| Comparative Analysis Regression Evaluation | Regression Equation: y = 0.983x + 0.6
Correlation Coefficient of 0.9997
Sample Range: 12-1363 U/L | Regression Equation: y = 0.984x -1.3
Correlation Coefficient: 0.9995
Sample Range: 14-1139 U/L |
| Expected Values | 44-147 U/L | 35-123 U/L |
| Sample Stability | Serum ALP is stable for 7 days at 2-8°C and for 3 months at -20°C | Serum ALP is stable for 7 days at 2-8°C and for 3 months at -20°C |
| Detection Wavelength | 408/486 nm | 408/486 nm |
| Reagent Stability | Unopened ACE Alkaline Phosphatase Reagent is stable until the expiration date shown on the box and bottle labels when stored in the refrigerator at 2-8°C. | Unopened ACE Alkaline Phosphatase Reagent is stable until the expiration date shown on the box and bottle labels when stored in the refrigerator at 2-8°C. |
| Testing Environment | Clinical laboratories or physician office laboratories | Clinical laboratories or physician office laboratories |
| | Candidate Device | Predicate Device |
| --- | --- | --- |
| 510(k) # | k113436 | k931786 |
| Similarities | | |
| Intended Use/Indications | Same | ACE Amylase Reagent is intended for the quantitative determination of |
{5}
| | Candidate Device | Predicate Device |
| --- | --- | --- |
| 510(k) # | k113436 | k931786 |
| for Use | | α-amylase activity in serum |
| Calibration | Same | Enzyme activity directly determined by multiplying the change in absorbance per minute of the unknown samples by a constant factor based on the molar absorptivity of 2-chloro-p-nitrophenol |
| Method Traceability | Same | Tietz, N.W. (Ed.), Fundamentals of Clinical Chemistry, W.B. Saunders Co., Philadelphia, PA (1986); Sarber, R.L., Lishvin, L., Ramussen, J. and Blair, H.E., Clin. Chem. 32, 1136 (1986). |
| Use of Controls | Same | Two levels of control per day |
| Basic Principle | Same | Enzymatic assay for α-amylase |
| Measurement Type | Same | Reaction of α-amylase with chromogenic substrate (2-chloro-p-nitrophenyl-α-D-maltotrioside results in formation of 2-chloro-p-nitrophenol measured spectrophotometrically at 408/647 nm |
| Reactive Ingredients | Same | 2-Chloro-p-nitrophenyl-α-D-maltotrioside
Potassium thiocyanate
Sodium chloride
Calcium acetate
MES buffer (pH 6.0) |
| Non-reactive Ingredients | Same | Preservative |
| Dimensions | Same | Bottles with total volumes of 12 mL of reagent |
| Analysis Temperature | Same | 37°C |
| Reaction Type | Same | Kinetic |
| Sample Type | Same | Serum |
| Sample Volume | Same | 3 μL |
| Reaction Volume | Same | 168 μL |
{6}
| | Candidate Device | Predicate Device |
| --- | --- | --- |
| 510(k) # | k113436 | k931786 |
| (total) | | |
| Sample Stability | Same | Serum amylase is stable for 7 days at room temperature (18-26°C) and at 2-8°C for one month. Recommended storage is at 2-8°C. |
| Detection Wavelength | Same | 408/647 nm |
| Testing Environment | Same | Clinical laboratories or physician office laboratories |
| Differences | | |
| Instrument Platforms | ACE Axcel Clinical Chemistry System | ACE, ACE Alera® and NExCT™ Clinical Chemistry Systems |
| Detection Limit | 8.5 U/L | 0 U/L |
| Reportable Range | 9 to 1900 U/L | 0 to 1900 U/L |
| Endogenous Interferences | Bilirubin: For the low pool, no significant interference occurred below 27.95 mg/dL. Positive interference (19%) occurred at 55.9 mg/dL. For the high pool, no significant interference occurred.
Hemolysis: For the low pool, no significant interference occurred below 62.5 mg/dL. Negative interference (≥21%) occurred at ≥125 mg/dL. For the high pool, no significant interference occurred.
Lipemia (Intralipid): For low and high pools, no significant interference occurred below 1000 mg/dL. A MXINIT flag occurred at 2000 mg/dL.
Ascorbic Acid: No significant interference. | Bilirubin: No significant interference below 16.6 mg/dL. Concentrations greater than 16.6 mg/dL may cause interference. Samples with 33.2 mg/dL were found to positively interfere.
Hemolysis: No significant interference.
Lipemia (Intralipid): No significant interference. |
| Precision (U/L) | Within run:
Sample A: Mean 50.7, SD 1.7, CV 3.4%
Sample B: Mean 849.2, SD 15.6, CV 1.8%
Sample C: Mean 1619.6, SD 24.2, CV 1.5%
Sample D: Mean 64.1, SD 1.7, CV 2.6% | Within run:
Sample A: Mean 53, SD 1.3, CV 2.4%
Sample B: Mean 112, SD 3.2, CV 2.9%
Sample C: Mean 444, SD 10.9, CV 2.4%
Total:
Sample A: Mean 53, SD 2.2, CV |
{7}
Comparison for Alkaline Phosphatase (ALP):
| | Candidate Device | Predicate Device |
| --- | --- | --- |
| 510(k) # | k113436 | k931786 |
| Similarities | | |
| Intended Use/Indications for Use | Same | ACE LDH-L Reagent is intended for the quantitative determination of lactate dehydrogenase activity in serum. |
| Calibration | Same | Enzyme activity directly determined by multiplying the change in absorbance per minute of the unknown samples by a constant factor based on the molar absorptivity of NADH. |
| Calibration Stability | Same | Not a calibrated test |
| Method Traceability | Same | Wacker, W.E.C., Ulmer, D.D, Vallee, B.L., New Engl. J. Med., 255, 449 (1956). |
| Use of Controls | Same | Two levels of control per day |
| Basic Principle | Same | Conversion of L-lactate to pyruvate wherein NAD is converted to NADH |
| Measurement Type | Same | The rate of formation of NADH product is measured bichromatically at 340/647 nm. |
{8}
| | Candidate Device | Predicate Device |
| --- | --- | --- |
| 510(k) # | k113436 | k931786 |
| Reactive Ingredients | Same | L-lactate
Nicotinamide adenine dinucleotide |
| Non-reactive Ingredients | Same | Buffer, preservatives, stabilizers |
| Analysis Temperature | Same | 37°C |
| Reaction Type | Same | Kinetic |
| Sample Type | Same | Serum |
| Sample Volume | Same | 5 μL |
| Reaction Volume (total) | Same | 170 μL |
| Expected Values | Same | 100-190 U/L |
| Sample Stability | Same | Separated from cells, lactate dehydrogenase is stable for three days at both 2-8°C and room temperature. |
| Detection Wavelength | Same | 340/647 nm |
| Testing Environment | Same | Clinical laboratories or physician office laboratories |
| Differences | | |
| Instrument Platforms | ACE Axcel Clinical Chemistry System | ACE and ACE Alera® Clinical Chemistry Systems |
| Detection Limit | 8.3 U/L | 17 U/L |
| Reportable Range | 11 to 850 U/L | 17 to 850 U/L |
{9}
| | Candidate Device | Predicate Device |
| --- | --- | --- |
| 510(k) # | k113436 | k931786 |
| Endogenous Interferences | Bilirubin: No significant interference.
Hemolysis: For the low pool, significant interference occurred at all levels tested. A HI LIN flag occurred at 1000 mg/dL. For the high pool, no significant interference occurred below 31.25 mg/dL. Positive interference (≥12%) occurred at ≥62.5 mg/dL. A HI LIN flag occurred at 500 mg/dL. A DEPL flag occurred at 1000 mg/dL. Hemolysis of red cells release lactate dehydrogenase into the sample. Do not use hemolyzed samples.
Triglycerides: No significant interference.
Ascorbic Acid: No significant interference. | Bilirubin: No significant interference.
Hemolysis: Positive interference at 31 mg/dL
Lipemia (Intralipid): Positive interference at 500 mg/dL
Triglycerides: No significant interference below 460 mg/dL
Ascorbic Acid: No significant interference
Lactic Acid: No significant interference |
| Precision (U/L) | Within run:
Sample A: Mean 130.5, SD 3.0, CV 2.3%
Sample B: Mean 437.4, SD 7.1, CV 1.6%
Sample C: Mean 720.6, SD 11.8, CV 1.6%
Sample D: Mean 94.7, SD 2.9, CV 3.1%
Total:
Sample A: Mean 130.5, SD 4.1, CV 3.2%
Sample B: Mean 437.4, SD 10.2, CV 2.3%
Sample C: Mean 720.6, SD 16.5, CV 2.3%
Sample D: Mean 94.7, SD 4.3, CV 4.6% | Within Run:
Sample A: Mean 82, SD 5.1, CV 6.2%
Sample B: Mean 122, SD 5.7, CV 4.7%
Sample C: Mean 282, SD 4.6, CV 1.6%
Sample D: Mean 680, SD 9.7, CV 1.4%
Total:
Sample A: Mean 82, SD 7.0, CV 8.5%
Sample B: Mean 122, SD 7.7, CV 6.3%
Sample C: Mean 282, SD 10.6, CV 3.8%
Sample D: Mean 680, SD 18.4, CV 2.7% |
| Comparative Analysis Regression Evaluation | Regression Equation: y = 1.046x + 4.9
Correlation Coefficient of 0.9986
Sample Range: 22-829 U/L | Regression Equation: y = 0.965x + 0.7
Correlation Coefficient: 0.9994
Sample Range: 20-800 U/L |
{10}
11
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition
CLSI EP6-A: Evaluation of Linearity of Quantitative Measurement Procedures, A Statistical Approach; Approved Guideline
CLSI EP7-A2: Interference Testing in Clinical Chemistry; Approved Guideline-Second Edition
CLSI EP9-A2-IR: Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline-Second Edition
CLSI EP10-A3: Preliminary Evaluation of Quantitative Clinical Laboratory Measurement Procedures; Approved Guideline-Third Edition
CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline
L. Test Principle:
The ACE Alkaline Phosphatase Reagent for the Axcel Clinical Chemistry System is an enzymatic photometric test; alkaline phosphatase in serum catalyzes the hydrolysis of the p-nitrophenyl phosphate substrate to produce inorganic phosphate and p-nitrophenol product which is measured at 408 nm. The rate of increase of absorbance is directly proportional to the amount of alkaline phosphatase activity in the serum sample.
$$
\mathrm{p}\text{-nitrophenyl phosphate} + \mathrm{H}_2\mathrm{O} \longrightarrow {}^{(\mathrm{ALP}/\mathrm{Mg}^2+)} \longrightarrow \mathrm{p}\text{-nitrophenol} + \mathrm{H}_3\mathrm{PO}_4
$$
The ACE Amylase Reagent for the Axcel Clinical Chemistry System is an enzymatic photometric test; α-amylase in serum catalyzes the reaction of the maltotrios-linked 2-chloro-p-nitrophenol phosphate substrate to produce 2-chloro-p-nitrophenol product which is measured at 408 nm. The rate of increase of absorbance is directly proportional to the amount of amylase activity in the serum sample.
$$
10\mathrm{CNPG}_3 \longrightarrow {}^{(\mathrm{a}\text{-Amylase})} \longrightarrow 9\mathrm{CNP} + \mathrm{CNPG}_2 + 9\mathrm{G}_3 + \text{glucose}
$$
The ACE LDH-L Reagent for the Axcel Clinical Chemistry System is an enzymatic photometric test; LDH in serum catalyzes the conversion of the L-lactate and NAD substrates to pyruvate and NADH, and the NADH product which is measured at 340 nm. The rate of increase of absorbance from the formation of NADH is directly proportional to the amount of LDH activity in the serum sample.
$$
\mathrm{L}\text{-lactate} + \mathrm{NAD}^+ \longrightarrow {}^{(\mathrm{LDH})} \longrightarrow \text{pyruvate} + \mathrm{NADH} + \mathrm{H}^+
$$
{11}
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
In-house precision
| Precision studies were conducted by testing human serum pools at four levels. The samples were run 2 times per run, 2 runs per day, for a total of 20 days using one instrument. Results are summarized below. Alkaline Phosphatase | | Sample 1 | Sample 2 | Sample 3 | Sample 4 |
| --- | --- | --- | --- | --- | --- |
| | Mean (U/L) | 50.0 | 690.0 | 1020.6 | 60.9 |
| Within Run | SD | 1.6 | 9.8 | 13.2 | 1.6 |
| | %CV | 3.2 | 1.4 | 1.3 | 2.6 |
| Between Run | SD | 0.0 | 0.0 | 7.0 | 1.2 |
| | %CV | 0.0 | 0.0 | 0.7 | 2.0 |
| Between Day | SD | 1.5 | 17.4 | 24.0 | 2.0 |
| | %CV | 2.9 | 2.5 | 2.4 | 3.3 |
| Total | SD | 2.2 | 2.5 | 28.3 | 2.9 |
| | %CV | 4.4 | 2.9 | 2.8 | 4.7 |
| Amylase | | Sample 1 | Sample 2 | Sample 3 | Sample 4 |
| --- | --- | --- | --- | --- | --- |
| | Mean (U/L) | 50.7 | 849.2 | 1619.6 | 64.1 |
| Within Run | SD | 1.7 | 15.6 | 24.2 | 1.7 |
| | %CV | 3.4 | 1.8 | 1.5 | 2.6 |
| Between Run | SD | 0.0 | 0.0 | 9.4 | 0.0 |
| | %CV | 0.0 | 0.0 | 0.6 | 00.0 |
| Between Day | SD | 0.6 | 6.3 | 7.1 | 0.4 |
| | %CV | 1.2 | 0.7 | 0.4 | 0.7 |
| Total | SD | 1.8 | 16.8 | 26.9 | 1.7 |
| | %CV | 3.6 | 2.0 | 1.7 | 2.7 |
| LDH | | Sample 1 | Sample 2 | Sample 3 | Sample 4 |
| --- | --- | --- | --- | --- | --- |
| | Mean (U/L) | 130.5 | 437.4 | 720.6 | 94.7 |
| Within Run | SD | 3.0 | 7.1 | 11.8 | 2.9 |
| | %CV | 2.3 | 1.6 | 1.6 | 3.1 |
| Between Run | SD | 0.7 | 2.7 | 4.2 | 1.9 |
| | %CV | 0.6 | 0.6 | 0.6 | 2.0 |
{12}
| Between Day | SD | 2.7 | 6.7 | 10.9 | 2.6 |
| --- | --- | --- | --- | --- | --- |
| | %CV | 2.1 | 1.5 | 1.5 | 2.8 |
| Total | SD | 4.1 | 10.2 | 16.5 | 4.3 |
| | %CV | 3.2 | 2.3 | 2.3 | 4.6 |
# Point-of-Care precision
Precision studies were also conducted at three Physician Office Laboratories (POL) with four trained operators typically found in these settings. Human serum pools at three concentrations were tested three times a day for five days on three instruments (one at each lab). The results are presented below:
| Alkaline Phosphatase | | | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| Lab | Sample | Mean (U/L) | SD | %CV | SD | %CV |
| POL 1 | 1 | 49.9 | 2.4 | 4.8 | 2.9 | 5.7 |
| POL 2 | 1 | 49.4 | 1.4 | 2.8 | 1.7 | 3.5 |
| POL 3 | 1 | 48.3 | 0.6 | 1.3 | 1.3 | 2.7 |
| | | | | | | |
| POL 1 | 2 | 679.5 | 7.1 | 1.0 | 31.8 | 4.7 |
| POL 2 | 2 | 664.7 | 8.5 | 1.3 | 23.8 | 3.6 |
| POL 3 | 2 | 676.1 | 7.8 | 1.2 | 27.5 | 4.1 |
| | | | | | | |
| POL 1 | 3 | 1340.9 | 20.7 | 1.5 | 43.7 | 3.3 |
| POL 2 | 3 | 1293.4 | 15.9 | 1.2 | 56.3 | 4.4 |
| POL 3 | 3 | 1041.0 | 12.4 | 1.2 | 21.2 | 2.0 |
| Amylase | | | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| Lab | Sample | Mean (U/L) | SD | %CV | SD | %CV |
| POL 1 | 1 | 53.3 | 2.5 | 4.7 | 2.5 | 4.7 |
| POL 2 | 1 | 50.1 | 1.9 | 3.8 | 2.9 | 5.7 |
| POL 3 | 1 | 55.0 | 2.3 | 4.2 | 2.3 | 4.2 |
| | | | | | | |
| POL 1 | 2 | 908.9 | 10.5 | 1.2 | 16.1 | 1.8 |
| POL 2 | 2 | 872.7 | 30.9 | 3.5 | 41.6 | 4.8 |
| POL 3 | 2 | 945.6 | 8.6 | 0.9 | 18.0 | 1.9 |
| | | | | | | |
| POL 1 | 3 | 1749.3 | 25.0 | 1.4 | 41.7 | 2.4 |
| POL 2 | 3 | 1669.1 | 13.6 | 0.8 | 30.9 | 1.9 |
| POL 3 | 3 | 1782.5 | 16.9 | 0.9 | 16.9 | 0.9 |
| LDH | | | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| Lab | Sample | Mean (U/L) | SD | %CV | SD | %CV |
| POL 1 | 1 | 10.0 | 1.0 | 0.0 | 10.0 | 0.0 |
| POL 2 | 1 | 10.0 | 0.0 | 0.0 | 10.0 | 0.0 |
| POL 3 | 1 | 10.0 | 0.0 | 0.0 | 10.0 | 0.0 |
{13}
| POL 1 | 1 | 126.1 | 1.8 | 1.4 | 3.1 | 2.5 |
| --- | --- | --- | --- | --- | --- | --- |
| POL 2 | 1 | 132.3 | 3.0 | 2.2 | 3.0 | 2.2 |
| POL 3 | 1 | 130.9 | 2.3 | 1.8 | 2.8 | 2.2 |
| | | | | | | |
| POL 1 | 2 | 420.8 | 12.6 | 3.0 | 13.8 | 3.3 |
| POL 2 | 2 | 442.1 | 8.3 | 1.9 | 8.3 | 1.9 |
| POL 3 | 2 | 442.1 | 6.8 | 1.5 | 9.0 | 2.0 |
| | | | | | | |
| POL 1 | 3 | 701.1 | 9.6 | 1.4 | 12.4 | 1.8 |
| POL 2 | 3 | 727.2 | 15.2 | 2.1 | 18.9 | 2.6 |
| POL 3 | 3 | 738.3 | 8.4 | 1.1 | 12.3 | 1.7 |
b. Linearity/assay reportable range:
Linearity across the assay range was confirmed by spiking serum samples to a high concentration of analyte, then diluting the sample to obtain between 12 and 15 levels to cover the measuring range of each assay. The assigned value of the highest sample was set to its mean value. The assigned values of the other levels were calculated by multiplying the mean value by the dilution ratios obtained from the manufacturer. Each level was tested in replicates of four. Data was analyzed to show linear regression equations and also the $2^{\text{nd}}$ and $3^{\text{rd}}$ polynomial equations, and all data demonstrated that the 3 devices were linear across the claimed measuring range. Results are presented below:
Alkaline Phosphatase
Linear Regression: $y = 0.973x - 1.3$, $r^2 = 0.9976$
$2^{\text{nd}}$ Order: $y = 0.000036x^2 + 0.925x + 4.28$
$3^{\text{rd}}$ Order: $y = 0.0000001x^3 - 0.00013x^2 + 1.01x - 0.260$
Claimed measuring range: 9 - 1400 U/L
Amylase
Linear Regression: $y = 1.006x + 4.8$, $r^2 = 0.9995$
$2^{\text{nd}}$ Order: $y = 0.00001x^2 + 1.041x - 1.62$
$3^{\text{rd}}$ Order: $y = -0.0x^3 + 0.0000013x^2 + 1.03x - 0.460$
Claimed measuring range: 9 - 1900 U/L
LDH
Linear Regression: $y = 1.015x + 7.36$, $r^2 = 0.9980$
$2^{\text{nd}}$ Order: $y = 0.000112x^2 + 1.11x - 1.51$
$3^{\text{rd}}$ Order: $y = -0.0000002x^3 + 0.000194x^2 + 1.01x + 2.65$
Claimed measuring range: 11 - 850 U/L
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
{14}
Traceability:
Calibration of the LDH-L assay is traceable to a frozen Master Pool of verification material utilized by the reagent supplier. Each lot of reagent is tested by running the Master Pool and verifying that results of the Master Pool levels are within an acceptable percentage of the assigned values of the Master Pool. For value assignment, each new verification Master Pool is made by gravimetrically adding quantities of lactate dehydrogenase to a serum pool to target concentrations. Five levels of Master Pool are prepared, aliquoted and stored at $\leq -70^{\circ}\mathrm{C}$ . The final values of the Master Pool are assigned for each level by testing at least 3 replicates on multiple instruments. The activity levels of the new Master Pool are verified using a previously approved Master Pool lot as a control.
The ACE Alkaline Phosphatase and Amylase reagents are traceable to an IFCC traceable method, using a linearity verification set with various levels run in triplicate and assessed for linearity versus the assigned values from the linearity set.
# d. Detection limit:
The limit of detection and the limit of the blank were determined by assaying five low samples (serum samples) and five true blanks (human serum albumin in saline). Testing was carried out over three days on two ACE Axcel Clinical Chemistry Analyzers. Serum samples and true blanks were assayed every day for a total of 60 measurements. The limit of quantitation was determined with 40 replicates of 3 low samples, and was determined to be the mean when the $\% \mathrm{CV}$ was $\leq 20\%$ .
| Analyte | LoB (U/L) | LoD (U/L) | LoQ (U/L) |
| --- | --- | --- | --- |
| Alkaline Phosphatase | 1.1 | 1.3 | 6 |
| Amylase | 7.9 | 8.5 | 9 |
| LDH | 7.8 | 8.3 | 11 |
# e. Analytical specificity:
Interference studies were performed to determine the effects from potential interferents. The various concentrations of interferent were spiked into serum pools containing alkaline phosphatase, amylase and LDH at normal and abnormal concentrations. Hemolysis was simulated using a freeze-thaw method to lyse the red cells. All samples were tested in triplicate. Seven levels were tested for each interferent. Significant interference was defined as a difference in analyte recovery of more than $\pm 10\%$ .
Alkaline Phosphatase:
| Interferent Compound | Concentration with No Interference Up To |
| --- | --- |
| Ascorbic Acid | 6 mg/dL |
{15}
| Unconjugated Bilirubin | 28 mg/dL |
| --- | --- |
| Hemolysis (hemoglobin) | 62.5 mg/dL* |
| Intralipid | 500 mg/dL |
Amylase:
| Interferent Compound | Concentration with No Interference Up To |
| --- | --- |
| Ascorbic Acid | 6 mg/dL |
| Unconjugated Bilirubin | 28 mg/dL |
| Hemolysis (hemoglobin) | 62.5 mg/dL* |
| Intralipid | 1000 mg/dL |
LDH:
| Interferent Compound | Concentration with No Interference Up To |
| --- | --- |
| Ascorbic Acid | 6 mg/dL |
| Unconjugated Bilirubin | 62 mg/dL |
| Hemolysis (hemoglobin) | Interference at All Levels* |
| Triglycerides** | 2620 mg/dL |
*The package insert contains the following statement: Do not use hemolyzed samples.
**Triglycerides were used in the study with LDH as the sponsor suspected matrix interferences from intralipid with this analyte.
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
An in-house method comparison study to the predicate device was performed with serum patient samples. A total of 113 alkaline phosphatase (103 native, 5 diluted and 5 spiked) serum samples, 111 amylase (101 native, 4 diluted and 6 spiked) serum sample, and 121 LDH (109 native, 5 dilutes and 7 spiked) serum samples covering the assay range were tested. The results are presented in the table below:
| Analyte | n | Regression Equation | r2 | Standard Error | Sample range (U/L) |
| --- | --- | --- | --- | --- | --- |
| Alkaline Phosphatase | 112 | y=0.983x+0.6 | 0.9997 | 5.1 | 12-1363 |
| Alkaline Phosphatase | 112 | y=0.983x+0.6 | 0.9997 | 5.1 | 12-1363 |
{16}
Additional method comparison studies were performed at three Physician Office Laboratories, with four operators. Operators assayed serum samples ranging from 11-1388 U/L alkaline phosphatase, 12-1856 U/L amylase, and 18-819 LDH on the Ace Axcel clinical chemistry analyzer and the ACE clinical chemistry System. The results are presented in the tables below:
Alkaline Phosphatase
| POL | n | Regression Equation | r² | Standard Error | Sample range (U/L) |
| --- | --- | --- | --- | --- | --- |
| 1 | 68 | y=1.040x+3.5 | 0.9957 | 25.3 | 11-1311 |
| 2 | 53 | y=0.972x+1.5 | 0.9998 | 6.0 | 49-1261 |
| 3 | 49 | y=1.000x+8.7 | 0.9983 | 16.8 | 26-1388 |
Amylase
| POL | n | Regression Equation | r² | Standard Error | Sample range (U/L) |
| --- | --- | --- | --- | --- | --- |
| 1 | 56 | y=0.997x-2.5 | 0.9998 | 7.6 | 12-1819 |
| 2 | 49 | y=0.984x-0.5 | 0.9985 | 22.3 | 19-1856 |
| 3 | 47 | y=1.019x-1.5 | 1.0000 | 3.4 | 18-1797 |
LDH
| POL | n | Regression Equation | r² | Standard Error | Sample range (U/L) |
| --- | --- | --- | --- | --- | --- |
| 1 | 60 | y=1.010x-1.1 | 0.9983 | 13.1 | 33-819 |
| 2 | 53 | y=1.042x-5.7 | 0.9993 | 6.3 | 18-773 |
| 3 | 47 | y=1.011x+2.6 | 0.9988 | 7.9 | 22-787 |
b. Matrix comparison:
The device is being cleared for serum use only.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
{17}
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable. Clinical studies are not typically submitted for this device type.
5. Expected values/Reference range:
The reference range for each analyte was verified according to CLSI C28-A3. 50 normal healthy patient samples from a diverse population with an age range from 20 to 60, were each analyzed for ALP, Amylase, and LDH. The 95% confidence intervals were calculated and the reference range was shown to validate the reference range stated in the literature (for amylase and LDH: Tietz Clinical Guide to Laboratory Tests, 4th Ed, Wu et al; for ALP: Medline Plus reference range data base, U.S. National Library of Medicine, National Institutes of Health).
ALP: 44 – 147 U/L
Amylase: 20 – 104 U/L
LDH: 100 – 190 U/L
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
18
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.